K Number
K150612

Validate with FDA (Live)

Date Cleared
2015-11-16

(251 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.

Device Description

Annuy Latex Patient Examination Glove (powdered and powder-free) is a glove made from natural rubber that covers the hand up to the wrist. It is cuffed and equally wearable on either hand, free from differentiation between the left hand and the right. It has three sizes, i.e., large, medium, and small. All variations share the same natural color, creamy white. The glove is non-sterile and is for single use only, to be discarded after each examination is performed. It acts as a barrier between the examiner and the subject being examined in order to prevent contamination between them.

Annuy Nitrile Patient Examination Glove (powdered and powder-free) is a glove made from nitrile butadiene rubber Latex that covers the hand up to the wrist. It is cuffed and equally wearable on either hand, free from differentiation between the left hand and the right. It has three sizes, i.e. large, medium, and small. All variations share the same blue color. The glove is non-sterile and is for single use only, to be discarded after each examination is performed. It acts as a barrier between the examiner and the subject being examined in order to prevent contamination between them.

AI/ML Overview

The provided documents detail the acceptance criteria and performance study for four types of examination gloves: Powdered Latex, Powder-free Latex, Powdered Nitrile, and Powder-free Nitrile.

1. Table of Acceptance Criteria and Reported Device Performance

Device: Annuy Latex Patient Examination Gloves (Powdered)

CharacteristicsStandardAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM D3578-10
Palm widthSmall: 80±10 mmMedium: 95±10 mm
Overall LengthSmall: 220 min mmMedium: 230 min mm
Finger Thickness0.08 min mm0.09~0.12 mm (across sizes)
Palm Thickness0.08 min mm0.09~0.12 mm (across sizes)
Physical PropertiesASTM D3578-10
Before Aging
- Tensile Strength18 Mpa min20.3~35.1 Mpa
- Ultimate Elongation650% min961~1233%
- Stress at 500% Elongation5.5 Mpa max4.6~5.4 Mpa
After aging at 70±2 °C, 166±2 hrs
- Tensile Strength14 Mpa min16.7~24.0 Mpa
- Ultimate Elongation500% min748~1021%
Freedom from HolesASTM D5151-11Inspection Level I, AQL 2.5Passed Standard Acceptance Criteria
Powder AmountASTM D6124-11≤ 10 mg/dm²2.34 mg/dm²
Protein ContentASTM D5712-10≤ 200 µg/dm² Aqueous extractable protein72.1 mcg/dm²
BiocompatibilityISO 10993-10:2010-08-01
Primary Skin IrritationNot an irritantNot an irritant
Dermal sensitizationNot a sensitizerNot a sensitizer

Device: Annuy Latex Patient Examination Gloves (Powder-free)

CharacteristicsStandardAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM D3578-10
Palm widthSmall: 80±10 mmMedium: 95±10 mm
Overall LengthSmall: 220 min mmMedium: 230 min mm
Finger Thickness0.08 min mm0.09~0.12 mm (across sizes)
Palm Thickness0.08 min mm0.08~0.12 mm (across sizes)
Physical PropertiesASTM D3578-10
Before Aging
- Tensile Strength18 Mpa min18.4~32.3 Mpa
- Ultimate Elongation650% min955~1148%
- Stress at 500% Elongation5.5 Mpa max4.8~5.4 Mpa
After aging at 70±2 °C, 166±2 hrs
- Tensile Strength14 Mpa min18.7~24.5 Mpa
- Ultimate Elongation500% min797~976%
Freedom from HolesASTM D5151-11Inspection Level I, AQL 2.5Passed Standard Acceptance Criteria
Powder residueASTM D6124-11≤ 2 mg/glove0.68 mg/glove
Protein ContentASTM D5712-10≤ 200 µg/dm² Aqueous extractable protein70.8 mcg/dm²
BiocompatibilityISO 10993-10:2010-08-01
Primary Skin IrritationNot an irritantNot an irritant
Dermal sensitizationNot a sensitizerNot a sensitizer

Device: Annuy Nitrile Patient Examination Gloves (Powdered)

CharacteristicsStandardAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM D6319-10
Palm widthSmall: 80±10 mmMedium: 95±10 mm
Overall LengthSmall: 220 min mmMedium: 230 min mm
Finger Thickness0.05 min mm0.08 mm (across sizes)
Palm Thickness0.05 min mm0.08 mm (across sizes)
Physical PropertiesASTM D6319-10
Before Aging
- Tensile Strength14 Mpa min17.5~31.1 Mpa
- Ultimate Elongation500% min761~948%
After aging at 70±2 °C, 166±2 hrs
- Tensile Strength14 Mpa min15.7~18.5 Mpa
- Ultimate Elongation400% min421~839%
Freedom from HolesASTM D5151-11Inspection Level G-1, AQL 2.5Passed Standard Acceptance Criteria
Powder AmountASTM D6124-11≤ 10 mg/dm²1.86 mg/dm²
BiocompatibilityISO 10993-10:2010-08-01
Primary Skin IrritationNot an irritantNot an irritant
Dermal sensitizationNot a sensitizerNot a sensitizer

Device: Annuy Nitrile Patient Examination Gloves (Powder-free)

CharacteristicsStandardAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM D6319-10
Palm widthSmall: 80±10 mmMedium: 95±10 mm
Overall LengthSmall: 220 min mmMedium: 230 min mm
Finger Thickness0.05 min mm0.08 mm (across sizes)
Palm Thickness0.05 min mm0.08 mm (across sizes)
Physical PropertiesASTM D6319-10
Before Aging
- Tensile Strength14 Mpa min18.8~29.5 Mpa
- Ultimate Elongation500% min745~945%
After aging at 70±2 °C, 166±2 hrs
- Tensile Strength14 Mpa min16.2~21.0 Mpa
- Ultimate Elongation400% min445~773%
Freedom from HolesASTM D5151-11Inspection Level G-1, AQL 2.5Passed Standard Acceptance Criteria
Powder residueASTM D6124-11≤ 2 mg/glove0.72 mg/glove
BiocompatibilityISO 10993-10:2010-08-01
Primary Skin IrritationNot an irritantNot an irritant
Dermal sensitizationNot a sensitizerNot a sensitizer

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the exact "sample size used for the test set" (i.e., the number of individual gloves tested for each characteristic). However, it references specific ASTM and ISO standards that govern testing procedures for these types of gloves. These standards typically define sampling plans, AQL (Acceptable Quality Level) for certain tests (like freedom from holes), or specify the number of specimens for physical property testing. For example, for "Freedom from Holes," AQL values are given (2.5 for Inspection Level I for latex, 2.5 for Inspection Level G-1 for nitrile), implying a sampling plan was followed.

The data provenance is Anhui Anyu Latex Products Co., Ltd in Bengbu, Anhui, China, as they are the manufacturer and submitter of the 510(k) premarket notification. The data appears to be from retrospective testing conducted by the manufacturer to demonstrate compliance with the referenced standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This section is not applicable as the described tests are laboratory-based, objective performance measures (e.g., dimensions, physical properties, chemical residue, biocompatibility) of physical devices (gloves) against established engineering and safety standards. They do not involve interpretation of complex data (like medical images) by human experts to establish ground truth in the way medical AI or diagnostic devices would.

4. Adjudication Method for the Test Set

This is not applicable for the same reason as point 3. The assessment criteria are derived from objective measurements against specified standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document explicitly states: "Clinical data was not included in this submission." The evaluation focused entirely on technical and physical performance metrics of the gloves against recognized standards.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a patient examination glove, a physical product, not an algorithm or AI system. Therefore, the concept of a "standalone" algorithmic performance is irrelevant. The tests performed are standalone in the sense that they assess the intrinsic properties of the glove material and manufacturing.

7. The Type of Ground Truth Used

The ground truth used for performance evaluation is based on established industry standards and regulatory requirements. Specifically, these include:

  • ASTM (American Society for Testing and Materials) Standards:
    • ASTM D3578-10 (Standard Specification for Rubber Examination Gloves)
    • ASTM D6319-10 (Standard Specification for Nitrile Examination Gloves for Medical Application)
    • ASTM D5151-11 (Standard Test Method for Detection of Holes in Medical Gloves)
    • ASTM D6124-11 (Standard Test Method for Residual Powder on Medical Gloves)
    • ASTM D5712-10 (Standard Test Method for the Analysis of Aqueous Extractable Protein in Natural Rubber Products Using the Modified Lowry Method)
  • ISO (International Organization for Standardization) Standard:
    • ISO 10993-10 (Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization)

These standards define the acceptable ranges or thresholds for various physical, chemical, and biological properties of examination gloves.

8. The Sample Size for the Training Set

This is not applicable. The device is a physical product (medical glove), not an AI/ML model that requires a training set. The manufacturing process of the gloves would involve quality control and process monitoring, but this is distinct from training a software algorithm.

9. How the Ground Truth for the Training Set was Established

This is not applicable as there is no training set for a physical medical device like an examination glove.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, represented by three overlapping profiles, suggesting a sense of community and support. The profiles are arranged in a way that creates a sense of depth and movement. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 16, 2015

Anhui Anyu Latex Products Co., Ltd. c/o Doris Dong Shanghai CV Technology Co., Ltd. Room 1706. No. 128 Songle Rd., Songjiang Area Shanghai 201600 CHINA

Re: K150612 Trade/Device Name: "Annuy Latex Patient Examination Glove (Powdered and Powder-free)" & "Annuy Nitrile Patient Examination Glove (Powdered and Powder-free)" Regulation Number: 21 CFR 880.6250 Regulation Name: Patient examination glove Regulatory Class: Class I Product Code: LYY, LYZ Dated: October 4, 2015 Received: October 16, 2015

Dear Ms. Dong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K150612

Device Name

Annuy Latex Patient Examination Glove (powdered and powder-free); Annuy Nitrile Patient Examination Glove (powdered and powder-free)

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a logo with a circular design. The top half of the circle is blue, while the bottom half is green. There are several curved lines in the green section that resemble blades of grass or leaves. The logo is set against a light blue background.

