Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    , V-PRO maX 2 Low Temperature Sterilization System, V-PRO s2 Low Temperature Sterilization System, V-PRO
    60 Low Temperature Sterilization System, RAS 12 Rack, RAS 12 Long Rack, RAS Cycle of the AMSCO 7052HP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AMSCO 600 Steam Sterilizers are designed for sterilization of heat and moisture- stable materials used in healthcare facilities and are equipped with the following factory-programmed cycles.
    The V-PRO maX 2 Low Temperature Sterilization System using VAPROX HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and temperature, suitable for processing medical devices without leaving toxic residues.
    The V-PRO s2 Low Temperature Sterilization System using VAPROX® HC Sterilant are intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and low temperature, suitable for processing medical devices without leaving toxic residues.
    The RAS Racks are used in the RAS Cycle of the AMSCO 7052HP Single-Chamber Washer/Disinfector and the AMSCO 7053HP Single-Chamber Washer/Disinfector for the effective cleaning, rinsing, intermediate level disinfection and drying of reusable da Vinci® X/Xi and S/Si EndoWrist® instruments.
    The V-PRO® 60 Low Temperature Sterilization System using VAPROX® HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and low temperature, suitable for processing medical devices without leaving toxic residues.

    Device Description

    The AMSCO 600 Steam Sterilizer uses saturated steam, generated from a house steam utility (e.g. boiler system) or from a steam generator, to sterilize heat-stable health care products.
    The V-PRO Low Temperature Sterilization System is a vaporized hydrogen peroxide sterilizer. It has the following pre-programmed cycles (the Lumen Cycle, the Non Lumen Cycle, the Flexible Cycle and the Fast Non Lumen Cycle). The V-PRO Low Temperature Sterilization System is intended for terminal sterilization of cleaned, rinsed, dried and packaged reusable surgical instruments used in healthcare facilities.
    The V-PRO s2 Sterilizer executes four sterilization cycles (the Lumen, Non Lumen, Flexible Cycles and Fast Cycle). The V-PRO s2 Sterilizer contains a cabinetry modification and is free-standing.
    The RAS Racks are designed to enable the mechanical cleaning, rinsing, and disinfection of up to twelve (12) robotic-assisted surgery instruments in a compatible washer-disinfector.
    The V-PRO s2 Sterilizer executes four sterilization cycles (the Lumen, Non Lumen, Flexible Cycles and Fast Cycle). The V-PRO s2 Sterilizer contains a cabinetry modification and is free-standing.

    AI/ML Overview

    The provided FDA 510(k) summary describes several sterilization and washer-disinfector systems. The acceptance criteria and testing detailed are for modifications related to component obsolescence (oscillator replacement, coin cell battery to super capacitor, flash memory component alternates and associated software updates). The document does not describe a study related to AI or human-in-the-loop performance. Therefore, questions related to expert consensus, MRMC studies, effect size, and standalone algorithm performance are not applicable.

    Here's an analysis of the acceptance criteria and supporting studies for the component modifications across the various devices:

    1. Table of Acceptance Criteria and Reported Device Performance

    Across all devices (AMSCO 600 Steam Sterilizer, V-PRO maX 2, V-PRO s2, V-PRO 60 Low Temperature Sterilization Systems, and RAS Racks/Cycle), the acceptance criteria and performance for the specific modifications are consistent:

    TestAcceptance CriterionReported Device Performance
    ½ Cycle sterility assurance Test (for Sterilizers)All biological indicators must show no growth after a ½ Cycle exposure with a worst-case load.Pass
    Critical parameters test (for RAS Racks/Cycle)Compare the cycle data of RAS cycle between modified and new Kodiak controllers to confirm the cleaning efficiency.Pass
    Software testVerify sterilizer can set and read RTC, bootloader can load and run software application and verify proper operation of applications using serial flash.Pass
    Software confirmation test (for RAS Racks/Cycle)Ensure proper version, proper parameters are in the cycles for proper operation of cycles.Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes (e.g., number of cycles, number of biological indicators) used for these non-clinical performance tests. It refers to "worst-case load" for the sterility assurance tests.

    The provenance of the data is non-clinical performance testing conducted by the manufacturer, STERIS Corporation. The country of origin is implicitly the United States, given STERIS's address in Mentor, Ohio, and the FDA submission. The studies are retrospective in the sense that they are proving the modified device performs equivalently to the original validated design.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This question is not applicable as the studies described are non-clinical performance tests verifying device functionality and sterility/cleaning efficacy, not diagnostic or interpretive tasks requiring human expert ground truth. The "ground truth" for these tests is defined by established sterilization and cleaning standards (e.g., "no growth" for biological indicators, confirmed cleaning efficiency).

    4. Adjudication Method for the Test Set

    This question is not applicable for non-clinical performance tests. The results (e.g., growth/no growth, successful software operation) are typically objective and determined by laboratory analysis or automated system checks.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states that the modifications "has no impact on the device performance" compared to the predicate device because the changes are related to component obsolescence, not a change in the fundamental operating principle or intended use that would require a comparison of human reader effectiveness.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. The devices are sterilization and cleaning equipment, not algorithms that perform diagnostic or interpretive functions. The software tests performed are focused on verifying the proper functioning of the embedded system software controlling the sterilizers/washer-disinfectors, not standalone interpretive performance.

