    Device Name :

    ViewFlex™ Xtra ICE Catheter; ViewFlex™ Eco Reprocessed ICE Catheter; Advisor™ HD Grid Mapping Catheter

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ViewFlex™ Xtra ICE Catheter
    The ViewFlex™ Xtra ICE Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac, structures, blood flow and other devices within the heart.

    ViewFlex™ Eco Reprocessed ICE Catheter
    The ViewFlex™ Eco Reprocessed Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac structures, blood flow and other devices within the heart.

    Advisor™ HD Grid Mapping Catheter, Sensor Enabled™
    The Advisor™ HD Grid Mapping Catheter, Sensor Enabled™, is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.

    Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™
    The Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™, is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.

    Agilis™ NxT Steerable Introducer
    The Agilis™ NxT Steerable Introducer is indicated for the introduction of various cardiovascular catheters into the heart, including the left side of the heart, during the treatment of cardiac arrhythmias.

    Agilis™ NxT Steerable Introducer Dual-Reach™
    The Agilis™ NxT Steerable Introducer Dual-Reach™ is indicated for the introduction of various cardiovascular catheters into the heart, including the left side of the heart, during the treatment of cardiac arrhythmias.

    Device Description

    The Agilis™ NxT Steerable Introducer Dual-Reach™ is a sterile, single-use device that con-sists of a dilator and steerable introducer, which is designed to provide flexible catheter positioning in the cardiac anatomy. The inner diameter of the steerable introducer is 13F. The steerable introducer includes a hemostasis valve to minimize blood loss during catheter intro-duction and/or exchange. It has a sideport with three-way stopcock for air or blood aspiration, fluid infusion, blood sampling, and pressure monitoring. The handle is equipped with a rotating collar to deflect the tip clockwise ≥180° and counterclockwise ≥90°. The steerable introducer features distal vent holes to facilitate aspiration and minimize cavitation and a radiopaque tip marker to improve fluoroscopic visualization.

    AI/ML Overview

    This FDA 510(k) clearance letter (K251211) and its accompanying 510(k) summary pertain to a change in workflow for several existing cardiovascular catheters, specifically allowing for a "Zero/Low Fluoroscopy Workflow."

    The key phrase here is "Special 510(k) – Zero/Low Fluoroscopy Workflow". This type of submission is for modifications to a previously cleared device that do not significantly alter its fundamental technology or intended use, but rather introduce a change in how it's used or processed.

    Crucially, this submission does NOT describe a new AI/software device that requires extensive performance testing against acceptance criteria in the manner you've outlined for AI/ML devices. Instead, it's about demonstrating that existing devices, when used with a new, less-fluoroscopy-dependent workflow, remain as safe and effective as before.

    Therefore, many of the questions you've asked regarding acceptance criteria, study details, ground truth, and expert adjudication are not applicable to the information provided in this 510(k) document. The document explicitly states:

    • "Bench-testing was not necessary to validate the Clinical Workflow modifications."
    • "Substantial Equivalence of the subject devices to the predicate devices using the zero/low fluoroscopy workflow has been supported through a summary of clinical data across multiple studies in which investigators used alternative visualization methods."

    This indicates that the "study" proving the device (or rather, the new workflow) meets acceptance criteria is a summary of existing clinical data where alternative visualization methods were already employed, rather than a prospective, controlled study of a new AI algorithm.

    Based on the provided document, here's what can be answered:

