    Device Name :

    Tosoh Automated Glycohemoglobin Analyzer HLC-723G8

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is intended for in vitro diagnostic use for the measurement of % hemoglobin A1c (HbA1c) (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in venous whole blood specimens using ion-exchange high-performance liquid chromatography (HPLC). This test is an aid in diagnosis of diabetes and identifying patients who may be at risk for developing diabetes, and for monitoring of long-term blood glucose control in individuals with diabetes mellitus.

    Device Description

    The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is an automated High-Performance Liguid Chromatography (HPLC) system that separates and reports stable hemoglobin A1c (sA1c) percentage in venous whole blood. The operational portion of the G8 is composed of a sampling unit, liquid pump, degasser, column, detector, microprocessors, sample loader, smart media card, operation panel, and a printer. The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 uses ion-exchange HPLC for rapid, accurate, and precise separation of the stable form of HbA1c (sA1c) from other hemoglobin fractions. The G8 uses a non-porous cation exchange column and separates the hemoglobin components in the blood. Separation is achieved by utilizing differences in ionic interactions between the cation and exchange group on the column resin surface and the hemoglobin components in a step gradient elution. The hemoglobin fractions (designated as A1a. A1b. F. LA1c+, SA1c, A0, and, if present, H-V0, H-V2, H-V2 and H-V3) are subsequently removed from the column by performing a step-wise elution gradient using the varied salt concentrations in the Variant Elution Buffers HSi 1, 2 and 3. The peaks, H-V0, H-V1, H-V2 and H-V3 are typically presumptive HbAD, HbAS, HbAC and HbAE respectively. The software compares the retention times of hemoglobin fractions in a sample to the expected "windows of retention" and labels each fraction that correctly elutes within a defined expected window of retention. The software designates a hemoglobin fraction as POX (where X is the order of the peak as it elutes from the column) if it does not match a defined window of retention. All automated processes in the G8 are controlled by internal microprocessors, using software downloaded via a smart media card. The result report is printed and can be stored on the instrument. The data can be transmitted to a host computer through a bi-directional interface. The result report includes the sample ID, date, percentage and retention time of each fraction of hemoglobin, sA1c percentage and total A1 percentage, along with a chromatogram of the elution pattern of the hemoglobin fractions. If a sample contains a hemoglobin variant, the column elutes the fraction depending upon its charge.

    AI/ML Overview

    The provided text describes the non-clinical performance testing of the Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 (subject device) to support its substantial equivalence to a predicate device. This document focuses on the analytical performance of a diagnostic device rather than an AI/ML powered device, so some of the specific questions regarding AI/ML study design (e.g., number of experts, adjudication methods, MRMC studies) are not applicable.

    Here's the information extracted from the document:

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the statement "All performance testing results met their pre-determined acceptance criteria." While explicit numerical acceptance criteria for each test are not listed in a consolidated table, the discussion throughout the "Summary of Non-Clinical Performance Testing" implicitly defines them through the methodology and results. For example, for precision/repeatability, the claim of "imprecision at ≤ 2%" was a pre-established criterion. Similarly, for hemoglobin variant interference, "Non-clinically significant interference was defined as