    7. The Type of Ground Truth Used

    The ground truth used for these non-clinical tests is based on:

    • Biological indicator results: "No growth" post-sterilization, indicating successful killing of microorganisms. This is a direct measure of sterility.
    • Physical and chemical parameters verification: Ensuring cycle data (time, temperature, pressure, sterilant injection weight) for sterilizers and cleaning/thermal disinfection parameters for the washer-disinfector (temperature, time, chemical dosing, pump pressure) meet pre-defined specifications.
    • Software functionality: Verification that the embedded software correctly sets and reads Real-Time Clock (RTC), loads and runs applications, and ensures proper operation using serial flash memory.
    • Cleaning efficiency: For the RAS Racks/Cycle, comparison of cycle data between modified and original controllers to confirm equivalent cleaning efficiency.

    These "ground truths" are derived from validated operational parameters and expected biological/physical outcomes inherent to sterilization and cleaning processes, as per industry standards and internal validation protocols.

    8. The Sample Size for the Training Set

    This question is not applicable. These are hardware/software modifications to existing medical devices, not AI/ML algorithms that require a separate training set. The "associated software updates" are to accommodate new hardware components (flash memory, super capacitor) and ensure the existing validated sterilization/cleaning cycles continue to operate as intended.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable, as there is no "training set" in the context of these device modifications.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182568
    Manufacturer
    Date Cleared
    2019-01-03

    (107 days)

    Product Code
    Regulation Number
    880.6860
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    V-PRO s2 Low Temperature Sterilization System, V-PRO 60 Low Temperature Sterilization System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V-PRO® s2 Low Temperature Sterilization System using VAPROX® HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and low temperature, suitable for processing medical devices without leaving toxic residues.

    Each Cycle can sterilize non-luments and instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors.

    The V-PRO s2 Sterilizer Non Lumen Cycle can sterilize: † Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes.

    The V-PRO s2 Sterilizer Fast Cycle can sterilize: * Medical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations:

    • Single or dual channeled devices with stainless steel lumens that are ≥ 0.77 mm (~1/32”) internal diameter (ID) and ≤ 410 mm (~16-9/64”) in length
    • Triple channeled devices with stainless steel lumens that are either: ≥ 1.2 mm (~3/64”) ID and ≤ 275 mm (~10-53/64”) in length ≥ 1.8 mm (~5/64”) ID and ≤ 310 mm (~12-13/64”) in length or ≥ 2.8 mm (~7/64”) ID and ≤ 317 mm (12-31/64”) in length
      Non-lumened instruments including non-lumened general medical instruments. non-lumened rigid, semi-rigid and flexible endoscopes.

    The V-PRO s2 Sterilizer Flexible Cycle can sterilize: @ One surgical flexible endoscope (such as those used in ENT, Urology and Surgical Care) or bronchoscope with light cord (if not integral to endoscope), mat, and additional load.

    • The flexible endoscope may be a single or dual lumen device with lumens that are ≥ 1 mm (~3/64”) ID and ≤ 990 mm (38-63/64”) in length.
    • Additional load, up to 11 lb (5 kg) can include stainless steel lumens with the following dimensions ≥ 2 mm (~5/64”) ID and ≤ 400 mm (~15 3/4”) in length ≥ 0.76 mm (~1/32”) ID and ≤ 233 mm (~9 11/64”) in length ≥ 1.0 mm (~3/64”) ID and ≤ 254 mm (~10”) in length

    The V-PRO s2 Sterilizer Lumen Cycle can sterilize: ^ Medical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations:

    • Single or dual channeled devices with stainless steel lumens that are ≥ 0.77 mm (~1/32”) internal diameter (ID) and ≤ 410 mm (16-9/64”) in length
    • Triple channeled devices with stainless steel lumens that are either: ≥ 1.2 mm (~3/64") ID and ≤ 275 mm (~10-53/64") in length ≥ 1.8 mm (~5/64") ID and ≤ 310 mm (~12-13/64") in length or ≥ 2.8 mm (~7/64") ID and ≤ 317 mm (12-31/64") in length

    The V-PRO® s 60 Low Temperature Sterilization System using VAPROX® HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and low temperature, suitable for processing medical devices without leaving toxic residues.

    Each Cycle can sterilize non-luments and instruments with diffusionrestricted spaces such as the hinged portion of forceps and scissors.

    The V-PRO 60 Sterilizer Non Lumen Cycle can sterilize: t Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes.

    The V-PRO 60 Sterilizer Flexible Cycle can sterilize: @ One surgical flexible endoscope (such as those used in ENT, Urology and Surgical Care) or bronchoscope with light cord (if not integral to endoscope), mat, and additional load.

    • The flexible endoscope may be a single or dual lumen device with lumens that are ≥ 1 mm (~3/64”) ID and ≤ 990 mm (38-63/64”) in length.
    • Additional load, up to 11 lb (5 kg) can include stainless steel lumens with the following dimensions ≥ 2 mm (~5/64”) ID and ≤ 400 mm (~15 3/4”) in length ≥ 0.76 mm (~1/32”) ID and ≤ 233 mm (~9 11/64”) in length ≥ 1.0 mm (~3/64”) ID and ≤ 254 mm (~10”) in length

    The V-PRO 60 Sterilizer Lumen Cycle can sterilize: ^ Medical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations:

    • Single or dual channeled devices with stainless steel lumens that are ≥ 0.77 mm (~1/32”) internal diameter (ID) and ≤ 410 mm (16-9/64”) in length
    • Triple channeled devices with stainless steel lumens that are either: ≥ 1.2 mm (~3/64") ID and ≤ 275 mm (~10-53/64") in length ≥ 1.8 mm (~5/64") ID and ≤ 310 mm (~12-13/64") in length or ≥ 2.8 mm (~7/64") ID and ≤ 317 mm (12-31/64") in length
    Device Description

    The V-PRO s2 Low Temperature Sterilization System is a new vaporized hydrogen peroxide sterilizer model to be added to the STERIS V-PRO family of sterilizers. The V-PRO product line currently consists of the V-PRO 1, V-PRO 1 Plus, V-PRO maX, V-PRO maX 2 and V-PRO 60 Sterilizers.

