Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    THE PHILIPS MP20, MP30, MP40, MP50, MP60, MP70, AND MP90 INTELLIVUE PATIENT MONITORS, WITH 802.11 WIRELESS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in healthcare environments. The MP20, MP30, MP40 and MP50 are additionally intended for use in transport situations within healthcare environments. ST Segment monitoring is restricted to adult patients only. The transcutaneous gas measurement (tcp02 / tcpCQ2) is restricted to neonatal patients only.

    Device Description

    The Philips MP20, MP30, MP40, MP50, MP60, MP70, and MP90 IntelliVue Patient Monitors with 802.11 Wireless LAN. The modification is a change that creates, an optional, 802.11 wireless network adapter for the Philips MPxx IntelliVue patient monitors.

    AI/ML Overview

    The provided 510(k) summary for the Philips IntelliVue Patient Monitors with 802.11 Wireless LAN does not contain detailed information about specific acceptance criteria and a study proving the device meets those criteria, as typically found for AI/ML-based medical devices.

    Instead, this submission is for a modification (the addition of an 802.11 wireless network adapter) to existing patient monitors. The focus is on demonstrating substantial equivalence to previously cleared predicate devices, rather than a de novo evaluation of novel performance.

    Here's a breakdown of what can be extracted and what is not present in the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    This information is not explicitly detailed in the provided document in the context of specific quantitative metrics for the wireless LAN performance.

    The document states: "Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence."

    This indicates that:

    • Acceptance Criteria: Were based on "specifications cleared for the predicate device." These specific criteria (e.g., latency, throughput, signal strength, reliability) are not listed.
    • Reported Device Performance: The summary states "test results showed substantial equivalence" and "The results demonstrate that the Philips IntelliVue Patient Monitor meets all reliability requirements and performance claims." No quantitative performance data for the wireless LAN is given.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided for the specific wireless LAN testing. The document generally refers to "system level tests, performance tests, and safety testing" without detailing the sample sizes (e.g., number of test cases, duration of tests, number of devices tested) or data provenance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable/not provided. The device is an integrated patient monitor with a wireless component, not an AI/ML diagnostic or prognostic tool that requires expert-established ground truth for its core function. The "ground truth" here would relate to engineering specifications and performance standards for wireless communication and patient monitoring, not clinical expert consensus on interpretations of data.

    4. Adjudication Method for the Test Set

    This information is not applicable/not provided. Similar to point 3, adjudication methods like 2+1 or 3+1 statistical consensus are typically used for clinical image or signal interpretation ground truth, not for evaluating the performance of a wireless network adapter in a medical device.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    This information is not applicable/not provided. MRMC studies are used to assess the diagnostic performance of human readers, often with and without AI assistance, usually for interpreting medical images or complex data. This submission is for a hardware modification (wireless LAN) to a patient monitor, not a diagnostic AI system.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable/not provided. The device is a patient monitor, which operates with human interaction and is not solely an "algorithm" in the conventional AI sense. The wireless LAN is a component of a larger system.

    7. The Type of Ground Truth Used

    The "ground truth" for the testing described would be based on engineering specifications and established performance standards for patient monitoring devices and wireless communication. For example, standards for signal integrity, data transmission rates, latency, security, and reliability in a medical environment would serve as the ground truth against which the wireless LAN's performance was measured. No clinical outcomes data or pathology reports would typically be relevant for this type of device modification.

    8. The Sample Size for the Training Set

    This information is not applicable/not provided. The document describes a hardware modification and associated performance testing, not the development or training of an AI/ML algorithm. Therefore, there is no "training set" in this context.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable/not provided for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    THE PHILIPS MP20, MP30, MP40, MP50, MP60, MP70 AND MP90 INTELLIVUE PATIENT MONITORS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The MP20, MP30, MP40 and MP50 are intended for use in transport situations within additionally hospital environments.

    ST Segment monitoring is restricted to adult patients only.

    The transcutaneous gas measurement (tcp02 / tcpCO2) is restricted to neonatal patients only.

    Device Description

    The Philips MP20, MP30, MP40, MP50, MP60, MP70, and MP90 IntelliVue Patient Monitors. The modification is the introduction of Release B.1.2 software for the IntelliVue patient monitor devices and incorporating a ITS wireless interface.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and the supporting study:

    The provided document is a 510(k) summary for the Philips IntelliVue Patient Monitors, specifically for the introduction of Release B.1.2 software and a wireless interface. It focuses on demonstrating substantial equivalence to previously cleared devices rather than establishing novel performance claims with specific acceptance criteria that would require a detailed standalone study.

    Key takeaway: This submission relies on proving the updated device functions similarly to its predicates, implying that the predicate devices already met established performance standards. Therefore, the information typically requested for a de novo device's acceptance criteria and studies (like specific accuracy metrics, sensitivity/specificity, etc.) is not explicitly present in this summary.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of numerical acceptance criteria (e.g., sensitivity, specificity, accuracy targets) or specific reported device performance metrics in the way a new device would for novel claims. Instead, it states that:

    • "Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate."
    • "Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence."
    • "The results demonstrate that the Philips IntelliVue Patient Monitor meets all reliability requirements and performance claims."

    This means the "acceptance criteria" were essentially that the modified device performed at least as well as the predicate devices for their established specifications. The "reported device performance" is implicitly that it achieved this equivalence across all tested aspects.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a sample size for a test set in terms of patient data or specific clinical cases. The testing described is more focused on system-level validation and verification rather than clinical performance against a diagnostic gold standard in a typical study.