Section 5 510(k) Summary

[As required by 21 CFR 807.92]

1. Submission Information:

510(k) Number:K150612
Date:Sept 30th, 2015
Type of 510(k) Submission:Traditional
Basis for 510(k) Submission:New device
Submitter/Manufacturer:Anhui Anyu Latex Products Co., LtdEast Qinji Road & North Xinghua Road, Bengbu, Anhui, 233010 China
Contactor:Doris Dong (Consultant)Shanghai CV Technology Co., Ltd.Room 1706, No. 128 Songle Rd., Songjiang Area, Shanghai, 201600 ChinaE-mail: doris_d@126.comTel: 86 21-31261348 / Fax: 86 21-37824346

2. Device Description:

Proprietary Name:Annuy Latex Patient Examination Glove (powdered and powder-free);Annuy Nitrile Patient Examination Glove (powdered and powder-free)
Common Name:Latex Examination Gloves (powdered and powder free);Nitrile Examination Gloves (powdered and powder free)
Classification Name:Glove, patient examination, latex;Glove, patient examination, poly
Product Code:LYY & LZA
Device Class:I
Regulation Number:21 CFR 880.6250
Review Panel:General Hospital
Indications for use:A patient examination glove is a disposable device intended for medical purposes thatis worn on the examiner's hand or fingers to prevent contamination between patient andexaminer

Device Description:

Annuy Latex Patient Examination Glove (powdered and powder-free) is a glove made from natural rubber that covers the hand up to the wrist. It is cuffed and equally wearable on either hand, free from differentiation between the left hand and the right. It has three sizes, i.e., large, medium, and small. All variations share the same natural color, creamy white. The glove is non-sterile and is for single use only, to be discarded after each examination is performed. It acts as a barrier between the examiner and the subject being examined in order to prevent contamination between them.

Annuy Nitrile Patient Examination Glove (powdered and powder-free) is a glove made from nitrile butadiene rubber Latex that covers the hand up to the wrist. It is cuffed and equally wearable on either hand, free from differentiation between the left hand and the right. It has three sizes, i.e. large, medium, and small. All variations share the same blue color. The glove is non-sterile and is for single use only, to be discarded after each examination is performed. It acts as a barrier between the examiner and the subject being examined in order to prevent contamination between them.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a logo with a circular shape. The top half of the circle is blue, and the bottom half is green. There are several curved lines in the center of the circle, resembling a plant or a bird's wing. The logo is simple and clean, with a focus on natural elements.

3. Predicate Device Identification

510(k) Number: K072145

Product Name:

Latex Examination Gloves (Powdered and Powder-Free)

Nitrile Examination Gloves (Powdered and Powder-Free)

Manufacturer: Zhanjiang Jiali Glove Products Co., LTD

4. Performance Test Data Summary

• The Powdered Latex Patient Examination Gloves

CharacteristicsStandardTest Data
DimensionASTM D3578-10
Palm widthSize(mm)(mm)
Small80±1080~84
Medium95±1093~97
Large111±10107~109
Overall LengthSize(mm)(mm)
Small220 min240~246
Medium230 min241~248
Large230 min240~249
Finger ThicknessSize(mm)(mm)
Small0.08 min0.09~0.11
Medium0.08 min0.09~0.11
Large0.08 min0.09~0.12
Palm ThicknessSize(mm)(mm)
Small0.08 min0.10~0.12
Medium0.08 min0.10~0.12
Large0.08 min0.09~0.12
Physical PropertiesASTM D3578-10
Before AgingAfter aging at 70±2 °C166±2 hrsBefore AgingAfter aging at 70±2 °C166±2 hrs
- Tensile Strength18Mpa min14Mpa min20.3~35.1Mpa16.7~24.0Mpa
- Ultimate Elongation650% min500% min961~1233%748~1021%
- Stress at 500% Elongation5.5Mpa max--4.6~5.4Mpa--
Freedom from HolesASTM D5151-11Passed Standard Acceptance CriteriaHoles at Inspection Level I, AQL 2.5
Powder AmountASTM D6124-11, ≤ 10mg/dm²2.34 mg/dm², Meets ASTM D6124-11Results generated values below 10mg/dm² ofresidual powder
Protein ContentASTM D5712-10, ≤200µg/dm² Aqueousextractable proteinAqueous extractable protein: 72.1 mcg /dm²,Meets ASTM D5712-10 Results generated

values below 200µg/dm² of aqueous extractable

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows a logo with a circular design. The top half of the circle is light blue, while the bottom half is green. Within the green portion, there are several curved lines that resemble leaves or blades of grass. The overall design is simple and clean, with a focus on natural elements.

protein content
BiocompatibilityPrimary Skin Irritation in rabbits ISO10993-10:2010-08-01Under the condition of the test, not an irritant.
Dermal sensitization in the guinea pig ISO10993-10:2010-08-01Under the condition of the test, not a sensitizer.