    The V-PRO s2 Sterilizer contains the same three sterilization cycles as the predicate device (K172319), the Lumen, Non Lumen and Flexible Cycles, with a new cycle, the Fast Cycle. In addition, compared to the claimed predicate device, the proposed device contains a cabinetry modification, adds additional claims for devices that may be processed in the Flexible Cycle, and increases the maximum claimed load weight for the Non Lumen Cycle.

    The V-PRO 60 Low Temperature Sterilization System is a vaporized hydrogen peroxide sterilizer model that currently exists in the STERIS V-PRO family of sterilizers. The current proposed device claims are intended to modify the indications for use statement.

    As with the predicate device (K172319), the V-PRO 60 Sterilizer has three preprogrammed cycles: the Lumen Cycle, the Non Lumen Cycle and the Flexible Cycle, none of which are being altered. The V-PRO 60 Low Temperature Sterilization System is intended for terminal sterilization of cleaned, rinsed, dried and packaged surgical instruments used in healthcare facilities.

    The V-PRO s2 and V-PRO 60 Low Temperature Sterilization Systems are intended for terminal sterilization of cleaned, rinsed, dried and packaged surgical instruments used in healthcare facilities and utilizes VAPROX® HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). The pre-programmed cycles all utilize a conditioning phase, a sterilize phase and an aeration phase. The packaged sterilized devices are ready for use at the completion of the cycle, no cool down or aeration period is required following completion of the cycle.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA, detailing the substantial equivalence of the STERIS V-PRO s2 Low Temperature Sterilization System and modifications to the V-PRO 60 Low Temperature Sterilization System to a legally marketed predicate device. This document describes the acceptance criteria and the study that proves a medical device (sterilizer) meets these criteria, rather than an AI/ML powered device.

    Therefore, the requested information points related to AI/ML device studies (sample size for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, type of ground truth, etc.) do not apply to this document as it pertains to a physical sterilization device and not a diagnostic AI/ML system.

    However, I can extract the acceptance criteria for the sterilization device and the reported performance based on the provided text.


    Acceptance Criteria and Reported Device Performance (for a Sterilization Device)

    The "acceptance criteria" for a sterilizer device are typically demonstrated through various validation studies to ensure the device effectively sterilizes medical instruments. The "reported device performance" is the conclusion that these studies meet the established criteria.

    Here's a table based on the "Summary of Nonclinical Tests" section (Page 17-18) which outlines the tests conducted (functioning as acceptance criteria) and their successful results (reported performance):

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Test Name)Reported Device Performance (Result and Conclusion)
    Determination of D-value and Total Kill EndpointResult: Greater than a 12 log reduction of the most resistant organism is achieved within Fast, Non Lumen and Flexible Cycles of the V-PRO s2 Sterilizer.
    Conclusion: PASS
    ½ Cycle Modified Total Kill Endpoint VerificationResult: Modified total kill end point analysis was demonstrated for Non Lumen and Flexible V-PRO s2 and V-PRO 60 Sterilizer cycles and for the Fast V-PRO s2 Sterilizer cycle. The standard injection weight of 1.1 g and at least one lower injection weight resulted in all sterile results for all cycles.
    Conclusion: PASS
    ½ Cycle Sterilization Verification of SurfacesResult: Verified surface sterilization efficacy for the most resistant material for VHP within a worst-case validation load in the V-PRO s2 Sterilizer Fast ½ Cycle and the V-PRO 60 Non Lumen Cycle.
    Conclusion: PASS
    ½ Cycle Verification of Mated SurfacesResult: Sterile efficacy was demonstrated for mated surfaces under worst case conditions in the Fast and Non Lumen V-PRO s2 Sterilizer cycles.
    Conclusion: PASS
    Simulated Use TestResult: Simulated use testing verified the ability of the Sterilizer cycles to sterilize medical devices under worst case processing conditions in the V-PRO s2 and V-PRO 60 Non Lumen and Flexible Cycles and in the V-PRO s2 Fast Cycle.
    Conclusion: PASS
    In Use TestResult: The in use investigation demonstrated the ability of the V-PRO s2 and V-PRO 60 Sterilizer Flexible Cycle and the V-PRO s2 Fast Cycle to sterilize patient-soiled, clinically-cleaned, medical instruments.
    Conclusion: PASS
    BiocompatibilityResult: Cytotoxicity and residue analysis of worst case materials have demonstrated biocompatibility after processing in the V-PRO s2 Sterilizer's Fast Cycle.
    Conclusion: PASS
    Final Process QualificationResult: The V-PRO s2 Sterilizer final process qualification was successful for the Fast Cycle.
    Conclusion: PASS
    Compliance with Safety Standards (IEC 61010-1:2010)Conclusion: Certified
    Compliance with Safety Standards (IEC 61010-2-040: 2015)Conclusion: Certified
    Compliance with Safety Standards (IEC 61326-1:2006/2012 (EMC))Conclusion: Certified

