Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K150612
    Date Cleared
    2015-11-16

    (251 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.

    Device Description

    Annuy Latex Patient Examination Glove (powdered and powder-free) is a glove made from natural rubber that covers the hand up to the wrist. It is cuffed and equally wearable on either hand, free from differentiation between the left hand and the right. It has three sizes, i.e., large, medium, and small. All variations share the same natural color, creamy white. The glove is non-sterile and is for single use only, to be discarded after each examination is performed. It acts as a barrier between the examiner and the subject being examined in order to prevent contamination between them.

    Annuy Nitrile Patient Examination Glove (powdered and powder-free) is a glove made from nitrile butadiene rubber Latex that covers the hand up to the wrist. It is cuffed and equally wearable on either hand, free from differentiation between the left hand and the right. It has three sizes, i.e. large, medium, and small. All variations share the same blue color. The glove is non-sterile and is for single use only, to be discarded after each examination is performed. It acts as a barrier between the examiner and the subject being examined in order to prevent contamination between them.

    AI/ML Overview

    The provided documents detail the acceptance criteria and performance study for four types of examination gloves: Powdered Latex, Powder-free Latex, Powdered Nitrile, and Powder-free Nitrile.

    1. Table of Acceptance Criteria and Reported Device Performance

    Device: Annuy Latex Patient Examination Gloves (Powdered)

    CharacteristicsStandardAcceptance Criteria (Standard)Reported Device Performance
    DimensionASTM D3578-10
    Palm widthSmall: 80±10 mmMedium: 95±10 mm
    Overall LengthSmall: 220 min mmMedium: 230 min mm
    Finger Thickness0.08 min mm0.09~0.12 mm (across sizes)
    Palm Thickness0.08 min mm0.09~0.12 mm (across sizes)
    Physical PropertiesASTM D3578-10
    Before Aging
    - Tensile Strength18 Mpa min20.3~35.1 Mpa
    - Ultimate Elongation650% min961~1233%
    - Stress at 500% Elongation5.5 Mpa max4.6~5.4 Mpa
    After aging at 70±2 °C, 166±2 hrs
    - Tensile Strength14 Mpa min16.7~24.0 Mpa
    - Ultimate Elongation500% min748~1021%
    Freedom from HolesASTM D5151-11Inspection Level I, AQL 2.5Passed Standard Acceptance Criteria
    Powder AmountASTM D6124-11≤ 10 mg/dm²2.34 mg/dm²
    Protein ContentASTM D5712-10≤ 200 µg/dm² Aqueous extractable protein72.1 mcg/dm²
    BiocompatibilityISO 10993-10:2010-08-01
    Primary Skin IrritationNot an irritantNot an irritant
    Dermal sensitizationNot a sensitizerNot a sensitizer

    Device: Annuy Latex Patient Examination Gloves (Powder-free)

    CharacteristicsStandardAcceptance Criteria (Standard)Reported Device Performance
    DimensionASTM D3578-10
    Palm widthSmall: 80±10 mmMedium: 95±10 mm
    Overall LengthSmall: 220 min mmMedium: 230 min mm
    Finger Thickness0.08 min mm0.09~0.12 mm (across sizes)
    Palm Thickness0.08 min mm0.08~0.12 mm (across sizes)
    Physical PropertiesASTM D3578-10
    Before Aging
    - Tensile Strength18 Mpa min18.4~32.3 Mpa
    - Ultimate Elongation650% min955~1148%
    - Stress at 500% Elongation5.5 Mpa max4.8~5.4 Mpa
    After aging at 70±2 °C, 166±2 hrs
    - Tensile Strength14 Mpa min18.7~24.5 Mpa
    - Ultimate Elongation500% min797~976%
    Freedom from HolesASTM D5151-11Inspection Level I, AQL 2.5Passed Standard Acceptance Criteria
    Powder residueASTM D6124-11≤ 2 mg/glove0.68 mg/glove
    Protein ContentASTM D5712-10≤ 200 µg/dm² Aqueous extractable protein70.8 mcg/dm²
    BiocompatibilityISO 10993-10:2010-08-01
    Primary Skin IrritationNot an irritantNot an irritant
    Dermal sensitizationNot a sensitizerNot a sensitizer

    Device: Annuy Nitrile Patient Examination Gloves (Powdered)

