Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K173194
    Manufacturer
    Date Cleared
    2018-10-10

    (373 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Orantech Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable and designed for use with monitors of Philips Model MP30 and Nihon Kohden Model BSM-6301A. These devices are indicated for used by qualified medical personnel only.

    Device Description

    Reusable temperature probes are used during patient temperature measurement for multi-patient use. These probes consist of a phone plug connector on the monitor end and a thermistor on the patient end. Temperature probes measure temperature by a resistor that is sensitive to temperature changes. These probes are connected to the patient monitor either directly by using a phono plug or by an interconnect cable. These probes have a skin or core contact with a patient.

    These temperature probes are typically used with legacy Philips monitor model MP30 and Nihon Kohden monitor model BSM-6301A.

    AI/ML Overview

    The medical device in question is the Orantech Inc. Temperature Probe, models TS-Y400-AG30, TS-Y400-AS30, TS-PH-AG30, and TS-PH-AS30.

    Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Set by Standard or Performance Goal)Reported Device PerformanceOutcome
    Biocompatibility
    Cytotoxicity (ISO 10993 series)No cytotoxicity detectedPass
    Sensitization (ISO 10993 series)No sensitization detectedPass
    Irritation (ISO 10993 series)Negligible irritationPass
    Safety
    Electrical Safety (IEC 60601-1)Meets requirements of IEC 60601-1Pass
    Performance (Continuous Clinical Thermometer)
    Laboratory Accuracy (25-45°C range)Not greater than 0.3°C (Accuracy: ± 0.1°C stated below)Pass
    Time ResponseHeating transient time < 150sPass
    Use Life TestingMeets requirements for expected service lifePass
    Cleaning and Disinfection EffectivenessCleaning and disinfection method is effectivePass

    Note on Accuracy: The 'Product Specification' table explicitly states an accuracy of "± 0.1°C" for all models within the 25-45°C range. The "Laboratory accuracy" test result states "Not greater than 0.3°C", which is a broader allowance than the stated ± 0.1°C. Assuming the ± 0.1°C is the target, 'Not greater than 0.3°C' still indicates compliance.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not specify the sample size used for the test set for any of the performance studies (biocompatibility, safety, accuracy, time response, use life, cleaning/disinfection).

    The document does not specify the data provenance (e.g., country of origin of the data, retrospective or prospective). The testing was conducted by or for Orantech Inc., which is based in China.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The type of device (temperature probe) and the nature of the tests (laboratory accuracy, electrical safety, biocompatibility) typically do not involve establishing ground truth through human experts in the same way, for example, an AI diagnostic imaging device would. The "ground truth" for these tests would be derived from calibrated equipment and established scientific methods, rather than expert consensus on interpretation.

    4. Adjudication Method for the Test Set

    This information is not applicable/provided. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation (e.g., radiology reads) to resolve discrepancies in ground truth establishment. For the types of tests conducted for a temperature probe, such methods are not relevant.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There was no MRMC comparative effectiveness study done. This type of study is relevant for AI-powered diagnostic devices that assist human readers (e.g., radiologists interpreting images). The Orantech Temperature Probe is a physical medical device for direct measurement, not an AI software.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not applicable to the Orantech Temperature Probe. It is a physical device, not an algorithm. The performance tests (laboratory accuracy, time response, etc.) represent its "standalone" performance in a controlled environment, but it's not an algorithm whose performance can be evaluated independently of human interaction in the clinical use context that AI devices are.

    7. The Type of Ground Truth Used

    The ground truth for the various tests was established through:

    • Biocompatibility: Adherence to established international standards (ISO 10993 series) using laboratory testing for cytotoxicity, sensitization, and irritation. The "ground truth" is the scientific determination of biological response to the materials.
    • Safety: Compliance with electrical safety standard (IEC 60601-1) through testing. The "ground truth" is the objective measurement of electrical parameters against the standard's limits.
    • Performance (Laboratory Accuracy, Time Response, Use Life): Measurement against calibrated reference instruments and established performance criteria defined by medical device standards or internal specifications. For accuracy, the "ground truth" is the actual temperature measured by a highly accurate reference thermometer.
    • Cleaning and Disinfection: Validation against AAMI TIR documents and FDA guidance, likely involving microbiological testing to confirm the reduction/elimination of pathogens. The "ground truth" is the demonstrably effective reduction of microbial load.