    Regarding the AI/ML-specific questions, as stated previously, this document does not contain information on an AI/ML powered device. Therefore, the following points are not applicable:

    1. Sample size used for the test set and the data provenance: Not applicable. The document discusses "validation studies" and "validation loads" for a sterilizer, which are physical materials and instruments, not data samples for an AI test set.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a sterilizer is established through scientific validation (e.g., microbial kill, chemical indicators), not human expert review.
    3. Adjudication method for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used: For a sterilizer, the ground truth is the demonstrable absence of viable microorganisms ("sterility") and material compatibility, established through laboratory methods (e.g., biological indicators, chemical indicators, and material analysis) under worst-case conditions, as opposed to expert consensus, pathology, or outcomes data in the context of diagnostic AI.
    7. The sample size for the training set: Not applicable. This refers to a physical device, not an AI model.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    V-PRO 1 Plus Low Temperature Sterilization System, V-PRO maX Low Temperature Sterilization System, V-PRO
    60 Low Temperature Sterilization System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V-PRO 1, V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems using VAPROX HC Sterilant are intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and temperature, suitable for processing medical devices without leaving toxic residues.

    Each Cycle can sterilize non-luments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. Only stainless steel or titanium diffusion-restricted spaces should be processed in the Non Lumen Cycle.

    The V-PRO 1 Plus and V-PRO maX Sterilizers' Non Lumen Cycle can sterilize: t

    Non-lumened instruments including non-luments, non-luments, non-lumened rigid, semi-rigid, semi-rigid and flexible endoscopes.

    • The validation studies were conducted using a validation load consisting of two instrument trays for a total weight of 50 lbs (22.7 kg).

    The V-PRO maX Sterilizer's Flexible Cycle can sterilize:

    Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of the two configurations:

    1. Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load. * The flexible endoscopes may contain either:
    • A single lumen that is ≥ 1 mm internal diameter (ID) and ≤ 1050 mm in length
      • Or two lumens with:

    • One lumen that is ≥ 1 mm ID and ≤ 990 mm in length

    • And the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length

    • The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope).
    1. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments. ++ The flexible endoscope may contain either:

    · A single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length

    • Or two lumens with:
      • One lumen that is ≥ 1 mm ID and ≤ 990 mm in length
      • And the other lumen is ≥ 1 mm ID and ≤ 850 mm in length.

    ** The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (if not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total weight of 24 lbs (11 kg).

    The V-PRO 1, V-PRO 1 Plus and V-PRO maX Sterilizers' Lumen Cycle can sterilize: *

    Medical devices (including single, dual and triple channeled rigid endoscopes) with the following configurations:

    • Single channeled devices with a stainless lumen that is > 0.77 mm ID and ≤ 500 mm in length
    • Dual channeled devices with stainless lumens that are ≥ 0.77 mm ID and ≤ 527 mm in length
    • · Triple channeled devices with stainless lumens that are either:

    ≥ 1.2 mm ID and ≤ 275 mm in length ≥ 1.8 mm ID and ≤ 310 mm in length or

    2.8 mm ID and 1 mm (~3/64") ID and 1.8 mm (~5/64") ID and

    Device Description

    The V-PRO 1, V-PRO 1 Plus, V-PRO maX and V-PRO 60 Low Temperature Sterilization Systems, using VAPROX® HC Sterilant, are vaporized hydrogen peroxide sterilizers intended for use in the terminal sterilization of cleaned, rinsed and dried metal and nonmetal medical devices used in healthcare facilities. The Sterilizers use VAPROX® HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). The packaged sterilized devices are ready for use at the completion of the cycle; no cool down or aeration period is required following completion of the cycle.

    The V-PRO 1, V-PRO 1 Plus, and the V-PRO maX and V-PRO 60 Low Temperature Sterilization Systems are programmed to enact one (the Lumen Cycle), two (the Lumen Cycle and the Non Lumen Cycle) or three (the Lumen Cycle, the Non Lumen Cycle and the Flexible Cycles, respectively.

    AI/ML Overview

    The provided text describes the acceptance criteria and supporting studies for the STERIS Amsco V-PRO Low Temperature Sterilization Systems.

    Here's the breakdown:

    1. A table of acceptance criteria and the reported device performance

    TestAcceptance CriteriaResults
    Determination of Worst Case (Most Resistant) Material for VHPTest: The most resistant material (MRM) is the material that is most difficult to sterilize (the material with the largest number of non-sterile results)
    Controls: Controls shall perform as intended.Pass: As identified within previous V-PRO Sterilizer submissions, the most resistant material (MRM) for surface sterilization is Kraton.
    Medical Device Material CompatibilityThe devices shall be processed for 50 V-PRO 60 Sterilizer Lumen Cycles (worst case). The devices shall show no structural defects (crazing, cracks or chipping). The devices shall function following processing.Pass: This testing verified compatibility of the tested 20 materials with the V-PRO Sterilizers.
    Cytotoxicity Evaluation of Medical DevicesAll Worst Case Cycles shall proceed to completion. The cytotoxic endpoint for each medical device shall be determined. Cytotoxicity Assay Controls shall perfrom as intended.Pass: Testing demonstrates that various medical devices containing [unspecified, but likely referring to the materials tested in compatibility study] were determined to be non-cytotoxic following processing in three (3) V-PRO 60 Sterilizer Lumen Cycles.
    Hydrogen Peroxide Residual Analysis of Processed Medical DevicesThe residual hydrogen peroxide level for each medical device shall be determined and compared to the allowable hydrogen peroxide limits for mucosal and internal tissue contact. Each device shall demonstrate exhaustive extraction by showing less than a 10% increase for two (2) consecutive sample time points. Hydrogen Peroxide Residual Analysis Controls shall perform as intended.Pass: Testing demonstrates that the hydrogen peroxide residuals from the medical devices composed of the 19 proposed compatible materials are from 2660-fold lower to greater than 13,000-fold lower than the tolerable exposure limits for mucosal and internal tissue contact when processed in three (3), worst case, back-to-back Lumen Cycles in the V-PRO 60 Sterilizer.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document primarily describes validation studies for the V-PRO sterilization systems, which are typically performed in a prospective manner.

