Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K203539
    Manufacturer
    Date Cleared
    2021-12-10

    (372 days)

    Product Code
    Regulation Number
    878.4040
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unico Global Level 3 Surgical Gown (Film-reinforced SMS) and Unico Global Level 3 Surgical Gown (PP+PE) are sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the healthcare worker from the transfer of microorganisms, body fluids and particulate matter.

    Device Description

    The subject device offers the two material models which are Unico Global Level 3 Surgical Gown (Film-reinforced SMS) and Unico Global Level 3 Surgical Gown (PP+PE). The two models use different base materials but use the same PE film layer in addition to the fabric. They have slightly different design elements in neck, body, and belt straps.

    The Unico Global Level 3 Surgical Gowns (Film-reinforced SMS and PP+PE) have been tested according to AATCC Test Method 127:2017, ANSI/AAMI PB70:2012, and ISO 811, and meet the AAMI Level 3 barrier level protection for a surgical gown.

    The Unico Global Level 3 Surgical Gowns are single use that will be provided in both sterile and non-sterile packaging configurations and offer a variety of sizes ranging from medium to XX-large.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device: Unico Global Level 3 Surgical Gown. It is a premarket notification to the FDA to demonstrate substantial equivalence to a predicate device. This document does NOT contain information about an AI/ML powered device, nor does it contain a comparative effectiveness study with human readers (MRMC study) or a standalone algorithm performance study.

    Therefore, many of the requested details about acceptance criteria and study design for an AI/ML device cannot be extracted from this document. The document describes the performance testing of a physical product (surgical gowns) against established standards for physical properties and safety.

    However, I can extract the acceptance criteria and performance for the surgical gown itself as presented in the document.

    Detailed Breakdown based on the provided document:

    The product in question is a physical medical device (surgical gown), not an AI/ML powered device. As such, the study design and acceptance criteria relate to the physical properties, safety, and performance of the gown, not to AI model performance, human reader improvement, or ground truth establishment in the context of AI/ML.

    Here's an attempt to answer your questions based on the available information in the provided document, acknowledging the mismatch with a typical AI/ML device request:


    1. A table of acceptance criteria and the reported device performance

    Test ItemAcceptance CriteriaReported Device Performance
    Sterilization ValidationSterility Assurance Level: 10^-6Pass
    Shelf-Life ValidationBioburden test: ISO 11737-1 ('Sterilization of medical devices – Microbiological methods – Part 1: Determination of a population of microorganisms on products' Sterility test: No microbial growth Seal Strength Test (Blister): Maximum seal strength (N/15mm) ->>1.2 Dye Penetration test: No leakagePass
    Acute systemic toxicityAdverse no effects occurring at any time within 72h after single, multiple or continuous exposures of a test sample for 24h.Pass
    Sensitization & Irritation TestsSensitization test: The observations are scored between 0 (no visible change) to 3 (intense erythema and swelling). More than grade 0 indicates sensitization. Irritation tests: The test result obtained is a score between 0 and 8 calculated from the various observations, considered as negligible (0-0.4), slight (0.5-1.9), moderate (2-4.9), or severe (5-8).Pass
    Water Resistance: Impact PenetrationVisual Penetration: None seen Amount of Penetration (g) : ≤ 1.0 gPass
    Water Resistance: Hydrostatic PressurePressure: ≥ 50 cm (Specifically for Level 3 Surgical Gowns based on ANSI/AAMI PB70:2012)Pass
    Tear ResistanceNo acceptance criteria currently exist for this test.Pass
    Basis WeightTest articles should be weighed, and the basis weight should be calculated as the mass divided by the area. all should weigh the same.Pass
    Sterilant Gas Residue AnalysisMaximum levels of sterilant residuals: ETO
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1