• The Powder-free Latex Patient Examination Gloves

CharacteristicsStandardTest Data
DimensionASTM D3578-10
Palm widthSize(mm)(mm)
Small80±1080~84
Medium95±1093~97
Large111±10107~109
Overall LengthSize(mm)(mm)
Small220 min240~246
Medium230 min241~248
Large230 min241~248
Finger ThicknessSize(mm)(mm)
Small0.08 min0.09~0.12
Medium0.08 min0.09~0.11
Large0.08 min0.09~0.12
Palm ThicknessSize(mm)(mm)
Small0.08 min0.10~0.12
Medium0.08 min0.10~0.12
Large0.08 min0.08~0.12
Physical PropertiesASTM D3578-10
Before AgingAfter aging at 70±2 °C166±2 hrsBefore AgingAfter aging at 70±2 °C166±2 hrs
- Tensile Strength18Mpa min14Mpa min18.4~32.3Mpa18.7~24.5Mpa
- Ultimate Elongation650% min500% min955~1148%797~976%
- Stress at 500%Elongation5.5Mpa max--4.8~5.4Mpa--
Freedom from HolesASTM D5151-11Passed Standard Acceptance CriteriaHoles at Inspection Level I, AQL 2.5
Powder residueASTM D6124-11, ≤2mg/glove0.68 mg/glove, Meets ASTM D6124-11Results generated values below 2mg/glove ofresidual powder
Protein ContentASTM D5712-10, ≤200µg/dm2 Aqueousextractable proteinAqueous extractable protein: 70.8 mcg/dm2,Meets ASTM D5712-10, Results generatedvalues below 200µg/dm² of Aqueous extractableprotein content
BiocompatibilityPrimary Skin Irritation in rabbits ISOUnder the condition of the test, not an irritant.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows a logo with a circular design. The top half of the circle is blue, while the bottom half is green. Within the green portion, there are several curved lines that resemble blades of grass or leaves. The logo has a clean and modern look.

10993-10:2010-08-01
Dermal sensitization in the guinea pig ISO10993-10:2010-08-01Under the condition of the test, not a sensitizer.

• The Powdered Nitrile Patient Examination Gloves

• The Powdered Nitrile Patient Examination Gloves
CharacteristicsStandardTest Data
DimensionASTM D6319-10
Palm widthSize(mm)(mm)
Small80±1081~85
Medium95±1095~99
Large110±10110~114
Overall LengthSize(mm)(mm)
Small220 min251~255
Medium230 min250~262
Large230 min259~264
Finger ThicknessSize(mm)(mm)
Small0.05 min0.08
Medium0.05 min0.08
Large0.05 min0.08
Palm ThicknessSize(mm)(mm)
Small0.05 min0.08
Medium0.05 min0.08
Large0.05 min0.08
Physical PropertiesASTM D6319-10
Before AgingAfter aging at 70±2 °C166±2 hrsBefore AgingAfter aging at 70±2 °C166±2 hrs
- Tensile Strength14Mpa min14Mpa min17.5~31.1Mpa15.7~18.5Mpa
- Ultimate Elongation500% min400% min761~948%421~839%
Freedom from HolesASTM D5151-11Passed Standard Acceptance Criteria
Holes at Inspection Level G-1, AQL 2.5
Powder AmountASTM D6124-11, ≤ 10mg/dm²1.86 mg/dm², Meets ASTM D6124-11Results generated values below 10mg/dm² ofresidual powder
BiocompatibilityPrimary Skin Irritation in rabbits ISO10993-10:2010-08-01Under the condition of the test, not an irritant.
Dermal sensitization in the guinea pig ISO10993-10:2010-08-01Under the condition of the test, not a sensitizer.

• The Powder-free Nitrile Patient Examination Gloves

CharacteristicsStandardTest Data
DimensionASTM D6319-10
Palm widthSize (mm)(mm)

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows a logo with a circular shape. The top half of the circle is blue, and the bottom half is green. Inside the circle, there are several curved lines that resemble leaves or blades of grass. The lines are also green and are arranged in a fan-like pattern.

Small80±1081~85
Medium95±1094~99
Large110±10110~114
Overall LengthSize(mm)(mm)
Small220 min252~257
Medium230 min255~260
Large230 min257~263
Finger ThicknessSize(mm)(mm)
Small0.05 min0.08
Medium0.05 min0.08
Large0.05 min0.08
Palm ThicknessSize(mm)(mm)
Small0.05 min0.08
Medium0.05 min0.08
Large0.05 min0.08
Physical PropertiesASTM D6319-10
Before AgingAfter aging at 70±2 °C166±2 hrsBefore AgingAfter aging at 70±2 °C166±2 hrs
- Tensile Strength14Mpa min14Mpa min18.8~29.5Mpa16.2~21.0Mpa
- Ultimate Elongation500% min400% min745~945%445~773%
Freedom from HolesASTM D5151-11Passed Standard Acceptance CriteriaHoles at Inspection Level G-1, AQL 2.5
Powder residueASTM D6124-11, ≤2mg/glove0.72 mg/glove, Meets ASTM D6124-11Results generated values below 2mg/glove ofresidual powder
BiocompatibilityPrimary Skin Irritation in rabbits ISO10993-10:2010-08-01Dermal sensitization in the guinea pig ISO10993-10:2010-08-01Under the condition of the test, not an irritant.Under the condition of the test, not a sensitizer.

5. Substantial Equivalent Based on Assessment of Non-Clinical Performance Data:

Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  1. ASTM D3578-10, Standard Specification for Rubber Examination Gloves.