    • Determination of Worst Case Material for VHP: This refers to previous submissions, but the current document specifically highlights Kraton as the "most resistant material." No specific sample size is given for how many materials were evaluated to determine Kraton as the worst case in previous studies.
    • Medical Device Material Compatibility: "20 materials" were tested.
    • Cytotoxicity Evaluation of Medical Devices: "various medical devices containing [unspecified material components]" were tested.
    • Hydrogen Peroxide Residual Analysis of Processed Medical Devices: "19 proposed compatible materials" were tested across "three (3), worst case, back-to-back Lumen Cycles" in the V-PRO 60 Sterilizer.

    The provenance of the data (country of origin) is not explicitly stated, but as this is an FDA submission for a U.S. company (STERIS Corporation, Mentor, OH), it's highly probable the studies were conducted in the U.S. or at facilities compliant with U.S. regulatory standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This type of information is not applicable to the studies described. The studies are engineering and chemical validation tests for a sterilization system, not diagnostic device studies that would require expert human review or interpretation of medical images or patient data to establish ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not applicable as the studies are not diagnostic assessments requiring adjudication of human reader interpretations. The acceptance criteria are based on measurable physical and chemical properties and biological effects (sterilization efficacy, material integrity, cytotoxicity, residuals).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is a sterilization system, not an AI-powered diagnostic tool used by human readers. Therefore, no MRMC study or AI assistance improvement metrics are relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a sterilization system and there is no algorithm or AI component involved in its operation or performance evaluation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in these validation studies is established by objective scientific and engineering measurements and standards:

    • Sterilization effectiveness: Implicit in the "Worst Case Material" determination, where the goal is sterilization. Sterilization is typically confirmed by biological indicators demonstrating the inactivation of a specific resistant microorganism (e.g., Geobacillus stearothermophilus) to a defined sterility assurance level (SAL). While not explicitly detailed here, this is a standard method for sterilization validation.
    • Material compatibility: Visual inspection for defects (crazing, cracks, chipping) and functional testing of devices after processing.
    • Cytotoxicity: In vitro assays to assess cellular toxicity, with controls to ensure assay validity.
    • Hydrogen peroxide residual analysis: Chemical analytical methods to quantify residual hydrogen peroxide, compared against established biological tolerance limits for mucosal and internal tissue contact.

    8. The sample size for the training set

    This is not applicable. The device is a sterilization system, not a machine learning model. There is no concept of a "training set."

    9. How the ground truth for the training set was established

    This is not applicable as there is no training set.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    V-PRO 1 Plus Low Temperature Sterilization System, V-PRO maX Low Temperature Sterilization System, V-PRO
    60 Low Temperature Sterilization System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V-PRO 1, V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems, with VAPROX® HC Sterilant, are vaporized hydrogen peroxide sterilizers intended for use in the terminal sterilization of cleaned, rinsed and dried metal and nonmetal medical devices used in healthcare facilities. The pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and are thus suitable for processing medical devices sensitive to heat and moisture.

    The V-PRO 1, V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems' Lumen Cycle can sterilize:(a)

    • Lumened and non-luments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • · Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations: (a)
      • o single channeled devices with a stainless lumen that is ≥ 0.77 mm internal diameter (ID) and ≤ 500 mm in length
      • o dual channeled devices with stainless steel lumens that are ≥ 0.77 mm ID and ≤ 527 mm in length
      • o triple channeled devices with stainless steel lumens that are
        • ≥ 1.2 mm ID and ≤ 275 mm in length
        • ≥ 1.8 mm ID and ≤ 310 mm in length
        • or
        • 2.8 mm ID and 1 mm ID and 1 mm ID and 1 mm ID and 1 mm (~3/64") ID and

    Device Description

    The V-PRO 1, V-PRO 1 Plus, V-PRO maX and V-PRO 60 Low Temperature Sterilization Systems, with VAPROX® HC Sterilant, are vaporized hydrogen peroxide sterilizers intended for use in the terminal sterilization of cleaned, rinsed and dried metal and nonmetal medical devices used in healthcare facilities. The Sterilizers use VAPROX® HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). The packaged sterilized devices are ready for use at the completion of the cycle; no cool down or aeration period is required following completion of the cycle.

    The V-PRO 1, V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems are programmed to enact one (the Lumen Cycle), two (the Lumen Cycle and the Non Lumen Cycle) or three (the Lumen Cycle, the Non Lumen Cycle and the Flexible Cycle) cycles, respectively.

