Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K093965
    Manufacturer
    Date Cleared
    2010-03-17

    (84 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ULTRAVERSE 014 AND 018 BALLOON DILATATION CATHETERS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters are recommended for use in Percutaneous Transluminal Angioplasty of the renal, tibial, popliteal, femoral and peroneal arteries. These catheters are not for use in coronary arteries.

    Device Description

    The subject device, the Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters are small vessel balloon catheters consisting of an over the wire catheter with a balloon fixed at the distal tip. The semi-compliant, low profile balloon has two radiopaque marker bands to delineate the working length of the balloon and aid in balloon placement. The coaxial, over the wire catheter is compatible with 0.014" and 0.018" guidewires.

    AI/ML Overview

    The Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters underwent a series of non-clinical in vitro tests to demonstrate substantial equivalence to a predicate device and ensure they met predetermined acceptance criteria.

    1. Acceptance Criteria and Reported Device Performance:

    The document describes a comprehensive set of in vitro tests performed to evaluate the physical and performance characteristics of the Ultraverse® catheters. The acceptance criteria for these tests are generally implied to be conformity with established standards, guidance documents, and internal risk assessments. The reported performance is that the devices "met all predetermined acceptance criteria" and "demonstrate that the technological characteristics and performance criteria of the Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters are comparable to the predicate device and that it can perform in a manner equivalent to devices currently on the market for the same intended use."

    Here's a table summarizing the tested parameters and the reported outcome:

    Acceptance Criteria (Implied)Reported Device Performance
    Material/Design Specifications:
    Catheter LengthMet acceptance criteria
    Guidewire Inner Lumen DiameterMet acceptance criteria
    Shaft Outer DiameterMet acceptance criteria
    Balloon Outer Diameter at Operating PressureMet acceptance criteria
    Balloon LengthMet acceptance criteria
    Marker Band Spacing/AlignmentMet acceptance criteria
    Tip LengthMet acceptance criteria
    Tip TaperMet acceptance criteria
    Guidewire CompatibilityMet acceptance criteria
    Mechanical Performance:
    DistensibilityMet acceptance criteria
    Balloon FatigueMet acceptance criteria
    Balloon Inflation and DeflationMet acceptance criteria
    Balloon Burst StrengthMet acceptance criteria
    Balloon to Catheter Shaft Bond StrengthMet acceptance criteria
    Joint Strength of Hub to ShaftMet acceptance criteria
    Hub Torsion TestingMet acceptance criteria
    Hub Stress TestingMet acceptance criteria
    Catheter Shaft LeaksMet acceptance criteria
    Catheter ElongationMet acceptance criteria
    Crossing ProfileMet acceptance criteria
    Rewrapping with StyletMet acceptance criteria
    Usability/Visibility:
    Simulated Use: Introducer Sheath CompatibilityMet acceptance criteria
    Simulated Use: TrackabilityMet acceptance criteria
    Marker Band VisibilityMet acceptance criteria
    Compatibility with Standard Luer LocksMet acceptance criteria
    Packaging/Sterility:
    Package IntegrityMet acceptance criteria
    Dye Penetration of PackagingMet acceptance criteria
    Pouch Seal StrengthMet acceptance criteria

    2. Sample Size for the Test Set and Data Provenance:

    The document describes in vitro tests, which are laboratory-based tests. Therefore, the concept of a "test set" in the context of clinical data (like patient data with a specific provenance) does not directly apply here. The sample sizes for each specific in vitro test (e.g., number of balloons tested for burst strength) are not explicitly stated in this document but would have been defined in the individual test protocols. The data provenance is from in vitro laboratory testing, conducted according to FDA Guidance Documents and internal Risk Assessment procedures.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. For in vitro engineering and performance tests of medical devices, "ground truth" is typically established by physical measurements, established engineering standards, and pre-defined acceptance limits, rather than expert consensus on interpretive data (like in image analysis).

    4. Adjudication Method for the Test Set:

    Not applicable. As described above, the acceptance was based on meeting pre-defined physical and engineering specifications and performance criteria, not on human adjudication of medical findings.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No. This document describes in vitro engineering and performance testing for a medical device (balloon catheters), not a diagnostic algorithm or imaging system requiring human interpretation. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant or mentioned.

    6. Standalone (Algorithm Only) Performance Study:

    No. This document does not describe the evaluation of an algorithm or AI system. It focuses on the physical and mechanical performance of a medical device.

    7. Type of Ground Truth Used:

    For the in vitro tests, the "ground truth" was based on established engineering specifications, physical measurements, and performance benchmarks derived from FDA Guidance Documents and internal risk assessments. For example, a balloon's burst strength would be compared against a pre-defined minimum pressure it must withstand.

    8. Sample Size for the Training Set:

    Not applicable. There is no mention of a "training set" as this study is not for a machine learning or AI algorithm. It involves direct physical and functional testing of device prototypes/samples.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. As there is no training set for an AI algorithm mentioned, the process of establishing ground truth for a training set is irrelevant to this submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1