Search Results
Found 31 results
510(k) Data Aggregation
(524 days)
Urea Nitrogen2
The Urea Nitrogen2 assay is used for the quantitation of Urea Nitrogen in human serum, plasma, or urine on the ARCHITECT c System.
The Urea Nitrogen2 assay is to used as an aid in the diagnosis and treatment of certain renal and metabolic diseases.
The Urea Nitrogen2 assay is an automated clinical chemistry assay. The Urea Nitrogen2 assay is a modification of a totally enzymatic procedure. The test is performed as a kinetic assay in which the initial rate of the reaction is linear for a limited period of time. Urea in the sample is hydrolyzed by urease to ammonia and carbon dioxide. The second reaction, catalyzed by glutamate dehydrogenase (GLDH), converts ammonia and a-ketoglutarate to glutamate and water with the concurrent oxidation of reduced nicotinamide adenine dinucleotide (NADH) to nicotinamide adenine dinucleotide (NAD). Two moles of NADH are oxidized for each mole of urea present. The initial rate of decrease in absorbance at 340 nm is proportional to the urea concentration in the sample.
The provided document is a 510(k) premarket notification for a new in vitro diagnostic device, the Urea Nitrogen2 assay. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than comprehensive clinical effectiveness studies as might be seen for novel devices. Therefore, the information provided primarily concerns non-clinical performance characteristics of the assay itself, rather than human-in-the-loop performance or diagnostic accuracy evaluated in a clinical setting with patient outcomes.
Based on the provided document, here's an analysis of the acceptance criteria and the study that proves the device meets them:
Core Purpose of the Submission: To demonstrate that the Urea Nitrogen2 assay is substantially equivalent to the predicate device (Urea Nitrogen assay, K981918) for the quantitative measurement of urea nitrogen in human serum, plasma, and urine on the ARCHITECT c System. This means proving the new device performs similarly and is as safe and effective as the predicate.
Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly derived from established clinical laboratory standards (CLSI guidelines) and comparison to the predicate device's known performance. The performance metrics evaluated are standard for in vitro diagnostic assays.
1. Table of Acceptance Criteria and Reported Device Performance
Performance Characteristic | Acceptance Criteria (Implicit) | Reported Device Performance (Urea Nitrogen2) |
---|---|---|
Reportable Interval | Must cover clinically relevant range and be supported by analytical performance (LoD, LoQ, linearity, imprecision, bias). It should be comparable to or improve upon the predicate device. Predicate: Serum 2-125 mg/dL, Urine 2-1991 mg/dL. | Serum/Plasma: AMI: 2-125 mg/dL, EMI: 125-625 mg/dL, Reportable Interval: 2-625 mg/dL. |
Urine: AMI: 16-1991 mg/dL, Reportable Interval: 11-1991 mg/dL. | ||
(Meets/Exceeds predicate in serum/plasma upper limit, comparable in urine.) | ||
Within-Laboratory Precision (Imprecision) | Based on CLSI EP05-A3 guidelines. Low %CV (Coefficient of Variation) and SD (Standard Deviation) demonstrating consistent results. Comparable to or better than predicate. Predicate: Serum %CV 1.8-2.0% (15.5-48.0 mg/dL), Urine %CV 3.1-3.8% (504.8-896.4 mg/dL). | Serum/Plasma: Samples (4-102 mg/dL) demonstrated SDs ≤ 0.4 mg/dL and %CV ≤ 2.7%. |
Urine: Samples (55-1605 mg/dL) demonstrated SDs ≤ 11.7 mg/dL and %CV ≤ 2.1%. | ||
(Meets/Exceeds predicate with broader range tested and generally lower %CV.) | ||
Accuracy (Bias) | Low percentage bias relative to a recognized standard reference material (NIST SRM 912b). Bias should be within acceptable limits for clinical utility. | Serum: Bias ranged from 1.6% to 4.2%. |
Urine: Bias ranged from -1.3% to 3.0%. | ||
(Demonstrates acceptable accuracy.) | ||
Lower Limits of Measurement (LoB, LoD, LoQ) | Determined by CLSI EP17-A2 guidelines. LoB, LoD, and LoQ should be low enough to be clinically useful and comparable to the predicate. Predicate Serum: LoD 0.7 mg/dL, LoQ 1.4 mg/dL. Predicate Urine: LoD 15.0 mg/dL, LoQ 40.0 mg/dL. | Serum: LoB 1 mg/dL, LoD 2 mg/dL, LoQ 2 mg/dL. |
Urine: LoB 6 mg/dL, LoD 11 mg/dL, LoQ 16 mg/dL. | ||
(Comparable to predicate's analytical sensitivity; slight differences but within typical ranges for this type of assay.) | ||
Linearity | Device should show a linear response across its stated analytical measuring interval as per CLSI EP06-A. | Serum: Linear across 2 to 125 mg/dL. |
Urine: Linear across 16 to 1991 mg/dL. | ||
(Demonstrates linearity across its Analytical Measuring Interval.) | ||
Interference | No significant interference (typically defined as ±10% bias) from common endogenous and exogenous substances at specified levels. The scope and levels tested should address potential clinical interferences. | Serum/Plasma (Endogenous): No significant interference (within ±10%) for Bilirubin (60 mg/dL), Hemoglobin (2000 mg/dL), Triglycerides (1500 mg/dL). Interference was observed for Total Protein (11 g/dL at 10 mg/dL analyte level: 11% (9%, 14%) CI). |
Serum/Plasma (Exogenous): No significant interference (within ±10%) for a broad list of drugs. Interference was observed for Cefoxitin (6600 mg/L at 10 mg/dL analyte level: 10% (6%, 14%) CI). | ||
Urine (Endogenous): No significant interference (within ±10%) for Ascorbate (200 mg/dL), Glucose (1000 mg/dL), Protein (50 mg/dL). | ||
Urine (Exogenous): No significant interference (within ±10%) for a broad list of drugs/substances. (Generally good performance, with identified interferences clearly reported.) | ||
Method Comparison | High correlation and acceptable agreement (slope and intercept near 1 and 0, respectively) when compared to the predicate device on the same platform. Expected correlation coefficient near 1.00. | Serum: N=124, Correlation Coefficient=1.00, Intercept=0.74, Slope=1.02 (Concentration Range 4-123 mg/dL). |
Urine: N=121, Correlation Coefficient=1.00, Intercept=8.95, Slope=1.03 (Concentration Range 41-1754 mg/dL). | ||
(Excellent correlation and agreement with the predicate.) | ||
Tube Type Suitability | Demonstration that the device performs acceptably with specified blood collection tube types. | Serum: Serum tubes, Serum separator tubes. |
Plasma: Lithium heparin tubes, Lithium heparin separator tubes, Sodium heparin tubes. | ||
(Acceptable for specified tube types.) | ||
Dilution Verification | Demonstration that the automated dilution protocol yields results comparable to manual dilution. Accuracy of diluted results compared to undiluted or expected values. | % difference values for automated dilution vs. manual dilution ranged from -2.8% to -1.3%, demonstrating acceptable performance. |
(Acceptable performance for automated dilution.) |
Study Details
The studies described are primarily analytical performance studies, characteristic of a 510(k) submission for an in vitro diagnostic device, especially a chemical analyzer assay. They demonstrate the device's technical specifications and how it performs compared to a reference method or the predicate device.
2. Sample Size Used for the Test Set and Data Provenance
- Precision Studies (Within-Laboratory):
- Serum/Plasma: 3 human serum panels + 2 controls. Each tested in duplicate, twice per day, for 20 days on 3 reagent lot/calibrator lot/instrument combinations. For a representative combination, n=80 per sample/control.
- Urine: 3 human urine panels + 2 controls. Each tested in duplicate, twice per day, for 20 days on 3 reagent lot/calibrator lot/instrument combinations. For a representative combination, n=80 per sample/control.
- Data Provenance: Not explicitly stated but inferred to be laboratory-based analytical studies, likely from the manufacturer's R&D facilities. No country of origin is specified. The studies are retrospective analytical evaluations of manufactured samples and controls.
- Accuracy (Bias): 3 concentrations of standard across 3 reagent lots, 2 calibrator lots, and 1 instrument. (NIST SRM 912b is the standard).
- Lower Limits of Measurement (LoB, LoD, LoQ): n ≥ 60 replicates of zero-analyte samples for LoB, n ≥ 60 replicates of low-analyte level samples for LoD/LoQ. Conducted using 3 reagent lots on 2 instruments over a minimum of 3 days.
- Linearity: Not explicitly stated sample count, but typically involves preparing a dilution series of samples across the range.
- Interference: "Each substance was tested at 2 levels of the analyte (approximately 10 mg/dL and 30 mg/dL for serum/plasma; 700 mg/dL and 1500 mg/dL for urine)." No specific N for how many replicates or individual samples are run per interferent level, but implied to be sufficient for statistical analysis (e.g., 95% CI).
- Method Comparison:
- Serum: n=124 samples.
- Urine: n=121 samples.
- Data Provenance: Not explicitly stated, but these would be clinical or proficiency samples analyzed side-by-side with the predicate.
- Tube Type: Samples collected from a minimum of 40 donors.
- Dilution Verification: 5 human serum samples (spiked with urea). Each sample tested with automated dilution and 3 manual dilutions (by 2 technicians). Tested in replicates of 5.
3. Number of Experts Used to Establish Ground Truth and Qualifications
- For this type of in vitro diagnostic device (a quantitative chemical assay), "ground truth" is established by:
- Reference materials: e.g., NIST SRM 912b for accuracy. This is a primary standard, not established by human experts.
- Predicate device measurements: For method comparison, the predicate device provides the comparative 'truth' (or established method performance).
- Clinical laboratory professional consensus/guidelines: Standards like CLSI (Clinical and Laboratory Standards Institute) guidelines (EP05-A3, EP17-A2, EP06-A, EP07, EP09-A3, EP34) serve as the "expert consensus" on how to conduct and interpret these analytical studies. These are published by committees of experts in laboratory medicine, clinical chemistry, and statistics.
- No "expert readers" in the traditional sense (e.g., radiologists interpreting images) are involved in establishing ground truth for this type of device. The validation is based on metrological traceability to standards and comparison to an established analytical method.
4. Adjudication Method for the Test Set
- Not applicable as this is an analytical performance study of a chemical assay, not a diagnostic accuracy study relying on human interpretation of subjective data (like imaging or pathology). Results are quantitative measurements read by the instrument.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No. An MRMC study is relevant for diagnostic devices that involve human interpretation (e.g., AI-assisted image interpretation) and aim to show an improvement in human reader performance. This device is a quantitative chemical assay that provides a numerical result; there is no human "reader" to assist in the primary measurement. The comparison is between the new assay's performance and the predicate assay's performance, as well as against analytical standards.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, effectively. The entire submission details the standalone analytical performance of the Urea Nitrogen2 assay system (reagents + ARCHITECT c System instrument). Its performance (precision, accuracy, linearity, etc.) is evaluated independently of human interpretation of the final numerical result, beyond the standard operation and quality control typically performed in a clinical lab.