  2. ASTM D5151-11, Standard Test Method for Detection of Holes in Medical Gloves.

  3. ASTM D6124-11, Standard Test Method for Residual Powder on Medical Gloves.

  4. ASTM D5712-10, Standard Test Method for the Analysis of Aqueous Extractable Protein in Natural Rs

Products Using the Modified Lowry Method.

  1. ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

  2. ISO 10993-10, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.

6. Substantial Equivalent Based on Assessment of Clinical Performance Data:

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows a logo with a circular design. The logo features a blue circle with a green, stylized image of a plant or bird inside. The green image has several curved lines that resemble leaves or feathers. The overall design is simple and modern.

Clinical data was not including in this submission.

7. Substantially Equivalent Comparison Conclusion

• For Powdered Latex Patient Examination Gloves

ParametersNew DevicePredicate Device
1510(k) NumberK150612K072145
2Marketingclearance date--Feb 26th, 2008
3Device NameAnnuy Powdered Latex Patient ExaminationGlovePowdered Latex Patient Examination Glove
4ManufacturerAnhui Anyu Latex Products Co., LtdZhanjiang Jiali Glove Products Co., LTD
5Product CodeLYYLYY
6Regulation No.21 CFR 880.625021 CFR 880.6250
7ClassII
8Intended useA patient examination glove is a disposabledevice intended for medical purposes that isworn on the examiner's hand or finger toprevent contamination between patient andexaminer.A patient examination glove is a disposabledevice intended for medical purposes that isworn on the examiner's hand or finger toprevent contamination between patient andexaminer.
9MaterialNatural Rubber LatexNatural Rubber Latex
10DesignSingle useNon-sterilePowderedAmbidextrousCuffedSingle useNon-sterilePowderedAmbidextrousCuffed
11DescriptionA device covering the hand and wrist area.Gloves have separate sheaths or openings foreach finger and the thumb.A device covering the hand and wrist area.Gloves have separate sheaths or openingsfor each finger and the thumb.
12SizeProvided in sizes small, medium and largeProvided in sizes small, medium and large
13SpecificationsPowdered Latex Examination Gloves MeetASTM D3578-10Powdered Latex Examination Gloves MeetASTM D3578-05
14Dimensions- Overall LengthMeets ASTM D3578-10Small: 220mm minMedium: 230mm minLarge: 230mm minMeets ASTM D3578-05Small: 220mm minMedium: 230mm minLarge: 230mm min
15Dimensions- Palm WidthMeets ASTM D3578-10Small: 80±10mmMedium: 95±10mmLarge: 111±10mmMeets ASTM D3578-05Small: 80±10mmMedium: 95±10mmLarge: 111±10mm
16Dimensions- Finger ThicknessMeets ASTM D3578-100.08mm minMeets ASTM D3578-050.08mm min
Dimensions- Palm ThicknessMeets ASTM D3578-100.08mm minMeets ASTM D3578-050.08mm min
17Physical PropertiesMeets ASTM D3578-10Meets ASTM D3578-05
Before Aging:Before Aging:
- Tensile Strength: 18Mpa min- Tensile Strength: 18Mpa min
- Ultimate Elongation: 650% min- Ultimate Elongation: 650% min
- Stress at 500% Elongation: 5.5Mpa max- Stress at 500% Elongation: 5.5Mpa max
Meets ASTM D3578-10Meets ASTM D3578-05
After aging at 70±2 °C 166±2 hrs:After aging at 70±2 °C 166±2 hrs:
- Tensile Strength: 14Mpa min- Tensile Strength: 14Mpa min
- Ultimate Elongation: 500% min- Ultimate Elongation: 500% min
18Freedom fromMeets ASTM D3578-10 and ASTM D5151-11Meets ASTM D3578-05 and ASTM
HolesHoles at Inspection Level I, AQL 2.5D5151-06
19Powder AmountMeets ASTM D6124-11Meets ASTM D6124-06
Below 10mg/dm² of residual powderBelow 10mg/dm² of residual powder
20Protein ContentMeets ASTM D5712-10,Meets ASTM D5712-05
Below 200µg/dm² of aqueous extractableBelow 200µg/dm² of aqueous extractable
protein contentprotein content
21Absorbable donningCornstarch,Cornstarch,
or dusting powderPowdered with absorbable dusting powder,Powdered with absorbable dusting powder,
U.S.P.U.S.P.
22BiocompatibilityMeets ISO 10993-10:2010;Meets ISO 10993-10:2006;
Skin Irritation: Under the condition of theSkin Irritation: Under the condition of the
test, not an irritant.test, not an irritant.
Skin Sensitization:Under the condition of theSkin Sensitization:Under the condition of
test, not a sensitizer.the test, not a sensitizer.
23ColorCreamy white, no colorantSame
24Labeling FeaturesInclude the required labeling:Include the required labeling:
• Examination Gloves• Examination Gloves
• Non-sterile• Single Use Only
• Single Use Only• Powdered
• Powdered• Size, Serial No., Manufacturing Date and
• AmbidextrousQuantity
• Natural rubber latex• Natural rubber latex
• Device color: Creamy white• Protein Label Claims
• Allergy warning• Allergy warning
• Contents, Size, Lot No. and Manufacturing• Manufacturer Name and Address
Date
• Storage conditions
• Manufacturer Name and Address

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows a logo with a circular design. The top half of the circle is blue, while the bottom half is green. Within the green portion, there are several curved lines that resemble blades of grass or stylized leaves. The overall design is simple and modern, suggesting a connection to nature or the environment.

