Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K230587
    Date Cleared
    2023-11-30

    (273 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Wrist Pulse Oximeter

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MD300W628 Wrist Pulse Oximeter is a portable, non-invasive device intended to measure, display, store and transfer functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate (PR) of adult, adolescent and child in hospitals, clinics, long-term care, and home use.

    The device intended for spot-checking and can be reused, not intended for continuous monitoring, use during motion or use with low perfusion.

    Device Description

    The subject device Pulse Oximeter is an internally powered device. The main functions of the devices include hemoglobin oxygen saturation (SpO2), pulse rate (PR) measurements and Pulse amplitude index (PAI), data storage and transmission.

    Place one fingertip into the sensor for diagnosis and the oxygen saturation (SpO2), pulse rate (PR) measurements and pulse amplitude index (PAI) will appear on the display. The device is normally applied to adult and pediatric patients in hospital and home care environment.

    The subject device is composed of following components to achieve the above detection process: power supply module, detector and emitter, signal collection and process module (MCU), OLED display screen and Bluetooth module.

    Principle of the oximeter is as follows: The pulse oximeter works by applying a sensor to a fingertip. The sensor contains a dual light source and photo detector. Skin, bone, tissue and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.

    AI/ML Overview

    This FDA 510(k) summary provides information on the Beijing Choice Electronic Technology Co., Ltd. MD300W628 Wrist Pulse Oximeter, including its acceptance criteria and the studies conducted to demonstrate its performance.

    Here's the breakdown:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document relies on compliance with recognized standards for its acceptance criteria, particularly ISO 80601-2-61 for pulse oximeter equipment accuracy.

    Performance MetricAcceptance Criteria (Standard/Range)Reported Device Performance (MD300W628)
    SpO2 Accuracy$\text{ARMS} \leq 3.0%$ (standard for pulse oximeters)$\text{ARMS} = 1.68%$ for SpO2 range 70-100%
    SpO2 Display RangeVerified according to IEC 60601-1 and ISO 80601-2-61 requirements$70% \sim 100%$
    PR Display RangeVerified according to IEC 60601-1 and ISO 80601-2-61 requirements$30 \sim 250 \text{bpm}$
    PAI Accuracy$0.3 \sim 1.0% (\pm 0.2 \text{digits})$; $1.1 \sim 20.0% (\pm 20%)$Test results available in System Test Report (specific values not detailed in summary)
    Electrical SafetyCompliance with IEC60601-1, IEC 60601-1-11Conformed
    EMCCompliance with IEC60601-1-2Conformed
    BiocompatibilityCompliance with ISO10993-1, ISO10993-5, ISO10993-10Met standard requirements
    Software ValidationCompliance with FDA Guidance for Content of Premarket Submissions for Software Contained in Medical DevicesCompliant (Moderate level of concern)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Test (SpO2 Accuracy): 11 healthy adult volunteer subjects.
    • Data Provenance: The clinical study was conducted at Yue Bei People's Hospital. The origin is inferred to be China, as Beijing Choice Electronic Technology Co., Ltd. is based in Beijing, China. The study appears to be prospective as it involved recruiting subjects and performing specific tests.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not explicitly state the number of experts used or their qualifications for establishing ground truth. It mentions that the clinical study evaluated SpO2 accuracy "as compared to arterial blood CO-Oximetry." CO-Oximetry is the gold standard for measuring arterial oxygen saturation, and its results are considered highly accurate, typically not requiring expert interpretation for ground truth establishment in this context.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the test set. Given that the ground truth for SpO2 accuracy was established using arterial blood CO-Oximetry, which is an objective measurement, an adjudication method for disagreements on ground truth would generally not be necessary for this specific performance metric.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study being conducted or of any effect size related to human readers improving with or without AI assistance. This device is a pulse oximeter, which provides direct physiological measurements, not interpretations that would typically involve multiple human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The SpO2 accuracy performance study described is essentially a standalone performance study. The device's measurement of SpO2 is directly compared to an objective gold standard (arterial blood CO-Oximetry), without human intervention in the interpretation of the device's output. The device itself (algorithm) provides the SpO2 reading.

    7. Type of Ground Truth Used

    The ground truth used for the SpO2 accuracy testing was arterial blood CO-Oximetry.

    8. Sample Size for the Training Set

    The document does not provide information on the sample size for a training set. Pulse oximeters typically rely on established optical principles and signal processing algorithms rather than machine learning models that require large training datasets in the same way AI algorithms for image analysis might. The "algorithm" here refers to the signal processing rather than a deep learning model.