    The V-PRO 60 Low Temperature Sterilization System, with VAPROX® HC Sterilant, is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried metal and nonmetal medical devices used in healthcare facilities. The three pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for processing medical devices sensitive to heat and moisture.

    AI/ML Overview

    The STERIS V-PRO Low Temperature Sterilization Systems were studied to verify two modifications: the removal of eye-contacting limitations and the addition of polyurethane as a compatible material for the Lumen Cycle.

    Here's a breakdown of the acceptance criteria and study details:

    1. Acceptance Criteria and Reported Device Performance

    Device ModificationVerification / Validation ActivityAcceptance Criteria (Implied)Reported Device Performance
    Removal of Eye-Contacting Limitations
    Compare the residual hydrogen peroxide level for materials after processing under worst-case conditions in the V-PRO Sterilizers and the STERRAD 100NX Sterilizer.Residual hydrogen peroxide levels in V-PRO processed samples should be statistically less than or equal to materials similarly processed in the predicate device (STERRAD 100NX Sterilizer).Pass: "V-PRO 60 Sterilizer (worst-case) processed samples hydrogen peroxide residues were statistically less than or equal to materials similarly processed in the STERRAD 100NX Sterilizer."
    Compare the dilution required to achieve non-cytotoxic results for materials after processing under worst-case conditions in the V-PRO Sterilizers and the STERRAD 100NX Sterilizer.The cytotoxic response for materials processed in the V-PRO Sterilizer cycles (worst-case) should be equivalent to or less than the cytotoxic response for the same materials processed in the predicate device (STERRAD 100NX Sterilizer Cycles).Pass: "The cytotoxic response for all materials processed in the V-PRO 60 Sterilizer cycles (worst-case) was equivalent to or less than the cytotoxic response for the same materials processed in STERRAD 100NX Sterilizer Cycles."
    Confirm that the cytotoxic component present in each extract (if the neat sample is cytotoxic) is hydrogen peroxide (H2O2) by the addition of catalase to the extracts.The cytotoxic component should be identified as hydrogen peroxide (H2O2)."The cytotoxic component for all V-PRO 60 Sterilizer (worst-case) processed materials was identified as hydrogen peroxide (H2O2)." (This is an observation supporting the mechanism, not a pass/fail criteria itself, but confirms understanding of the cytotoxic agent).
    Determine the cytotoxicity for medical devices after processing in three (3) back-to-back V-PRO 60 Sterilizer (worst-case) Lumen Cycles.Following exposure to three (3) back-to-back cycles, devices should be non-cytotoxic at the neat dilution.Pass: "Following exposure to three (3) back-to-back cycles, devices are non-cytotoxic at the neat dilution."
    Compare the residual hydrogen peroxide level for medical devices after processing in three (3) back-to-back V-PRO 60 Sterilizer (worst-case) Lumen Cycles to hydrogen peroxide levels known to cause eye biocompatibility concerns.Following exposure to three (3) back-to-back cycles, device hydrogen peroxide residues should be below levels known to cause biocompatibility concerns (specifically eye biocompatibility concerns).Pass: "Following exposure to three (3) back-to-back cycles, device hydrogen peroxide residues are below levels known to cause biocompatibility concerns."
    Evaluate V-PRO Sterilizer materials processed for 3 consecutive cycles for ocular irritation per ISO 10993-10.Materials coupons processed for 3 consecutive cycles in the V-PRO maX Lumen (worst-case) Cycle should be non-irritants per ISO 10993-10 evaluation.Pass: "Materials coupons processed for 3 consecutive cycles in the V-PRO maX Lumen (worst-case) Cycle are non-irritants per ISO 10993-10 evaluation."
    Addition of Polyurethane as a compatible material for V-PRO Sterilizers' Lumen Cycle
    Evaluate polyurethane material compatibility after processing the worst-case V-PRO Sterilizer Cycle.No material degradation or cosmetic changes should be observed for the polyurethane-containing device.Pass: "No material degradation or cosmetic changes were observed for the polyurethane-containing device."
    Sterilization efficacy for polyurethane in Lumen Cycle is demonstrated by previously completed testing in the V-PRO maX and V-PRO 60 Sterilizers’ (worst-case) Non Lumen Cycle.Sterilization efficacy for polyurethane should be demonstrated to meet the required sterility assurance level (SAL). (Implied from the "Pass" result, as specific SAL is not stated here but is a standard requirement for sterilization validation.)Pass: (Specific sterility efficacy data is not provided in this table, but the "Pass" indicates that prior testing supports this. This would typically involve bacterial spore inactivation.)

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document does not explicitly state the exact "sample size for the test set" in terms of number of unique devices or materials for each specific test. However, it indicates:

    • For residual hydrogen peroxide and cytotoxicity comparison: "materials" were processed.
    • For in-device cytotoxicity and tissue residue: "medical devices" were processed.
    • For ocular irritation: "Materials coupons" were processed.
    • For polyurethane compatibility: "polyurethane-containing device" and "polyurethane material" were evaluated.

    Data Provenance: The studies were conducted by STERIS Corporation for a 510(k) premarket notification. This implies the data is prospective for the purpose of demonstrating the safety and effectiveness of the proposed changes. The country of origin of the data is not specified but is presumably the USA, where STERIS Corporation is based.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. The tests described are primarily laboratory-based evaluations of chemical residues, cytotoxicity, material compatibility, and biological irritation, rather than assessments requiring interpretation by clinical experts. For biocompatibility tests like ocular irritation per ISO 10993-10, trained toxicologists or evaluators would interpret the results.

