Search Results
Found 1 results
510(k) Data Aggregation
(110 days)
Trident**®** II Acetabular System
The Trident® II Acetabular System is intended for use in total hip arthroplasty and is intended for either primary or revision Total Hip Arthroplasty.
Indications for Use:
• Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
• Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
• Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
• Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
When used with MDM Liners:
• Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
- Dislocation risks
When used with Constrained Liners:
• The Trident Constrained Acetabular Insert is indicated for use in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability.
The Trident II Acetabular Shells are indicated for cementless use only.
The Trident® II Acetabular Shells, 6.5mm Low Profile Hex Screws and Hex Dome Hole Plug are sterile, single-use devices that are intended for cementless fixation into a prepared acetabulum for either primary or revision Total Hip Arthroplasty. The Trident® II Acetabular System is an extension of the Trident System product line and features the same locking mechanism as the current Trident® and Trident® II Tritanium® product lines. The Trident® II Acetabular Shells are intended to be used with existing Trident® polyethylene inserts, Trident Constrained Acetabular Inserts, MDM® liners and both new and existing surgical instruments.
The Trident® II Tritanium® implant consists of a unique configuration of both solid and porous sections that are simultaneously built using a Laser Rapid Manufacturing (LRM) method of additive manufacturing, applying Stryker's proprietary Tritanium® In-Growth Technology. Tritanium® is a novel highly porous titanium material designed for biological fixation. The Trident® II PSL® Clusterhole HA and Trident® II Clusterhole HA Acetabular Shells feature hydroxylapatite coating over a Commercially Pure Titanium plasma sprayed surface for cementless fixation.
This document describes a 510(k) premarket notification for the "Trident® II Acetabular System," a medical device for total hip arthroplasty. The submission asserts substantial equivalence to previously marketed predicate devices, and therefore, clinical testing was not required. The information provided focuses on non-clinical testing.
Here's an analysis of the provided text in relation to your request:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present acceptance criteria in a table format with corresponding reported performance for each criterion. Instead, it lists a series of non-clinical tests performed, implying that the device "meets the requirements" or was "characterized according to methods" outlined in various ASTM standards and FDA guidance documents.
Here's a partial reconstruction of what could be inferred as acceptance criteria and "device performance" based on the text:
Acceptance Criteria (Inferred from Test Objective/Guidance) | Reported Device Performance (Implied) |
---|---|
Mechanical Stability (Cementless Fixation): | |
Resist push-out forces (ASTM F1820) | Testing performed. Implied to meet requirements. |
Resist lever-out forces (ASTM F1820) | Testing performed. Implied to meet requirements. |
Resist torque-out forces (ASTM F1820) | Testing performed. Implied to meet requirements. |
Durability/Fatigue: | |
Withstand acetabular shell fatigue (Acetabular Shell Fatigue Testing) | Testing performed. Implied to meet requirements. |
Wear/Interface Performance: | |
Minimal fretting at MDM liner interface (Fretting Evaluation of the MDM liner) | Testing performed. Implied to meet requirements. |
Material Properties: | |
Resistance to plastic deformation (Plastic Deformation) | Testing performed. Implied to meet requirements. |
Material chemistry meets specifications (Material Chemistry) | Testing performed. Implied to meet requirements. |
Porous surface meets FDA guidance for orthopedic implants with modified metallic surfaces (Characterization of Chemistry, Physical & Mechanical Properties of porous surface) | Established that porous surface meets requirements. |
Titanium plasma spray characterized per guidance (Characterization of Chemistry, Physical & Mechanical Properties of porous surface) | Characterized according to methods. |
Biocompatibility/Pyrogenicity: | |
Endotoxin limit |
Ask a specific question about this device
Page 1 of 1