    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of a training set in the context of typical machine learning, there is no information on how ground truth for a training set was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183556
    Date Cleared
    2019-02-15

    (57 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    oCare Wrist Pulse Oximeter

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The oCare Wrist Pulse Oximeter, Model Pro 100, is a wrist-worn device indicated for use in noninvasive measuring, displaying, and storing functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse rate (PR). The intended measuring site of this device is the wrist skin surface. It is intended for spot-checking or continuous monitoring of adult patients during no motion conditions, in hospital-type facilities, and home environments.

    Device Description

    The oCare™ Wrist Pulse Oximeter, Model Pro 100 is a wrist-worn device for noninvasive measuring, displaying, and storing functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse rate. The intended measuring site of this device is the lateral side of wrist and slightly above the wrist bone (ulnar styloid processus).

    The oCare™ Wrist Pulse Oximeter, Model Pro 100 is a watch-like device with a reflectance pulse oximetry sensor located at the bottom of the watch case and on top of the wrist. The reflectance pulse oximetry sensor includes three light emitting diodes (LEDs) of red, infrared and green wavelength and one photodiode light detector placed next to each other. Light beams are emitted from LEDs through the skin to the arteriolar bed of the tissue. Changes in light absorption during the pulsing cycle are measured by the photodiode light detector as scattered lights are reflected back from the pulsating arteriolar bed. The functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse rate are measured by the well-established non-invasive pulse oximetry technology where the red and infrared light is absorbed in different amounts depending on the oxygenation of the blood during the arterial pulsing. The maximum optical output power is less than 2 mW.

    The oCare™ Wrist Pulse Oximeter, Model Pro 100 is a single-patient use, non-sterile pulse oximeter. It is available in one configuration as a standalone device with a wrist pulse oximeter and a detachable watchband for wearing the pulse oximeter on the wrist.

    The oCare™ Wrist Pulse Oximeter, Model Pro 100 is a compact and light weight device which consists of a reflectance pulse oximetry sensor, a color graphic OLED display, a lithium ion polymer rechargeable battery, a memory, an analog and digital unit, a microprocessor and an operating software.

    AI/ML Overview

    The provided text describes the performance data for the oCare™ Wrist Pulse Oximeter, Model Pro 100, focusing on its accuracy in measuring functional oxygen saturation of arterial hemoglobin (% SpO2) and pulse rate (PR).

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Specification)Reported Device Performance (Test Results)
    SpO2 Accuracy±3% SpO2 (over 70-100% SaO2 range), non-motionWithin ±3% SpO2 (over 70-100% SaO2 range), non-motion
    Pulse Rate±3 bpm (from 40-240 bpm)±3 bpm (from 40-240 bpm)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The exact number of human subjects is not explicitly stated. The text mentions "human subjects screened for this testing" and "each human subject," implying multiple individuals, but a specific number (e.g., n=X) is not provided.
    • Data Provenance: The study was conducted at an "independent research university hospital." The subjects were "health, male, female, no-smoking and light to dark-skinned adult subjects of 20-40 years of age."
      • Country of Origin: Not explicitly stated, but the submission is from Taiwan Biophotonic Corporation, suggesting the study might have been conducted in Taiwan or an affiliated institution.
      • Retrospective or Prospective: Prospective. The study "was designed in accordance with ISO 80601-2-61" to "verify the SpO2 accuracy" where hypoxia was induced in a "stair-stepped manner" and arterial blood samples were taken simultaneously. This indicates a pre-planned, forward-looking study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The text does not indicate that human experts were used to establish ground truth for the device's performance. The ground truth for SpO2 was established invasively using co-oximetry with arterial blood samples. Therefore, this section is not applicable in the context of expert review for ground truth.

    4. Adjudication Method for the Test Set

    Not applicable. The ground truth for SpO2 accuracy was established through direct
    invasive measurement (co-oximetry) of arterial blood samples, not through human
    adjudication of medical images or observations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not conducted. This study focused on the standalone accuracy of the pulse oximeter device against an invasive gold standard (co-oximetry), not on evaluating how human readers (e.g., clinicians) perform with or without AI assistance. The device is a diagnostic tool, not an AI-assisted diagnostic aid for clinicians.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, a standalone study was performed. The "Clinical Studies" section describes a study testing the oCare™ Wrist Pulse Oximeter's accuracy in measuring SpO2 directly against a gold standard (co-oximetry with arterial blood samples). This is a purely algorithmic/device performance evaluation without any human interpretation of the device's output being part of the primary performance metric.

    7. The Type of Ground Truth Used

    The ground truth used for SpO2 accuracy was invasive functional oxygen saturation of arterial hemoglobin (% SaO2) analyzed by the co-oximetry method with arterial blood samples taken simultaneously.