7. The Type of Ground Truth Used
- Analytical Standards and Reference Methods/Predicate Device:
- NIST SRM 912b: Used as the true value for accuracy determination.
- Predicate Device (Urea Nitrogen assay): Used as the comparative reference for method comparison studies, demonstrating substantial equivalence.
- CLSI Guidelines: Act as the "ground truth" for the methodologies and acceptance criteria applied to evaluate the analytical performance (e.g., how LoQ is defined and determined, how precision is calculated).
8. The Sample Size for the Training Set
- For this type of in vitro diagnostic assay, there isn't a "training set" in the machine learning sense (where an algorithm learns from data). The "training" for such an assay primarily refers to:
- Reagent formulation and optimization: This involves extensive R&D to achieve desired chemical reactions and stability.
- Instrument calibration: The ARCHITECT c System is calibrated using specific calibrators (Consolidated Chemistry Calibrator mentioned, which is itself traceable to standards) to ensure accurate measurement across the range.
- The "training" is inherent in the chemical and engineering development of the assay and the instrument, rather than an algorithmic learning process on a large dataset. Therefore, a specific "training set sample size" as one would discuss for an AI model is not applicable.
9. How the Ground Truth for the Training Set Was Established
- As explained above, there isn't a "training set" with ground truth in the AI/ML context for this device. Instead, the assay's performance characteristics (calibration, linearity, reaction kinetics, etc.) are optimized and confirmed through:
- Chemical principles and R&D: The enzymatic reaction (urease, GLDH kinetics) is based on established biochemical mechanisms.
- Quality control materials and calibrators: These materials have assigned values, often traceable to international standards (like NIST SRM), and are used to "train" or calibrate the instrument system.
- Iterative laboratory testing and optimization: The assay's components and instrument parameters are refined through repeated experiments to meet performance specifications.
Ask a specific question about this device
(536 days)
Medicon Hellas Albumin, Medicon Hellas Calcium, Medicon Hellas Creatinine, Medicon Hellas Urea Nitrogen
Medicon Hellas Albumin: Reagent for the quantitative measurement of albumin in serum. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
Medicon Hellas Calcium: Reagent for the quantitative measurement of calcium in serum or urine. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).
Medicon Hellas Creatinine: Reagent for the quantitative measurement of creatinine in serum and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis.
Medicon Hellas Glucose: Reagent for the quantitative measurement of glucose in serum and urine. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Medicon Hellas Direct Bilirubin; Reagent for the quantitative measurement of direct bilirubin (conjugated) in serum. Measurements of the level of direct bilirubin is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall blader block.
Medicon Hellas Total Bilirubin: Reagent for the quantitative measurements of total bilirubin in serum. Measurements of the levels of total bilirubin is used in the diagnosis and treatment of liver. hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.
Medicon Hellas Urea Nitrogen: Reagent is for the quantitative measurement of urea nitrogen in serum and urine. Measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.
The Medicon Hellas Albumin, Medicon Hellas Calcium, Medicon Hellas Creatinine, Medicon Hellas Glucose, Medicon Hellas Direct Bilirubin, Medicon Hellas Total Bilirubin, and Medicon Hellas Urea Nitrogen are reagents for use with Diatron Pictus 500 Clinical Chemistry Analyzers. They are test systems for the quantitative measurement of albumin, calcium, creatinine, glucose, direct and total bilirubin, and urea nitrogen in human serum and urine where clinically applicable. The methods employed are photometric, utilizing reactions between the sample and reagents to produce a colored chromophore or a change in absorbance that is proportional to the concentration of the analyte. The analyzer photometer reads the absorbances at time intervals dictated by the method application stored in the analyzer memory, and the change in absorbance is calculated automatically.
The provided text describes the performance of several Medicon Hellas assays (Albumin, Calcium, Creatinine, Glucose, Direct Bilirubin, Total Bilirubin, and Urea Nitrogen) when run on the Diatron Pictus 500 Clinical Chemistry Analyzer, demonstrating their substantial equivalence to predicate devices (Beckman Coulter AU reagents on AU2700 analyzer, and Abbott Architect Direct Bilirubin on Architect c8000 analyzer).
Here's an analysis of the provided information, structured to address your specific points regarding acceptance criteria and study details:
1. A Table of Acceptance Criteria and the Reported Device Performance:
The document doesn't explicitly state "acceptance criteria" in a single, overarching table with pass/fail remarks. Instead, it describes each performance characteristic and then presents the results. The "Summary" sections for each study type imply that the results met the pre-defined acceptance criteria for demonstrating substantial equivalence. For instance, for accuracy, it states "Accuracy studies completed on at least three lots of each candidate reagent confirm that Medicon albumin... are substantially equivalent to the related predicate devices." This implies that the statistical analyses (Deming regression, R2, slope, intercept) fell within acceptable ranges. Similarly, for precision, it states "All lots passed acceptance criteria for each applicable sample type at each level."
Since explicit acceptance criteria are not presented, they are inferred from the demonstrated performance and the statement that the devices "passed acceptance criteria" or "met statistical acceptance criteria." Below is a table summarizing the reported device performance for each analyte. The "Acceptance Criteria" column will reflect the general statements of success or the implied ranges from the results themselves, as explicit numerical targets for individual tests are not given.
Implied Acceptance Criteria and Reported Device Performance
Analyte | Performance Characteristic | Implied/General Acceptance Criteria | Reported Device Performance |
---|---|---|---|
Medicon Hellas Albumin | |||
Accuracy (Method Comparison) | R2 Correlation | Values suggesting substantial equivalence (>0.98 is generally good) | R2 = 0.9862 |
Slope | Values close to 1 | Slope = 1.0180 | |
Intercept | Values close to 0 | Intercept = 0.05 | |
Reportable Range (Serum) | Comparable to predicate device | 1.50 - 6.00 g/dL | 1.50 - 6.00 g/dL |
LOD (Serum) | Acceptable for clinical use | 0.40 g/dL | 0.40 g/dL |
LOQ (Serum) | Acceptable for clinical use | 0.50 g/dL | 0.50 g/dL |
Precision (Serum) | CV% within acceptable clinical limits (e.g., typically |
Ask a specific question about this device
(348 days)
CALCIUM-CRESOLPHTHALEIN, GLUCOSE, UREA/BUN-UV
CALCIUM-CRESOLPHTHALEIN: Reagent for the measurement of calcium concentration in human serum, plasma or urine. The obtained values are useful as an aid in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory.
GLUCOSE: Reagent for the measurement of glucose concentration in human serum and plasma. The obtained values are useful as an aid in the diagnosis and treatment of the diabetes mellitus. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory.
UREA/BUN - UV: Reagent for the measurement of urea concentration in human serum, plasma or urine. The obtained values are useful as an aid in the diagnosis of certain renal and metabolic diseases. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory.
Not Found
This document is an FDA 510(k) clearance letter for three BioSystems S.A. reagents: CALCIUM-CRESOLPHTHALEIN, GLUCOSE, and UREA/BUN-UV. As such, it does not contain the information requested regarding acceptance criteria and study details for an AI/ML medical device.
The information typically provided in a 510(k) summary for in vitro diagnostic (IVD) reagents like these focuses on analytical performance characteristics (e.g., linearity, precision, accuracy, interference, stability) and correlation with a predicate device, rather than the AI/ML-specific criteria requested in the prompt.
Therefore, it is impossible to answer the questions based on the provided text. The document is about chemical reagents for laboratory testing, not an AI-powered diagnostic device.
Ask a specific question about this device
(29 days)
VITROS XT Chemistry Products UREA-CREA Slides
For in vitro diagnostic use only
The UREA test within the VITROS XT Chemistry Products UREA-CREA Slides quantitatively measures urea concentration, reported either as urea nitrogen or as urea (UREA), in serum, plasma, and urine using the VITROS XT 7600 Integrated System. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases
The CREA test within the VITROS XT Chemistry Products UREA-CREA Slides quantitatively measures creatinine (CREA) concentration in serum, plasma, and urine using the VITROS XT 7600 Integrated System. Creatinine measurements are used in the diagnosis and treatment of renal dialysis, and as a calculation basis for measuring other urine analytes.
Special conditions for use statement: For prescription use only.
The new device, the VITROS XT Chemistry Products UREA-CREA Slide is a single device that contains both a UREA test and a CREA test multilayered, analytical element coated on a polyester support separated by a plastic barrier sealed within a single slide frame. In this format, individual reactions occur and test results are generated for each analyte independently of the other analyte.
To perform the UREA test, a drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. Water and nonproteinaceous components then travel to the underlying reagent layer, where the urease reaction generates ammonia. The semipermeable membrane allows only ammonia to pass through to the colorforming layer, where it reacts with the indicator to form a dye. The reflection density of the dye is measured and is proportional to the concentration of urea in the sample.
To perform the CREA test, a drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. Creatinine diffuses to the reagent layer, where it is hydrolyzed to creatine in the rate-determining step. The creatine is converted to sarcosine and urea by creatine amidinohydrolase. The sarcosine, in the presence of sarcosine oxidase, is oxidized to glycine, formaldehyde, and hydrogen peroxide. The final reaction involves the peroxidase-catalyzed oxidation of a leuco dye to produce a colored product. Following addition of the sample, the slide is incubated. During the initial reaction phase, endogenous creatine in the sample is oxidized. The resulting change in reflection density is measured at 2 time points. The difference in reflection density is proportional to the concentration of creatinine present in the sample.
This document describes the analytical performance of the VITROS XT Chemistry Products UREA-CREA Slides for quantitatively measuring urea and creatinine concentrations. The information provided is for an in-vitro diagnostic device and does not involve AI assistance, human readers, or image analysis, thus many of the requested elements are not applicable.
1. Table of Acceptance Criteria and Reported Device Performance
The device performance is primarily assessed through method comparison (against predicate devices), precision, detection limits, and linearity. The "acceptance criteria" are implied by the measured performance demonstrating substantial equivalence to the predicate devices and meeting clinical laboratory standards for accuracy and precision.