• For Powder-free Latex Patient Examination Gloves

ParametersNew DevicePredicate Device
1510(k) NumberK150612K072145
2Marketingclearance date--Feb 26th, 2008
3Device NameAnnuy Powder free Latex Patient Examination GlovePowder free Latex Patient Examination Glove
4ManufacturerAnhui Anyu Latex Products Co., LtdZhanjiang Jiali Glove Products Co., LTD
5Product CodeLYYLYY
6Regulation No.21 CFR 880.625021 CFR 880.6250
7ClassII
8Intended useA patient examination glove is a disposabledevice intended for medical purposes that isworn on the examiner's hand or finger toprevent contamination between patient andexaminer.A patient examination glove is a disposabledevice intended for medical purposes that isworn on the examiner's hand or finger toprevent contamination between patient andexaminer.
9MaterialNatural Rubber LatexNatural Rubber Latex
10DesignSingle useNon-sterilePowder freeAmbidextrousCuffedSingle useNon-sterilePowder freeAmbidextrousCuffed
11DescriptionA device covering the hand and wrist area.Gloves have separate sheaths or openings foreach finger and the thumb.A device covering the hand and wrist area.Gloves have separate sheaths or openingsfor each finger and the thumb.
12SizeProvided in sizes small, medium and largeProvided in sizes small, medium and large
13SpecificationsPowdered Latex Examination Gloves MeetASTM D3578-10Powdered Latex Examination Gloves MeetASTM D3578-05
14Dimensions- Overall LengthMeets ASTM D3578-10Small: 220mm minMedium: 230mm minLarge: 230mm minMeets ASTM D3578-05Small: 220mm minMedium: 230mm minLarge: 230mm min
15Dimensions- Palm WidthMeets ASTM D3578-10Small: 80±10mmMedium: 95±10mmLarge: 111±10mmMeets ASTM D3578-05Small: 80±10mmMedium: 95±10mmLarge: 111±10mm
16Dimensions- Finger ThicknessMeets ASTM D3578-100.08mm minMeets ASTM D3578-050.08mm min
Dimensions- Palm ThicknessMeets ASTM D3578-100.08mm minMeets ASTM D3578-050.08mm min
17Physical PropertiesMeets ASTM D3578-10Before Aging:- Tensile Strength: 18Mpa min- Ultimate Elongation: 650% min- Stress at 500% Elongation: 5.5Mpa maxMeets ASTM D3578-10After aging at 70±2 °C 166±2 hrs:- Tensile Strength: 14Mpa min- Ultimate Elongation: 500% minMeets ASTM D3578-05Before Aging:- Tensile Strength: 18Mpa min- Ultimate Elongation: 650% min- Stress at 500% Elongation: 5.5Mpa maxMeets ASTM D3578-05After aging at 70±2 °C 166±2 hrs:- Tensile Strength: 14Mpa min- Ultimate Elongation: 500% min
18Freedom fromHolesMeets ASTM D3578-10 and ASTM D5151-11Holes at Inspection Level I, AQL 2.5Meets ASTM D3578-05 and ASTMD5151-06
ANHUI YU LATEX PRODUCTS CO., LTDEAST QINJI ROAD & NORTH XINGHUA ROAD, BENGBU, ANHUI, 233010 CHINA
19Powder residueMeets ASTM D6124-11Below 2mg/glove of residual powderMeets ASTM D6124-06Below 2mg/glove of residual powder
20Protein ContentMeets ASTM D5712-10,Below 200µg/dm² of aqueous soluble proteincontentMeets ASTM D5712-05,Below 200µg/dm² of aqueous solubleprotein content
21Polymer CoatingIt is coated with a polyurethane based coatingon the inner side to facilitate donning.Polymer name: polyurethane & polyacrylicacidIt is coated with a polyurethane basedcoating on the inner side to facilitatedonning.Polymer name: polyurethane & polyacrylicacid
22BiocompatibilityMeets ISO 10993-10:2010;Skin Irritation: Under the condition of thetest, not an irritant.Skin Sensitization: Under the condition of thetest, not a sensitizer.Meets ISO 10993-10:2006;Skin Irritation: Under the condition of thetest, not an irritant.Skin Sensitization: Under the condition ofthe test, not a sensitizer.
23ColorCreamy white, no colorantCreamy white, no colorant
24Labeling FeaturesInclude the required labeling:• Examination Gloves• Non-sterile• Disposable• Powder-free• Ambidextrous• Natural rubber latex• Device color: Creamy white• Allergy warning• Contents, Size, Lot No. and ManufacturingDate• Storage conditionsInclude the required labeling:• Examination Gloves• Single Use Only• Powder-free• Size, Serial No., Manufacturing Date andQuantity• Natural rubber latex• Protein Label Claims• Allergy warning• Manufacturer Name and Address

{10}------------------------------------------------

{11}------------------------------------------------

ANHUI ANYU LATEX PRODUCTS CO., LTD EA

• For Powdered Nitrile Patient Examination Gloves

• Manufacturer Name and Address

ParametersNew DevicePredicate Device
1510(k) NumberK150612K072145
2Marketing clearance date--Feb 26th, 2008
3Device NameAnnuy Powdered Nitrile Patient Examination GlovePowdered Nitrile Patient Examination Glove
4ManufacturerAnhui Anyu Latex Products Co., LtdZhanjiang Jiali Glove Products Co., LTD
5Product CodeLZALZA
6Regulation No.21 CFR 880.625021 CFR 880.6250
7ClassII
8Intended useA patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient andA patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and
examiner.examiner.
9MaterialNitrile Butadiene Rubber LatexNitrile Butadiene Rubber Latex
10DesignSingle useSingle use
Non-sterileNon-sterile
PowderedPowdered
AmbidextrousAmbidextrous
CuffedCuffed
11DescriptionA device covering the hand and wrist area.A device covering the hand and wrist area.
Gloves have separate sheaths or openings forGloves have separate sheaths or openings
each finger and the thumb.for each finger and the thumb.
12SizeProvided in sizes small, medium and largeProvided in sizes small, medium and large
13SpecificationsPowdered Nitrile Examination Gloves MeetPowdered Nitrile Examination Gloves Meet
ASTM D6319-10ASTM D6319-05
14DimensionsMeets ASTM D6319-10Meets ASTM D6319-05
- Overall LengthSmall: 220mm minSmall: 220mm min
Medium: 230mm minMedium: 230mm min
Large: 230mm minLarge: 230mm min
15DimensionsSmall: 80±10mmSmall: 80±10mm
- Palm WidthMedium: 95±10mmMedium: 95±10mm
Large: 110±10mmLarge: 110±10mm
Meets ASTM D6319-10Meets ASTM D6319-05
16DimensionsMeets ASTM D6319-10Meets ASTM D6319-05
- Finger Thickness0.05mm min0.05mm min
DimensionsMeets ASTM D6319-10Meets ASTM D6319-05
- Palm Thickness0.05mm min0.05mm min
17Physical PropertiesMeets ASTM D6319-10Meets ASTM D6319-05
Before Aging:Before Aging:
- Tensile Strength: 14Mpa min- Tensile Strength: 14Mpa min
- Ultimate Elongation:500% min- Ultimate Elongation:500% min
Meets ASTM D6319-10Meets ASTM D6319-05
After aging at 70±2 °C 166±2 hrs:After aging at 70±2 °C 166±2 hrs:
- Tensile Strength: 14Mpa min- Tensile Strength: 14Mpa min
- Ultimate Elongation: 400% min- Ultimate Elongation: 400% min
18Freedom fromMeets ASTM D6319-10 and ASTMMeets ASTM D6319-05 and ASTM
HolesD5151-11;D5151-06
Holes at Inspection Level G-1, AQL 2.5
19Powder AmountMeets ASTM D6124-11,Meets ASTM D6124-06
Below 10mg/dm²Below 10mg/dm² of residual powder
20Absorbable donningCornstarch,Cornstarch,
or dusting powderPowdered with absorbable dusting powder,Powdered with absorbable dusting powder,
U.S.P.U.S.P.
21BiocompatibilityMeets ISO 10993-10:2010;Meets ISO 10993-10:2006;
Skin Irritation: Under the condition of theSkin Irritation: Under the condition of the
test, not an irritant.test, not an irritant.
Skin Sensitization:Under the condition of theSkin Sensitization:Under the condition of
test, not a sensitizer.the test, not a sensitizer.
22ColorBlueChemical characterization:Copper Phthalocyanine β modificationBlueChemical characterization:Copper Phthalocyanine β modification
23Labeling FeaturesInclude the required labeling:• Nitrile Examination Gloves• Non-sterile• Disposable• Powdered• Ambidextrous• Device color: Blue• Contents, Size, Lot No. and Manufacturing Date• Storage conditions• Manufacturer Name and AddressInclude the required labeling:• Nitrile Examination Gloves• Single Use Only• Powdered• Size, Serial No., Manufacturing Date and Quantity• Manufacturer Name and Address

{12}------------------------------------------------

{13}------------------------------------------------

Image /page/13/Picture/1 description: The image shows a logo with a circular shape. The left side of the circle is blue, while the right side is green. Inside the circle, there are curved lines that resemble a plant or a wave. The logo has a clean and modern design.

