Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K222299
    Date Cleared
    2022-11-30

    (121 days)

    Product Code
    Regulation Number
    862.1675
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Steripath Micro Blood Collection System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Steripath® Micro Blood Collection System is a system to draw blood for in vitro diagnostic testing.

    The Steripath® Micro Blood Collection System is indicated for use as a blood collection system that diverts and sequesters the initial specimen prior to collection of a subsequent test sample to reduce the frequency of blood culture contamination when contaminants are present in the initial blood cultures drawn with standard procedure without manual diversion.

    Additionally, components of the system may be used for infusion following sample collection after discomection of the Initial Specimen Diversion Device® (ISDD®).

    Device Description

    The device, Steripath® Micro Blood Collection System, is a system to draw blood for in vitro diagnostic testing. The purpose of this submission is to expand the product line to include inlet components with the BD Venipuncture® UltraTouch™ Push Button Needle, outlet components with a transfer adapter, and universal models with no inlet component.

    The Steripath® Micro Blood Collection System is indicated for use as a blood collection system that diverts and sequesters the initial specimen prior to collection of a subsequent test sample to reduce the frequency of blood culture contamination when contaminants are present in the initial blood sample compared to blood cultures drawn with standard procedure without manual diversion.

    The device sequesters and diverts 0.5 - 1.0 mL of the initial specimen of blood (potentially contaminated) into a diversion chamber. Once diversion is complete, a subsequent blood sample flows through a second pathway within the device and is collected either directly into a culture bottle (not provided by Magnolia Medical Technologies), or into a syringe that is used to inoculate culture bottles. Additionally, components of the system may be used for infusion after disconnection of the Initial Specimen Diversion Device® (ISDD®).

    The Steripath® Micro Blood Collection System, needle configurations, for the subject device are manufactured using non-sterile BD UltraTouch® needles, which are equivalent to the commercially available BD UltraTouch® Push Button™ Blood Collection Set (K212724), except for the sterilization step. The Steripath® Micro Blood Collection System and UltraTouch® needles have the same intended use. Incorporation of the needle with the ISDD® functions on the same technological characteristics and principles of operation with the difference that the subject device leverages the Initial Specimen Diversion Technology® (ISDT) and mechanically sequesters the initial specimen of blood. Inclusion of this technology does not raise any added questions of safety or effectiveness.

    The different configurations of the Steripath® Micro Blood Collection System are made available by using different combinations of the inlet and outlet accessories. Proposed configurations are listed in Table 5.1.

    AI/ML Overview

    The provided text describes a medical device, the Steripath® Micro Blood Collection System, and its 510(k) premarket notification for substantial equivalence to a predicate device. The information details the device's characteristics and various types of testing performed to demonstrate its safety and effectiveness.

    Here's an analysis of the acceptance criteria and the study information as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    RequirementDescriptionAcceptance Criteria (Implied)Reported Device Performance
    SequestrationThe ISDD® shall sequester the diversion volume.Effective sequestration of the diversion volume.PASS
    Minimum Vacuum performance.The ISDD® shall meet its performance requirements under minimum vacuum conditions.Performance within specifications under minimum vacuum.PASS
    Maximum Vacuum performance.The ISDD® shall meet its performance requirements under maximum vacuum conditions.Performance within specifications under maximum vacuum.PASS
    Diversion Volume.The ISDD® shall meet the minimum and maximum diversion volume requirements.Diversion volume between 0.5mL and 1.0mL.PASS
    Positive Pressure MaximumThe ISDD® shall remain functionally intact and safe under maximum positive pressure conditions.Remaining functionally intact and safe under maximum positive pressure.PASS
    SterilizationCompliance with gamma radiation process.Sterility Assurance Level (SAL) of 10^-6 at 25kGy dose.Complies
    Aging/Shelf-Life (12 months)Maintenance of functional properties over time.Functional (needle insertion, blood collection, needle retraction, hub attachment, culture bottle septum puncture, blood collection, removal), diversion volume, and pressure rating maintained.Validated to 12 months
    BiocompatibilityMeets requirements for biological evaluation.Compliance with ISO 10993-1 for indicated contact duration.Meets requirements
    Packaging Integrity / ShippingPackaging maintain integrity during transport.Compliance with ASTM D4169-16, Distribution Cycle 13, Assurance Level II; seal strength, package leaks, drop tests met.Meets requirements
    Functional (General)Meets all functional and performance requirements for safe and effective performance.Safe and effective performance for intended use.Successfully meets all

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not explicitly state the sample sizes used for the individual performance tests (sequestration, vacuum, diversion volume, etc.). It describes the types of tests performed (e.g., Sterilization, Aging/Shelf-Life, Biocompatibility, Packaging Integrity, and Performance Testing) and states that "all design verification and validation activities were performed."

