510(k) Data Aggregation

    K Number
    K062698
    Manufacturer
    Date Cleared
    2006-10-11

    (30 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TRIO® and TRIO®+ Spinal Systems are pedicle screw systems intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocations, scoliosis, spinal tumor and failed previous fusion (pseudarthrosis).

    The TRIO® and TRIO®+ Spinal Systems are also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) at the LS-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the development of a solid fusion mass.

    The TRIO® and TRIO®+ Spinal Systems are sacral/iliac screw fixation systems indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, kyphosis and/or lordosis), tumor, pseudoarthrosis and review of failed fusion attempts.

    The TRIO® Spinal System is intended to be used in conjunction with the OSS Diapason Rods, Opus Spinal System Rods, and the Multi-Axis Cross Connectors.

    The TRIO®+ Spinal System is intended to be used in conjunction with the OSS Diapason Rods, Opus Spinal System Rods, Xia® Pre-bent Rods, and the Multi-Axis Cross Connectors.

    Device Description

    The TRIO® and TRIO®+ Spinal Systems are pedicle screw systems intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. The TRIO® and TRIO®+ Spinal Systems are sacral/iliac screw fixation systems indicated for degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, deformities, tumor, pseudoarthrosis and review of failed fusion attempts. The TRIO® and TRIO®+ Spinal Systems are comprised of bone screws, rods, and connectors. The components are available in a variety of lengths in order to accommodate patient anatomy and are fabricated from titanium alloy.

    AI/ML Overview

    This 510(k) premarket notification for the TRIO®+ Spinal System is for a line extension to an existing device, the TRIO® Spinal System. As such, the submission primarily relies on demonstrating substantial equivalence to predicate devices through material, design, and intended use comparisons, along with mechanical testing. It is not a submission for an AI/ML powered device, nor does it describe a clinical study of the type that would typically involve acceptance criteria, ground truth establishment, or human reader performance comparisons.

    Therefore, many of the requested details are not applicable or cannot be extracted from the provided text for this specific type of medical device submission.

    Here's an attempt to answer the questions based on the provided document, noting where the information is not present or relevant:


    Acceptance Criteria and Device Performance Study for TRIO®+ Spinal System

    This submission pertains to a line extension (TRIO®+ Spinal System) to an existing spinal fixation system. The primary method for demonstrating device performance and meeting acceptance criteria is through mechanical testing to confirm comparable mechanical properties to predicate devices, and through demonstrating substantial equivalence in materials, design, and intended use. The document does not specify quantitative acceptance criteria in terms of clinical outcomes or performance metrics typically seen for diagnostic AI/ML devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (as implied by submission type)Reported Device Performance (as implied by submission type)
    Mechanical PropertiesComparable strength, fatigue life, and other relevant mechanical characteristics to predicate devices (e.g., TRIO®, Xia® Spinal Systems)."Mechanical testing also demonstrated comparable mechanical properties to the predicate devices..." (Details of specific tests and numerical results are not provided in this summary, but would have been in the full submission).
    Material CompositionUse of the same material (Titanium alloy) as predicate devices."The subject system consists of Titanium alloy bone screws, rods, and connectors..."
    Basic Design ConceptsShares basic design concepts with predicate devices."The subject components share the same intended use, material, and basic design concepts as that of the predicate device..."
    Intended UseSame intended use as predicate devices."The subject components share the same intended use..." (Indications for use are explicitly listed and are consistent with predicate devices).

    2. Sample Size for Test Set and Data Provenance

    • Sample Size (Test Set): Not applicable in the context of this device and submission type. The "test set" for this device would refer to the samples used in mechanical testing, not a clinical data set. The document does not specify the number of implants/components tested.
    • Data Provenance: Not applicable in the context of clinical data for this device. Mechanical testing data would typically be generated in a laboratory setting.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not applicable. This device is a mechanical implant, not a diagnostic tool requiring expert interpretation for ground truth establishment on a test set.

    4. Adjudication Method for the Test Set

    • Not applicable. Adjudication methods are relevant for clinical studies involving expert disagreement, which is not described for this device.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of diagnostic or AI-assisted tools on human reader performance, which is not applicable to a spinal implant line extension.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • No, a standalone study of an algorithm was not done. This device is a physical medical implant, not an algorithm.

    7. Type of Ground Truth Used

    • Not applicable in the clinical sense. For mechanical testing, the "ground truth" would be established by validated test methods and engineering standards.

    8. Sample Size for the Training Set

    • Not applicable. This is not an AI/ML device requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable.