Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K222679
    Date Cleared
    2022-11-07

    (62 days)

    Product Code
    Regulation Number
    870.1210
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Vericor Support Catheter

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vericor Support Catheter is intended for use in small vessel or superselective anatomy for diagnostic and interventional procedures, including peripheral use.

    Device Description

    The Vericor Support Catheter with hydrophilic coating is a braided, kink-resistant catheter designed to facilitate wire guide exchange, wire guide support and to provide a conduit for the delivery of saline solutions or diagnostic contrast agents. The Vericor Support Catheter will be available in three catheter lengths which are 90 cm, 135 cm and 150 cm. The proposed device is a single-lumen catheter. The hub of the catheter incorporates standard luer adapter to facilitate the attachment of accessories. The outer surface of the distal catheter has a hydrophilic coating. The catheter has three radiopaque markers at the distal end to facilitate fluoroscopic visualization. The device is provided in single use and sterile.

    AI/ML Overview

    This document, a 510(k) Summary, describes the Vericor Support Catheter and its substantial equivalence to a predicate device. This is a medical device, not an AI/ML product, so the questions regarding AI/ML clinical studies are not applicable.

    Here's the breakdown of the acceptance criteria and the study (non-clinical) that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) Summary details numerous non-clinical tests. The column "Test Method Summary" indirectly describes the acceptance criteria by outlining what the test aims to demonstrate (e.g., "To demonstrate that the device meets the corrosion resistance"). The "Results and Conclusion" column reports that the device met these criteria.

    TestAcceptance Criteria (Implied from "Test Method Summary")Reported Device Performance
    Visual inspectionNo structural or mechanical damage under x2.5 magnificationThe catheter met acceptance criteria.
    Corrosion ResistanceMeets ISO 10555-1:2013 corrosion resistance requirementsCorrosion resistance met acceptance criteria.
    Dimensional verificationMeets specified dimensionsSize verification met acceptance criteria.
    Flowrate determination testMeets ISO 10555-1:2013 flowrate requirementsThe flowrate of the Vericor catheter size verification met acceptance criteria.
    Compatibility testCan be used as intended with compatible devices in a vascular modelThe device can be used as intended.
    Simulated useCan be used as intended in a vascular modelThe device can be used as intended.
    Power injectionFree of leakage, rupture, or other failure modes during power injectionThe catheters were free of leakage, rupture or other failure modes during power injection test, and the test result met acceptance criteria.
    Torque transmissionMeets torque transmission requirementsTorque transmission met acceptance criteria.
    Air leakageNo air leakage during hub aspirationNo air leakage
    Liquid leakageNo liquid leakage under pressureNo liquid leakage
    Static burst pressureMeets burst pressure requirements (greater than maximum injection pressures)Burst pressure met acceptance criteria.
    Friction of coatingAcceptable change in coating friction before and after useFriction met acceptance criteria
    Coating integrityNo coating defects with magnification after simulated useCoating integrity met acceptance criteria.
    FlexibilityMeets ASTM F2606-2008 flexibility requirementsTip flexibility met acceptance criteria.
    Kink resistanceMeets kink resistance requirements (implied by method)Kink resistance met acceptance criteria
    Peak tensile forceMeets maximum tensile force acceptance criteriaPeak tensile force met acceptance criteria.
    Catheter tip twist to damage testMeets tip twist acceptance criteria (e.g., no deformation/failure under specified rotation)Tip twist met acceptance criteria
    RadiopacityRadiopaque marker on the catheter tip divisible under X-rayThe radiopaque marker on the catheter tip is visible under X - ray.
    Catheter body axial compression testMeets catheter body axial force acceptance criteriaCatheter body axial force met acceptance criteria.
    Delivery and retrieval forces testMeets FDA Guidance delivery and retrieval requirementsDeliver and retrieval force met acceptance criteria.
    Particulate testingQuantity and size of generated particles meet acceptance criteria after simulated useThe number and size of the particles met acceptance criteria.
    Connector performanceMeets ISO 80369-7 requirements for small bore connectorsConnector performance met acceptance criteria.
    CytotoxicityNo cytotoxicityNo cytotoxicity
    Skin SensitizationNo skin sensitizationNo skin sensitization
    Intracutaneous ReactivityNo irritationNo irritation
    Systemic ToxicityNo systemic toxicityNo systemic toxicity
    PyrogenNo pyrogenNo pyrogen
    Complement ActivationNo significant difference to control groupNo significant difference to control group
    In Vivo ThromboresistanceMinimal thrombosisMinimal thrombosis
    HemolysisNo hemolysisNo hemolysis
    Partial Thromboplastin TimeNo significant difference to control groupNo significant difference to control group

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the sample sizes for each non-clinical test. It generally states that "Non clinical tests were conducted to verify that the proposed device met all design specifications." The data provenance is not applicable in terms of country of origin or retrospective/prospective for these non-clinical, laboratory-based tests. The tests were performed on the Vericor Support Catheter itself.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This is not applicable as the studies are non-clinical, laboratory-based performance tests, not studies involving human interpretation or clinical data. Ground truth in this context is established by adherence to specified standards and measurement techniques.

    4. Adjudication Method for the Test Set

    Not applicable. The tests are based on objective measurements and adherence to established standards, not expert adjudication of subjective interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    Not applicable. This is a conventional medical device (support catheter), not an AI/ML diagnostic or assistive device.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    Not applicable. This device does not involve an algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For these non-clinical tests, the "ground truth" is defined by the objective criteria within the referenced international and national standards (e.g., ISO 10993, ISO 10555-1, ASTM standards, USP). Performance is measured against these established technical specifications and benchmarks.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI/ML system that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This device does not involve a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1