    4. Adjudication Method for the Test Set

    This information is not applicable as the studies described are laboratory-based experimental validations, not clinical studies requiring independent interpretation or adjudication of outcomes by multiple experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A MRMC comparative effectiveness study was not done. This type of study is typically performed to evaluate the diagnostic accuracy of imaging or other interpretive devices, often with and without AI assistance to human readers. The STERIS V-PRO System is a sterilizer, and its performance is evaluated through physical, chemical, and biological efficacy studies, not reader interpretation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    A standalone study was not done because the device itself is a sterilizer, not an algorithm. Its performance is inherent in its operation (sterilization efficacy, material compatibility, residue levels), not in an algorithmic output.

    7. Type of Ground Truth Used

    The ground truth used in these studies is based on:

    • Quantitative measurements: Residual hydrogen peroxide levels (chemical analysis).
    • Biological assays: Cytotoxicity tests (e.g., using live cells to detect toxic effects), ocular irritation tests (per ISO 10993-10 guidelines, likely involving animal models or in vitro alternatives).
    • Visual assessment: Material degradation or cosmetic changes.
    • Sterilization efficacy: Implied to be based on inactivation of biological indicators with a known sterility assurance level (SAL).

    8. Sample Size for the Training Set

    This information is not applicable as the V-PRO Sterilization System is a physical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this device type.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140498
    Manufacturer
    Date Cleared
    2014-07-24

    (147 days)

    Product Code
    Regulation Number
    880.6860
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    V-PRO 60 LOW TEMPERATURE STERILIZATION SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V-PRO 60 Low Temperature Sterilization System, with VAPROX® HC Sterilant, is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in healthcare facilities. The pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for processing medical devices sensitive to heat and moisture.

    The V-PRO 60 Sterilizer's Lumen Cycle can sterilize:

    • Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • Non-lumened devices including non-lumened rigid and semi-rigid endoscopes
    • Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations:
      • single or dual lumen devices with stainless steel lumens that are ≥ 0.77 mm (~1/32") internal diameter (ID) and ≤ 410 mm (16-9/64") in length
      • triple lumen devices with stainless steel lumens that are
        • ≥ 1.2 mm (~3/64") ID and ≤ 275 mm (~10-55/64") in length
        • ≥ 1.8 mm (~5/64") ID and ≤ 310 mm (~12-13/64") in length
        • or ≥ 2.8 mm (~7/64") ID and ≤ 317 mm (12-31/64") in length

    The V-PRO 60 Sterilizer's Non Lumen Cycle can sterilize

    • Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with stainless steel or titanium diffusion-restricted spaces such as the hinged portion of forceps and scissors.

    The V-PRO 60 Sterilizer's Flexible Cycle can sterilize:

    • One flexible surgical endoscope or bronchoscope with a light cord (if not integral to endoscope) and mat without any additional load. The flexible endoscope may be a:
      • single or dual lumen device with lumens that are ≥ 1 mm (~3/64") ID and ≤ 990 mm (38-63/64") in length
    Device Description

    The V-PRO 60 Low Temperature Sterilization System is a new vaporized hydrogen peroxide sterilizer model to be added to the STERIS V-PRO family of sterilizers. The V-PRO product line currently consists of the Amsco V-PRO 1, Amsco V-PRO 1 Plus and Amsco V-PRO maX Sterilizers. As with the predicate device (K102330), the V-PRO 60 Sterilizer has three preprogrammed cycles: the Lumen Cycle, the Non Lumen Cycle and the Flexible Cycle. The V-PRO 60 Low Temperature Sterilization System is intended for terminal sterilization of cleaned, rinsed, dried and packaged reusable surgical instruments used in healthcare facilities. The V-PRO 60 Sterilizer uses VAPROX® HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). Its three preprogrammed cycles all utilize a conditioning phase, a sterilize phase and an aeration phase. The packaged sterilized devices are ready for use at the completion of the cycle, no cool down or aeration period is required following completion of the cycle.

