Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K192401
    Manufacturer
    Date Cleared
    2020-04-06

    (216 days)

    Product Code
    Regulation Number
    872.3630
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Straumann Screw-Retained Abutments

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Prosthetic components directly connected to the endosseous dental implant are intended for use as an aid in prosthetic rehabilitations. Temporary components can be used prior to the final components to maintain, stabilize and shape the soft tissue during the healing phase; they are to be placed out of occlusion. Final abutments may be placed into occlusion for implants with sufficient primary stability or for implants that are fully osseointegrated.

    Temporary Abutments have a maximum duration of usage of 180 days.

    Device Description

    The subject devices described in the submission are prosthetic components that are intended for use in prosthetic rehabilitations, directly or indirectly connected to the endosseous dental implant. These devices are intended for use as an aid in prosthetic rehabilitations. The subject devices described in this submission are to be added Straumann's existing prosthetic portfolio.

    AI/ML Overview

    The provided text is a 510(k) summary for Straumann® Screw-Retained Abutments. This document primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices through comparisons of technological characteristics and performance data, mainly bench testing and biocompatibility assessments. It explicitly states that no device-specific clinical data has been submitted.

    Therefore, the information required to describe acceptance criteria and a study proving the device meets those criteria, especially in the context of an AI/human reader study or performance metrics typical for AI devices (like sensitivity, specificity, AUC), is not present in this document.

    This document describes a traditional medical device (dental implants and abutments) cleared via the 510(k) pathway, which fundamentally relies on demonstrating substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria through novel clinical studies or AI algorithm validation.

    To address the user's request, I must state that the information is not available in the provided document, as it pertains to a different type of device and regulatory submission pathway.

    Specifically, the document states:

    • "No device specific clinical data has been submitted to demonstrate substantial equivalence." (Page 12, Section 5.7.4)

    Here's an explanation with respect to the user's request:

    1. A table of acceptance criteria and the reported device performance: Not applicable. The submission focuses on substantial equivalence through comparison with predicates, not on meeting predefined performance metrics from a study in the way an AI/software device would. The "performance data" refers to bench testing (dynamic fatigue, static strength, torque) and biocompatibility, which are compared to predicate devices rather than against specific numerical acceptance criteria.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable, as no clinical study (with test sets of patient data) was conducted or submitted. The 'samples' referred to in bench testing are physical abutment components.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for clinical data involving human perception (like AI for image analysis) is not relevant here as no such clinical study was performed.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-powered device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI-powered device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable.

    8. The sample size for the training set: Not applicable. This is not an AI-powered device with a training set.

    9. How the ground truth for the training set was established: Not applicable.

    In summary: The provided 510(k) summary for Straumann® Screw-Retained Abutments does not contain the type of AI/clinical performance study information requested. The device is a traditional medical device (dental implant components) and its clearance is based on substantial equivalence to predicate devices, verified through bench testing and biocompatibility assessments, not through human reader studies or AI algorithm performance evaluations on clinical datasets.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171757
    Date Cleared
    2017-11-27

    (166 days)

    Product Code
    Regulation Number
    872.3630
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Straumann Screw Retained Abutments

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Straumann® Screw Retained Abutments are indicated to be placed into the implants of the Straumann® Dental Implant System to provide support for prosthetic reconstructions such as crowns, bridges and bars. The final processed devices have the purpose of restoring chewing function. Straumann® Screw Retained Abutments are indicated for screw-retained restorations.

    Device Description

    Straumann® Screw Retained Abutments include one-piece straight and angled (17° and 30°) abutments, basal screws and abutment carrier pin. The subject device is identical to the primary predicate, K133421, and subsequent premarket notification submissions K141871, K150814 and K151247 with the exception that the subject device will be delivered to the user sterile as opposed to non-sterile and a change in packaging to ensure sterility of the devices.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Straumann® Screw Retained Abutments. It aims to demonstrate substantial equivalence to previously cleared predicate devices. Since this is not a diagnostic device and doesn't involve an AI algorithm, the information requested in your prompt (acceptance criteria, study details, expert input, MRMC study, standalone performance, training set) typically applies to AI/ML-based medical devices or diagnostic tools.

    This submission focuses on engineering, material, and sterilization equivalence for a dental implant component. Therefore, the detailed AI/ML-specific questions in your prompt are not applicable to this document.

    However, I can extract information related to the device's technical characteristics and the claim of substantial equivalence.

    Here's a breakdown of what is applicable from your prompt, based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" in the format typically used for performance metrics of diagnostic algorithms. Instead, it relies on demonstrating substantial equivalence to predicate devices. This means that the device's characteristics must be similar enough to the predicate to ensure equivalent safety and effectiveness.

