Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K070060
    Manufacturer
    Date Cleared
    2007-12-18

    (347 days)

    Product Code
    Regulation Number
    870.1120
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FlexiPort™ Blood Pressure Cuff is used in conjunction with noninvasive blood pressure measurement systems. The cuff is nonsterile and is available in pediatric through adult sizes. The device is not intended for neonatal applications. The FlexiPort™ Blood Pressure Cuffs are not designed, sold, or intended for use, except as indicated.

    Device Description

    FlexiPort™ Blood Pressure Cuffs are designed with the FlexiPort™, a standardized connector for quick exchange of tubing and fittings that interface with manual or electronic sphygmomanometers and inflation devices. This allows the interface tubing (either one- or two-tube) to remain with the blood pressure measurement device, while the cuff can be easily removed and moved from one location to another to be used universally within a facility on multiple devices. The port connector can also rotate 360 degrees to reduce tube kinking.

    AI/ML Overview

    The Welch Allyn FlexiPort™ Blood Pressure Cuff is designed to measure blood pressure in conjunction with noninvasive blood pressure measurement systems. The device is intended for pediatric through adult sizes and is not for neonatal applications.

    1. Table of Acceptance Criteria & Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    BiocompatibilityConforms to ISO 10993-1
    Compatibility with other devicesTested and demonstrated compatible with other devices
    ReprocessingTested and demonstrated effective reprocessing
    PackagingTested and demonstrated effective packaging
    SterilizabilityCapable of being sterilized using ETO
    EnvironmentalNot explicitly detailed, but implied to meet standards through overall testing
    Mechanical StrengthNot explicitly detailed, but implied to meet standards through overall testing
    DurabilityTested for repeated inflations; implied to meet durability standards
    AccuracyConforms to AAMI SP10, 2002 requirements for Non-Automated Sphyg.
    AAMI/AHA bladder sizesConforms to AAMI/AHA bladder sizes

    2. Sample size used for the test set and the data provenance:

    The provided summary does not explicitly state a specific sample size for the test sets.
    The data provenance is not explicitly mentioned but would be considered retrospective as the testing was done to demonstrate conformance to existing standards and predicate device characteristics, not as an ongoing clinical trial. The testing was conducted by Welch Allyn, Inc. in the USA.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the given document. The study relies on established industry standards (AAMI SP10, ISO 10993-1, AAMI/AHA bladder sizes) and bench/laboratory testing, rather than expert interpretation of results from a specific test set requiring ground truth establishment by human experts.

    4. Adjudication method for the test set:

    An adjudication method is not applicable as the evaluation relies on objective measurements against predetermined standards and predicate device comparisons, not subjective interpretations requiring expert consensus.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device is a blood pressure cuff and does not involve AI or human "readers" in the context of image interpretation or diagnostic assistance. It is a physical medical device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This is not applicable as the device is a physical blood pressure cuff, not an algorithm.

    7. The type of ground truth used:

    The "ground truth" for this device's performance is based on:

    • Established industry standards: AAMI SP10, ISO 10993-1, and AAMI/AHA bladder sizes.
    • Bench and laboratory testing: Physical measurements and evaluations to demonstrate conformance to these standards and equivalence to predicate devices.

    8. The sample size for the training set:

    This is not applicable as there is no mention of a "training set" in the context of this physical medical device.

    9. How the ground truth for the training set was established:

    This is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1