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria: The implicit acceptance criterion is that the devices, when used with "zero/low fluoroscopy workflow," maintain substantial equivalence to their predicate devices in terms of safety and effectiveness. This means they must continue to perform as intended for visualizing cardiac structures, blood flow, mapping, or introducing catheters.
    • Reported Device Performance: The document states that "Substantial Equivalence... has been supported through a summary of clinical data across multiple studies in which investigators used alternative visualization methods." This implies that the performance (e.g., adequate visualization, successful mapping, successful catheter introduction) was maintained. Specific quantitative metrics of performance (e.g., accuracy, sensitivity, specificity, or inter-reader agreement for a diagnostic AI) are not provided or applicable here as this is not an AI/ML diagnostic clearance.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified. The document refers to "a summary of clinical data across multiple studies." This suggests an aggregation of results from existing (likely retrospective) patient data where alternative visualization techniques (allowing for "zero/low fluoroscopy") were already utilized clinically. It's not a new, single, prospectively designed test set for an AI algorithm.
    • Data Provenance: Not specified regarding country of origin or specific patient demographics. It is implied to be clinical data collected from studies where these types of procedures were performed using alternative visualization. The data would be retrospective as it's a "summary of clinical data" that already exists.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable in the context of this 510(k). Ground truth in an AI/ML context typically refers to adjudicated labels for images or signals. Here, the "ground truth" is inferred from standard clinical practice and outcomes in the historical data summarized. There's no mention of a specific expert panel for new ground truth establishment for a diagnostic AI.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a study requiring adjudication of diagnostic outputs by multiple readers.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI-assisted diagnostic device. The workflow change is about using alternative non-fluoroscopic imaging modalities (e.g., intracardiac echocardiography, electro-anatomical mapping systems), not about AI improving human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is not an AI algorithm. The predicate devices are physical catheters.

    7. The type of ground truth used:

    • The "ground truth" is inferred from clinical outcomes and established clinical practice using the devices with alternative visualization methods in real-world scenarios. It's not a specific, adjudicated dataset for an AI algorithm. The performance of the devices (such as successful navigation, visualization, and mapping) under the "zero/low fluoroscopy" workflow is assumed to be equivalent to their performance under full fluoroscopy, as demonstrated by prior clinical use where such methods were employed.

    8. The sample size for the training set:

    • Not applicable. There is no AI model being trained discussed in this document.

    9. How the ground truth for the training set was established:

    • Not applicable. No training set for an AI model.

    In summary:

    This 510(k) is for a workflow modification for existing medical devices (catheters), not for an AI/ML diagnostic or assistive software. Therefore, the detailed data performance evaluation typically required for AI models against specific acceptance criteria (as requested in your template) is not presented or relevant in this clearance letter. The "proof" relies on the concept of substantial equivalence to previously cleared predicate devices, supported by a summary of existing clinical data that used alternative visualization methods, implying that the devices function safely and effectively even with reduced fluoroscopy.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231588
    Manufacturer
    Date Cleared
    2023-06-30

    (29 days)

    Product Code
    Regulation Number
    870.1200
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ViewFlex™ Eco Reprocessed ICE Catheter

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ViewFlex™ Eco Reprocessed ICE Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac structures, blood flow and other devices within the heart.

    Device Description

    The ViewFlex™ Eco Reprocessed ICE Catheter is a temporary intracardiac ultrasound catheter intended for use in patients to accurately visualize cardiac structures, blood flow and other devices within the heart when connected to a compatible intracardiac ultrasound console via the compatible ViewFlex™ Catheter Interface Module. Examples of the types of other devices that can be visualized include, and are not limited to, intracardiac catheters, septal occluders, delivery wires, delivery sheaths, sizing balloons and transseptal needles. The use of these images is limited to visualization with no direct or indirect diagnostic use. The ViewFlex™ Eco Reprocessed ICE Catheter has a useable length of 90 cm, with a 9 French (F) shaft with an ultrasound transducer. A 10F introducer is recommended for use with this catheter for insertion into the femoral or jugular veins. The catheter tip has four-directional deflection allowing for Left-Right and Posterior-Anterior deflection, with an angle of at least 120 degrees in each direction.
    The ViewFlex™ Eco Reprocessed ICE Catheter is compatible with the ViewMate™ Z, ViewMate™, ViewMate™ Multi and Philips CX50 ultrasound consoles.
    The ViewFlex™ Eco Reprocessed ICE Catheter is reprocessed by Abbott no more than two (2) times. Each catheter includes marking on the proximal handle and connector that identify the catheter status. The device is taken out of service after reaching the maximum number of reprocessing cycles. Abbott restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided FDA 510(k) summary for the ViewFlex™ Eco Reprocessed ICE Catheter:

    Summary of Acceptance Criteria and Device Performance:

    Acceptance Criteria CategoryReported Device Performance
    BiocompatibilityMet established performance specifications.
    Design ValidationMet established performance specifications.
    Design VerificationMet established performance specifications.
    Cleaning ValidationMet established performance specifications.
    Risk ManagementNo new or modified hazards identified as a result of the proposed modification (additional reprocessing cycle).