    The table provided (pages 5-6) serves as the primary evidence for this. It compares the Subject Device (Straumann® Screw Retained Abutments) to several Predicate Devices across various technical characteristics. The "performance" in this context is demonstrating that these characteristics are either identical or sufficiently similar to the legally marketed predicates.

    FeatureSubject Device PerformancePredicate Device Performance (K133421, K141871, K150814, K151247)
    Indications for UseStraumann® Screw Retained Abutments are indicated to be placed into the implants of the Straumann® Dental Implant System to provide support for prosthetic reconstructions such as crowns, bridges and bars. The final processed devices have the purpose of restoring chewing function. Straumann® Screw Retained Abutments are indicated for screw-retained restorations. (Identical to predicates)Straumann® Magellan™ Screw Retained Abutments are indicated to be placed into the implants of the Straumann® Dental Implant System to provide support for prosthetic reconstructions such as crowns, bridges and bars. The final processed devices have the purpose of restoring chewing function. Straumann® Screw Retained Abutments are indicated for screw-retained restorations. (For K133421, K141871, K150814, K151247)
    Implant-to-Abutment ConnectionNarrow CrossFit (NC), Regular CrossFit (RC) (Identical to K133421, K151247; broader than K141871, K150814 which only list NC)Narrow CrossFit (NC), Regular CrossFit (RC) (K133421, K151247)
    Narrow CrossFit (NC) (K141871, K150814)
    Interface TypeEngaging (Identical to all predicates)Engaging (All predicates)
    Platform Diameter(s)NC Straight and Angled: Ø3.5 mm, Ø4.6 mm
    RC Straight and Angled: Ø4.6 mm (Comparable ranges; Subject device appears to cover the full range of predicates)NC Straight: Ø3.5 mm, Ø4.6 mm
    RC Straight and Angled: Ø4.6 mm (K133421)
    NC Angled: Ø3.5 mm, Ø4.6 mm (K141871)
    NC Straight: Ø3.5 mm, Ø4.6 mm (K150814)
    NC Angled: Ø3.5 mm, Ø4.6 mm
    RC Angled: Ø4.6 mm (K151247)
    Abutment Angulation(s)0°, 17°, 30° (Identical to K133421, K151247; broader than K141871 (17°, 30°) and K150814 (0°))0°, 17°, 30° (K133421)
    17°, 30° (K141871, K151247)
    (K150814)
    Gingival Height(s)NC and RC Straight: 1.0, 2.5 and 4.0 mm
    NC and RC Angled: 2.5, 4.0 and 5.5 mm (Comparable and comprehensive ranges)NC and RC Straight: 1.0, 2.5 and 4.0 mm
    RC Angled: 2.5 and 4.0 mm (K133421)
    NC Angled: 2.5 and 4.0 mm (K141871)
    NC Straight: 1.0 mm (K150814)
    NC and RC Angled: 5.5 mm (K151247)
    Orientation of Angulation to Engagement FeaturesType A (45°), Type B (0°) (Identical to K133421, K141871, K151247; not applicable for K150814 as it's 0°)Type A (45°), Type B (0°) (K133421, K141871, K151247)
    N/A (K150814)
    Device MaterialTi-6Al-7Nb titanium alloy (Identical to all predicates)Ti-6Al-7Nb titanium alloy (All predicates)
    Sterilization StatusEnd user receives product sterilized per Gamma Irradiation, 25 kGy minimum, Validated per ISO 11137-1 and ISO 11137-2 to an SAL of 10 x 10-6 (Key difference: Subject device is sterile upon delivery)End user to sterilize product per IFU Method: Autoclave moist heat fractionated vacuum or gravity displacement Conditions: 134°C (273°F) for 5 minutes (All predicates are non-sterile; user is responsible for sterilization)
    PackagingChanged to ensure sterility. Equivalent to reference predicate device, Straumann® Healing Abutments, Healing Caps, Closure Screws (K130808) (Difference from non-sterile predicates, similarity to sterile reference predicate)Not explicitly detailed for non-sterile predicates, but implied to be non-sterile packaging. K130808 (sterile reference predicate) has equivalent packaging.

    The study proving the device meets its "acceptance criteria" (i.e., substantial equivalence) is the comparison to predicate devices and the verification/validation activities for the changes (primarily sterilization and packaging). The document explicitly states: "The subject device is identical to the primary predicate, K133421, and subsequent premarket notification submissions K141871, K150814 and K151247 with the exception that the subject device will be delivered to the user sterile as opposed to non-sterile and a change in packaging to ensure sterility of the devices." And "Packaging materials and configuration is equivalent to reference predicate device, Straumann® Healing Abutments, Healing Caps, Closure Screws, K130808."

