Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K133280
    Date Cleared
    2014-06-12

    (230 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    TERUMO SURFLO I.V. CATHETER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TERUMO® SURFLO® I.V. Catheter is inserted into the patient's vascular system for short term use (less than 30 days) to withdraw blood samples, monitor blood pressure, or administer fluids intravenously.

    Device Description

    The TERUMO® SURFLO® I.V. Catheter is a device consisting of a catheter assembly (catheter, caulking pin, and catheter hub) and needle assembly (needle, needle hub, filter cap and filter). The device is an over-the needle, peripheral catheter made of a slender, flexible, radio-opaque plastic with a hub that is inserted into the patient's vascular system for short term (

    AI/ML Overview

    Here's an analysis of the acceptance criteria and supporting studies for the TERUMO® SURFLO® I.V. Catheter, based on the provided 510(k) summary:

    This device is not an AI device, therefore, some of the requested information (like number of experts, adjudication methods, MRMC studies, training set details) are not applicable. The provided document details a traditional medical device submission for substantial equivalence.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the TERUMO® SURFLO® I.V. Catheter are based on conformity to recognized consensus ISO standards and specific internal performance tests. The reported performance indicates that the device Meets Standard or Meets acceptance criteria for all tests.

    Test CategorySpecific TestStandard/Acceptance CriteriaReported Device Performance
    ISO StandardsForce at breakISO 10555-1Meets standard
    Corrosion resistanceISO 10555-1Meets standard
    Liquid leakage under pressureISO 10555-1Meets standard
    Air leakage into hub assembly during aspirationISO 10555-1Meets standard
    Vent fittingISO 10555-5Meets standard
    Flow rateISO 10555-5Meets standard
    Strength of union between needle hub and needle tubeISO 10555-5Meets standard
    GaugingISO 594-1Meets standard
    Liquid leakageISO 594-1Meets standard
    Air leakageISO 594-1Meets standard
    Separation forceISO 594-1Meets standard
    Stress crackingISO 594-1Meets standard
    Internal TestsCannula and needle hub fitting forceAcceptance criteriaMeets acceptance criteria
    Catheter tube and catheter hub fitting forceAcceptance criteriaMeets acceptance criteria
    Catheter tube and catheter hub leakageAcceptance criteriaMeets acceptance criteria
    Catheter tube and needle fitting forceAcceptance criteriaMeets acceptance criteria
    FlashbackAcceptance criteriaMeets acceptance criteria
    Needle and needle hub leakageAcceptance criteriaMeets acceptance criteria
    Needle heel and catheter tip distanceAcceptance criteriaMeets acceptance criteria
    Needle penetrationAcceptance criteriaMeets acceptance criteria

    Biocompatibility Tests:

    TestReported Performance
    CytotoxicityBiocompatible
    SensitizationBiocompatible
    Intracutaneous reactivityBiocompatible
    Systemic toxicity (Acute)Biocompatible
    PyrogenicityBiocompatible
    Subchronic toxicity (subacute toxicity)Biocompatible
    HemocompatibilityBiocompatible
    GenotoxicityBiocompatible
    ImplantationBiocompatible
    Physicochemical CharacterizationBiocompatible

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes (number of units tested) for the performance tests. It states that "Performance testing was conducted to ensure the safety and effectiveness... and demonstrate substantial equivalence to the predicate device."
    The data provenance is from the manufacturer, Terumo (Philippines) Corporation, and the tests were performed on the new device, comparing it to the legally marketed predicate device (K891087) manufactured by Terumo Medical Corporation, Elkton, Maryland. These are considered prospective tests performed on the device being submitted for clearance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable as the device is a physical medical instrument (intravascular catheter), not an AI/software device that relies on expert interpretation for ground truth. The "ground truth" for this device's performance is objective measurements against established engineering and biological standards.

    4. Adjudication Method for the Test Set

    This information is not applicable because the performance tests involve objective measurements and comparisons against established standards, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically used for diagnostic or screening devices where human readers interpret medical images or data, and AI assistance can impact their performance. This is not relevant for an intravascular catheter.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, a standalone performance test in the context of an algorithm's performance was not done. The "standalone" performance here refers to the device's physical and biological characteristics as it would be used, independent of human interaction other than proper insertion and operation.

    7. The Type of Ground Truth Used

    The ground truth used for this device is based on:

    • Established consensus standards: International Organization for Standardization (ISO) standards (e.g., ISO 10555 series, ISO 594 series, ISO 11137 for sterilization, ISO 10993-1 for biocompatibility).
    • Predefined acceptance criteria: Internal specifications for physical properties and functionality (e.g., fitting forces, leakage, flashback, needle penetration).
    • Demonstrated performance of a legally marketed predicate device: The new device's performance is compared to that of the predicate device (K891087) to establish substantial equivalence.

    8. The Sample Size for the Training Set

    This is not applicable. This submission is for a physical medical device, not an AI or machine learning algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set was Established

    This is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1