Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K210106
    Date Cleared
    2021-04-07

    (82 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nitrile Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The subject device is powder free nitrile examination gloves. The subject device is blue. The subject device is non-sterile.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for Nitrile Examination Gloves submitted by Zhenjiang Huayang Latex Products Co., Ltd. Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test ItemAcceptance Criteria (Predicate Device)Reported Device Performance (Subject Device)Remark
    Physical Properties
    Before Aging
    Tensile Strength14MPa, min14MPa, minSame
    Ultimate Elongation500% min500% minSame
    After Aging
    Tensile Strength14MPa, min14MPa, minSame
    Ultimate Elongation400%min400%minSame
    Overall ComplianceComply with ASTM D6319Comply with ASTM D6319Same
    Freedom from HolesBe free from holes when tested in accordance with ASTMD5151 AQL=2.5Be free from holes when tested in accordance with ASTMD5151 AQL=2.5Same
    Powder ContentMeet the requirements of ASTM D61240.02 mg per gloveSame
    Biocompatibility
    IrritationNot an irritant(Implicitly meets standard ISO 10993-10:2010)Same
    SensitizationNot a sensitizer(Implicitly meets standard ISO 10993-10:2010)Same
    CytotoxicityDevice extract is not cytotoxic(Implicitly meets standard ISO 10993-5:2009)Same
    Dimensions
    Length (mm)230 min (for S, M, L, XL sizes)230 min (for S, M, L sizes)Similar (although subject device only tested S, M, L)
    Width (mm)75, 85, 95, 105, 115 ±5 (for XS, S, M, L, XL)80, 95, 110 ±10 (for S, M, L)Similar (Physical dimensions are different but meet ASTM D6319-19 requirements)
    Finger Thickness (mm)0.05 min0.05 minSimilar
    Palm Thickness (mm)0.05 min0.05 minSimilar

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the sample sizes used for the non-clinical tests (physical properties, freedom from holes, powder content, and biocompatibility). The provenance of the data is retrospective, as it refers to testing conducted on the subject device to demonstrate substantial equivalence to a legally marketed predicate device. The country of origin of the device manufacturer is China (Zhenjiang Huayang Latex Products Co., Ltd. in Jiangsu Province, P.R.China).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The "ground truth" here is established by standardized test methods (e.g., ASTM, ISO) performed in a laboratory setting, rather than through expert consensus in a medical imaging or diagnostic context.

    4. Adjudication method for the test set

    This information is not applicable to this type of device comparison. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies involving human interpretation (e.g., radiologists reading images) where disagreement among experts needs resolution. For a medical glove, performance is evaluated against defined test standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. An MRMC study is relevant for AI-assisted diagnostic tools. This device is a nitrile examination glove, which does not involve human readers or AI assistance in a diagnostic capacity.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable. This refers to the performance of an algorithm without human input, which is not relevant for a medical glove.

    7. The type of ground truth used

    The ground truth used for this device is based on standardized test methods and specifications (e.g., ASTM D6319, ASTM D5151, ASTM D6124, ISO 10993-10, ISO 10993-5) and comparison to a legally marketed predicate device.

    8. The sample size for the training set

    This is not applicable. This device is not an AI/ML algorithm that requires a "training set." Its performance is evaluated through physical and chemical testing.

    9. How the ground truth for the training set was established

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1