    The substantial equivalence is primarily based on:

    • Identical Indications for Use.
    • Identical fundamental operating principles.
    • Identical materials (Ti-6Al-7Nb titanium alloy).
    • Comparable design features (Implant-to-Abutment Connection, Interface Type, Platform Diameter(s), Abutment Angulation(s), Gingival Height(s), Orientation of Angulation to Engagement Features).
    • Validation of the sterilization process according to recognized standards (ISO 11137-1 and ISO 11137-2) and equivalence of packaging to a previously cleared sterile device (K130808).

    The following points are not applicable to this 510(k) submission for a non-AI/ML dental abutment:

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    8. The sample size for the training set
    9. How the ground truth for the training set was established

    These points would be relevant for devices that involve algorithms, image interpretation, or diagnostics where a "ground truth" needs to be established and performance metrics (e.g., sensitivity, specificity, AUC) are calculated based on a test set. This submission is for a physical medical device (dental abutment) where the primary evaluation is based on engineering design, material properties, and manufacturing processes, demonstrated through comparison to legally marketed predicate devices and validation of changes like sterilization.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151247
    Manufacturer
    Date Cleared
    2015-08-10

    (91 days)

    Product Code
    Regulation Number
    872.3630
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    NC Straumann Screw Retained Abutments

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Straumann® Screw Retained Abutments are indicated to be placed into the NC and RC Bone Level implants of the Straumann® Dental Implant System to provide support for prosthetic reconstructions such as crowns, bridges and bars.

    The final processed devices have the purpose of restoring chewing function.

    Straumann® Screw Retained Abutments are indicated for screw-retained restorations.

    Device Description

    The proposed Bone Level Screw Retained Abutments are a line extension of the Screw Retained Abutments previously cleared to market per K133421 and K141871. The proposed Screw Retained Abutment devices include one-piece straight and angled (17° and 30°) abutments, having a gingival height of 5.5 mm.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and studies, with the understanding that this document is a 510(k) summary for a medical device and not a report for an AI/ML device. Therefore, many of the requested fields related to AI/ML studies will not be applicable.

    The document discusses the Straumann® Screw Retained Abutments, which are dental implant components.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in the traditional sense of numerical thresholds for performance. Instead, it demonstrates substantial equivalence to predicate devices through material and technological characteristics, and performance data from bench studies.

    Feature / CriterionAcceptance Criteria (Implied)Reported Device Performance
    Material (Composition)Conformance to ISO 5832-11 (Ti-6Al-7Nb)Ti-6Al-7Nb titanium alloy conforming to ISO 5832-11
    Manufacturing ProcessPrecision machining from solid material (one-piece construction)Precision machining from solid material (one-piece construction)
    Mechanical Strength/Durability (Fatigue)Consistent with FDA guidance and ISO 14801Dynamic fatigue test data consistent with FDA guidance and ISO 14801 was referenced. (Specific numerical results not provided in this summary)
    Implant-to-Abutment ConnectionIdentical to predicate devices (NC, RC Cross Fit)NC & RC Cross Fit
    Platform Diameter(s)Equivalent/Comparable to predicate devicesNC: Ø4.6 mm, RC: Ø4.6 mm (Note: Predicate had NC: Ø3.5 mm, Ø4.6 mm, RC: Ø4.6 mm)
    Abutment Angulation(s)Equivalent/Comparable to predicate devicesNC: 17°, 30°, RC: 17°, 30° (Note: Predicate had 0°, 17°, 30°)
    Gingival Height(s)Equivalent/Comparable to predicate devicesNC: 5.5 mm, RC: 5.5 mm (Note: Predicate had 1.0, 2.5 and 4.0 mm)
    Orientation of Angulation to Engagement FeaturesIdentical to predicate devices (Type A (45°), Type B (0°))NC: Type A (45°), Type B (0°), RC: Type A (45°), Type B (0°)
    Primary PackageMedical grade polyethylene blister with a sealing lidMedical grade polyethylene blister with a sealing lid
    SterilizationNon-sterile; intended for terminal sterilization via moist heat (autoclave)Non-sterile; intended for terminal sterilization via moist heat (autoclave)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document refers to "bench studies" and "dynamic fatigue test data" but does not explicitly state the sample size (number of abutments tested) for these tests. There's no information about the country of origin of the data or whether it was retrospective or prospective, as these are typically not relevant for benchtop mechanical testing of physical devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the device is a physical dental implant abutment, not an AI/ML imaging analysis tool. "Ground truth" in this context would come from engineering specifications, material standards, and physical measurements/tests, not expert consensus on interpretations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable for a physical device undergoing bench testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for showing substantial equivalence relies on:

    • Engineering specifications and material standards: Conformance to ISO standards for material composition (ISO 5832-11) and mechanical testing (ISO 14801).
    • Physical measurements: Dimensions (platform diameter, gingival height), angles, and connection types for comparison with predicate devices.
    • Bench test results: Data from dynamic fatigue tests to demonstrate mechanical performance under simulated conditions.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1