Search Results
Found 1 results
510(k) Data Aggregation
(59 days)
The Wattson temporary pacing guidewire is intended to introduce and position catheters and other interventional devices within the chambers of the heart, including those used within transcatheter aortic valve replacement (TAVR) procedures and balloon aortic valvuloplasty (BAV), while transmitting an electrical signal from an external pulse generator to the heart. The temporary pacing guidewire is not intended to remain in place following the clinical procedure.
The Wattson temporary pacing guidewire is a dual-purpose 0.035" guidewire designed for the delivery of devices and for temporary pacing of the heart. The distal end of the device has an atraumatic pigtail shape. The shaft of the device has a hydrophilic coating. Pacing is achieved through a bipolar electrode configuration integrated in the guidewire. The electrodes terminate in a single positive electrode at the distal tip of the device and three discrete negative electrodes located proximal to the tip within the pigtail section. The entire device is visible using standard fluoroscopic methods. The device is packaged with a proprietary adapter that locks on to the proximal end of the guidewire and converts it to discrete positive and negative terminal pins. The terminal pins of the adapter are compatible with standard external pulse generators. Unshrouded adaptor pins are provided to ensure compatibility with all connecter cable terminals. The Wattson temporary pacing guidewire is sterilized with ethylene oxide.
This document is a 510(k) Summary for the Wattson Temporary Pacing Guidewire (Model 2250). It details the device's indications for use, its characteristics, and a comparison to a predicate device (K192454, also a Wattson temporary pacing guidewire). The summary also outlines the performance data gathered to support the claim of substantial equivalence.
Based on the provided text, the device is a medical guidewire used in cardiac procedures, specifically Transcatheter Aortic Valve Replacement (TAVR) and Balloon Aortic Valvuloplasty (BAV), which also has the capability to transmit an electrical signal for temporary cardiac pacing.
Here's an analysis of the acceptance criteria and study proving the device meets them, based solely on the provided text:
Important Note: The provided document is a 510(k) Summary, which is a regulatory submission for medical devices to demonstrate substantial equivalence to a legally marketed predicate device. It typically summarizes testing, but does not provide the full detailed study reports. Therefore, some information requested in your prompt (e.g., number of experts, specific effect sizes for MRMC studies, sample sizes for training sets for AI models) may not be present because this device is a traditional guidewire, not an AI/ML-driven diagnostic device. The questions you've asked are largely applicable to AI/ML device submissions, which is not what this document describes.
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a submission for a physical medical device (guidewire) and not an AI/ML diagnostic tool, "acceptance criteria" are typically defined by engineering specifications, material biocompatibility, and functional performance benchmarks rather than diagnostic metrics like sensitivity/specificity. The document doesn't present a formal table of quantitative acceptance criteria and performance results in the way one would for an AI model.
However, the "Performance Data" section outlines the types of testing performed, which implicitly demonstrates "acceptance" based on successful completion of these tests.
| Category | Acceptance Criteria (Implicit from testing type) | Reported Device Performance |
|---|---|---|
| Biocompatibility | Meets ISO 10993-1 standards for an externally communicating device in contact with circulating blood and tissue for limited time (<24 hours). | - Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogenicity, Hemolysis, Complement Activation, and Thrombogenicity tests were conducted. (Implies successful completion for acceptance). |
| Bench Performance | Device design verified for: Performance, Patient Safety, Structural Integrity, Infection, Microbial & Particulate Contamination, Device Interface. | - Device design was verified through these tests. (Implies successful completion against internal specifications). |
| Animal Study (GLP Safety) | Performance similar to predicate device, no new safety concerns, no sustained arrhythmia, return to normal sinus rhythm, no histological differences. | - Performed similarly to the predicate regarding adverse events and other animal health concerns. - No sustained arrhythmia in post-rapid pacing monitoring period. - All animals returned to normal sinus rhythm after rapid pacing was turned off. - No histological findings indicating differences from predicate. |
2. Sample Sizes Used for the Test Set and Data Provenance
- Animal Study (GLP Safety): The study was performed in a porcine model. The document states "No animals experienced sustained arrhythmia...", implying a cohort of animals, but the exact number of animals (sample size) is not specified.
- Data Provenance: The animal study was a prospective GLP Safety study. The country of origin for the data is not specified but given the US FDA submission, it's likely a US-based study or conducted to US regulatory standards. Biocompatibility and Bench tests are laboratory-based and generally don't have "data provenance" in the same regional sense.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This question is not applicable to this device submission. The Wattson Temporary Pacing Guidewire is a physical medical device, not an AI-based diagnostic tool that relies on expert interpretation for "ground truth" to validate an algorithm's performance on image data. For an AI/ML device, ground truth would typically be established by expert radiologists or pathologists.
For this device, "ground truth" (or verification) is established through technical performance specifications, material science, and observed biological/physiological responses in animal models and bench testing.
4. Adjudication Method for the Test Set
Not applicable. As noted above, this device is not an AI/ML diagnostic. Adjudication methods (like 2+1 or 3+1) are used in studies where multiple human readers interpret data to create a consensus "ground truth" for evaluating an AI algorithm. For a physical device, performance is measured against engineering specifications, biological responses, and comparative performance to a predicate.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and Effect Size
No. An MRMC study is relevant for evaluating the impact of an AI algorithm on human reader performance (e.g., radiologists interpreting images with or without AI assistance). This submission is for a guidewire, which does not involve human readers interpreting images in a way that would necessitate an MRMC study for its primary function.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI algorithm; it is a physical medical device. Therefore, the concept of "standalone performance" for an algorithm does not apply. The device's performance is inherently "standalone" in its primary function as a guidewire and pacing tool.
7. The Type of Ground Truth Used
The "ground truth" for this device's validation is based on:
- Physiological/Biological Outcomes: Observed responses in the porcine model (e.g., return to normal sinus rhythm, absence of sustained arrhythmia, lack of adverse events, histological findings).
- Bench Test Specifications: Performance against internal engineering parameters, safety margins, structural integrity, and material properties.
- Biocompatibility Standards: Conformance to recognized ISO standards (ISO 10993-1).
- Comparison to Predicate: The ultimate "ground truth" for substantial equivalence is demonstrating that the study device performs "similarly" and raises "no new questions of safety or effectiveness" compared to the predicate device.
8. The Sample Size for the Training Set
Not applicable. As this device is a physical medical device and not an AI/ML algorithm, there is no "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable. Since there is no AI/ML algorithm and therefore no training set, the concept of establishing ground truth for a training set does not apply.
Ask a specific question about this device
Page 1 of 1