Test Parameter | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Method Comparison | High correlation (e.g., r > 0.975) and acceptable bias compared to predicate. | UREA Serum: N=124, Slope=1.04, Intercept=0.00, Corr. Coeff.=0.999 (Test Range 3-106 mg/dL, Measuring Range 2.0-120.0 mg/dL) |
UREA Urine: N=128, Slope=1.05, Intercept=-13.21, Corr. Coeff.=0.999 (Test Range 105-2451 mg/dL, Measuring Range 67-2520 mg/dL) | ||
CREA Serum: N=130, Slope=1.00, Intercept=-0.01, Corr. Coeff.=1.000 (Test Range 0.20-13.49 mg/dL, Measuring Range 0.15-14.0 mg/dL) | ||
CREA Urine: N=116, Slope=1.01, Intercept=-0.93, Corr. Coeff.=0.998 (Test Range 13.0-336.6 mg/dL, Measuring Range 3.2-346.5 mg/dL) | ||
Precision | Demonstrated repeatability, within-day, and within-lab precision within acceptable CV% and SD limits for various concentration levels. | Detailed tables provided showing SD and CV% for Repeatability, Within Day, and Within Lab precision for multiple pools/controls across Urea Serum, Urea Urine, CREA Serum, and CREA Urine. E.g., UREA Serum Pool 1: Repeatability SD 0.1, CV% 4.0; Within Lab SD 0.2, CV% 8.7. CREA Serum Pool 1: Repeatability SD 0.007, CV% 1.1; Within Lab SD 0.011, CV% 1.7. All measured values are generally low, indicating good precision. |
Detection Limits (LoQ) | Measured LoQ values no greater than claimed LoQ, with Total Error goal met. | UREA Serum: LoQ 1.7 mg/dL (Claimed 2.0 mg/dL); Total Error goal ≤ 1.2 mg/dL |
UREA Urine: LoQ 41 mg/dL (Claimed 67 mg/dL); Total Error goal ≤ 21 mg/dL Urea N | ||
CREA Serum: LoQ 0.11 mg/dL (Claimed 0.15 mg/dL); Total Error goal ≤ 0.06 mg/dL | ||
CREA Urine: LoQ 2.3 mg/dL (Claimed 3.2 mg/dL); Total Error goal ≤ 1.2 mg/dL | ||
Linearity | Linear range supporting the claimed measuring range. | UREA Serum: Linear Range 1.93-148.79 mg/dL (Claimed Measuring Range 2.0-120.0 mg/dL) |
UREA Urine: Linear Range 62.46-3198.85 mg/dL (Claimed Measuring Range 67-2520 mg/dL) | ||
CREA Serum: Linear Range 0.04-14.86 mg/dL (Claimed Measuring Range 0.15-14.0 mg/dL) | ||
CREA Urine: Linear Range 1.1-418.3 mg/dL (Claimed Measuring Range 3.2-346.5 mg/dL) |
2. Sample Size and Data Provenance
- Method Comparison Test Set:
- UREA Serum: 124 samples
- UREA Urine: 128 samples
- CREA Serum: 130 samples
- CREA Urine: 116 samples
- Precision Test Set: Not specified as a separate patient sample set, but rather "patient pools and quality control materials." Each precision study included "a minimum of 80 observations (2 replicates per run, 2 runs per day over 20 days)" for serum and urine pools/controls.
- Detection Limits (LoQ) Test Set: 72 determinations for serum (UREA and CREA) and 64 for urine (CREA). 72 for urine (UREA).
- Linearity Test Set: Twenty proportionally related admixtures of low and high test fluids, each tested in quadruplicate. Specific number of patient samples not stated, as it uses contrived samples.
- Specificity (Interference) Test Set: Not explicitly stated how many patient samples were used, but rather "96 test substances" and "88 test substances" were spiked into serum samples, and "two substances" and "nineteen test substances" for urine.
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, the studies used standard clinical laboratory protocols (e.g., CLSI EP09-A3, EP05-A3, EP17-A2, EP06-A) which typically involve fresh or appropriately preserved clinical samples. The reference interval studies mention "an internal study of 3160 apparently healthy adults" and "an external study" (for UREA) or "an external study of apparently healthy adults (serum: 180 males and 180 females)" and "a separate external study" (for CREA).
3. Number of Experts and Qualifications for Ground Truth
- This is an in-vitro diagnostic device (chemistry analyzer) measuring chemical concentrations. The "ground truth" is established by the reference methodology (predicate devices on a different system) and gravimetric/analytical preparation of calibrators and quality controls. It does not involve human expert interpretation of images or other subjective assessments. Therefore, the concept of "experts establishing ground truth" in the context of radiology or pathology (e.g., radiologists interpreting images) is not applicable here.
4. Adjudication Method for the Test Set
- Not applicable. This is not a study involving human reader interpretation requiring adjudication. Performance is assessed analytically against quantitative measurements.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done
- No, an MRMC study was not done. This device is an automated in-vitro diagnostic assay, not an AI-assisted diagnostic tool for human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, this study represents a standalone analytical performance evaluation of the device's ability to accurately measure urea and creatinine concentrations. "Algorithm" in this context refers to the chemical reactions and measurement principles of the assay, not a software algorithm that processes medical images or data for interpretation. The performance metrics (method comparison, precision, detection limits, linearity, specificity) directly quantify the device's analytical capabilities without human intervention in the measurement process itself.
7. The type of ground truth used
- The ground truth is established through:
- Reference Methods/Predicate Devices: For method comparison, the results generated by the predicate devices (VITROS BUN/UREA Slides and VITROS CREA Slides on the VITROS 5600 Integrated System) serve as the comparative ground truth.
- Analytically Prepared Materials: For precision, detection limits, and linearity, the "ground truth" concentrations of control materials, patient pools, and serially diluted samples are established through precise gravimetric and volumetric preparations, often traceable to certified reference materials.
- Clinical Laboratory Standards (CLSI Protocols): The studies adhered to widely accepted CLSI guidelines for analytical performance evaluation, implying that the methodologies used to establish "ground truth" for each parameter meet industry standards for accuracy and rigor in a clinical lab setting.
8. The Sample Size for the Training Set
- This is not an AI/machine learning study that involves "training sets" in the conventional sense. The device's performance is based on the inherent chemical and optical principles of the dry-slide technology. There's no AI model being trained with data.
9. How the ground truth for the training set was established
- Not applicable, as there is no AI training set.
Ask a specific question about this device
(264 days)
epoc Blood Urea Nitrogen Test, epoc Total Carbon Dioxide Test
The Blood Urea Nitrogen and Total Carbon Dioxide tests, as part of the epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.
Blood Urea Nitrogen measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal and metabolic diseases.
Total Carbon Dioxide measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance.
The epoc Blood Analysis System is an in vitro diagnostic device system for the quantitative testing of blood gases, electrolytes, and metabolites in venous, arterial, and capillary whole blood samples. The epoc System is comprised of 3 major subsystems: epoc Host, epoc Reader and epoc BGEM Test Card. The main accessory used with the epoc System includes the epoc Care-Fill Capillary Tubes used to collect and introduce capillary blood samples into the epoc Test Card.
The epoc Blood Analysis System was previously cleared for prescription use to quantitate pH, pCO2, pO2, Na, K, iCa, Cl, Glu, Lact, Crea, and Hct in arterial, venous, and capillary blood samples per K061597, K090109, K092849, K093297, and K113726. This premarket notification submission adds blood urea nitrogen (BUN) and total carbon dioxide (TCO2) quantitation to the epoc BGEM Test Card and Blood Analysis System.
The epoc Blood Urea Nitrogen Test
and epoc Total Carbon Dioxide Test
, as part of the epoc Blood Analysis System
, are intended for use by trained medical professionals as an in vitro diagnostic device for quantitative testing of heparinized or un-anticoagulated arterial, venous or capillary whole blood.
The acceptance criteria and device performance are described in several studies:
Acceptance Criteria and Device Performance:
Study | Acceptance Criteria | Reported Device Performance |
---|---|---|
Analytical Sensitivity (LoB, LoD, LoQ per CLSI EP17-A2) | Not explicitly stated as acceptance criteria, but demonstrates detection limits. | BUN: LoB 2 mg/dL, LoD 3 mg/dL, LoQ 3 mg/dL |
TCO2: LoB 4.0 mM, LoD 4.3 mM, LoQ 4.3 mM | ||
Linearity (per CLSI EP06-A) | Not explicitly stated as acceptance criteria, but demonstrates linearity across reportable range. | BUN (4-119 mg/dL): Slope 1.020, Intercept 0.4, R 0.9989 |
TCO2 (4-49 mmol/L): Slope 0.903, Intercept 3.32, R 0.9997 | ||
Precision (Aqueous Controls) (CLSI EP05-A3) | Not explicitly stated as acceptance criteria, but demonstrates precision. | BUN High Level (51.7 mg/dL): SWR 1.01 (2.0% CV), ST 1.16 (2.3% CV) |
BUN Low Level (7.1 mg/dL): SWR 0.30 (4.2% CV), ST 0.32 (4.5% CV) | ||
TCO2 High Level (30.7 mmol/L): SWR 0.82 (2.7% CV), ST 0.92 (3.0% CV) | ||
TCO2 Low Level (16.2 mmol/L): SWR 0.88 (5.4% CV), ST 1.02 (6.3% CV) | ||
Interference (CLSI EP07-A2) | Unacceptable interference bias defined as producing a significant error more than 5% of the time. | Clinically significant interfering substances for BUN and TCO2 are itemized and reported. Various exogenous and endogenous interferences were tested and found to be clinically insignificant below certain thresholds. |
Clinical Field Precision (Aqueous Controls) (CLSI EP05-A3) | Not explicitly stated as acceptance criteria, but demonstrates precision in a clinical setting. | BUN Level 1 (52.1 mg/dL): SWR 1.06 (2.0%), Total Reproducibility 1.54 (3.0%) |
BUN Level 2 (17.7 mg/dL): SWR 0.45 (2.5%), Total Reproducibility 1.11 (6.3%) | ||
BUN Level 3 (7.1 mg/dL): SWR 0.24 (3.4%), Total Reproducibility 0.26 (3.7%) | ||
TCO2 Level 1 (15.9 mM): SWR 0.44 (2.8%), Total Reproducibility 0.50 (3.1%) | ||
TCO2 Level 2 (19.7 mM): SWR 0.66 (3.4%), Total Reproducibility 0.78 (3.9%) | ||
TCO2 Level 3 (30.4 mM): SWR 0.58 (1.9%), Total Reproducibility 1.05 (3.4%) | ||
Clinical Field Precision (Whole Blood) | Not explicitly stated as acceptance criteria, but demonstrates precision in a clinical setting. | BUN Hi-Syringe (57.4 mg/dL): %CV 2.3% |
BUN Lo-Cap Tube (7.6 mg/dL): %CV 7.0% | ||
TCO2 Hi-Syringe (36.5 mM): %CV 1.5% | ||
TCO2 Lo-Cap Tube (13.5 mM): %CV 3.5% | ||
Method Comparison (BUN) (CLSI EP09-A3) | Not explicitly stated as a numerical acceptance criterion, but implies a high correlation with the reference method. | Comparing epoc BUN to Roche Cobas 8000 : Slope 0.985, Intercept 0.3, R 0.998, Mean Bias at 26 mg/dL -0.1+0.2 |
Method Comparison (TCO2) | Not explicitly stated as a numerical acceptance criterion, but implies a high correlation with the reference method. | Comparing epoc TCO2 to i-STAT-CHEM8+ : Slope 1.039, Intercept -0.8, R 0.974, Mean Bias at 20 mM 0.0+0.2 |
Matrix Comparison: Anticoagulant | No significant difference between results in Li-heparinized, Na-heparinized, and non-anticoagulated blood samples | Concluded no significant difference in BUN and TCO2 results. |
Study Information:
-
Sample sizes used for the test set and the data provenance:
- Analytical Sensitivity (LoB, LoD, LoQ): Test samples were prepared from dialyzed whole blood. The specific number of samples or runs is not explicitly stated, but the study was conducted according to CLSI EP17-A2.
