Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K234106
    Device Name
    Venue Fit
    Date Cleared
    2024-06-10

    (167 days)

    Product Code
    Regulation Number
    892.1550
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Venue Fit is a general purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals or practitioners that are legally authorized or licensed by law in the country, state or other local municipality in which he or she practices, for ultrasound imaging, measurement, display and analysis of the human body and fluid. The users may or may not be working under supervision or authority of a physician. Users may also include medical students working under the supervision or authority of a physician during their education / training. Venue Fit is intended to be used in a hospital or medical clinic. Venue Fit clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), Vascular/Peripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transrectal, Transvaginal, Transesophageal, Intraoperative (vascular) and interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, CW Doppler, Color M Doppler, Color M Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD, B/CWD. B/Color/CWD.

    Device Description

    The Venue Fit is a general purpose diagnostic ultrasound system. It is a portable system with a touch screen interface. It can be powered through an electrical wall outlet or an internal battery. It utilizes linear, convex, and phased array transducers and supports standard acquisition modes. Compatible biopsy kits can be used for needle-guidance procedures. The system can display the patient's ECG trace synchronized to the scanned image. A barcode reader and RFID scanner are available as additional input devices. A roller bag is available for transport. It is capable of wired or wireless internet connection and meets DICOM requirements for image storage and archiving.

    AI/ML Overview

    The provided text describes specific features and functionalities of the GE Venue Fit ultrasound system but does not contain information about acceptance criteria, reported device performance, or a specific study proving it meets acceptance criteria as typically outlined for AI/CADe devices.

    The document is a 510(k) Premarket Notification Submission for the Venue Fit ultrasound system. It primarily focuses on demonstrating substantial equivalence to a predicate device (K220848 Venue Fit) regarding general imaging capabilities, technological characteristics, safety, and effectiveness.

    Here's what can be extracted and what is explicitly stated as not applicable:

    1. A table of acceptance criteria and the reported device performance
    Not provided in the document. The document focuses on performance testing in the context of safety and effectiveness, rather than specific diagnostic accuracy metrics.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    Not applicable. The document states, "The subject of this premarket submission, Venue Fit, did not require clinical studies to support substantial equivalence." Therefore, there is no test set in the context of a clinical study for diagnostic performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    Not applicable, as no clinical studies were required.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not applicable, as no clinical studies were required.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This document does not describe an AI/CADe device that assists human readers, nor does it present results from an MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. The Venue Fit is a diagnostic ultrasound system, not a standalone algorithm. Its capabilities, like "Auto Volume Flow (AVF)" or "Bladder Volume Tool", are functionalities of the ultrasound system itself, not standalone AI algorithms with reported diagnostic performance metrics.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    Not applicable, as no clinical studies requiring ground truth establishment were mentioned.

    8. The sample size for the training set
    Not applicable. The document does not describe a machine learning algorithm that would have a training set. The "Venue Fit" is a general-purpose diagnostic ultrasound system. While it has "semi-automated tools," the document does not detail their development or any AI-specific training sets.

    9. How the ground truth for the training set was established
    Not applicable, as there is no mention of a training set for an AI algorithm.


    Summary of what is present:

    The document establishes substantial equivalence through a comparison to a predicate device (K220848 Venue Fit) and a list of reference devices (K231301 Vscan Air, K161047 LOGIQ P9 and LOGIQ P7, K231989 LOGIQ E10s/LOGIQ Fortis, K180374 Voluson S8/ S10/ S10 Expert, K202035 Vscan Air).

    It outlines several non-clinical tests performed to ensure safety and compliance with standards:

    • Acoustic output evaluation
    • Biocompatibility
    • Cleaning and disinfection effectiveness
    • Thermal, electrical, electromagnetic, and mechanical safety

    The system states compliance with various voluntary standards including:

    • Marketing Clearance of Diagnostic Ultrasound Systems and Transducers - Guidance for Industry and Food and Drug Administration Staff (February 21, 2023)
    • AAMI/ANSI ES60601-1 (Medical Electrical Equipment - Part 1: General Requirements for Safety)
    • IEC 60601-1-2 (Electromagnetic Compatibility Requirements and Tests)
    • IEC 60601-2-37 (Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment)
    • IEC 62359 (Ultrasonics Field characterization Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields)
    • ISO 10993-1 (Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing Within a Risk Management Process)
    • ISO 14971 (Application of risk management to medical devices)
    • NEMA PS 3.1 3.20e (Digital Imaging and Communications in Medicine (DICOM) Set)
    • AAMI TIR69 (Technical Information Report Risk management of radio-frequency wireless coexistence for medical devices and systems)