    8. The Sample Size for the Training Set

    Not applicable. The device is a pulse oximeter that uses well-established non-invasive pulse oximetry technology (based on light absorption by hemoglobin) and does not appear to employ machine learning that would require a "training set" in the conventional sense (e.g., for AI model development). Its operation is based on physiological principles and signal processing, not a data-driven learning algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention or indication of a "training set" for this
    device's underlying technology.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172366
    Date Cleared
    2018-03-16

    (224 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Wrist Pulse Oximeter

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MD300W512 / MD300W314/ MD300W314B4 is a wrist pulse oximeter indicated for use in measuring, displaying, storing and transmitting functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adult, adolescent, child and infant patients. It is intended for spot-check and / or data collection, recording and transmitting. It can be used in sleep labs, long-term care, hospitals and home use.

    Device Description

    The proposed devices Wrist Pulse Oximeter MD300W512 /MD300W314/MD300W314B4 are internally powered devices. The main functions of the devices include hemoglobin oxygen saturation (SpO2) and pulse rate (PR) measurements, visual and audible indication, data storage and transmission. The proposed device MD300W512/MD300W314/MD300W314B4 consists of power supply module, detector and emitter LED, signal collection and process module, display module, user interface, button control circuit, data storage and transmission module.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria & Reported Device Performance

    Acceptance Criteria (Stated)Reported Device Performance (Proposed Device)
    SpO2 measurement range: 70%~100%70%~100%
    SpO2 accuracy:70%~100%, ±2%;
    Ask a Question

    Ask a specific question about this device

    K Number
    K152089
    Date Cleared
    2016-08-17

    (387 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Wrist Pulse Oximeter

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wrist Pulse Oximeter is intended for spot checking in measuring and displaying oxygen saturation of functional arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patients in hospital facilities and home healthcare environments.

    Device Description

    This device is a small, lightweight, portable device intended for use in measuring and displaying oxygen saturation of functional arterial haemoglobin (%SpO2) and pulse rate (PR).The device measures SpO2 and PR with a SpO2 senror and displays on the OLED after certain further processing. The device is mainly composed of the motherboard, SpO2 transducer and built-in battery. The power supply is 4.2V Li-battery and also the subject device can be powered by adapter. The subject device has a detachable sensor as an accessory to measure the SpO2 and pulse rate (PR). The subject device is not for life-supporting or life-sustaining and it is not an implantable device. The subject device owns software which has been validated by the manufacturer.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study findings for the Wrist Pulse Oximeter, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The relevant acceptance criteria and performance data for the Wrist Pulse Oximeter primarily come from the "Substantial Equivalence" section and the "Clinical Test" section.

    FeatureAcceptance Criteria (Predicate or Standard)Reported Device Performance (SONOSAT-W01T)
    SpO2 Accuracy70 to 100% ±3% (Predicate MD300W4)±2% in the range of 70% to 100%
    SpO2 Measurement Range70%-100% (Predicate MD300W4)0%-100% (Subject Device)
    Pulse Rate Accuracy25bpm~250bpm, ±3bpm (Predicate MD300W4)25bpm~250bpm, ±3bpm (Subject Device)
    SpO2 Resolution1% (Predicate MD300W4)1% (Subject Device)
    Pulse Rate Resolution1 bpm (Predicate MD300W4)1 bpm (Subject Device)
    General RequirementCompliance with IEC60601-1 (Predicate & Subject)Compliance with IEC60601-1
    Special RequirementCompliance with ISO 80601-2-61, IEC 60601-1-11Compliance with ISO 80601-2-61, IEC 60601-1-11
    EMCCompliance with IEC60601-1-2 (Predicate & Subject)Compliance with IEC60601-1-2
    BiocompatibilityCompliance with ISO10993-1 (Predicate & Subject)Compliance with ISO10993-1
    Biocompatibility Test ResultsNo toxicity, no delayed contact sensitization, no irritation to skin (Predicate)No toxicity to cells, No delayed contact sensitization, No irritation to skin (Subject Device)
    Software ValidationCompliance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical DevicesIn compliance with FDA Guidance
    Clinical Study ComplianceCompliance with ISO 80601-2-61:2011 and FDA guidance - "Pulse Oximeter Premarket Notification Submissions"In compliance with standards and guidance

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The clinical evaluation was conducted on a total of 24 healthy adult male and female volunteers.
    • Data Provenance: The document does not explicitly state the country of origin for the clinical study participants beyond mentioning "2 white people, and 15 yellow-skin people." It is implied to be a prospective study as it describes a clinical evaluation conducted to generate data for the submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number of experts used or their qualifications for establishing ground truth in the clinical study. For pulse oximetry, the ground truth is typically established by arterial blood gas (ABG) analysis, which is a laboratory measurement, not directly by "experts" in the sense of human readers. The clinical study compares the device's readings against these reference measurements.