    CharacteristicsStandardAcceptance Criteria (Standard)Reported Device Performance
    DimensionASTM D6319-10
    Palm widthSmall: 80±10 mmMedium: 95±10 mm
    Overall LengthSmall: 220 min mmMedium: 230 min mm
    Finger Thickness0.05 min mm0.08 mm (across sizes)
    Palm Thickness0.05 min mm0.08 mm (across sizes)
    Physical PropertiesASTM D6319-10
    Before Aging
    - Tensile Strength14 Mpa min17.5~31.1 Mpa
    - Ultimate Elongation500% min761~948%
    After aging at 70±2 °C, 166±2 hrs
    - Tensile Strength14 Mpa min15.7~18.5 Mpa
    - Ultimate Elongation400% min421~839%
    Freedom from HolesASTM D5151-11Inspection Level G-1, AQL 2.5Passed Standard Acceptance Criteria
    Powder AmountASTM D6124-11≤ 10 mg/dm²1.86 mg/dm²
    BiocompatibilityISO 10993-10:2010-08-01
    Primary Skin IrritationNot an irritantNot an irritant
    Dermal sensitizationNot a sensitizerNot a sensitizer

    Device: Annuy Nitrile Patient Examination Gloves (Powder-free)

    CharacteristicsStandardAcceptance Criteria (Standard)Reported Device Performance
    DimensionASTM D6319-10
    Palm widthSmall: 80±10 mmMedium: 95±10 mm
    Overall LengthSmall: 220 min mmMedium: 230 min mm
    Finger Thickness0.05 min mm0.08 mm (across sizes)
    Palm Thickness0.05 min mm0.08 mm (across sizes)
    Physical PropertiesASTM D6319-10
    Before Aging
    - Tensile Strength14 Mpa min18.8~29.5 Mpa
    - Ultimate Elongation500% min745~945%
    After aging at 70±2 °C, 166±2 hrs
    - Tensile Strength14 Mpa min16.2~21.0 Mpa
    - Ultimate Elongation400% min445~773%
    Freedom from HolesASTM D5151-11Inspection Level G-1, AQL 2.5Passed Standard Acceptance Criteria
    Powder residueASTM D6124-11≤ 2 mg/glove0.72 mg/glove
    BiocompatibilityISO 10993-10:2010-08-01
    Primary Skin IrritationNot an irritantNot an irritant
    Dermal sensitizationNot a sensitizerNot a sensitizer

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact "sample size used for the test set" (i.e., the number of individual gloves tested for each characteristic). However, it references specific ASTM and ISO standards that govern testing procedures for these types of gloves. These standards typically define sampling plans, AQL (Acceptable Quality Level) for certain tests (like freedom from holes), or specify the number of specimens for physical property testing. For example, for "Freedom from Holes," AQL values are given (2.5 for Inspection Level I for latex, 2.5 for Inspection Level G-1 for nitrile), implying a sampling plan was followed.

    The data provenance is Anhui Anyu Latex Products Co., Ltd in Bengbu, Anhui, China, as they are the manufacturer and submitter of the 510(k) premarket notification. The data appears to be from retrospective testing conducted by the manufacturer to demonstrate compliance with the referenced standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This section is not applicable as the described tests are laboratory-based, objective performance measures (e.g., dimensions, physical properties, chemical residue, biocompatibility) of physical devices (gloves) against established engineering and safety standards. They do not involve interpretation of complex data (like medical images) by human experts to establish ground truth in the way medical AI or diagnostic devices would.

    4. Adjudication Method for the Test Set

    This is not applicable for the same reason as point 3. The assessment criteria are derived from objective measurements against specified standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document explicitly states: "Clinical data was not included in this submission." The evaluation focused entirely on technical and physical performance metrics of the gloves against recognized standards.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a patient examination glove, a physical product, not an algorithm or AI system. Therefore, the concept of a "standalone" algorithmic performance is irrelevant. The tests performed are standalone in the sense that they assess the intrinsic properties of the glove material and manufacturing.

    7. The Type of Ground Truth Used

    The ground truth used for performance evaluation is based on established industry standards and regulatory requirements. Specifically, these include:

    • ASTM (American Society for Testing and Materials) Standards:
      • ASTM D3578-10 (Standard Specification for Rubber Examination Gloves)
      • ASTM D6319-10 (Standard Specification for Nitrile Examination Gloves for Medical Application)
      • ASTM D5151-11 (Standard Test Method for Detection of Holes in Medical Gloves)
      • ASTM D6124-11 (Standard Test Method for Residual Powder on Medical Gloves)
      • ASTM D5712-10 (Standard Test Method for the Analysis of Aqueous Extractable Protein in Natural Rubber Products Using the Modified Lowry Method)
    • ISO (International Organization for Standardization) Standard:
      • ISO 10993-10 (Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization)

    These standards define the acceptable ranges or thresholds for various physical, chemical, and biological properties of examination gloves.

    8. The Sample Size for the Training Set

    This is not applicable. The device is a physical product (medical glove), not an AI/ML model that requires a training set. The manufacturing process of the gloves would involve quality control and process monitoring, but this is distinct from training a software algorithm.

    9. How the Ground Truth for the Training Set was Established

    This is not applicable as there is no training set for a physical medical device like an examination glove.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1