    • Sample Size: Not specified.
    • Data Provenance: Not specified. Since the testing is described as "system level tests, performance tests, and safety testing from hazard analysis," it's likely a mix of internal engineering tests, simulated data, and potentially some real-world operational testing, but not a large-scale clinical study with patient data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. This type of submission, focused on software and interface modifications for an existing patient monitor, primarily involves engineering and regulatory assessment, rather than expert clinical review of novel diagnostic outputs.

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified.

    4. Adjudication Method for the Test Set

    Not applicable. As no human-in-the-loop diagnostic task or a specific clinical test set requiring expert adjudication is described, there's no mention of an adjudication method like 2+1 or 3+1.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The document does not describe any study involving human readers or AI assistance, as the device is a patient monitor, not an AI-powered diagnostic tool.

    6. If a Standalone Performance Study Was Done

    A standalone performance study with explicit metrics (like sensitivity/specificity against a clinical gold standard) for a new algorithm was not performed or described. The performance assessment was centered on validating that the modified device maintains the performance, functionality, and reliability characteristics of its predicate devices. This includes "performance tests" at a system level.

    7. The Type of Ground Truth Used

    Given the nature of the device (patient monitor) and the submission's focus on substantial equivalence through software and interface updates:

    • Ground Truth Type: For the "performance, functionality, and reliability," the ground truth would be the specifications and established performance of the predicate devices. This implies that the internal workings and outputs (e.g., heart rate readings, blood pressure measurements, alarm triggers) of the modified device were compared against the known, validated behavior of the older versions. It's an engineering and functional ground truth rather than a clinical diagnostic ground truth like pathology or outcome data.

    8. The Sample Size for the Training Set

    Not applicable. This document describes a medical device (patient monitor) with software updates, not an AI/Machine Learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of an AI/Machine Learning model or a training set.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    THE PHILIPS MP20, MP30, MP40, MP50, MP60, MP70, AND MP90 INTELLIVUE PATIENT MONITORS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters patients. Intended for monitoring, recording and alarming of of multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The MP20, MP30, MP40 and MP50 are additionally intended for use in transport situations within hospital environments.

    EASI 12-lead ECG is only for use on adult and pediatric patients.

    ST Seqment monitoring is restricted to adult patients only.

    The transcutaneous gas measurement (tcp02/tcpCO2) is restricted to neonatal patients only.

    Device Description

    The names of the devices are the Philips MP20, MP30, MP40, MP50, MP60, MP70, and MP90 IntelliVue Patient Monitor. The modification is the introduction of Release B.1.1 software for the IntelliVue patient monitor devices, MP20, MP30, MP40, MP50 interfacing the Gas Analyzer model M1013A EGM (K041956). The modified devices have the same technological characteristics as the legally marketed predicate devices.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Philips IntelliVue Patient Monitor, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state specific quantitative acceptance criteria (e.g., "accuracy must be >95%") or detailed reported device performance in a numerical format for individual parameters. Instead, it makes a general statement about meeting specifications and reliability.

    Acceptance Criteria (Implied)Reported (General) Device Performance
    Conformance to specifications cleared for the predicate devicesMeets all reliability requirements
    Performance of the modified device with respect to the predicateMeets all performance claims
    Functionality of the modified device with respect to the predicateEstablished
    Reliability of the modified device with respect to the predicateEstablished
    Safety testing pass/fail based on hazard analysisPass/Fail criteria met

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. The document mentions "system level tests, performance tests, and safety testing from hazard analysis" but doesn't quantify the number of test cases, patient data, or specific scenarios used.
    • Data Provenance: Not explicitly stated. It's unclear if the testing involved real patient data, simulated data, or a combination, nor the country of origin. The regulatory submission is from Germany, suggesting testing might have occurred there or within Phillips' global network. It is implied to be prospective testing, as it's for a new software release.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not mentioned in the document. The type of device (patient monitor) suggests that ground truth would likely be based on established physiological measurement standards and potentially clinical expert review, but this is not detailed.

    4. Adjudication Method for the Test Set

    Not mentioned in the document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No, an MRMC comparative effectiveness study is not applicable and was not performed. This device is a physiological patient monitor, not an AI-assisted diagnostic tool for human readers. Its primary function is to measure and display physiological parameters, and alarm if they fall outside set limits. There's no AI component for improving human interpretation in the way one would see in an imaging AI application.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the testing described appears to be a standalone evaluation of the device's performance, functionality, and reliability, without explicitly involving human-in-the-loop performance for its core functions. The software Release B.1.1 itself is the "algorithm only" component that was tested.

    7. The Type of Ground Truth Used

    The document does not specify the exact "type of ground truth" in the way one might for an imaging algorithm (e.g., pathology, outcomes data). For a patient monitor, ground truth would be established through a combination of:

    • Known physical inputs/simulations: Using calibration equipment or simulators to generate precise physiological signals (e.g., a known ECG rhythm, a specific blood pressure waveform) and verifying the device's output against these known inputs.
    • Comparison to reference devices: Comparing the monitor's readings to those of other validated and highly accurate reference monitoring devices.
    • Technical specifications: Verifying that the device's measurements fall within the specified accuracy and precision limits for each parameter.

    8. The Sample Size for the Training Set

    Not applicable. This device is a patient monitor, not an AI/machine learning algorithm that requires a "training set" in the conventional sense. The "training" would involve the development and engineering process of the software itself, not a data-driven training pipeline for a predictive model.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of an AI/machine learning algorithm for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1