- Linearity: Multiple whole blood samples were used, spanning the reportable range. Conducted per CLSI EP06-A. Specific number not provided.
- Precision (Aqueous Controls): 320 replicates for each level of both BUN and TCO2. These were in-house measurements.
- Clinical Field Precision (Aqueous Controls): N=170 for BUN Level 1, 171 for Level 2, 168 for Level 3. N=172 for TCO2 Level 1, 170 for Level 2, 169 for Level 3. Data provenance is from "three different clinical sites."
- Clinical Field Precision (Whole Blood): N=134-136 for BUN samples, N=134-139 for TCO2 samples, depending on the type (syringe/cap tube) and level (high/NB/low). Data provenance is from "three different clinical sites."
- Precision (Duplicate Epoc Test Results): Over 430 patient tests run in duplicate. "Approximately equal numbers of venous, arterial and capillary samples." Data provenance not explicitly stated (e.g., country of origin), assumed to be from clinical sites in the context of "Clinical Field Precision." This is prospective clinical data.
- Method Comparison (BUN): N=433 venous, arterial, and capillary blood samples. Performed at "three clinical sites." This is prospective clinical data.
- Method Comparison (TCO2): N=574 venous, arterial, and capillary patient samples. Performed at "three clinical sites." This is prospective clinical data.
- Matrix Comparison: Anticoagulant: Over 60 volunteer donors, with samples further aliquoted into 3 vacutainers each. Data provenance not explicitly stated.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This device is a quantitative diagnostic test for chemical analytes (BUN, TCO2), not an imaging or qualitative diagnostic device requiring expert interpretation for ground truth. The ground truth for analytical performance studies is typically established using reference methods (e.g., IDMS-traceable laboratory system) or prepared reference materials.
-
Adjudication method for the test set: Not applicable. The ground truth for quantitative chemical analytes is established by reference methods or gravimetric preparation, not through human adjudication.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a diagnostic testing system for chemical analytes, not an AI-assisted diagnostic imaging or qualitative interpretation tool for human readers.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Yes, the entire performance characterization (analytical sensitivity, linearity, precision, interference, method comparison, and matrix comparison) represents standalone algorithm/device performance. The device provides quantitative results directly. Human-in-the-loop performance is about accuracy of human readers, and the clinical field precision study assesses the precision of the device in the hands of intended users, not the interpretative performance of those users.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Analytical Sensitivity, Linearity, Precision: Ground truth established via prepared reference materials (dialyzed whole blood, gravimetric mixtures of high/low samples) and aqueous controls with known concentrations.
- Method Comparison (BUN): Ground truth established by an "IDMS-traceable plasma/serum-based laboratory system" (
Roche Cobas 8000
). - Method Comparison (TCO2): Ground truth established by a "whole blood point-of-care system" (
i-STAT-CHEM8+
), which is also a predicate device. - Interference and Matrix Comparison: Comparisons were made against control samples (e.g., solvent added, or anticoagulant-free) to assess the impact of interfering substances or different matrices.
-
The sample size for the training set: Not applicable. This document describes the performance of a chemical analyte detection system, not a machine learning or AI model that requires a training set.
-
How the ground truth for the training set was established: Not applicable, as there is no training set for this device.
Ask a specific question about this device
(207 days)
ACE BUN/UREA REAGENT, ACE CREATININE REAGENT, ACE URIC ACID REAGENT, ACE CK REAGENT
The ACE BUN/Urea Reagent is intended for the quantitative determination of blood urea nitrogen (BUN) concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. BUN measurements are used in the diagnosis and treatment of certain renal and metabolic diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
The ACE Creatinine Reagent is intended for the quantitative determination of creatinine concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
The ACE Uric Acid Reagent is intended for the quantitative determination of uric acid concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Uric acid measurements are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions and of patients receiving cytotoxic drugs. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
The ACE CK Reagent is intended for the quantitative determination of creatine kinase activity in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Measurement of creatine kinase is used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
In the ACE BUN/Urea Reagent assay, urea in serum is hydrolyzed in the presence of urease to yield ammonia and carbon dioxide. The ammonia formed then reacts in the presence of glutamate dehydrogenase with 2-oxoglutarate and NADH to yield glutamate and NAD. NADH absorbs strongly at 340 nm, whereas NAD+ does not. The initial rate of decrease in absorbance, monitored bichromatically at 340 nm/647 nm, is proportional to the urea concentration in the sample.
In the ACE Creatinine Reagent assay, creatinine reacts with picric acid in an alkaline medium to form a red-orange colored complex, which absorbs strongly at 505 nm. The rate of complex formation, determined by measuring the increase in absorbance bichromatically at 505 nm/573 nm during a fixed time interval, is directly proportional to the creatinine concentration in the sample.
In the ACE Uric Acid Reagent assay, uric acid in serum is oxidized by uricase to allantoin and hydrogen peroxide. The hydrogen peroxide then acts to oxidatively couple dichlorohydroxybenzene sulfonic acid and 4-aminoantipyrine in a reaction catalyzed by peroxidase, producing a red colored quinoneimine complex, which absorbs strongly at 505 nm. The amount of chromogen formed is determined by measuring the increase in absorbance bichromatically at 505 nm/610 nm, and is directly proportional to the uric acid concentration in the sample.
In the ACE CK Reagent assay, serum creatine kinase initiates the conversion of creatine phosphate to creatine with the transfer of a phosphate group to adenosine diphosphate (ADP), forming ATP. The ATP is then used in the phosphorylation of D-glucose to form D-glucose-6-phosphate and ADP. This reaction is catalyzed by hexokinase. The enzyme glucose-6-phosphate dehydrogenase catalyzes the reduction of D-glucose-6-phosphate and nicotinamide adenine dinucleotide phosphate (NADP+). The series of reactions triggered by serum creatine kinase and ending in the formation of NADPH. NADPH strongly absorbs at 340 nm, whereas NADP+ does not. Therefore, the rate of conversion of NADP+ to NADPH can be determined by monitoring the increase in absorbance bichromatically at 340 nm/378 nm. This rate of conversion from NADP+ to NADPH is a function of the activity of CK in the sample.
Here's a summary of the acceptance criteria and supporting studies for the Alfa Wassermann ACE Reagents (BUN/Urea, Creatinine, Uric Acid, CK), based on the provided 510(k) summary.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly derived from comparisons to a predicate device (Alfa Wassermann ACE K930104 reagents) and performance characteristics such as precision, accuracy (correlation/regression with predicate), linearity, detection limits, and interference. The reported device performance is from in-house studies and Point-of-Care (POL) studies.
Note: The document does not explicitly state "acceptance criteria" numerical targets. Instead, it presents performance data for the candidate device, implying that the data's comparability to the predicate and established analytical standards is the basis for acceptance. I will present the reported performance, which demonstrates the device's meeting the necessary equivalency.
Characteristic | Acceptance Criteria (Implied) | Reported Device Performance (Candidate Device) |
---|---|---|
Intended Use | Same as predicate (quantitative determination in serum) | BUN: Quantitative determination in serum and lithium heparin plasma. |
Creatinine: Quantitative determination in serum and lithium heparin plasma. | ||
Uric Acid: Quantitative determination in serum and lithium heparin plasma. | ||
CK: Quantitative determination in serum and lithium heparin plasma. | ||
(Extended to lithium heparin plasma compared to predicate, requiring performance studies in this matrix) | ||
Platforms | Compatible with ACE Clinical Chemistry System | ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. (Expanded platforms compared to predicate) |
Method | Photometric (Same as predicate) | Photometric (Same as predicate) |
Calibration Stability | 7 days (BUN), 2 days (Creatinine), 30 days (Uric Acid) | Same |
On-Board Stability | 30 days (BUN), 10 days (Creatinine), 30 days (Uric Acid), 25 days (CK) | Same |
Sample Type | Serum (per predicate) | Serum and lithium heparin plasma (Candidate device demonstrates equivalence in both) |
Sample Volume | 3 µL (BUN, Uric Acid), 20 µL (Creatinine), 5 µL (CK) | Same |
Reaction Volume | 333 µL (BUN), 240 µL (Creatinine), 243 µL (Uric Acid), 170 µL (CK) | Same |
Expected Values | Same as predicate | Same |
Measuring Range | 3-100 mg/dL (BUN), 0.33-25.0 mg/dL (Creatinine), 1.5-16.0 mg/dL (Uric Acid), 11-1350 U/L (CK) | Same |
Sample Stability | Same as predicate (storage conditions) | Same |
Precision | Low, Mid, High %CV and SD comparable to predicate/clinical needs | In-House Serum/Plasma: Generally 0.98, Slope ~1, Intercept ~0) |
Creatinine: R > 0.99, Slope 1.003-1.050, Intercept -0.077 to 0.005. | ||
Uric Acid: R > 0.98, Slope 1.008-1.028, Intercept -0.29 to -0.09. | ||
CK: R > 0.99, Slope 0.978-1.006, Intercept -0.5 to 0.1. (See pages 8-9) | ||
Method Comparison (POL) | Comparison to In-House ACE results: Slope, Intercept, Correlation (R) and Std Error Est. demonstrating equivalence to predicate system (e.g., R > 0.98, Slope ~1, Intercept ~0). | BUN: R > 0.99, Slope 0.989-1.039, Intercept -0.1 to 1.4. |
Creatinine: R > 0.99, Slope 0.977-1.051, Intercept -0.085 to 0.037. | ||
Uric Acid: R > 0.99, Slope 0.936-1.034, Intercept 0.02 to 0.58. | ||
CK: R > 0.99, Slope 0.962-1.053, Intercept -16.5 to 1.1. (See pages 14-15) | ||
Detection Limits (LoB, LoD, LoQ) | Low values demonstrating capability to measure analytes at clinically relevant low concentrations. | BUN: LoB 1.53, LoD 1.97, LoQ 3.0 mg/dL. |
Creatinine: LoB 0.14, LoD 0.18, LoQ 0.33 mg/dL. | ||
Uric Acid: LoB 1.11, LoD 1.34, LoQ 1.50 mg/dL. | ||
CK: LoB 4.68, LoD 8.30, LoQ 11.0 U/L. (See page 16) | ||
Linearity | Wide linear range covering clinical needs, with high correlation. | BUN: Linear to 100.0 mg/dL, R² 0.9991. |
Creatinine: Linear to 25.0 mg/dL, R² 0.9981. | ||
Uric Acid: Linear to 16.0 mg/dL, R² 0.9939. | ||
CK: Linear to 1350.0 U/L, R² 0.9975. (See page 16) | ||
Interferences | No significant interference at specified levels of common interferents. | Demonstrated no significant interference from icterus, hemolysis, lipemia/triglycerides, and ascorbic acid at clinically relevant concentrations for all four analytes. (See page 17) |
Studies Proving Acceptance Criteria:
The studies are described under "Performance Data" and "Device Comparison with Predicate" sections of the 510(k) summary. These studies aim to demonstrate substantial equivalence to the previously cleared predicate device (Alfa Wassermann ACE BUN/Urea Reagent, ACE Creatinine Reagent, ACE Uric Acid Reagent, and ACE CK Reagents, K930104).
