Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K210712
    Date Cleared
    2021-09-08

    (182 days)

    Product Code
    Regulation Number
    880.5860
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes. The safety shield is intended to reduce the risk of sharp injuries.

    Device Description

    Verifine Mechanical Safety Insulin Syringe is a sterile device consisting of a calibrated barrel with plunger intended to be used to administer an injection of insulin to a patient subcutaneously. It includes an attached needle with a safety mechanism, i.e, after the injection, the protective shield will permanently be locked in place by pushing forward till click, providing protection against needle sticks and rendering the device unusable.

    AI/ML Overview

    This document is a 510(k) summary for the Verifine Mechanical Safety Insulin Syringe. It outlines the device's characteristics, its intended use, and a comparison to a predicate device to demonstrate substantial equivalence.

    Here's the information extracted and structured as requested:

    1. Table of Acceptance Criteria and the Reported Device Performance:

    The document refers to various ISO and ASTM standards, implying that compliance with these standards serves as the acceptance criteria for different aspects of the device's performance. The "reported device performance" is generally stated as "Complied" or "met the USP acceptance criteria," indicating successful adherence to these standards.

    Performance CharacteristicAcceptance Criteria (Standard)Reported Device Performance
    Needle Tubing (Stainless Steel)ISO 9626:2016Bench testing performed verifies performance
    Sterile Hypodermic NeedlesISO 7864:2016Bench testing performed verifies performance
    Sterile Single-Use Syringes for InsulinISO 8537:2016Bench testing performed verifies performance
    Sharps Injury ProtectionISO 23908:2011Bench testing performed verifies performance
    Sharps Injury Prevention FeaturesFDA guidance, Medical Devices with Sharps Injury Prevention FeaturesSimulated clinical use testing conducted
    Biocompatibility (Cytotoxicity)ISO 10993-5:2009No cytotoxicity
    Biocompatibility (Skin Irritation)ISO 10993-10:2010No evidence of skin irritation
    Biocompatibility (Skin Sensitization)ISO 10993-10:2010No evidence of sensitization
    Biocompatibility (Acute Systemic Toxicity)ISO 10993-11:2006No systemic toxicity
    Biocompatibility (Subacute Systemic Toxicity)ISO 10993-11:2017No systemic toxicity
    Biocompatibility (Hemocompatibility)ISO 10993-4:2017No evidence of hemolysis
    Biocompatibility (Pyrogen)USP42-NF37Non pyrogenic
    Particulate MatterUSPMet the USP acceptance criteria
    Sterilization (Ethylene Oxide)ISO 11135:2014Complied with standard
    Package IntegrityASTM D4169-16All packaging deemed acceptable for protection of product and sterility maintenance
    Seal StrengthASTM F88/F88-15Performed
    Dye PenetrationASTM F1929-15Performed
    Sterility TestUSP38-NF33 C71Performed
    Shelf LifeASTM F1980-165 years validated
    Hub/Needle bond strength (≥0.33mm)Implicit: Minimum shearing strength 22NImplied compliance via "Same" to predicate
    Hub/Needle bond strength (
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1