Search Results
Found 3 results
510(k) Data Aggregation
(59 days)
The uMR 570 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and that display internal anatomical structure and/or function of the head, body and extremities.
These images and the physical parameters derived from the interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan.
The uMR 570 is a 1.5T superconducting magnetic resonance diagnostic device with a 70cm size patient bore. It consists of components such as magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer, equipment cabinets, power distribution system, internal communication system, and vital signal module etc. The uMR 570 Magnetic Resonance Diagnostic Device is designed to conform to NEMA and DICOM standards.
The modification performed on the uMR 570 (K201540) in this submission is due to the following changes that include:
- (1) Addition and modification of pulse sequences
- (2) Addition of MR imaging processing methods: Inline T2 mapping, Cardiac T1 mapping, Cardiac T2 mapping, Cardiac T2* mapping, Flow Quantification, Arterial Spin Labeling (3D ASL).
- (3) Addition of Spectroscopy Sequences and Post Processing Features: SVS MRS (Liver), Prostate MRS, SVS MRS (Breast).
- (4) Addition of New function: Implant mode, Remote Assistance.
- (5) Addition of New Workflow Features: EasyScan.
This FDA 510(k) summary for the uMR 570 Magnetic Resonance Diagnostic Device details hardware and software modifications compared to a predicate device (K201540). The document asserts substantial equivalence without providing detailed acceptance criteria or study results for novel features.
Here's the breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state acceptance criteria in a quantitative manner for most of the new features. It generally confirms that the device "performs as expected" and is "substantially equivalent" to the predicate. For the specific item where a standard is mentioned:
| Acceptance Criteria (from predicate/standard alignment) | Reported Device Performance |
|---|---|
| NEMA MS 14-2019 (RF coil heating) | "perform as well as predicate devices." |
For other features, the acceptance criterion is implicitly that the features function as intended and yield diagnostically useful information, consistent with the foundational technology of MRI.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not provide information on the sample size used for the test set or the data provenance (country of origin, retrospective/prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided in the document. The general statement is that "physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis."
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
The document does not specify any adjudication method for establishing ground truth or evaluating test results.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The document does not mention any multi-reader multi-case (MRMC) comparative effectiveness study. The device itself is a diagnostic imaging system, not an AI-assisted interpretation tool, so this type of study would not be applicable in this context. The enhancements are primarily in imaging sequences and post-processing capabilities.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is an MRI system, not an AI algorithm for standalone diagnosis. Therefore, a standalone (algorithm only) performance study as typically understood for AI devices would not be relevant or performed. The performance evaluation is for the imaging capabilities and derived parameters.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The document does not explicitly state the type of ground truth used. For an MRI system, the "ground truth" for image quality and quantitative parameter accuracy would typically involve phantom studies, comparison to established imaging techniques, and potentially correlation with other diagnostic findings or clinical outcomes, but this is not detailed. The phrase "physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis" implies clinical utility rather than a specific ground truth mechanism for validation.
8. The sample size for the training set
This information is not applicable as the device is an MRI diagnostic system and not an AI/ML model that would typically have a "training set" in the context of machine learning. The "training" for the system refers to its engineering and calibration.
9. How the ground truth for the training set was established
This information is not applicable for the reasons stated above (not an AI/ML model with a "training set").
Summary of the Study that Proves the Device Meets Acceptance Criteria:
The provided 510(k) summary indicates that non-clinical testing and a clinical performance evaluation were conducted.
- Non-clinical testing: This included evaluation against NEMA MS 14-2019 for RF coil heating, which demonstrated the proposed device performs "as well as predicate devices."
- Clinical performance evaluation: This was done for new features such as Flow Quantification, 3D ASL, Inline T2 Mapping, Cardiac T1/T2/T2* Mapping, MARS+, MultiBand, QScan, and Silicon-Only Imaging.
- Performance testing for Spectroscopy: Specific tests were performed for Prostate MRS, SVS MRS (Breast), and SVS MRS (Liver).
- System Validation: Reports for Implant mode, Remote Assistance, and EasyScan were also provided.
The conclusion states that "The test results demonstrated that the device performs as expected and thus, it is substantial equivalent to the predicate devices to which it has been compared." This implies that the performance was deemed acceptable in comparison to the previously cleared predicate device, without detailing specific metrics or thresholds for acceptance for each new feature beyond the general statement of "as expected" and "substantial equivalence."
Ask a specific question about this device
(105 days)
The uMR 570 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and that display internal anatomical structure and/or function of the head, body and extremities.
These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan.