2. Sample Size Used for the Test Set and Data Provenance
-
Test Set (Matrix Comparison: Serum vs. Plasma):
- BUN: 95 pairs (ACE), 96 pairs (Alera), 51 pairs (Axcel)
- Creatinine: 102 pairs (ACE), 102 pairs (Alera), 55 pairs (Axcel)
- Uric Acid: 97 pairs (ACE), 95 pairs (Alera), 55 pairs (Axcel)
- CK: 94 pairs (ACE), 96 pairs (Alera), 55 pairs (Axcel)
- Data Provenance: The document states "In-House Precision" and "In-House Matrix Comparison". This typically implies that the data was generated within the manufacturer's laboratory or a testing facility under their control. The country of origin is not explicitly stated but is implicitly the US, given the 510(k) submission to the FDA. The data is retrospective, as it's being used to characterize reagent performance.
-
Test Set (POL - Method Comparison):
- BUN: 53-54 samples per POL lab for comparison with In-House ACE.
- Creatinine: 51 samples per POL lab for comparison with In-House ACE.
- Uric Acid: 49 samples per POL lab for comparison with In-House ACE.
- Creatinine Kinase: 48-50 samples per POL lab for comparison with In-House ACE.
- Data Provenance: "POL - Method Comparison" indicates data from Physician Office Laboratories (POLs), likely external to the main testing facility but still considered part of the overall validation. The document refers to "In-House ACE (x) vs. POL 1 ACE (y)", "POL 2 ACE (y)", etc., indicating comparisons against internal reference methods. The data is retrospective.
-
Test Set (Detection Limits, Linearity, Interferences, Alera Precision): The sample sizes for these specific studies are not explicitly detailed in the provided summary beyond "Low level tested," "Upper level tested," and "number of replicates for precision measurements (i.e. '3.2, 4.0%') implies multiple measurements. These are likely in-house, retrospective studies.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not provided in the document. For in vitro diagnostic devices like these reagents, the "ground truth" is typically established by reference methods or validated comparative methods, often run on established clinical chemistry analyzers. The expertise lies in operating these reference instruments and ensuring proper laboratory practices, rather than expert interpretation of images or clinical cases.
4. Adjudication Method for the Test Set
This concept is not applicable to this type of device. Adjudication methods (like 2+1, 3+1) are common in studies involving subjective interpretations (e.g., medical image analysis by radiologists) where discrepancies among readers need to be resolved to establish ground truth. For quantitative IVD reagents, the reference method provides a direct numerical result, not a subjective interpretation requiring adjudication.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
This is not applicable to this type of device. MRMC studies are used to assess the effectiveness of an AI system (or any diagnostic aid) for human readers, particularly in medical imaging. The current device is a diagnostic reagent, which directly measures chemical concentrations, not an AI intended to assist human interpretation of cases.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
This is not applicable in the context of an IVD reagent. The "algorithm" here is the chemical reaction and photometric measurement itself. The performance data presented (precision, linearity, method comparison, etc.) is the standalone performance of the reagent on the specified analyzers, without human interpretive input altering the result.
7. Type of Ground Truth Used
The ground truth for all performance studies (precision, matrix comparison, method comparison, linearity) is established by comparison against a reference method or a substantially equivalent predicate method performed on existing, validated clinical chemistry analyzers (specifically, the predicate ACE Clinical Chemistry System and the candidate ACE, ACE Alera, and ACE Axcel systems themselves acting as the "reference" for their own performance claims, and for method comparisons, the "In-House ACE" results). This is a common and accepted approach for demonstrating substantial equivalence for IVD reagents.
8. Sample Size for the Training Set
This information is not provided and is generally not applicable in the way it is asked for AI/ML devices. These are chemical reagents, not AI/ML algorithms that require "training sets" in the conventional sense of machine learning. The development process would involve formulation, optimization, and internal testing to define assay parameters, which is a different concept than an AI training set.
9. How the Ground Truth for the Training Set Was Established
As stated above, the concept of a "training set" with established ground truth in the AI/ML sense is not applicable to these chemical reagents. The "ground truth" during their development and optimization would be based on established analytical chemistry principles and performance measurements against known standards or reference materials.
Ask a specific question about this device
(268 days)
ACE CK REAGENT,ACE BUN/UREA REAGENT,ACE URIC ACID REAGENT,ACE CREATININE REAGENT
The ACE BUN/Urea Reagent is intended for the quantitative determination of blood urea nitrogen (BUN) concentration in serum using the ACE Axcel Clinical Chemistry System. BUN measurements are used in the diagnosis and treatment of certain renal and metabolic diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
The ACE Creatinine Reagent is intended for the quantitative determination of creatinine concentration in serum using the ACE Axcel Clinical Chemistry System. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
The ACE Uric Acid Reagent is intended for the quantitative determination of uric acid concentration in serum using the ACE Axcel Clinical Chemistry System. Uric acid measurements are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions and of patients receiving cytotoxic drugs. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
The ACE CK Reagent is intended for the quantitative determination of creatine kinase activity in serum using the ACE Axcel Clinical Chemistry System. Measurement of creatine kinase is used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
The ACE Axcel Clinical Chemistry System consists of two major components, the chemistry instrument and an integrated Panel PC. The instrument accepts the physical patient samples, performs the appropriate optical or potentiometric measurements on those samples and communicates that data to an integral Panel PC. The Panel PC uses keyboard or touch screen input to manually enter a variety of data, control and accept data from the instrument, manage and maintain system information and generate reports relative to patient status and instrument performance. The Panel PC also allows remote download of patient requisitions and upload of patient results via a standard interface.
In the ACE BUN/Urea Reagent assay, urea in serum is hydrolyzed to yield ammonia and carbon dioxide in the presence of urease. The ammonia formed then reacts with 2-oxoglutarate and NADH in the presence of glutamate dehydrogenase to yield glutamate and NAD. Two moles of NADH are oxidized for each mole of urea present. NADH absorbs strongly at 340 nm, whereas NAD+ does not. The initial rate of decrease in absorbance, monitored bichromatically at 340 nm/647 nm, is proportional to the urea concentration in the sample.
In the ACE Creatinine Reagent assay, creatinine reacts with picric acid in an alkaline medium to form a red-orange colored complex, which absorbs strongly at 505 nm. The rate of complex formation, determined by measuring the increase in absorbance bichromatically at 505 nm/573 nm during a fixed time interval, is directly proportional to the creatinine concentration in the sample.
In the ACE Uric Acid Reagent assay, uric acid in serum is oxidized by uricase to allantoin and hydrogen peroxide. The hydrogen peroxide then acts to oxdatively couple dichlorohydroxybenzene sulfonic acid and 4-aminoantipyrine in a reaction catalyzed by peroxidase, producing a red colored quinoneimine complex, which absorbs strongly at 505 nm. The amount of chromogen formed, determined by measuring the increase in absorbance bichromatically at 505 nm/610 nm, is directly proportional to the uric acid concentration in the sample.
In the ACE CK Reagent assay, serum creatine kinase initiates the conversion of creatine phosphate to creatine with the transfer of a phosphate group to adenosine diphosphate (ADP), forming ATP. The ATP is then used in the phosphorylation of D-glucose to form D-glucose-6-phosphate and ADP. This reaction is catalyzed by hexokinase. The enzyme glucose-6-phosphate dehydrogenase catalyzes the reduction of D-glucose-6-phosphate and nicotinamide adenine dinucleotide phosphate (NADP+). The series of reactions triggered by serum creatine kinase and ending in the formation of NADPH. NADPH strongly absorbs at 340 nm, whereas NADP+ does not. Therefore, the rate of conversion of NADP+ to NADPH can be determined by monitoring the increase in absorbance bichromatically at 340 nm/378 nm. This rate of conversion from NADP+ to NADPH is a function of the activity of CK in the sample.
The ACE BUN/Urea Reagent consists of a single reagent bottle. The reagent contains alpha-ketoglutarate, urease, glutamate dehydrogenase, adenosine diphosphate (ADP), nicotinamide adenine dinucleotide and reduced (NADH).
The ACE Creatinine Reagent consists of two reagent bottles (Sodium Hydroxide Reagent and Picric Acid Reagent). The Sodium Hydroxide Reagent (R1) contains sodium hydroxide. The Picric Acid Reagent (R2) contains picric Acid.
The ACE Uric Acid Reagent consists of a single reagent bottle. The reagent contains 4-aminoantipyrine, dichlorohydroxybenzene sulfonic acid, peroxidase and uricase.
The ACE CK Reagent consists of two reagent bottles (Buffer and Substrate). The Buffer Reagent (R1) contains: imidazole buffer, glucose, N-acetyl-cysteine, magnesium acetate, EDTA, NADP and hexokinase. The Substrate Reagent (R2) contains: creatine phosphate, ADP, AMP, diadenosine pentaphosphate, EDTA and glucose-6-phosphate dehydrogenase.
This 510(k) summary describes the analytical performance of the Alfa Wassermann ACE BUN, Creatinine, Uric Acid, and CK Reagents when used with the ACE Axcel Clinical Chemistry System. The study aims to demonstrate substantial equivalence to a predicate device by evaluating precision, accuracy, and detection limits.
1. Table of Acceptance Criteria (Implied) and Reported Device Performance
The acceptance criteria for this type of device are generally understood to be that the performance of the new device (ACE Axcel System with new reagents) should be comparable to or better than a legally marketed predicate device (Alfa Wassermann ACE Clinical Chemistry System). While explicit numerical acceptance criteria are not strictly stated as "acceptance criteria" but rather as "reported performance," the goal is to show the device performs within acceptable analytical limits for clinical chemistry assays and is strongly correlated with the predicate.
