Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K153280
    Manufacturer
    Date Cleared
    2016-06-02

    (203 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerGlide Pro™ Midline Catheter is inserted into a patient's vascular system for short-term use (<30 days) to sample blood or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The PowerGlide Pro™ Midline Catheter is suitable for use with power injectors.

    Device Description

    Bard Access Systems, Inc.'s PowerGlide Pro™ Midline Catheter is a sterile, single use device designed to provide access to the patient's vascular system. The device is intended for short term use (<30 days) to sample blood and administer fluids intravenously, and employs a placement technique similar to the cited predicate device. The subject device consists of an introducer needle with a passive safety mechanism, guidewire, and single lumen catheter rated for power injection. The PowerGlide Pro™ Midline Catheter features device housings and insertion mechanisms that are different from the predicate device. The PowerGlide Pro™ Midline Catheter is offered in 18, 20, and 22 gauge sizes. The 18 and 20 gauge devices are offered in 8 cm or 10 cm lengths. The 22 gauge device is offered in only an 8 cm length.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the C. R. Bard, Inc. PowerGlide Pro™ Midline Catheter. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving performance against specific acceptance criteria in the manner typically seen for novel diagnostic or AI devices.

    Therefore, many of the requested categories (e.g., sample size for test set, number of experts, adjudication method, MRMC study, training set details, ground truth types) are not applicable or derivable from this type of regulatory submission. The document emphasizes mechanical and biological performance tests that are standard for this class of medical device.

    Here's the information that can be extracted or deduced from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "The subject device met all predetermined acceptance criteria derived from the above listed tests and demonstrated substantially equivalent performance as compared to the cited predicate device." However, the exact numerical acceptance criteria for each test are not explicitly detailed. The tests performed are as follows:

    Performance Test CompletedISO Standard UtilizedReported Device Performance
    Burst Pressure Post Power InjectionISO 10555 - 1: 2013, Sterile, single use intravascular catheters, Part 1: General requirementsMet all predetermined acceptance criteria
    Burst Pressure Without Power InjectionISO 10555 - 1: 2013, Sterile, single use intravascular catheters, Part 1: General requirementsMet all predetermined acceptance criteria
    Effective Needle LengthISO 11070: 2014, Sterile, single use intravascular catheter introducerMet all predetermined acceptance criteria
    Effective Guidewire LengthCoronary and Cerebrovascular Guidewire Guidance, January 1995Met all predetermined acceptance criteria
    Needle to Hub Tensile StrengthNot explicitly statedMet all predetermined acceptance criteria
    Guidewire Bond Tensile StrengthNot explicitly statedMet all predetermined acceptance criteria
    CytotoxicityISO 10993-5: 2009Met all predetermined acceptance criteria
    Sensitization/IrritationISO 10993-10: 2010Met all predetermined acceptance criteria

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable for this type of device and regulatory submission. The document describes engineering and biological performance tests, not a clinical study on a patient test set.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. Ground truth as typically defined for diagnostic or AI algorithm studies is not relevant here. The "ground truth" for these performance tests would be the established engineering and biocompatibility standards.

    4. Adjudication Method for the Test Set

    Not applicable.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is not relevant for the performance testing of an intravascular catheter to demonstrate substantial equivalence based on mechanical and biological properties.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is an intravascular catheter, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the device's acceptable performance is based on International Organization for Standardization (ISO) standards for sterile, single-use intravascular catheters and biological evaluation of medical devices, along with internal predetermined acceptance criteria for mechanical and biocompatibility tests.

    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" as this is not an AI/ML device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1