The uMR 570 is a 1.5T superconducting magnetic resonance diagnostic device with a 70cm size patient bore. It consists of components such as magnet. RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer, equipment cabinets, power distribution system, internal communication system, and vital signal module etc. The uMR 570 Magnetic Resonance Diagnostic Device is designed to conform to NEMA and DICOM standards.
The provided text describes a 510(k) premarket notification for the uMR 570 Magnetic Resonance Diagnostic Device. It details the device's technical specifications and compares it to a previously cleared predicate device (K180925). The focus of the submission is on modifications to the predicate device, including new RF coils, pulse sequences, and imaging processing methods.
However, the document does not describe, in detail, the acceptance criteria for specific performance metrics or the studies that prove the device meets those criteria with quantitative results. Instead, it generally states that "The test results demonstrated that the device performs as expected and thus, it is substantial equivalent to the predicate devices to which it has been compared." The testing listed is primarily technical and safety performance, as opposed to clinical or diagnostic performance with specific acceptance criteria.
Therefore, much of the requested information cannot be extracted from the provided text. I will populate the table and answer the questions based on the information available.
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Image Signal to Noise Ratio | "Performs as expected" (not quantified) |
| Image Uniformity | "Performs as expected" (not quantified) |
| Performance testing for Spectroscopy | "Performs as expected" (not quantified) |
| Performance testing for Computed DWI | "Performs as expected" (not quantified) |
| Surface Heating of RF Receive | "Performs as expected" (not quantified) |
| ES 60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance | "Comply with ES60601-1" |
| IEC 60601-1-2 Medical electrical equipment Part 1-2: General Requirements for basic safety and essential Performance | "Comply with IEC60601-1-2" |
| IEC 60601-2-33 Medical Electrical Equipment - Part 2-33: Particular Requirements For The Basic Safety And Essential Performance Of Magnetic Resonance Equipment For Medical Diagnostic | "Comply with IEC60601-2-33" |
| Clinical performance evaluation | "Performs as expected" (not quantified) |
| Biocompatibility | Patient Contact Materials were tested and demonstrated no cytotoxicity (ISO 10993-5), no evidence for irritation and sensitization (ISO 10993-10). |
| Performance testing for Inline T1/T2*Map | "Performs as expected" (not quantified) |
| Performance testing for SWI+ | "Performs as expected" (not quantified) |
| Performance testing for Easy-scan | "Performs as expected" (not quantified) |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the document. The document mentions "Clinical performance evaluation" but does not detail the nature of this evaluation, the sample size, or data provenance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided in the document. While it states that images are "interpreted by a trained physician," there is no detail on the number or qualifications of experts for ground truth establishment during testing.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided in the document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
A multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned in the document. This device is an MRI system, not explicitly a an AI/CAD device. The new features like Inline T1/T2*Map, PSIR, cDWI, and SWI+ are imaging processing methods, not AI assistance for human readers in the context of comparative effectiveness.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not provided in the document. The device itself is an MRDD that produces images to be interpreted by a trained physician; it does not describe a standalone algorithm for diagnostic output without human interpretation.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The document primarily focuses on technical performance and equivalence to a predicate device. For the "Clinical performance evaluation," the type of ground truth used is not specified. For the general use of the device, images are "interpreted by a trained physician," implying clinical diagnosis as the ultimate ground for interpretation, but not in the context of establishing ground truth for a formal evaluation of the device's diagnostic accuracy.
8. The sample size for the training set
The document describes the device, its modifications, and performance testing for regulatory clearance. It does not mention any training set as this is not a submission for an AI/ML algorithm that typically requires a training set. The "new application software features" are described as "substantially equivalent" to conventional methods, not as machine learning models that require training.
9. How the ground truth for the training set was established
As no training set is mentioned (see point 8), this information is not applicable.
Ask a specific question about this device
(58 days)
The uMR 570 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and that display internal anatomical structure and/or function of the head, body and extremities.
These images and the physical parameters derived from the interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan.
The uMR 570 is a 1.5T superconducting magnetic resonance diagnostic device with a 70cm size patient bore. It consists of components such as magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer, equipment cabinets, power distribution system, internal communication system, and vital signal module etc. The uMR 570 Magnetic Resonance Diagnostic Device is designed to conform to NEMA and DICOM standards.
I'm designed to only summarize and extract information from the provided text, and the given text does not contain acceptance criteria or study details needed to answer your request. The document is an FDA 510(k) clearance letter and summary for a Magnetic Resonance Diagnostic Device (uMR 570), which focuses on demonstrating substantial equivalence to a predicate device rather than detailing specific acceptance criteria and a study proving those criteria.
Therefore, I cannot provide the acceptance criteria and study details you've requested based on the provided input.
Ask a specific question about this device
Page 1 of 1