Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K032611
    Manufacturer
    Date Cleared
    2003-09-17

    (23 days)

    Product Code
    Regulation Number
    864.5425
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Lyphochek® Coagulation Control is intended for use as a quality control plasma to monitor the precision of citrated coagulation systems.

    Device Description

    Lyphochek® Coagulation Control is prepared from human plasma, with added purified biochemicals (tissue extract from animal origin), and preservatives. The control is provided in lyophilized form for increased stability.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Lyphochek® Coagulation Control Level 4, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Product Claim)Reported Device Performance
    Open Vial Stability: All analytes stable for 24 hours at 2-25°CImplied to meet this criterion for all analytes (Studies performed, data on file)
    Shelf Life: Three years when stored at 2-8°CThree years when stored at 2-8°C

    Note: The document states that stability studies were performed to determine these claims, and all supporting data is retained on file at Bio-Rad Laboratories. It does not explicitly present a table comparing measured values against a predefined acceptable range, but rather asserts that the product meets these stability claims based on the studies.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the test set in the stability studies. It only mentions that "Stability studies have been performed."

    Regarding data provenance:

    • Country of Origin: Not specified in the provided document.
    • Retrospective or Prospective: Not specified, but stability studies are typically prospective, involving monitoring the product over time.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not applicable to this type of device (quality control plasma). The "ground truth" for a coagulation control is its inherent stability and performance characteristics, which are determined through laboratory testing using various coagulation systems, not through expert consensus or interpretation of images/data by human experts.

    4. Adjudication Method

    This information is not applicable to this type of device. Adjudication methods are typically used in studies involving human interpretation or subjective assessments, where multiple reviewers' opinions need to be reconciled (e.g., in medical image analysis).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC comparative effectiveness study was not performed and is not applicable to this device. This type of study assesses how human readers' performance changes with AI assistance, which is irrelevant for a laboratory reagent like a coagulation control.

    6. Standalone Performance Study

    A standalone performance study (algorithm only without human-in-the-loop performance) was not explicitly detailed as such, but the stability studies performed on the Lyphochek® Coagulation Control Level 4 represent its "standalone" performance. The device itself is a control material, and its performance is evaluated directly through laboratory testing using established coagulation methods (PT, APTT, Fibrinogen, AT III, TT), not through an algorithm or human-in-the-loop assessment.

    7. Type of Ground Truth Used

    The ground truth used for evaluating the device would be the measured values of the coagulation analytes (Prothrombin Time, Activated Partial Thromboplastin Time, Fibrinogen, Antithrombin III, Thrombin Time) over time, compared against expected ranges or reference values for a stable product. This is essentially laboratory test results and established analytical methods.

    8. Sample Size for the Training Set

    This information is not applicable. This device is a biochemical reagent, not an AI or machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1