Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K122684
    Date Cleared
    2012-12-07

    (98 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reverse Medical Microcatheter is intended for use in neuro, peripheral and coronary vasculature for the infusion of diagnostic agents such as contrast media, and therapeutic agents such as occlusion coils.

    Device Description

    The Reverse Medical Microcatheter is a single lumen, flexible stiffness composite catheter. The catheter shaft has a hydrophilic coating to reduce friction during use. The Reverse Medical Microcatheter dimensions are included on the individual device labels. The Reverse Medical Microcatheter inner lumen can accommodate guidewires up to 0.018 inches in diameter to access distal tortuous vasculature. Dual radiopaque markers at the distal portion of the catheter facilitate fluoroscopic visualization. The Reverse Medical Microcatheter incorporates a standard luer adapter to facilitate the attachment of accessories. The catheter is provided sterile, non-pyrogenic, and is intended for single use only.

    AI/ML Overview

    The provided text describes the 510(k) submission for the Reverse Medical Microcatheter. Here's a breakdown of the requested information, based solely on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally not explicitly stated with specific numerical thresholds (e.g., "flexibility must be X N/m"). Instead, the report indicates that the device "Met established criteria" for each test. This implies qualitative or pre-defined quantitative criteria that were assessed for conformance.

    TestAcceptance CriteriaReported Device Performance
    Biocompatibility (2. of 7 tests listed)Materials must be biocompatibleMet established criteria
    CytotoxicityPass L929 MEM Elution TestMet established criteria
    SensitizationPass Kligman MaximizationMet established criteria
    Intracutaneous Reactivity (Irritation)Pass Intracutaneous Injection TestMet established criteria
    Systemic Toxicity (Acute)Pass ISO Acute Systemic Injection TestMet established criteria
    Hemocompatibility (Complement Activation)Pass Complement Activation TestMet established criteria
    Hemocompatibility (Hemolysis)Pass Hemolysis TestMet established criteria
    Hemocompatibility (Inactivated Partial Thromboplastin Time Test)Pass Inactivated Partial Thromboplastin Time TestMet established criteria
    Hemocompatibility (In vivo thrombogenicity)Pass in vivo thrombogenicity TestMet established criteria
    PyrogenicityPass USP Material Mediated Rabbit Pyrogen TestMet established criteria
    EtO ResidualsEthylene oxide and Ethylene chlorohydrins residualsMet established criteria
    SterilizationSterility Assurance Level (SAL) of 10⁻⁶Met established criteria
    Design Verification (Bench-Top Testing)
    Dimensional and Visual InspectionMeet established criteria (e.g., dimensions, visual integrity)Met established criteria
    Coating LubricityMeet established criteriaMet established criteria
    Tip BucklingMeet established criteriaMet established criteria
    Flexibility/Shaft Stiffness TestMeet established criteriaMet established criteria
    Flow Rate Test - 100 psi and 300 psiMeet established criteriaMet established criteria
    Priming Volume TestMeet established criteriaMet established criteria
    Kink Resistance TestMeet established criteriaMet established criteria
    Guidewire CompatibilityMeet established criteriaMet established criteria
    Guide Catheter CompatibilityMeet established criteriaMet established criteria
    Catheter Leak Test (Air leakage distal end plugged)Meet established criteriaMet established criteria
    Catheter Leak Test (Liquid leakage)Meet established criteriaMet established criteria
    Dynamic Pressure TestMeet established criteriaMet established criteria
    Static Burst TestMeet established criteriaMet established criteria
    Torque Strength (Turns to Failure)Meet established criteriaMet established criteria
    Tensile Strength TestMeet established criteriaMet established criteria
    Corrosion ResistanceMeet established criteriaMet established criteria
    USP Particulate TestMeet established criteriaMet established criteria
    Navigation Capabilities, Accessibility/Pushability Capabilities, Therapeutic Agents Deliverability (Coils and Stents)Meet established criteriaMet established criteria
    In vivo Tests
    System Deliverability, Compatibility, and VisibilityMeet established criteriaMet established criteria
    Biocompatibility TestingMeet established criteriaMet established criteria

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each bench-top or in-vivo test. It states, "All testing was performed on units that were sterilized and met all inspection criteria." The data provenance is not explicitly mentioned (e.g., country of origin, retrospective/prospective). This is a 510(k) submission, which typically relies on pre-clinical data (bench-top and potentially animal studies for in vivo testing).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This document describes a medical device's physical, mechanical, and biological performance testing, not an AI or diagnostic device that requires expert-established ground truth.

    4. Adjudication method for the test set

    Not applicable, as this is not an AI/diagnostic device and does not involve human readers for ground truth establishment.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device for physical intervention (microcatheter), not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical device; there is no "algorithm only" performance concept in this context.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance criteria in this context is defined by industry standards, engineering specifications, and established scientific methods for evaluating medical device properties (e.g., ISO standards, USP tests, internal design control procedures). For example, biocompatibility uses established biological assay methods, and mechanical tests rely on engineering principles.

    8. The sample size for the training set

    Not applicable. This is a medical device, not an AI model requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this type of medical device submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1