Reagent (Analyte) | Performance Metric | Implied Acceptance Criteria (Comparison to Predicate) | Reported Device Performance (ACE Axcel vs. ACE Clinical Chemistry System) |
---|---|---|---|
ACE BUN/Urea | Precision (Within-run CV) | 0.975 (strong correlation) | 0.9963 (lab), 0.9982 to 0.9988 (POL) |
Accuracy (Slope CI) | Close to 1 (e.g., 0.95-1.05) | 0.995 to 1.028 (lab), 0.983 to 1.039 (POL) | |
Accuracy (Intercept CI) | Close to 0 (e.g., -5 to 5) | -0.3 to 0.6 (lab), -0.7 to 1.6 (POL) | |
Detection Limit | Clinically relevant low level | 1.1 mg/dL | |
ACE Creatinine | Precision (Within-run CV) | 0.975 (strong correlation) | 0.9998 (lab), 0.9994 to 0.9998 (POL) |
Accuracy (Slope CI) | Close to 1 (e.g., 0.95-1.05) | 0.975 to 0.983 (lab), 0.961 to 1.027 (POL) | |
Accuracy (Intercept CI) | Close to 0 (e.g., -0.1 to 0.1) | -0.022 to 0.010 (lab), -0.136 to 0.001 (POL) | |
Detection Limit | Clinically relevant low level | 0.19 mg/dL | |
ACE Uric Acid | Precision (Within-run CV) | 0.975 (strong correlation) | 0.9958 (lab), 0.9858 to 0.9961 (POL) |
Accuracy (Slope CI) | Close to 1 (e.g., 0.95-1.05) | 1.023 to 1.060 (lab), 0.972 to 1.054 (POL) | |
Accuracy (Intercept CI) | Close to 0 (e.g., -0.5 to 0.5) | -0.18 to 0.07 (lab), -0.31 to 0.28 (POL) | |
Detection Limit | Clinically relevant low level | 1.13 mg/dL |
Note: Acceptance criteria are implied based on typical expectations for clinical chemistry assays and the intent to demonstrate substantial equivalence to a predicate device. Specific numerical targets for acceptance were not explicitly stated in the provided text, but the strong correlation and low CVs indicate meeting such criteria.
2. Sample Sizes Used for the Test Set and Data Provenance
-
ACE BUN/Urea Reagent:
- Accuracy (Correlation Study): 113 samples (clinical laboratory), and patient correlation studies at three Physician Office Laboratory (POL) sites (number of samples not explicitly stated for POL, but implied to be sufficient for regression analysis).
- Precision: Four BUN levels over 22 days (laboratory study), and three POL sites over 5 days (levels not specified for POL).
- Data Provenance: Not explicitly stated, but clinical laboratory and Physician Office Laboratory (POL) settings are mentioned, suggesting human serum samples. Whether these were retrospective or prospective is not specified, but typically, method comparison studies use prospective or collected retrospective clinical samples.
-
ACE Creatinine Reagent:
- Accuracy (Correlation Study): 136 samples (clinical laboratory), and patient correlation studies at three POL sites (number of samples not explicitly stated for POL).
- Precision: Four creatinine levels over 22 days (laboratory study), and three POL sites over 5 days (levels not specified for POL).
- Data Provenance: Not explicitly stated, but clinical laboratory and POL settings are mentioned, suggesting human serum samples.
-
ACE Uric Acid Reagent:
- Accuracy (Correlation Study): 106 samples (clinical laboratory), and patient correlation studies at three POL sites (number of samples not explicitly stated for POL).
- Precision: Four uric acid levels over 22 days (laboratory study), and three POL sites over 5 days (levels not specified for POL).
- Data Provenance: Not explicitly stated, but clinical laboratory and POL settings are mentioned, suggesting human serum samples.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
Not applicable. This is an in vitro diagnostic (IVD) device for quantitative measurement of chemical analytes (BUN, Creatinine, Uric Acid, CK) in serum. The 'ground truth' for such devices is established by a reference method or a legally marketed predicate device, not by expert interpretation of images or clinical findings.
4. Adjudication Method for the Test Set
Not applicable. As noted above, this is an IVD device for quantitative chemical analysis. Adjudication methods are typically used for qualitative or interpretive diagnostic devices where human expert disagreement might occur (e.g., radiology, pathology). Here, the comparison is directly numerical between the candidate device and the predicate device.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
Not applicable. This is an IVD device for laboratory chemical analysis, not an imaging or interpretive diagnostic device that involves human readers or AI assistance in interpretation.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, in a sense. The described studies evaluate the performance of the algorithm/system only (the ACE Axcel Clinical Chemistry System with the new reagents) in quantifying the analytes in serum. The performance data (precision, accuracy, detection limit) are intrinsic to the device's analytical capability, independent of human interpretation of the results for the purpose of generating the values themselves. While trained personnel operate the system, the analytical performance is measured as a standalone function of the device.
7. The Type of Ground Truth Used
The "ground truth" for the accuracy studies was established by comparing the results from the Alfa Wassermann ACE Axcel Clinical Chemistry System (the new device, 'y') to a legally marketed predicate device, the Alfa Wassermann ACE Clinical Chemistry System ('x'). This is a common method for IVD substantial equivalence, where the predicate is considered the accepted reference for performance. For detection limits, it would typically involve analyzing samples with known, very low concentrations of the analytes or diluting higher concentration samples to determine the lowest measurable level.
8. The Sample Size for the Training Set
The provided text describes performance validation studies, not the development or training of an algorithm in the machine learning sense. Therefore, there is no "training set" for an algorithm to learn from in this context. The study focuses on verifying the performance of the already-developed reagent and instrument system.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" in the machine learning sense for this type of IVD device submission.
Ask a specific question about this device
(197 days)
ACE AXCEL CLINICAL CHEMISTRY SYSTEM,ACE ALBUMIN REAGENT,ACE TOTAL PROTEIN REAGENT,ACE BUN/UREANITROGEN
The ACE Axcel Clinical Chemistry System is an automated, discrete, bench-top, random access analyzer that is intended for in vitro diagnostic use in the quantitative measurement of general chemistry assays for clinical use in physician office laboratories or clinical laboratories.
The ACE Axcel Clinical System includes an Ion Selective Electrode (ISE) module for the measurement of sodium, potassium and chloride in serum. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
- Sodium measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance
- Potassium measurements are used to monitor electrolyte balance and in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
- Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
The ACE Glucose Reagent is intended for the quantitative determination of glucose concentration in serum using the ACE Axcel Clinical Chemistry System. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
The ACE Axcel Clinical Chemistry System consists of two major components, the chemistry instrument and an integrated Panel PC. The instrument accepts the physical patient samples, performs the appropriate optical or potentiometric measurements on those samples and communicates that data to an integral Panel PC. The Panel PC uses keyboard or touch screen input to manually enter a variety of data, control and accept data from the instrument, manage and maintain system information and generate reports relative to patient status and instrument performance. The Panel PC also allows remote download of patient requisitions and upload of patient results via a standard interface.
In the ACE Glucose Reagent assay, glucose in serum reacts with adenosine triphosphate in the presence of hexokinase and magnesium with the formation of glucose-6-phosphate and adenosine diphosphate. Glucose-6-phosphate dehydrogenase catalyzes the oxidation of glucose-6-phosphate with NAD+ to form 6-phosphogluconate and NADH. NADH absorbs strongly at 340 nm, whereas NAD+ does not. The total amount of NADH formed is proportional to the concentration of glucose in the sample. The increase in absorbance is measured bichromatically at 340 nm/378 nm.
The ACE Ion Selective Electrode (ISE) Module is used with ACE CAL A and CAL B Calibration Solutions in the performance of a two-point calibration in order to measure concentrations of sodium, potassium and chloride in undiluted serum. The ISE module uses a potentiometric method to simultaneously measure sodium, potassium and chloride in undiluted serum. Each electrode uses an ion-specific membrane to measure the difference in ionic concentration between an inner electrolyte solution and the sample. This difference causes an electro-chemical potential to form on the membrane of the active electrode. The connection of the amplifier and ground (reference electrode) to the ion selective electrode forms the measuring system. The two-point calibration with CAL A and CAL B with precisely known ion concentrations (two-point calibration) and the measured voltage difference of the sample and CAL A are used to determine the ion concentration in the sample.
Here's a summary of the acceptance criteria and study information for the Alfa Wassermann ACE Axcel Clinical Chemistry System, ACE Ion Selective Electrode (ISE) Module, and ACE Glucose Reagent, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implied by the "Accuracy" data, where the device (y) is compared to a predicate device (x). The close correlation, with slopes near 1 and intercepts near 0, along with high correlation coefficients, indicates acceptable accuracy. Precision is evaluated by the Coefficient of Variation (CV%).