    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method in the context of human expert review for the test set. For a pulse oximeter, the "ground truth" is typically objective physiological measurements (ABG), not subjective expert interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. A MRMC study is typically performed for image-based diagnostic aids where the AI assists human readers. This document describes a medical device (pulse oximeter) that provides a direct physiological measurement, and its accuracy is validated against a gold standard (ABG), not via a comparative effectiveness study with human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes. The primary study detailed is a standalone performance validation of the Wrist Pulse Oximeter where its measurements (SpO2 and pulse rate) are compared directly against a reference method (likely arterial blood gas measurements, though not explicitly stated for SpO2, it's the standard for oximeter validation). The reported SpO2 accuracy of ±2% is an algorithm-only performance metric.

    7. The Type of Ground Truth Used

    The ground truth for the SpO2 accuracy validation was established using a physiological reference standard. While not explicitly named for SpO2 in the clinical test section, the widely accepted and regulatory-mandated method for validating pulse oximeter accuracy (specifically the Arms value) involves inducing various levels of hypoxemia and comparing the oximeter readings to co-oximeter measurements of arterial blood samples (Arterial Blood Gas - ABG).

    8. The Sample Size for the Training Set

    The document does not provide information regarding a separate training set or its sample size. The description of validation usually pertains to the final, locked algorithm or device. For devices like pulse oximeters, the core algorithm is often developed based on photometric principles and extensive internal testing, rather than a "training set" in the machine learning sense. The clinical test described is a validation test, not a training activity.

    9. How the Ground Truth for the Training Set was Established

    As no training set is described in the document, there is no information provided on how ground truth for such a set (if it existed) was established. Device development often involves internal calibration and verification, but these details are not typically part of a 510(k) summary focused on regulatory validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122046
    Date Cleared
    2013-02-04

    (207 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    WRIST PULSE OXIMETER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MD300W4 wrist pulse oximeter is a portable non-invasive device intended for spot checking ,data collection and recording of arterial oxygen saturation (SpO2) and pulse rate of adult and pediatric patient at home and hospital( including clinical use in internist/ surgery, Anesthesia, intensive care and etc).

    Device Description

    The proposed device MD300W4 consists of MCU circuit, power supply circuit, SpO2 module circuit, display circuit, Flash memory circuit, GPRS module circuit, charging circuit, real-time clock circuit, button circuit.
    It can measure, store, review and display the SpO2% and pulse rate value, time, ID number, pulse bar and battery power status, the connection of probe, and transmit data by GPRS or USB cable.
    The power supply is 4.2V Li-battery with capacity 1250mAh. The device can not be used to measure when it charges for the Li- battery.
    The Pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and a photodetector. The one wavelength of light source is 660 nm, which is red light; the other is 940 nm, which is infrared light.
    Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photodetector in probe collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.
    The proposed device has one model detachable sensor as the accessory of which the skin-contacting material is silicon. It is listed below:
    M-50G (510(k) number: K082487)
    The proposed device is not for life-supporting or life-sustaining, not for implant. The device or probe is not sterile and does not need sterilization or re-sterilization. The device is for prescription. The device does not contain drug or biological products.
    The device is software-driven and the software validation is provided in Section 14 Software on page 14-1.

    AI/ML Overview

    Here's an analysis of the provided text to fulfill your request:

    Acceptance Criteria and Device Performance Study for Wrist Pulse Oximeter MD300W4 (K122046)

    Based on the provided 510(k) summary, the device is a modified Wrist Pulse Oximeter, MD300W4, and its performance is evaluated against the predicate device, MD300W (K081125), and established standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the MD300W4 are largely aligned with its predicate device and industry standards for pulse oximeters. The reported device performance is presented as compliance with these specifications.

    CharacteristicAcceptance Criteria (Target/Standard)Reported Device Performance (MD300W4 Results)
    SpO2 Accuracy70%-100% ±3% (General)70%-100% ±3% (Claimed by manufacturer)
    90%-100% ±2%, 80%-90% ±3%, 70%-80% ±3% (Specific ranges)`90%-100% ±2%, 80%-90% ±3%, 70%-80% ±3%,
    Ask a Question

    Ask a specific question about this device

    K Number
    K081125
    Date Cleared
    2008-07-02

    (72 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    WRIST PULSE OXIMETER MD300W

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MD300W wrist oximeter is a portable non-invasive, spot-check, oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital ( including clinical use in internist/ surgery, Anesthesia, intensive care and etc). Not for continuously monitoring.

