Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K162186
    Date Cleared
    2017-03-31

    (239 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES 12", POWDER FREE, [STERLING/GREY] WITH CHEMOTHERAPY CLAIM is a disposable device intended for medical purpose, to be worn on the examiners hand or finger to prevent contamination between patient and examiner.

    The product has been tested with chemotherapy drugs in accordance with ASTM D6978. The breakthrough detection times are as follows:

    Test Chemotherapy Drug Name and Concentration Minimum Breakthrough Detection Time Carmustine (BCNU) (3.3mg/ml),(3,300 ppm) ● 4.3 Minutes >240 Minutes

    240 Minutes

    240 Minutes

    240 Minutes

    240 Minutes

    240 Minutes

    240 Minutes

    240 Minutes

    240 Minutes

    10.9 Minutes

    240 Minutes

    • Cisplatin(1.0mg/ml), (1,000 ppm)
    • Cyclophosphamide/Cytoxan(20mg/ml), (20,000 ppm)
    • Dacarbazine(DTIC) 10mg/ml, (10,000 ppm) ●
    • Doxorubicin HCL(2mg/ml),(2,000 ppm)
    • Etoposide /Toposar (20mg/ml), (20,000 ppm)
    • Fluorouracil(50mg/ml), (50,000 ppm)
    • Ifosfamide (50mg/ml), (50,000 ppm)
    • Mitoxantrone(2mg/ml),(2,000 ppm)
    • Paclitaxel(6.0mg/ml),(6,000 ppm)
    • Thiotepa (10mg/ml),(10,000 ppm)
    • . Vincristine Sulfate(1.0mg/ml),(1,000 ppm)

    Please note that Carmustine and Thiotepa have extremely low permeation times of 4.3 minutes and 10.9 minutes, respectively.

    Warning: Do not use with Carmustine and Thiotepa.

    Device Description

    The subject device in this 510(k) Notification is Sterling/Grey coloured Nitrile Examination gloves, with claims and tested for use with Chemotherapy drugs.

    The subject device is a patient examination glove made from nitrile compound, Sterling or Grey in colour, powder free and non sterile (as per 21 CFR 880.6250, class I).

    The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

    This device also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy drugs as per ASTM D6978-05

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves the BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES 12", POWDER FREE, [STERLING/GREY] WITH CHEMOTHERAPY CLAIM meets these criteria.

    Here's the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsStandardAcceptance Criteria (Predicate Device K081089)Reported Device Performance (Subject Device)Result of Comparison
    LabelingN/ANo special labeling claims, no claim of hypoallergenic.No special labeling claims, no claim of hypoallergenic.Same
    Device MaterialsN/ANitrile CompoundNitrile CompoundSame
    ColourN/ASterlingSterling / GreySame
    Tensile strength (Before and after ageing)ASTM 6319-10> 14 MPa> 14 MPaSame
    Ultimate Elongation (Before and after ageing)ASTM 6319-10> 500 %> 500 %Same
    Freedom from PinholesASTM 6319-10 (FDA 1000 ml water leak test)PassPassSame
    Dimensions: LengthASTM 6319-10> 295 mm> 295 mm (295 mm minimum - 325 mm)Same
    Dimensions: WidthASTM 6319-1070±10 mm to 120±10 mm (sizes XS to XL)70±10 mm to 120±10 mm (sizes XS to XL)Same
    Dimensions: ThicknessASTM 6319-10> 0.05 mm (palm & finger)> 0.05 mm (Palm: 0.075mm-0.08mm; Finger: 0.085mm-0.10mm)Same
    Residual PowderASTM 6319-10Less than 2 mg per glove; PASSLess than 2 mg per glove; PASSSame
    Primary Skin Irritation testISO 10993-10Under conditions of the study, not an irritantUnder conditions of the study, not an irritantSame
    Dermal sensitization assayISO 10993-10Under conditions of the study, not a sensitizerUnder conditions of the study, not a sensitizerSame
    Systemic ToxicityISO 10993-11PassPassSame
    Chemical Permeation / Chemotherapy Testing (Breakthrough Detection Time)ASTM D6978-05Carmustine and Thiotepa have extremely low permeation times of less than 30 minutesCarmustine: 4.3 Minutes; Cisplatin: >240 Minutes; Cyclophosphamide/Cytoxan: >240 Minutes; Dacarbazine: >240 Minutes; Doxorubicin HCL: >240 Minutes; Etoposide /Toposar: >240 Minutes; Fluorouracil: >240 Minutes; Ifosfamide: >240 Minutes; Mitoxantrone: >240 Minutes; Paclitaxel: >240 Minutes; Thiotepa: 10.9 Minutes; Vincristine Sulfate: >240 MinutesSame (for Carmustine and Thiotepa, other drugs exceed the predicate's general criteria)

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each test mentioned (e.g., how many gloves were tested for pinholes, tensile strength, or chemotherapy permeation). It references standards like ASTM D6319-10 and ASTM D6978-05, which would have their own defined sample size requirements.

    The data provenance is from Malaysia, as the submitter, BRIGHTWAY HOLDINGS SDN. BHD., is located in Klang, Selangor Darul Ehsan, Malaysia. The data appears to be prospective as it involves direct testing of the device against established standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The acceptance criteria and performance are based on recognized industry standards (ASTM, ISO), meaning the "ground truth" is defined by these standards rather than expert consensus on specific cases. The testing was performed according to these established protocols.

    4. Adjudication method for the test set

    This information is not applicable/not provided as the evaluation is based on objective measurements against predefined standards (ASTM, ISO) rather than subjective assessment requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is an examination glove, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study related to AI assistance is irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is an examination glove, not an algorithm.

    7. The type of ground truth used

    The ground truth used is based on established industry standards and regulatory requirements. Specifically:

    • Physical properties (tensile strength, elongation, dimensions, freedom from pinholes, residual powder) are compared against specifications in ASTM D6319-10.
    • Biocompatibility (primary skin irritation, dermal sensitization, systemic toxicity) is evaluated against ISO 10993-10 and ISO 10993-11 requirements.
    • Resistance to chemotherapy drug permeation is tested according to ASTM D6978-05.
    • The overall criteria for patient examination gloves are also linked to FDA 21 CFR 880.6250.

    8. The sample size for the training set

    This is not applicable. The device is a physical product (gloves), not a machine learning model that requires a training set. The "training" in this context refers to the manufacturing process and quality control, not data training for an algorithm.

    9. How the ground truth for the training set was established

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1