Measurement (Reagent/Module) | Performance Metric | Acceptance Criteria (Implied by Predicate Equivalence) | Reported Device Performance (ACE Axcel Clinical Chemistry System) |
---|---|---|---|
ACE Glucose Reagent | Precision (Within-run CV) | (Not explicitly stated, but low CV is desirable) | 1.0 to 1.4% (lab setting, 4 levels, 22 days); 0.3 to 2.2% (POL sites, 3 locations, 5 days) |
Precision (Total CV) | (Not explicitly stated, but low CV is desirable) | 1.0 to 1.9% (lab setting); 0.5 to 2.2% (POL sites) | |
Accuracy (Correlation Coefficient) | Close to 1.0 | 0.9998 (lab setting, 122 samples); 0.9992 to 0.9998 (POL sites, 3 locations) | |
Accuracy (Slope) | Close to 1.0 | 1.001 to 1.009 (lab setting); 0.972 to 1.021 (POL sites) | |
Accuracy (Intercept) | Close to 0 | -1.5 to 0.1 (lab setting); -3.0 to 4.1 (POL sites) | |
Detection Limit | (Not explicitly stated, but low is desirable) | 3.1 mg/dL | |
ACE Axcel Sodium ISE | Precision (Within-run CV) | (Not explicitly stated, but low CV is desirable) | 0.4 to 1.0% (lab setting, 4 levels, 21 days); 0.6 to 1.0% (POL sites, 3 locations, 5 days) |
Precision (Total CV) | (Not explicitly stated, but low CV is desirable) | 0.8 to 1.4% (lab setting); 0.8 to 1.4% (POL sites) | |
Accuracy (Correlation Coefficient) | Close to 1.0 | 0.9963 (lab setting, 113 samples); 0.9917 to 0.9995 (POL sites, 3 locations) | |
Accuracy (Slope) | Close to 1.0 | 0.992 to 1.024 (lab setting); 0.989 to 1.067 (POL sites) | |
Accuracy (Intercept) | Close to 0 | -3.60 to 0.92 (lab setting); -8.85 to 2.30 (POL sites) | |
ACE Axcel Potassium ISE | Precision (Within-run CV) | (Not explicitly stated, but low CV is desirable) | 0.6 to 3.5% (lab setting, 4 levels, 21 days); 1.0 to 1.6% (POL sites, 3 locations, 5 days) |
Precision (Total CV) | (Not explicitly stated, but low CV is desirable) | 1.3 to 3.5% (lab setting); 1.1 to 1.6% (POL sites) | |
Accuracy (Correlation Coefficient) | Close to 1.0 | 0.9974 (lab setting, 115 samples); 0.9973 to 0.9996 (POL sites, 3 locations) | |
Accuracy (Slope) | Close to 1.0 | 0.989 to 1.015 (lab setting); 0.960 to 1.035 (POL sites) | |
Accuracy (Intercept) | Close to 0 | -0.050 to 0.095 (lab setting); -0.194 to 0.216 (POL sites) | |
ACE Axcel Chloride ISE | Precision (Within-run CV) | (Not explicitly stated, but low CV is desirable) | 0.5 to 1.0% (lab setting, 4 levels, 21 days); 0.9 to 1.5% (POL sites, 3 locations, 5 days) |
Precision (Total CV) | (Not explicitly stated, but low CV is desirable) | 1.1 to 1.5% (lab setting); 1.1 to 2.6% (POL sites) | |
Accuracy (Correlation Coefficient) | Close to 1.0 | 0.9855 (lab setting, 111 samples); 0.9885 to 0.9996 (POL sites, 3 locations) | |
Accuracy (Slope) | Close to 1.0 | 0.939 to 1.002 (lab setting); 0.976 to 1.088 (POL sites) | |
Accuracy (Intercept) | Close to 0 | -1.07 to 5.63 (lab setting); -8.16 to 2.22 (POL sites) |
2. Sample Size Used for the Test Set and Data Provenance
-
ACE Glucose Reagent:
- Accuracy (Correlation Study): 122 samples (ranging from 6 to 729 mg/dL).
- Accuracy (Patient Correlation Studies): Data from three separate Physician Office Laboratory (POL) sites. Specific sample numbers per POL site are not provided, but the combined sites yielded multiple correlation coefficients, standard error estimates, and confidence intervals for slope and intercept.
- Provenance: Not explicitly stated, but likely from a laboratory setting and Physician Office Laboratories, presumably within the US given the submission to the FDA. The nature of the samples (e.g., patient samples, control materials) is not specified as prospective or retrospective.
-
ACE Axcel Sodium ISE:
- Accuracy (Correlation Study): 113 samples (ranging from 45.1 to 194.0 mmol/L).
- Accuracy (Patient Correlation Studies): Data from three separate Physician Office Laboratory (POL) sites.
- Provenance: Same as Glucose – likely US lab/POL, nature of samples not specified.
-
ACE Axcel Potassium ISE:
- Accuracy (Correlation Study): 115 samples (ranging from 1.57 to 14.20 mmol/L).
- Accuracy (Patient Correlation Studies): Data from three separate Physician Office Laboratory (POL) sites.
- Provenance: Same as Glucose – likely US lab/POL, nature of samples not specified.
-
ACE Axcel Chloride ISE:
- Accuracy (Correlation Study): 111 samples (ranging from 63.4 to 176.0 mmol/L).
- Accuracy (Patient Correlation Studies): Data from three separate Physician Office Laboratory (POL) sites.
- Provenance: Same as Glucose – likely US lab/POL, nature of samples not specified.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
Not applicable. For this type of in vitro diagnostic device, "ground truth" is established by comparing the performance of the new device to a legally marketed predicate device (Alfa Wassermann ACE Clinical Chemistry System and ACE Reagents) using quantitative measurements, not by expert interpretation. The predicate device itself acts as the reference method in these correlation studies.
4. Adjudication Method for the Test Set
Not applicable. As noted above, this is a quantitative comparison against a predicate device, not an interpretation-based ground truth requiring adjudication.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a clinical chemistry analyzer, an automated system for measuring analytes in samples. It does not involve human readers interpreting images or data where AI assistance would be relevant in the context of MRMC studies.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the performance studies described are inherently "standalone" in the sense that they evaluate the performance of the ACE Axcel Clinical Chemistry System (including its reagents and ISE module) in producing quantitative results independently. The accuracy studies compare its output directly against a predicate device's output. There isn't a "human-in-the-loop" aspect to the core measurement performance itself.
7. The Type of Ground Truth Used
The "ground truth" for the accuracy studies is the quantitative result obtained from the predicate device (Alfa Wassermann ACE Clinical Chemistry System and ACE Reagents). The studies are correlation studies comparing the new device's measurements (y) to the predicate device's measurements (x).
8. The Sample Size for the Training Set
Not applicable. This document describes a traditional medical device (clinical chemistry analyzer), not a machine learning or AI-based device that typically has a "training set." The device is intended to perform measurements based on established chemical and electrochemical principles.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" in the context of this device's development as described in the provided text.
Ask a specific question about this device
(273 days)
ACE UREA NITROGEN REAGENT, CALCIUM-ARSEAZO REAGENT, CREATININE REAGENT, INORGANIC PHOSPHORUS U.V.
The ACE Urea Nitrogen Reagent is intended for the quantitative determination of urea nitrogen concentration in urine using the ACE and ACE Alera Clinical Chemistry Systems. Urea nitrogen measurements are used in the diagnosis and treatment of certain renal and metabolic diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
The ACE Calcium-Arsenazo Reagent is intended for the quantitative determination of calcium in urine using the ACE and ACE Alera Clinical Chemistry Systems. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
The ACE Creatinine Reagent is intended for the quantitative determination of creatinine in urine using the ACE and ACE Alera Clinical Chemistry Systems. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis and as a calculation basis for measuring other urine analytes. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
The ACE Inorganic Phosphorus U.V. Reagent is intended for the quantitative determination of inorganic phosphorus concentration in urine using the ACE and ACE Alera Clinical Chemistry Systems. Measurements of inorganic phosphorus are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases and vitamin D imbalance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Alfa Wassermann Diagnostic Technologies, LLC ACE Urine Standard is intended for the calibration of quantitative urine reagents on Alfa Wassermann Clinical Chemistry Systems. For in vitro diagnostic use only.
In the ACE Urea Nitrogen Reagent assay, urea in urine is hydrolyzed to yield ammonia and carbon dioxide in the presence of urease. The ammonia formed then reacts with 2-oxoglutarate and NADH in the presence of glutamate dehydrogenase to yield glutamate and NAD. Two moles of NADH are oxidized for each mole of urea present. NADH absorbs strongly at 340 nm, whereas NAD+ does not. The initial rate of decrease in absorbance, monitored bichromatically at 340 nm/647 nm, is proportional to the urea concentration in the urine sample.
In the ACE Calcium-Arsenazo Reagent assay, calcium reacts with Arsenazo III in an acidic solution to form a blue-purple colored complex, which is measured bichromatically at 647 nm/692 nm. The intensity of color produced is directly proportional to the calcium concentration in the urine sample.
In the ACE Creatinine Reagent assay, creatinine reacts with picric acid in an alkaline medium to form a red-orange colored complex, which absorbs strongly at 505 nm. The rate of complex formation, determined by measuring the increase in absorbance bichromatically at 505 nm/573 nm during a fixed time interval, is directly proportional to the creatinine concentration in the urine sample.
In the ACE Inorganic Phosphorus U.V. Reagent assay, under acidic conditions, inorganic phosphorus reacts with ammonium molybdate to form an unreduced phosphomolybdate complex, which absorbs strongly at 340 nm The increase in absorbance measured bichromatically at 340 nm/378 nm, is directly proportional to the amount of phosphorus in the urine sample.
The ACE urine standard is a liquid, aqueous-based, ready-to-use calibration solution with known gravimetric concentrations of several analytes.
Here's an analysis of the acceptance criteria and supporting studies for the Alfa Wassermann ACE Reagents and Urine Standard, based on the provided 510(k) summary:
Acceptance Criteria and Device Performance
The acceptance criteria are not explicitly stated as distinct numerical targets in the provided document. Instead, the document presents performance data from studies and implies that these results demonstrate substantial equivalence to predicate devices, thus meeting acceptable performance standards. The reported device performance metrics are precision (within-run CV and total CV), accuracy (correlation coefficient, standard error estimate, confidence interval slope, and confidence interval intercept compared to predicate methods), and detection limit.
Table of Acceptance Criteria (Implied) and Reported Device Performance for Urine Analytes
Analyte | Performance Metric | Implied Acceptance Criteria (e.g., comparable to predicate) | Reported Device Performance (ACE System & ACE Alera System) |
---|---|---|---|
Urea Nitrogen | Precision (Within-run CV) | Low CV (e.g., ≤ 5%) | Lab: 1.1 - 2.3% (within-run), 2.0 - 3.5% (total CV) |
POL: 0.9 - 3.8% (within-run), 1.6 - 4.6% (total CV) | |||
Accuracy (Correlation Coefficient) | High correlation (e.g., ≥ 0.97) | ACE: 0.9769 (r) | |
ACE Alera: 0.9805 (r) | |||
POLs: 0.9853 - 0.9895 (r) | |||
- Standard Error Estimate | Low (e.g., |
Ask a specific question about this device
(463 days)
ELITECH CLINICAL SYSTEMS PHOSPHORUS, URIC ACID MONO SL AND UREA UV SL REAGENTS
ELITech Clinical Systems PHOSPHORUS reagent is for the quantitative in vitro diagnostic determination of inorganic phosphorus in human serum and plasma on the Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Measurements of phosphorus (inorganic) are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases, and vitamin D imbalance.
ELITech Clinical Systems URIC ACID MONO SL reagent is for the quantitative in vitro diagnostic determination of uric acid in human serum and plasma on the Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Measurements obtained by this device are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.
ELITech Clinical Systems UREA UV SL reagent is for the quantitative in vitro diagnostic determination of urea nitrogen in human serum and plasma on the Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases.
ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzer.
ELITech Clinical Systems ELITROL I is a multi-parametric control serum for in vitro diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzer. ELITech Clinical Systems ELITROL II is a multi-parametric control serum for in vitro diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior analyzer and the Vital Scientific Flexor Junior analyzers.
ELITech Clinical Systems ELICAL 2 is a lyophilized calibrator based on human serum containing constituents to ensure optimal calibration. ELICAL 2 is prepared exclusively from the blood of donors tested individually and found to be negative for HbsAg and to antibodies to HCV and HIV according to FDA-approved methods or methods in compliance with the European Directive 98/79/EC, Annex II, List A.
