YUWELL**®** Infrared Ear Thermometer (YHT100); YUWELL® Infrared Ear Thermometer (YHT107)
The Infrared Ear Thermometer is indicated for intermittent measurement of human body temperature from the ear canal. The device can be used by people of all ages except preterm babies or babies who are small for gestational age. The thermometer is intended for use in professional settings and the home environment. It is not for emergency clinical conditions.
YUWELL® Infrared Ear Thermometer is designed to measure human body temperature. This hand-held, no-contact device is battery-powered and utilizes infrared energy emitted in the subject's tympanic membrane to determine human body temperature. The thermometer provides temperature readings in just 2 seconds. It is suitable for individuals of all ages except preterm babies or babies who are small for gestational age. The Infrared Ear Thermometer YHT107 can transmit the temperature readings via Bluetooth, whereas the YHT100 does not have data transmission capability.
Here's a detailed breakdown of the acceptance criteria and study information for the YUWELL® Infrared Ear Thermometer, extracted from the provided FDA 510(k) clearance letter:
YUWELL® Infrared Ear Thermometer (YHT100; YHT107) Acceptance Criteria and Study Details
The YUWELL® Infrared Ear Thermometer (YHT100; YHT107) is indicated for intermittent measurement of human body temperature from the ear canal in professional and home environments, for people of all ages except preterm babies or babies who are small for gestational age.
1. Table of Acceptance Criteria and Reported Device Performance
The core performance acceptance criteria are related to temperature accuracy, as specified by the standards ISO 80601-2-56 and ASTM E1965.
Characteristic | Acceptance Criteria (Standard Specification) | Reported Device Performance (Subject Device) |
---|---|---|
Accuracy | For 35.0℃–42.0℃ (95.0℉–107.6℉): ±0.2℃ (±0.4℉) | Achieved: ±0.2℃ (±0.4℉) for 35.0℃–42.0℃ (95.0℉–107.6℉) |
For 34.0℃–34.9℃ (93.2℉–94.8℉) & 42.1℃–42.2℃ (107.8℉–108.0℉): ±0.3℃ (±0.5℉) | Achieved: ±0.3℃ (±0.5℉) for 34.0℃–34.9℃ (93.2℉–94.8℉) and 42.1℃–42.2℃ (107.8℉–108.0℉) | |
Temperature Range | Conforms with ASTM E1965 (Predicate range: 20℃-42.2℃) | 34.0-42.2℃ (93.2-108.0℉) |
Display Resolution | 0.1℃/0.1℉ (Matches predicate) | 0.1℃/0.1℉ |
Response Time | Not explicitly stated as a strict acceptance criterion with quantitative limits, but differences were evaluated. (Predicate: 2-3s) |