    AI/ML Overview

    This document describes the V-PRO® 60 Low Temperature Sterilization System, a vaporized hydrogen peroxide sterilizer. The acceptance criteria are based on achieving sterility and maintaining material compatibility.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (Implied)Reported Device Performance
    Sterilization Efficacy
    AOAC Sporicidal TestAll carriers processed should be sterile.All 720 carriers processed using 3 lots of EOSL sterilant were sterile.
    Determination of D-value and Total Kill EndpointGreater than a 12 log reduction of the most resistant organism.Greater than a 12 log reduction of the most resistant organism is achieved within all cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) of the V-PRO 60 Sterilizer.
    ½ Cycle Modified Total Kill Endpoint VerificationAll sterile results within the validation load.Modified total kill end point analysis was demonstrated for all three V-PRO 60 Sterilizer cycles. The standard injection weight of 1.1 g and at least one lower injection weight resulted in all sterile results within the validation load used to qualify each sterilizer cycle. Partial positives or all survive results were seen at lower injection weights.
    ½ Cycle Sterilization Verification of Cycle ClaimsReproducible sterilization of specified devices/lumens.• The Lumen Cycle reproducibly sterilizes single, dual and triple lumen devices under worst case conditions in ½ Cycle.
    • The Flexible Cycle reproducibly sterilizes 1 x 990 mm flexible endoscope lumens under worst case conditions in ½ Cycle.
    • The Non Lumen Cycle reproducibly sterilizes non-lumened devices under worst case conditions in ½ Cycle.
    ½ Cycle Verification of Mated SurfacesSterile efficacy for mated surfaces under worst case conditions.Sterile efficacy was demonstrated for mated surfaces under worst case conditions.
    Simulated Use TestSterilization of medical devices under worst case conditions.Simulated use testing verified the ability of the V-PRO 60 Sterilizer cycles to sterilize medical devices under worst case processing conditions.
    In Use TestSterilization of patient-soiled, clinically-cleaned instruments.The in use investigation demonstrated the ability of the V-PRO 60 Sterilizer cycles to sterilize patient-soiled, clinically-cleaned, medical instruments.
    Device Compatibility & Integrity
    BiocompatibilityBiocompatibility after processing.Cytotoxicity and residue analysis of 23 materials have demonstrated biocompatibility after processing in the V-PRO 60 Sterilizer.
    Medical Device Material CompatibilityCompatibility with materials of construction after multiple cycles.Evaluation of medical devices after multiple cycles in the V-PRO 60 Sterilizer has demonstrated compatibility with 23 materials of construction.
    Process Control & Verification
    Final Process QualificationSuccessful qualification for all cycles.The V-PRO 60 Sterilizer final process qualification was successful for all three (3) sterilizer cycles. All three lots of CI exhibited complete color change. All three SCBI PIs exhibited a passing color change and all SCBIs were negative for growth. Manual inspection of the process parameter data confirmed that all cycle specifications were met.
    Conformity to Standards (Safety and EMC)Compliance with EN 61010-1:2001 and EN61326-1:2006.The V-PRO 60 Low Temperature Sterilization System has been tested for conformity and is certified to EN 61010-1:2001 and EN61326-1:2006.

    Study Proving Device Meets Acceptance Criteria:

    The document summarizes the "Summary of Nonclinical Tests" performed on the V-PRO 60 Low Temperature Sterilization System to demonstrate its substantial equivalence to the predicate device, the V-PRO maX Low Temperature Sterilization System. The studies aim to prove the device's ability to sterilize various medical devices effectively and safely.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document provides specific sample sizes for particular tests:

    • AOAC Sporicidal Test: "All 720 carriers processed using 3 lots of EOSL sterilant."
    • Validation Loads:
      • Lumen Cycle: "a maximum of twelve (12) lumens per load... consisting of one instrument tray with instrument organizers and mat and two pouches for a total weight of 11 lbs (5.0 kg)."
      • Non Lumen Cycle: "a validation load consisting of one instrument tray with instrument organizers and mat and one pouch for a total weight of 12 lbs (5.4 kg)."
      • Flexible Cycle: "one flexible endoscope, packaged into a tray with silicone mat, instrument organizers and light cord (if not integral to scope) and no additional load."
    • Biocompatibility and Material Compatibility: "23 materials" were analyzed.

    Data Provenance: The document does not explicitly state the country of origin or whether the data is retrospective or prospective. However, based on the nature of medical device clearance submissions to the FDA (United States Food and Drug Administration), these are typically prospective laboratory and simulated-use studies conducted by the manufacturer or authorized testing facilities, aiming to meet regulatory requirements prior to market entry.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not specify the "number of experts" or their "qualifications" used to establish ground truth. The studies involve standardized microbiological and engineering tests rather than expert interpretation of results in a clinical setting. For sterilization studies, the ground truth is typically established by laboratory methods such as growth/no growth of biological indicators (e.g., Geobacillus stearothermophilus) following exposure to the sterilization process.

    4. Adjudication Method for the Test Set:

    Not applicable. The tests described are objective performance evaluations (e.g., sterility, log reduction, material compatibility) rather than interpretation-based assessments that would require an adjudication method like 2+1 or 3+1.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. The device is a sterilizer, and its performance is evaluated through objective physical and biological tests, not through human reader interpretation of cases.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Yes, the studies described are standalone performance evaluations of the sterilization device. There is no "human-in-the-loop" interaction in the output or interpretation of the sterilization cycle results beyond operating the device and reading standard indicators. The device itself is the "algorithm only" in the context of its function.

    7. The Type of Ground Truth Used:

    The primary type of ground truth used would be:

    • Microbiological Negativity/Positivity: For sterilization efficacy tests (AOAC Sporicidal, D-value/Total Kill, ½ Cycle Modified Total Kill Endpoint, ½ Cycle Sterilization Verification, Simulated Use, In Use), the ground truth is established by culturing biological indicators or test organisms and observing the presence or absence of growth. "Sterile" or "no growth" is the ground truth for effective sterilization.
    • Physical/Chemical Analysis: For biocompatibility and material compatibility, the ground truth is established through laboratory analyses (e.g., cytotoxicity assays, residue analysis, visual inspection for degradation) and quantitative measurements.
    • Process Parameter Verification: For Final Process Qualification, the ground truth is established by objective measurements of process parameters (time, temperature, pressure, sterilant injection weight) and the observed color change of chemical indicators.

    8. The Sample Size for the Training Set:

    Not applicable. This is a physical device (sterilizer) and not an AI/machine learning model that would require a 'training set' of data in the conventional sense. The development of the sterilizer's cycles and parameters is based on engineering principles and iterative testing, not on data training in the way an algorithm learns from a dataset.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no "training set" in the context of an AI/machine learning algorithm for this device. The physical and biological principles governing sterilization, along with extensive engineering development and testing, define the performance of the sterilizer.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1