• For Powder-free Nitrile Patient Examination Gloves

ParametersNew DevicePredicate Device
1510(k) NumberK150612K072145
2Marketing clearance date--Feb 26th, 2008
3Device NameAnnuy Powder-free Nitrile Patient Examination GlovePowder-free Nitrile Patient Examination Glove
4ManufacturerAnhui Anyu Latex Products Co., LtdZhanjiang Jiali Glove Products Co., LTD
5Product CodeLZALZA
6Regulation No.21 CFR 880.625021 CFR 880.6250
7ClassII
8Intended useA patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
9MaterialNitrile Butadiene Rubber LatexNitrile Butadiene Rubber Latex
10DesignSingle useNon-sterilePowder-freeAmbidextrousCuffedSingle useNon-sterilePowder-freeAmbidextrousCuffed
11DescriptionA device covering the hand and wrist area. Gloves have separate sheaths or openings for each finger and the thumb.A device covering the hand and wrist area. Gloves have separate sheaths or openings for each finger and the thumb.
12SizeProvided in sizes small, medium and largeProvided in sizes small, medium and large
13SpecificationsPowder-free Nitrile Examination Gloves Meet ASTM D6319-10Powder-free Nitrile Examination Gloves Meet ASTM D6319-05
14Dimensions- Overall LengthMeets ASTM D6319-10Small: 220mm minMedium: 230mm minLarge: 230mm minMeets ASTM D6319-05Small: 220mm minMedium: 230mm minLarge: 230mm min
15Dimensions- Palm WidthMeets ASTM D6319-10Small: 80±10mmMedium: 95±10mmLarge: 110±10mmMeets ASTM D6319-05Small: 80±10mmMedium: 95±10mmLarge: 110±10mm
16Dimensions- Finger ThicknessMeets ASTM D6319-100.05mm minMeets ASTM D6319-050.05mm min
Dimensions- Palm ThicknessMeets ASTM D6319-100.05mm minMeets ASTM D6319-050.05mm min
17Physical PropertiesMeets ASTM D6319-10Before Aging:- Tensile Strength: 14Mpa min- Ultimate Elongation:500% minMeets ASTM D6319-10After aging at 70±2 °C 166±2 hrs:- Tensile Strength: 14Mpa min- Ultimate Elongation: 400% minMeets ASTM D6319-05Before Aging:- Tensile Strength: 14Mpa min- Ultimate Elongation:500% minMeets ASTM D6319-05After aging at 70±2 °C 166±2 hrs:- Tensile Strength: 14Mpa min- Ultimate Elongation: 400% min
18Freedom fromHolesMeets ASTM D6319-10 and ASTMD5151-11;Holes at Inspection Level G-1, AQL 2.5Meets ASTM D6319-05 and ASTMD5151-06
19Powder residueMeets ASTM D6124-11Below 2mg/glove of residual powderMeets ASTM D6124-06Below 2mg/glove of residual powder
20Polymer CoatingIt is coated with a polyurethane based coatingon the inner side to facilitate donning.Polymer name: polyurethane & polyacrylicacidIt is coated with a polyurethane basedcoating on the inner side to facilitatedonning.Polymer name: polyurethane & polyacrylicacid
21BiocompatibilityMeets ISO 10993-10:2010;Skin Irritation: Under the condition of thetest, not an irritant.Skin Sensitization:Under the condition of thetest, not a sensitizer.Meets ISO 10993-10:2006;Skin Irritation: Under the condition of thetest, not an irritant.Skin Sensitization: Under the condition ofthe test, not a sensitizer.
22ColorBlueChemical characterization:Copper Phthalocyanine β modificationBlueChemical characterization:Copper Phthalocyanine β modification
23Labeling FeaturesInclude the required labeling:• Nitrile Examination Gloves• Non-sterile• Disposable• Powder-free• AmbidextrousInclude the required labeling:• Nitrile Examination Gloves• Single Use Only• Powder-free• Size, Serial No., Manufacturing Date andQuantity

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image shows a logo with a circular design. The top half of the circle is blue, while the bottom half is green. Within the green portion, there are several curved lines that resemble blades of grass or leaves. The overall design is simple and clean, with a focus on natural elements.

{15}------------------------------------------------

Image /page/15/Picture/1 description: The image shows a logo with a circular shape. The top half of the circle is blue, while the bottom half is green. Within the green portion, there are several white lines that resemble blades of grass or stylized leaves. The overall design is simple and clean, suggesting a focus on nature or the environment.

Contents, Size, Lot No. and ManufacturingDate Storage conditions Manufacturer Name and Address
------------------------------------------------------------------------------------------------------

Discussion: Based on the above comparison, the Annuy Latex Examination Gloves (Powdered and Powder-Free), Annuy Nitrile Examination Gloves (Powdered and Powder-Free) and the predicate device have same technology characteristics such as material, design, intended use, specification and performance features.

Both the subject device and predicate device meet the technology characteristics of ASTM D3578, ASTM D6319 and ISO10993-10 standards. From the tests, results generated values that,

Annuy Powdered Latex Examination Gloves contain no more than 10mg/dm² powder and no more than 200µg/dm² extractable protein.

Annuy Powder-free Latex Examination Gloves contain no more than 2mg/glove of residual powder and no more than 200ug/dm2 extractable protein.

Annuy Powdered Nitrile Examination Gloves contain no more than 10mg/dm² powder.

Annuy Powder-free Nitrile Examination Gloves contain no more than 2mg/glove of residual powder.

Conclusion: The Annuy Latex Examination Gloves (Powder-Free) and Nitrile Examination Gloves

(Powdered and Powder-Free) manufactured by Anhui Anyu Latex Products Co., Ltd are substantially equivalent to the predicate device (K072145).

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.