    8. The Sample Size for the Training Set

    This information is not applicable as the device is a physical temperature probe and not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as above (not an AI/ML algorithm).

    Ask a Question

    Ask a specific question about this device

    K Number
    K160091
    Manufacturer
    Date Cleared
    2016-05-20

    (126 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The temperature probe is intended for use for continuous blood temperature monitoring as measured at a temperature monitoring adapter located within a Medtronic extracorporeal circulation device as specified in the device's instructions for use. The temperature probe is designed for use with a YSI tele-thermometer to monitor and display temperature.

    Device Description

    The temperature probe devices are intended for use with the temperature monitoring adapter of compatible Medtronic devices and the YSI™ tele-thermometer 1. The probe has a thermistor sensor housed in a stainless steel sheath connected to a 3-m (10-ft) cable, terminating with a 6.35-mm (1/4-in) phono plug.

    AI/ML Overview

    This refers to a 510(k) premarket notification for a Class II medical device, specifically a "Temperature Probe" (Product Code: FLL). The document details the device's characteristics and its comparison to predicate devices, along with performance testing to demonstrate substantial equivalence.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" with numerical values. Instead, it lists performance tests and their "Pass" results, indicating the device met the requirements established for each test. The acceptance criteria for each test would be the specific requirements or standards defined prior to testing, which the device passed.

    Test DescriptionReference Standard/Requirement (Implicit Acceptance Criteria)Reported Device Performance
    IEC 60601Ensures compliance with IEC 60601 electrical safety and Electromagnetic compatibility (EMC).Pass
    Temperature RangeEnsures performance at extreme temperature ranges.Pass
    Ambient Temperature EnvironmentEnsures performance at normal temperature ranges.Pass
    AccuracyAccuracy over the entire specified temperature range.Pass
    Precision and RepeatabilityPrecision and repeatability of measurements over the specified temperature range with effects of air currents.Pass
    TimeTime required for the device to obtain a steady state reading.Pass
    Life TestingMeets requirements for the expected life of the probe.Pass
    Liquid Ingress and Chemical Exposure TestingMeets Ingress Protection and cleaning requirements.Pass

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not specify the exact sample size for each performance test. It states "Bench testing was used to demonstrate the performance characteristics," implying laboratory-based testing rather than clinical data from human subjects. There is no information regarding the country of origin of data or whether it was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. The device is a temperature probe, and its performance (e.g., accuracy, precision) is typically established through objective measurements against traceable standards in a laboratory setting, not by human expert opinion or interpretation of data in the same way an AI diagnostic tool would be evaluated.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation (e.g., medical image reading) where there might be disagreement among experts. For performance testing of a physical device like a temperature probe, objective measurements against established standards are used, not subjective adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study. This type of study is relevant for AI-assisted diagnostic tools involving human readers, which is not the case for this temperature probe.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable in the context of an AI algorithm. However, the performance tests listed in the table (e.g., accuracy, precision, life testing) assess the standalone performance of the device itself in a controlled environment, without human intervention beyond setting up the test conditions.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the performance tests mentioned, the "ground truth" would be established by:

    • Traceable standards: For accuracy and temperature range tests, the device's readings would be compared against highly accurate and calibrated reference thermometers.
    • Defined specifications/requirements: For other tests like life testing, ingress protection, or time to steady state, the ground truth is whether the device meets pre-defined engineering and safety specifications.
    • International standards: Compliance with IEC 60601 suggests adherence to internationally recognized electrical safety and EMC benchmarks.

    8. The sample size for the training set

    This information is not applicable. The temperature probe is a physical device, not an AI algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    This information is not applicable, as there is no training set for a physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K952800
    Date Cleared
    1996-04-04

    (290 days)

    Product Code
    Regulation Number
    880.5130
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K952803
    Date Cleared
    1996-04-04

    (290 days)

    Product Code
    Regulation Number
    880.5130
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K952804
    Date Cleared
    1996-04-04

    (290 days)

    Product Code
    Regulation Number
    880.5130
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K952807
    Date Cleared
    1996-04-04

    (290 days)

    Product Code
    Regulation Number
    880.5130
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K952801
    Date Cleared
    1996-04-04

    (290 days)

    Product Code
    Regulation Number
    880.5130
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K952805
    Date Cleared
    1996-04-04

    (290 days)

    Product Code
    Regulation Number
    880.5130
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1