Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K181823
    Device Name
    KardiaAI
    Manufacturer
    Date Cleared
    2019-03-11

    (245 days)

    Product Code
    Regulation Number
    870.1425
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KardiaAI is a software analysis library intended to assess ambulatory electrocardiogram (ECG) rhythms from adult subjects. The device supports analyzing data recorded in compatible formats from any ambulatory ECG devices such as event recorders, or other similar devices. The library is intended to be integrated into other device software. The library is not intended for use in life supporting, or sustaining systems, or cardiac alarm, or OTC use only devices.

    The KardiaAI library provides the following capabilities:

    • ECG noise filtering.
    • heart rate measurement from ECGs,
    • detection of noisy ECGs, and
    • ECG rhythm analysis for detecting the presence of normal sinus rhythm, atrial fibrillation, bradycardia (when prescribed or used under the care of a physician).
    Device Description

    KardiaAI is a software library that implements various ECG processing and analysis algorithms. This Software as a Medical Device (SaMD) computes various physiologic parameters from a 30-second ECG and provides these capabilities in the form of an Application Program Interface (API) library. ECG devices can incorporate the API library into ECG device ("target device") software to enable algorithmic analysis of ECGs to provide analytical capabilities. The device provides ECG noise filtering and detection of noisy ECGs as well as identifies normal sinus rhythm, atrial fibrillation, bradycardia, and tachycardia.

    AI/ML Overview

    The KardiaAI device, a software analysis library for assessing ambulatory ECG rhythms, underwent non-clinical testing to demonstrate its performance and substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text states that "The overlapping AF, NSR, and noise algorithms for KardiaAI and Kardia Band System met the same performance criteria." It also mentions that "Testing also ensured that differences in technological characteristics between the KardiaAI and the Kardia Band System (primary predicate) (i.e., bradycardia and tachycardia algorithms as well as multilead ambulatory ECG input) perform as intended and do not raise different questions of safety or effectiveness."

    However, specific numerical acceptance criteria (e.g., sensitivity, specificity thresholds for AF, NSR, bradycardia, tachycardia, or noise detection) and the corresponding reported device performance values are not explicitly detailed in the provided document. The document primarily describes that acceptance criteria were met and that the device performs as intended, rather than listing the criteria themselves with quantitative results.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that "Algorithm performance testing was assessed using ECG databases from the ANSI/AAMI EC57:2012 standard as well as AliveCor proprietary databases."

    • Sample Size: The specific sample sizes (number of ECGs or patients) from the ANSI/AAMI EC57:2012 standard databases and the AliveCor proprietary databases used for testing are not explicitly stated.
    • Data Provenance:
      • ANSI/AAMI EC57:2012 standard databases: These are standardized databases, typically containing diverse ECG recordings established for performance evaluation of ECG devices. The country of origin is not specified but these are internationally recognized standards.
      • AliveCor proprietary databases: The country of origin is not explicitly stated, nor is whether the data is retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number of experts used to establish the ground truth for the test set or their specific qualifications. It implicitly refers to "ECG databases from the ANSI/AAMI EC57:2012 standard" which typically have established ground truths, but the methodology for these specific tests is not detailed.

    4. Adjudication Method for the Test Set

    The document does not explicitly state the adjudication method used for the test set (e.g., 2+1, 3+1, none).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it provide an effect size for human readers improving with AI vs. without AI assistance. The testing described is focused on the standalone algorithm performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone study was done. The document explicitly states: "Non-clinical testing was conducted to assess algorithm performance and to verify that KardiaAI performs as intended. Algorithm performance testing was assessed using ECG databases..." This indicates a focus on the algorithm's performance in isolation.

    7. The Type of Ground Truth Used

    The ground truth for the test set seems to be derived from expert consensus embedded within the "ECG databases from the ANSI/AAMI EC57:2012 standard" and "AliveCor proprietary databases." While not explicitly stated, standardized ECG databases are typically annotated by cardiologists or other qualified experts, aligning with an expert consensus type of ground truth.

    8. The Sample Size for the Training Set

    The document does not explicitly state the sample size used for the training set. It only mentions the databases used for testing.

    9. How the Ground Truth for the Training Set Was Established

    The document does not explicitly state how the ground truth for the training set was established. While it implies the use of "AliveCor proprietary databases," the method for their ground truth annotation (e.g., expert review, pathology, outcomes data) is not detailed.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1