    Quality assurance measures applied during development include:

    • Risk Analysis
    • Requirements Reviews
    • Design Reviews
    • Testing on unit level (Module verification)
    • Integration testing (System verification)
    • Performance testing (Verification & Validation)
    • Safety testing (Verification)

    The conclusion states that based on equipment design similarities, conformance to recognized performance standards, and performance testing, the proposed Venue Fit is considered substantially equivalent in safety, effectiveness, and performance to the predicate device.

    Crucially, the document explicitly states: "The subject of this premarket submission, Venue Fit, did not require clinical studies to support substantial equivalence." This means that the information requested regarding acceptance criteria, device performance, sample sizes, ground truth, experts, adjudication, and MRMC/standalone studies in a diagnostic performance context is not available within this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220848
    Device Name
    Venue Fit
    Manufacturer
    Date Cleared
    2022-06-27

    (96 days)

    Product Code
    Regulation Number
    892.1550
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Venue Fit is a general purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Venue Fit is intended to be used in a hospital or medical clinical applications include: abdominal (GYN and Urology), thoracic pleural, ophthalmic. Fetal/OB. Small Organ (including breast, testes, thyroid), VascularPeripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatic), Transrectal, Transvaginal, Transesophageal, Intraoperative (vascular) and interventional guidance (includes tissue biopsy, fluid dramage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: BM, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD, B/CWD, B/Color/CWD.

    Device Description

    Venue Fit is a general-purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals to evaluate the body by ultrasound imaging and fluid flow analvsis. The Venue Fit is a compact, portable system with a small footprint. The system can be hand carried using the integrated handle, placed on a horizontal surface (if kickstand is attached), attached to a mobile cart or mounted on the wall. It has a high resolution color LCD monitor, with a simple, multi-touch user interface that makes the system intuitive. The system can be powered through an electrical wall outlet for long term use or from an internal battery for a short time with full functionality and scanning. The Venue Fit utilizes a variety of linear, convex, and phased array transducers which provide high imaging capability, supporting all standard acquisition modes. Compatible biopsy kits can be used for needle-guidance procedures. The system is capable of displaying the patient's ECG trace synchronized to the scanned image. This allows the user to view an image from a specific time of the ECG signal which is used as an input for gating during scanning. The ECG signal can be input directly from the patient or as an output from an ECG monitoring device. ECG information is not intended for monitoring or diagnosis. A barcode reader and RFID scanner are available as additional input devices. A roller bag will also be available for the customer to use when transporting the system. Venue Fit is capable of wired or wireless internet connection. The system meets DICOM requirements to support users image storage and archiving needs (local PACS or products such as Q-Path) and allows for output to printing devices. The user documentation is available electronically.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance (cNerve feature)

    Acceptance Criteria (cNerve)Reported Device Performance (cNerve)
    Sequence Accuracy Requirement:Achieved: Not explicitly stated as a percentage in the document, but the text states: "Success criteria were based on conformance of the cNerve detections to ground truth annotations of nerve bundles in individual frames. Since the intended use is nerve tracking during scouting rather than nerve segmentation accuracy, success criteria were derived via a preliminary survey. The target of the survey was to identify thresholds for pixel accuracy in frames and for frame accuracy in sequences that are appropriate for the intended use."
    At least 70% of the sequences are meaningfully detected.Achieved: Not explicitly stated as a percentage in the document. The study's focus was on meeting the success criteria derived from the preliminary survey which aimed to define "meaningfully detected" and "successfully detected" based on pixel and frame accuracy thresholds.
    At least 80% of the meaningfully detected sequences are successfully detected (meeting frame accuracy criteria).Achieved: Not explicitly stated as a percentage in the document. The study's focus was on meeting the success criteria derived from the preliminary survey which aimed to define "meaningfully detected" and "successfully detected" based on pixel and frame accuracy thresholds.

