VeriSight Intracardiac Echocardiography (ICE) Catheter, VeriSight Pro ICE Catheter
The VeriSight VeriSight Pro ICE Catheter is intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
The Intracardiac Echocardiography (ICE) System comprises of the VeriSight Pro ICE Catheters, the ICE PIM, and the EPIQ Diagnostic Ultrasound System (cleared under K202216).
The VeriSight/VeriSight Pro Intracardiac Echocardiography (ICE) catheters are sterile, disposable, and for single use only. The catheter's distal end has an ultrasound transducer providing 2D and/or 3D imaging capabilities. A handle, located at the proximal end of the catheter, has two steering wheels that can be manually operated to control four-way articulation of the distal segment in anterior, posterior, left and right directions. The catheter has a 9 French (F) shaft and a usable length of 90 cm.
The VeriSight and VeriSight Pro ICE Catheters are identical in all regards (material, processing, assembly and packaging). The VeriSight ICE catheter provides 2D ultrasound imaging capabilities. The VeriSight Pro ICE catheter provides 2D and/or 3D ultrasound imaging capabilities, depending on the model and configuration of the EPIQ ultrasound system it connects to. The catheters are compatible with ancillary equipment such as sheaths and introducers. The catheters are sterilized via Ethylene Oxide.
The catheters connect to the Philips EPIQ Diagnostic Ultrasound System via a patient interface module (PIM); the connection between the catheter and the PIM is located outside the sterile field. The catheters will not operate if connected to any other imaging system. These catheters are for exclusive use with Philips EPIQ 7C, CVx, and CVxi series of ultrasound systems, cleared under K202216.
The provided document, a 510(k) Summary for the Philips VeriSight Intracardiac Echocardiography (ICE) Catheter, indicates that clinical data was not required to establish substantial equivalence for this device. Therefore, there is no study detailing acceptance criteria or device performance in a clinical setting for this submission.
The manufacturer states: "The VeriSight/VeriSight Pro ICE catheters use the same scientific technology, operating principles and shares similar indications for use as the predicate device. Therefore, clinical data is not required to establish substantial equivalence." (Page 8).
Instead, the submission relies on the substantial equivalence to a predicate device (Siemens AcuNav Diagnostic Ultrasound Catheter 8F, 10F, cleared under K170263) and extensive non-clinical performance testing to demonstrate safety and effectiveness.
Here's a breakdown of the non-clinical testing performed:
1. Table of Acceptance Criteria and Reported Device Performance:
Since a clinical study was not conducted to establish performance metrics like sensitivity, specificity, or accuracy against an established ground truth, there is no direct table of acceptance criteria and reported device performance in the context of clinical outcomes. The "performance" in this context refers to the successful completion of engineering design verification and validation tests.
Test Category | Specific Test | Acceptance Criteria Implicit (e.g., meeting design specs, established safety limits) | Reported Device Performance (Successful completion) |
---|---|---|---|
Mechanical/Physical | Visual inspection | Device meets visual quality standards | Performed satisfactorily |
Dimensional | Device dimensions meet specifications | Performed satisfactorily | |
Actuation characteristics | Catheter steering and control meet specifications | Performed satisfactorily | |
Tip deflection | Tip deflection within design limits | Performed satisfactorily | |
Torsional shaft | Shaft torque resistance meets specifications | Performed satisfactorily | |
Tensile tests | Device structural integrity under tension meets specifications | Performed satisfactorily | |
Deliverability | Catheter navigability and deployment meet specifications | Performed satisfactorily | |
Imaging | Saline imaging | Demonstrates clear imaging in a simulated environment | Performed satisfactorily |
Safety | Particulate | Particulate matter within acceptable limits | Performed satisfactorily |
Biocompatibility | Meets biocompatibility standards for patient contact | Performed satisfactorily | |
Sterilization | Achieves and maintains sterility | Performed satisfactorily | |
Packaging and Shelf Life | Maintains integrity and sterility over shelf life | Performed satisfactorily | |
System Integration | Usability | Device is user-friendly and functions as intended | Performed satisfactorily |
Preclinical Animal testing | Demonstrates safe and effective function in an animal model | Performed satisfactorily | |
System testing with EPIQ System | Fully functional and compatible with the EPIQ System | Performed satisfactorily | |
Electrical Safety, EMC and acoustic output testing with EPIQ System | Complies with relevant electrical safety, EMC, and acoustic output standards (e.g., EN/IEC 60601-1, EN/IEC 60601-1-2, EN/IEC 60601-2-37, IEC 62359) | Performed satisfactorily |
2. Sample Size Used for the Test Set and Data Provenance:
Not applicable, as no clinical test set was used for the purpose of demonstrating device performance against clinical acceptance criteria. The testing listed above refers to engineering and bench testing, for which specific sample sizes are not detailed in this summary but are generally conducted on a representative number of devices.
3. Number of Experts Used to Establish Ground Truth and Qualifications:
Not applicable, as no ground truth was established by experts in a clinical context for this submission. The "ground truth" for the non-clinical tests would have been the established engineering specifications and regulatory standards.
4. Adjudication Method for the Test Set:
Not applicable, as no clinical test set requiring adjudication was used.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:
No, an MRMC comparative effectiveness study was not done. The submission explicitly states that clinical data was not required.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:
Not applicable. The device is a physical intracardiac echocardiography catheter, not an AI algorithm. Its function is to provide imaging capabilities for human interpretation.
7. The Type of Ground Truth Used:
For the non-clinical performance testing, the "ground truth" was defined by:
- Engineering design specifications and performance requirements for the device.
- Established national and international voluntary standards for medical devices, electrical safety, electromagnetic compatibility (EMC), and diagnostic ultrasound safety (e.g., EN/IEC 60601-1, EN/IEC 60601-1-2, EN/IEC 60601-2-37, IEC 62359).
- Results from preclinical animal testing for safety and function.
8. The Sample Size for the Training Set:
Not applicable, as this device does not involve a machine learning or AI algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as there is no training set for this device.