The device for this submission is available as kit only. It consists of 2 reagents: Reagent 1 contains Tris buffer (pH 7.60), Adenosine diphosphate potassium salt (ADP), alpha-Ketoglutarate, Urease (jack bean), Glutamate dehydrogenase (GIDH) (bovine liver) and sodium azide. Reagent 2 contains NADH and sodium azide
The device for this submission is available as kit only. It consists of 1 reagent R. Reagent R consists of sulfuric acid and ammonium molybdate.
The device for this submission is available as kit only. It consists of 1 reagent R. Reagent R consists of Phosphate buffer (pH 7.0), N-Ethyl-N-(2-Hydroxy-3-Sulfopropyl) m-Toluidine (EHSPT), Ferrocyanide, Amino-4-antipyrine (4-AAP), Uricase (microorganisms), Peroxidase (horseradich) and sodium azide.
ELITech Clinical Systems ELITROL I and ELITROL II are two level quality control products consisting of lyophilized human serum containing constituents at desired levels. Elitrol I and Elitrol II are prepared exclusively from the blood of donors tested individually and found to be negative for HbsAg and to antibodies to HCV and HIV according to FDA-approved methods or methods in compliance with the European Directive 98/79/EC, Annex II, List A.
The provided text describes several in vitro diagnostic (IVD) devices: ELITech Clinical Systems ELICAL 2 (calibrator), ELITech Clinical Systems UREA UV SL (reagent), ELITech Clinical Systems PHOSPHORUS (reagent), ELITech Clinical Systems URIC ACID MONO SL (reagent), and ELITech Clinical Systems ELITROL I and ELITROL II (controls).
The studies performed for these devices are primarily method comparison and precision studies, focusing on demonstrating substantial equivalence to predicate devices rather than proving a specific clinical utility through a multi-reader multi-case study or a standalone algorithm performance study.
Here's an analysis based on your requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by demonstrating equivalence to the predicate devices. The performance metrics presented are directly compared.
For ELITech Clinical Systems UREA UV SL (K100263):
Performance Metric | Acceptance Criteria (Predicate: ABX PENTRA UREA CP) | Reported Device Performance (ELITech Clinical Systems UREA UV SL) |
---|---|---|
Measuring Range | 1.03 to 140.3 mg/dL (Automatic post-dilution: 701.5 mg/dL) | 4.7 to 140.0 mg/dL (Extended measuring range: 140.0 to 670.0 mg/dL) |
Limit of Detection (LoD) | Not explicitly stated | 0.3 mg/dL |
Limit of Quantification (LoQ) | 0.9 mg/dL | 2.3 mg/dL |
Precision | ||
Within Run (Level 7.3 mg/dL) | 2.27% (for 18.7 mg/dL) | CV=2.1% |
Within Run (Level 29.2 mg/dL) | 1.66% (for 72.8 mg/dL) | CV=0.8% |
Within Run (Level 72.4 mg/dL) | 2.76% (for 6.0 mg/dL) etc. | CV=0.7% |
Total (Level 7.3 mg/dL) | 2.14% (for 18.5 mg/dL) | CV=2.8% |
Total (Level 29.2 mg/dL) | 1.93% (for 71.7 mg/dL) | CV=1.3% |
Total (Level 72.4 mg/dL) | 2.14% (for 19.2 mg/dL) etc. | CV=1.6% |
Method Comparison | y=0.99 x - 0.06 mg/dL, r²=0.996, range: 1.03 to 138.89 mg/dL | y=0.991 x + 0.6 mg/dL, r=0.999, range: 4.4 to 139.8 mg/dL |
Interference (Hemoglobin) | No significant influence up to 460 mg/dL | No significant interference up to 500 mg/dL |
Interference (Triglycerides) | No significant influence up to 612.5 mg/dL | No significant interference up to 614 mg/dL triglyceride equivalent |
Interference (Total Bilirubin) | No significant influence up to 22.23 mg/dL | No significant interference up to 30 mg/dL (unconjugated) / 29.5 mg/dL (conjugated) |
Calibration Frequency | 8 days | 7 days |
On-board Stability | 70 days | 14 days |
For ELITech Clinical Systems PHOSPHORUS (K100263):
Performance Metric | Acceptance Criteria (Predicate: ABX PENTRA PHOSPHORUS CP) | Reported Device Performance (ELITech Clinical Systems PHOSPHORUS) |
---|---|---|
Measuring Range | 0.30 to 24.18 mg/dL (Automatic post-dilution: 96.72 mg/dL) | 2.0 to 20.0 mg/dL |
Limit of Detection (LoD) | 0.28 mg/dL | 0.02 mg/dL |
Limit of Quantification (LoQ) | Not explicitly stated | 1.00 mg/dL |
Precision | ||
Within Run (Level 2.37 mg/dL) | 1.25% (for 4.08 mg/dL) | CV=1.1% |
Within Run (Level 4.80 mg/dL) | 0.77% (for 6.34 mg/dL) | CV=1.5% |
Within Run (Level 9.55 mg/dL) | 2.48% (for 2.39 mg/dL) etc. | CV=1.7% |
Total (Level 2.37 mg/dL) | 2.50% (for 4.01 mg/dL) | CV=1.9% |
Total (Level 4.80 mg/dL) | 1.82% (for 6.35 mg/dL) | CV=1.7% |
Total (Level 9.55 mg/dL) | 3.56% (for 2.50 mg/dL) etc. | CV=2.2% |
Method Comparison | y=1.04x + 0.15 mg/dL, r²=0.998, range: 0.30 to 24.08 mg/dL | y=0.999 - 0.09 mg/dL, r=0.999, range: 2.02 to 20.08 mg/dL |
Interference (Hemoglobin) | No significant influence up to 125 mg/dL | No significant interference up to 50 mg/dL |
Interference (Triglycerides) | No significant influence up to 262.5 mg/dL | No significant interference up to 732 mg/dL |
Interference (Total Bilirubin) | No significant influence up to 6 mg/dL | No significant interference up to 15 mg/dL (unconjugated) / 1.5 mg/dL (conjugated) |
Calibration Frequency | 34 days | 28 days |
On-board Stability | 70 days | 28 days |
For ELITech Clinical Systems URIC ACID MONO SL (K100263):
Performance Metric | Acceptance Criteria (Predicate: ABX PENTRA URIC ACID CP) | Reported Device Performance (ELITech Clinical Systems URIC ACID MONO SL) |
---|---|---|
Measuring Range | 0.18 to 25.00 mg/dL (Automatic post-dilution: 75.00 mg/dL) | 1.5 to 25.0 mg/dL (Extended measuring range: 25 to 78 mg/dL) |
Limit of Detection (LoD) | 0.19 mg/dL | 0.02 mg/dL |
Limit of Quantification (LoQ) | Not explicitly stated | 0.50 mg/dL |
Precision | ||
Within Run (Level 2.49 mg/dL) | 0.45% (for 4.62 mg/dL) | CV=0.8% |
Within Run (Level 5.19 mg/dL) | 0.34% (for 11.63 mg/dL) | CV=1.3% |
Within Run (Level 7.63 mg/dL) | 1.24% (for 2.53 mg/dL) etc. | CV=1.1% |
Total (Level 2.49 mg/dL) | 2.81% (for 4.64 mg/dL) | CV=2.6% |
Total (Level 5.19 mg/dL) | 1.39% (for 11.73 mg/dL) | CV=2.0% |
Total (Level 7.63 mg/dL) | 2.64% (for 4.67 mg/dL) etc. | CV=2.1% |
Method Comparison | y=0.95 x + 0.09 mg/dL, r²=0.996, range: 0.18 to 23.59 mg/dL | y=1.015 x + 0.03 mg/dL, r=0.999, range: 1.49 to 24.40 mg/dL |
Interference (Hemoglobin) | No significant influence up to 500 mg/dL | No significant interference up to 50 mg/dL |
Interference (Triglycerides) | No significant influence up to 612.5 mg/dL | No significant interference up to 1070 mg/dL |
Interference (Total Bilirubin) | No significant influence up to 36 mg/dL | No significant interference up to 30 mg/dL (unconjugated) / 14.8 mg/dL (conjugated) |
Calibration Frequency | 15 days | 28 days |
On-board Stability | 41 days | 28 days |
2. Sample size used for the test set and the data provenance
- UREA UV SL: The method comparison study gives a range of 4.4 to 139.8 mg/dL for the device. The R-value of 0.999 implies a good correlation across the study range. While a specific number of samples isn't explicitly stated, the R-value is usually calculated from a sufficiently large set of patient samples spanning the analytical range. The data provenance is not stated (e.g., country of origin, retrospective/prospective), but it is implied to be clinical samples used for method comparison.
- PHOSPHORUS: The method comparison study gives a range of 2.02 to 20.08 mg/dL for the device. The R-value of 0.999 implies a good correlation. Similar to UREA UV SL, the number of samples is not explicitly stated, and data provenance is not provided but implied to be clinical samples.
- URIC ACID MONO SL: The method comparison study gives a range of 1.49 to 24.40 mg/dL for the device. The R-value of 0.999 implies a good correlation. The number of samples is not explicitly stated, and data provenance is not provided but implied to be clinical samples.
For ELICAL 2, ELITROL I, ELITROL II: These are calibrators and controls; their performance is validated through stability studies and their intended use in calibrating and controlling assays. No "test set" in the sense of patient samples for diagnostic performance is applicable here. The device description for ELICAL 2, ELITROL I, and ELITROL II mentions they are "prepared exclusively from the blood of donors tested individually and found to be negative for HbsAg and to antibodies to HCV and HIV according to FDA-approved methods or methods in compliance with the European Directive 98/79/EC, Annex II, List A." This suggests human-derived materials.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. These are in vitro diagnostic devices for quantitative measurements, not imaging or subjective diagnostic interpretations requiring expert consensus for ground truth. "Ground truth" in this context refers to the measured values obtained by the predicate device or a reference method for the method comparison studies.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. This is not a study involving human interpretation or adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an in vitro diagnostic device, not an AI-assisted diagnostic tool requiring human reader studies.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This is an IVD reagent and calibrator kit, not an algorithm. The performance data presented (e.g., precision, method comparison, LoD, LoQ) represents the "standalone" analytical performance of the device on the specified analyzer (Selectra Junior/Flexor Junior).
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth for the method comparison studies is the quantitative result obtained from the predicate device (ABX PENTRA UREA CP, ABX PENTRA PHOSPHORUS CP, ABX PENTRA URIC ACID CP) or a reference method assumed to be the "truth" for those studies. For precision studies, there isn't an external "ground truth" but rather a measure of the device's own reproducibility.
8. The sample size for the training set
Not applicable. These are traditional IVD reagents and calibrators, not machine learning algorithms that require a "training set."
9. How the ground truth for the training set was established
Not applicable. As above, no training set for a machine learning algorithm is involved.
Ask a specific question about this device
Page 1 of 4