    Note: The document does not provide specific performance percentages against the 70% and 80% thresholds. It states that success criteria were based on conformance to ground truth and that preliminary surveys identified appropriate thresholds for pixel and frame accuracy for the intended use.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • Sequences: 124 sequences
      • Frames: 3776 frames
      • Individuals: 44 individuals
    • Data Provenance: USA, Japan, Israel (Retrospective, as it was a pre-existing dataset used for verification)

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: A single clinical expert.
    • Qualifications of Experts: "clinical expert" - further specific qualifications (e.g., years of experience, specialty) are not provided in the document.

    4. Adjudication Method for the Test Set

    • Adjudication Method: None explicitly stated. Ground truth annotations were performed by a single clinical expert.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done to evaluate human reader improvement with AI assistance. The testing focused on the standalone performance of the cNerve feature.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, a standalone study was performed. The "AI Summary of Testing" directly addresses the performance of the cNerve algorithm against ground truth annotations.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: Expert consensus (from a single clinical expert) on anatomical areas of nerve bundles within individual frames.

    8. The Sample Size for the Training Set

    • The document does not specify the sample size for the training set. It only states that the data used for verification is "completely distinct" from the training data.

    9. How the Ground Truth for the Training Set was Established

    • The document does not specify how the ground truth for the training set was established. It only ensures that the verification data was distinct from the training data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K203137
    Device Name
    Venue Fit
    Date Cleared
    2020-12-29

    (70 days)

    Product Code
    Regulation Number
    892.1550
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Venue Fit is a general purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Venue Fit is intended to be used in a hospital or medical clinic. Venue Fit clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, thyroid), VascularPeripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transrectal, Transvaginal, Transesophageal, Intraoperative (vascular) and interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD, B/CWD, B/Color/CWD.

    Device Description

    Venue Fit is a general-purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals to evaluate the body by ultrasound imaging and fluid flow analysis. The Venue Fit is a compact, portable system with a small footprint. The system can be hand carried using the integrated handle, placed on a horizontal surface (if kickstand is attached), attached to a mobile cart or mounted on the wall. It has a high resolution color LCD monitor, with a simple, multi-touch user interface that makes the system intuitive. The system can be powered through an electrical wall outlet for long term use or from an internal battery for a short time with full functionality and scanning.

    The Venue Fit utilizes a variety of linear, convex, and phased array transducers which provide high imaging capability. supporting all standard acquisition modes. Some transducers are compatible with OEM biopsy kits to support needle-guidance procedures.

    The system has the capability for displaying the patient's ECG trace synchronized to the scanned image. This allows the user to view an image from a specific time of the ECG signal. The ECG signal can be input directly from the patient leads or as an output from an ECG monitoring device. ECG information is not intended for monitoring or diagnosis.

    Venue Fit is capable of wired or wireless internet connection and a barcode reader is available to be used as an input device. The system meets DICOM requirements to support users image storage and archiving needs (local PACS or products such as O-Path) and allows for output to printing devices. The user documentation is available electronically. An additional accessory available for the customer is a roller bag.

    AI/ML Overview

    I am sorry, but the provided text does not contain the information required to answer your request. The document is a 510(k) premarket notification submission for a medical device (Venue Fit ultrasound system).

    It primarily focuses on:

    • Regulatory information: FDA letter, indications for use, and a 510(k) summary.
    • Device description: Explaining the Venue Fit's features, intended use, and technical specifications.
    • Comparison to predicate devices: Highlighting similarities and differences with previously cleared GE ultrasound systems.
    • Non-clinical tests: Listing the safety and performance standards the device complies with (e.g., acoustic output, biocompatibility, electrical safety).
    • A statement explicitly saying: "The subject of this premarket submission, Venue Fit, did not require clinical studies to support substantial equivalence."

    Therefore, there is no information regarding:

    • Acceptance criteria for AI/algorithm performance.
    • Results of a study proving the device meets acceptance criteria (as no clinical studies were required).
    • Sample sizes for test or training sets.
    • Data provenance.
    • Number of experts, their qualifications, or adjudication methods for ground truth.
    • MRMC studies or effect sizes.
    • Standalone algorithm performance.
    • Type of ground truth used.
    • How ground truth for training was established.

    This document indicates that the device's substantial equivalence was demonstrated through non-clinical testing and comparison to predicate devices, rather than a clinical study involving AI performance metrics.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1