    Study Information:

    1. Table of Acceptance Criteria and Reported Device Performance: This information is provided in the table above. The document states that all testing performed met the established performance specifications for each category.

    2. Sample Size Used for the Test Set and Data Provenance:

      • Sample Size: The document does not explicitly state the sample size for the test set used in the non-clinical performance evaluation. It only mentions that "Design verification activities were performed with their respective acceptance criteria."
      • Data Provenance: Not specified, but generally, non-clinical tests like these are conducted in a laboratory setting by the manufacturer (Abbott Medical, USA).
    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

      • Not applicable. This device is a medical catheter and the tests described are primarily engineering and biocompatibility evaluations, not diagnostic image interpretation studies requiring expert clinicians to establish ground truth.
    4. Adjudication Method for the Test Set:

      • Not applicable. The tests are objective measurements against predefined engineering and material specifications, not subjective interpretations requiring adjudication.
    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

      • No. The document explicitly states: "No clinical investigations have been performed for the subject or predicate devices. Clinical data is not required for the demonstration of substantial equivalent based on the risk assessment in Section 10 and verification in Section 12 and validation in Section 13 summarized in this 510(k)." Therefore, no MRMC study was conducted.
    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

      • Not applicable. This is a physical medical device (catheter), not an AI algorithm. Its performance is evaluated through engineering, material, and reprocessing verification and validation, not standalone algorithmic performance.
    7. The Type of Ground Truth Used:

      • The "ground truth" for the non-clinical studies (biocompatibility, design validation/verification, cleaning validation) would be defined by established engineering specifications, material standards, and regulatory guidance documents. For example, for biocompatibility, it would be adherence to ISO 10993 standards and a lack of adverse biological responses. For cleaning validation, it would be a verifiable reduction in contaminants below a specified threshold.
    8. The Sample Size for the Training Set:

      • Not applicable. As this is not an AI/algorithm-based device, there is no "training set" in the context of machine learning. The "development" and "testing" referred to are related to engineering design and physical performance testing.
    9. How the Ground Truth for the Training Set Was Established:

      • Not applicable, as there is no training set for an AI algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230934
    Manufacturer
    Date Cleared
    2023-05-03

    (30 days)

    Product Code
    Regulation Number
    870.1200
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ViewFlex™ Eco Reprocessed ICE Catheter

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ViewFlex™ Eco Reprocessed ICE Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac structures, blood flow and other devices within the heart.

    Device Description

    The ViewFlex™ Eco Reprocessed ICE Catheter is a temporary intracardiac ultrasound catheter intended for use in patients to accurately visualize cardiac structures, blood flow and other devices within the heart when connected to a compatible intracardiac ultrasound console via the compatible ViewFlex™ Catheter Interface Module. Examples of the types of other devices that can be visualized include, and are not limited to, intracardiac catheters, septal occluders, delivery wires, delivery sheaths, sizing balloons and transseptal needles. The use of these images is limited to visualization with no direct or indirect diagnostic use. The ViewFlex™ Eco Reprocessed ICE Catheter has a useable length of 90 cm, with a 9 French (F) shaft with an ultrasound transducer. A 10F introducer is recommended for use with this catheter for insertion into the femoral or jugular veins. The catheter tip has four-directional deflection allowing for Left-Right and Posterior-Anterior deflection, with an angle of at least 120 degrees in each direction. The ViewFlex™ Eco Reprocessed ICE Catheter is compatible with the ViewMate™ Z, ViewMate™, ViewMate™ Multi, and Philips CX50 ultrasound consoles.

    AI/ML Overview

    The provided text does not contain information about an AI/ML-driven device or study results proving a device meets acceptance criteria regarding AI/ML performance. The document is a 510(k) premarket notification for a medical device (ViewFlex™ Eco Reprocessed ICE Catheter) that appears to be entirely hardware-based, with no mention of AI or machine learning components.

    Therefore, I cannot provide the requested information about acceptance criteria and study results for an AI/ML device based on this document. The document primarily focuses on verifying the safety and effectiveness of a reprocessed catheter and its compatibility with an additional ultrasound console, through non-clinical testing (design validation and verification) and risk management.