    Device Description

    The applicant device of Wrist Pulse Oximeter MD300W is a wrist-worn device, which can display %SpO2, pulse rate value and vertical bar graph pulse amplitude,

    The applicant device consists of sensor, signal amplify unit, CPU, data display unit, data transmit unit, storage and power unit.

    The wrist oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660 nm, which is red light; the other is 940 nm, which is ultra red light.

    Skin, bone, tissue, and venous vessels normally absorb a constant amount of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.

    The applicant device has low battery voltage alarm function and automatically power of function. The power source of the applicant device is 1 AAA alkaline batteries.

    The applicant device is not for life-supporting or life-sustaining, not for implant. The device or sensor are not sterile and the sensor does not need sterilization and the sensor is reusable but does not need re-sterilization since it is not sterile. The device is for prescription. The device does not contain drug or biological products.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Wrist Pulse Oximeter MD300W:

    Please note: The provided text is a 510(k) Summary for a medical device. These summaries typically contain a high-level overview and refer to more detailed reports (like clinical test reports) in their appendices. Therefore, some detailed information might not be explicitly stated in this summary but would be found in the referenced documents.


    Acceptance Criteria and Device Performance

    CriteriaReported Device Performance
    SpO2 Accuracy Conformance"The accuracy of MD300W pulse oximeter equipment is compliance to the requirement" of ISO 9919:2005, Medical electrical equipment - Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use. (The specific accuracy range, e.g., ±2%, is not detailed in this summary but would be within the ISO standard.)
    Safety - ElectricalCompliance with IEC 60601-1 (General requirements for safety) and IEC 60601-1-2 (Electromagnetic compatibility).
    Safety - BiocompatibilityCompliance with ISO 10993 "Biological Evaluation of Medical Devices" for skin-contacting materials.
    Safety - Safety Mechanism (Current)Performance Test reports (MD300W-01-001) for safety mechanism preventing excess current from leading to burning injury. (Actual performance not explicitly stated beyond "reports regarding.")
    Safety - Low-Voltage Alarm SystemPerformance Test reports (MD300W-01-002) for Low-Voltage Alarm System. (Actual performance not explicitly stated beyond "reports regarding.")
    Software ValidationSoftware validation provided in Chapter VIII. (Specific acceptance criteria or detailed performance not provided in this summary.)

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not explicitly stated for the clinical test. The summary mentions "adult and pediatric patient," but the number of subjects is not provided.
      • Data Provenance: Clinical tests were "conducted in Wulanchabu City Center Hospital" and "Laboratory of Beijing Friendship Hospital." This suggests the data is prospective clinical trial data from China.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not explicitly provided in the 510(k) summary. For pulse oximetry clinical studies, ground truth is typically established using a co-oximeter measuring arterial blood gas samples, not expert consensus in the traditional sense of image interpretation.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This information is not applicable to a pulse oximetry accuracy study where ground truth is established through instrumental measurement (co-oximetry) rather than expert review. There's no "adjudication" necessary for the primary outcome.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This device is a standalone pulse oximeter, not an AI-assisted diagnostic tool that involves human readers interpreting results.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone performance assessment was done. The clinical test assesses the accuracy of the device itself (algorithm and hardware) against a reference standard without human interpretation. The summary states: "The Clinical Test reports following ISO 9919:2005... are conducted in Wulanchabu City Center Hospital provided in Attachment IV Clinical Test Reports."
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The ground truth for pulse oximeter accuracy studies is typically established by arterial blood gas analysis using a co-oximeter. While not explicitly named in the summary, conformance to ISO 9919:2005 (which specifies requirements for in vivo clinical testing) implies this method. ISO 9919:2005 requires comparing the device's SpO2 readings to simultaneously drawn arterial blood samples analyzed by a laboratory co-oximeter.
    7. The sample size for the training set:

      • This information is not applicable/not provided. Pulse oximeters of this type, especially from this era (2008), typically rely on established physiological models and calibration rather than machine learning "training sets" in the modern AI sense. While internal calibration data would be used, it's not a "training set" for an algorithm that learns from data.
    8. How the ground truth for the training set was established:

      • As noted above, the concept of a "training set" and associated ground truth is not directly applicable in the context of this device's type and the information presented. The device's underlying principles are based on the Beer-Lambert law and physiological light absorption, not a learned AI model.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1