Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K171873
    Date Cleared
    2017-11-15

    (145 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    These gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    The subject device in this 510(k) Notification is a Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs

    The subject device is a patient examination glove made from nitrile compound, blue in color, powder free and non-sterile (as per 21 CFR 880.6250, class I).

    The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

    AI/ML Overview

    This document describes the acceptance criteria and performance study for the Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs (K171873).

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
    DimensionASTM standard D 6319-10Meets
    Physical PropertiesASTM standard D 6319-10Meets
    Freedom from pinholes21 CFR 800.20, ASTM D5151-11Meets
    Powder ResidualASTM D6319-10 and D6124-06 (Reapproved 2011)Meets
    Biocompatibility (Skin Irritation)ISO 10993-10:2010 Primary Skin IrritationNot a primary skin irritant under the conditions of the study
    Biocompatibility (Sensitization)ISO 10993-10:2010 Dermal sensitization in the guinea pigNot a contact sensitizer under the conditions of the study
    Chemotherapy Drug PermeationASTM D6978-05 (Reapproved 2013) - Minimum Breakthrough Detection Time (drug specific)See table below for specific drug performance

    Chemotherapy Drug Permeation Performance:

    Drug and ConcentrationMinimum Breakthrough Detection Time in Minutes (Acceptance Criteria per ASTM D6978-05)Reported Device Performance (Minimum Breakthrough Detection Time in Minute)
    Carmustine (BCNU) 3.3 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.14.7
    Cisplatin 1.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
    Cyclophosphamide (Cytoxan) 20 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
    Cytarabine 100 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
    Dacarbazine (DTIC) 10.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
    Doxorubicin Hydrochloride 2.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
    Etoposide (Toposar) 20.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
    Fluorouracil 50.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
    Ifosfamide 50.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
    Methotrexate 25 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
    Mitomycin C 0.5 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
    Mitoxantrone 2.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
    Paclitaxel (Taxol) 6.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240
    Thiotepa 10.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.58.8
    Vincristine Sulfate 1.0 mg/mlNot explicitly stated as a minimum in the document, but assessed against predicate.>240

    Note: The document highlights that Carmustine (BCNU) and Thiotepa have "extremely low permeation times" and a warning statement is to be included in labeling for these specifically.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a distinct "test set" in terms of a dataset of items being classified or evaluated by an algorithm or human readers. Instead, it refers to bench tests conducted on the device itself.

    • Sample Size: Not explicitly stated for each specific test (e.g., number of gloves tested for pinholes, physical properties, or chemotherapy permeation). The general statement is "Bench tests were conducted to verify that the proposed device met all specifications."
    • Data Provenance: The tests are described as "Bench tests" which are typically laboratory-based and prospective in nature for a new device. The country of origin for the data is not specified, but the applicant company is Zhonghong Pulin Medical Products Co., Ltd. located in China.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable to this type of device submission. The "ground truth" for glove performance is established through adherence to recognized international standards (e.g., ASTM, ISO) and regulatory requirements (e.g., 21 CFR 800.20), not through expert consensus on a test set of data. The performance is objectively measured against predefined criteria in these standards.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, performance is measured against objective standards, not through consensus or adjudication of expert interpretations.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This type of study is not relevant for a medical device like an examination glove. MRMC studies are typically used for evaluating diagnostic imaging or AI algorithms where human interpretation plays a significant role.

    6. Standalone Performance Study (Algorithm Only Without HIL)

    No. This device is a physical medical glove and does not involve AI algorithms or human-in-the-loop performance evaluation in the context of an algorithm. The "device performance" refers to the physical and chemical properties of the glove itself.

    7. Type of Ground Truth Used

    The "ground truth" for evaluating the glove's performance is based on established industry standards and regulatory requirements. This includes:

    • ASTM D6319-10: Standard Specification for Nitrile Examination Gloves for Medical application (Dimension, Physical Properties, Powder Residual).
    • 21 CFR 800.20 & ASTM D5151-06 (2011): Freedom from pinholes.
    • ASTM D6124-06 (2011): Standard Test Method for Residual Powder on Medical Gloves.
    • ISO 10993-10:2010: Biocompatibility (Primary Skin Irritation, Dermal Sensitization).
    • ASTM D6978-05 (Reapproved 2013): Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    8. Sample Size for the Training Set

    Not applicable. This device is not an AI algorithm that requires a training set. Its performance is evaluated through physical and chemical testing.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1