    The data provenance is not specified in terms of country of origin or whether it was retrospective or prospective data. The testing appears to be primarily laboratory-based engineering and performance verification testing rather than large-scale clinical studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the text. The type of testing described (e.g., sterilization, biocompatibility, mechanical performance) typically relies on established standards and laboratory protocols rather than human expert interpretation for "ground truth" in the way a diagnostic imaging study would.

    4. Adjudication Method for the Test Set

    This information is not applicable or not provided. Since the testing focuses on objective mechanical and biological properties against established standards, an adjudication method for conflicting expert opinions is not relevant in the context of the described tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done based on the provided text. The submission focuses on device characteristics and performance verification for substantial equivalence, not comparative effectiveness of human readers with or without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, this device is a physical medical device (blood collection system), not an AI algorithm. Therefore, "standalone" algorithm performance is not applicable.

    7. The Type of Ground Truth Used

    The ground truth used for the device's performance is based on established industry standards and engineering specifications for medical devices. For example:

    • Sterilization: Standards like ISO 11137-1.
    • Aging/Shelf-Life: Standards like ASTM F1980-16.
    • Biocompatibility: Standards like ISO 10993-1.
    • Packaging: Standards like ASTM D4169-16.
    • Functional requirements (e.g., diversion volume, vacuum performance) are based on the device's design specifications.

    8. The Sample Size for the Training Set

    This device is not an AI/ML algorithm that requires a training set. Therefore, this information is not applicable.

    9. How the Ground Truth for the Training Set Was Established

    This device is not an AI/ML algorithm. Therefore, this information is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200661
    Date Cleared
    2020-10-08

    (210 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Steripath Micro Blood Collection System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Steripath® Micro Blood Collection System is a system to draw blood for in vitro diagnostic testing.

    The Steripath® Micro Blood Collection System is indicated for use as a blood collection system that diverts and sequesters the initial specimen prior to collection of a subsequent test sample to reduce the frequency of blood culture contamination when contaminants are present in the initial blood sample compared to blood cultures drawn with standard procedure without manual diversion.

    Additionally, components of the system may be used for infusion following sample collection after disconnection of the Initial Specimen Diversion Device® (ISDD®).

    Device Description

    The Steripath® Micro Blood Collection System diverts and sequesters the initial portion of the blood specimen (potentially contaminated blood) in the diversion reservoir. When diversion is complete, a subsequent blood sample flows through a second pathway within the device. The subsequent blood sample is collected into a syringe that is used to inoculate culture bottles. Upon removal of the Initial Specimen Diversion Device®, ISDD®, components of the system can be used for infusion per the included manufacturer's instructions for use.

    The subject device incorporates multiple syringe oulet accessory configurations that include and three sizes of syringes that are previously cleared as referenced below. The Steripath® Micro Blood Collection System are available:

    | Steripath®
    Micro Kit
    Model

    NumberISDD®Inlet AccessoryOutlet Accessory
    4005-ENP00353Luer Extension, 7"
    ICU Medical, Inc.
    Model B1754-NS
    K964435Syringe, 5ml
    Becton Dickinson
    Model 301027
    K980987
    4010-ENP00353Luer Extension, 7"
    ICU Medical, Inc.
    Model B1754-NS
    K964435Syringe, 10ml
    Becton Dickinson
    Model 301029
    K980987
    4020-ENP00353Luer Extension, 7"
    ICU Medical, Inc.
    Model B1754-NS
    K964435Syringe, 20ml
    Becton Dickinson
    Model 301031
    K980987
    AI/ML Overview

    The provided document is a 510(k) summary for the Steripath® Micro Blood Collection System (K200661). It details the device's indications for use, technology, comparison to a predicate device, and performance testing. However, it does not contain specific acceptance criteria, reported device performance metrics (e.g., sensitivity, specificity, accuracy, or contamination rates), sample sizes for test sets, data provenance, information about expert ground truth, adjudication methods, MRMC study details, or standalone performance metrics for the aspects related to the "acceptance criteria and study that proves the device meets the acceptance criteria" as requested in the prompt.

    The "Summary of Performance Testing" section lists some requirements and states "PASS" for each, but this is a high-level summary and doesn't provide the detailed numbers, methodologies, or study specifics that would typically characterize a study proving device performance against acceptance criteria. It primarily focuses on comparing the new device to its predicate and ensuring it meets safety and functionality standards.

    Therefore, many aspects of your request cannot be fulfilled by the provided document.

    Here's a breakdown of what can and cannot be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document mentions several requirements and functional tests:

    • Sequestration
    • Minimum Vacuum performance
    • Maximum Vacuum performance
    • Diversion Volume
    • Positive Pressure Maximum

    For each of these, the "Verification Test Result" is reported as "PASS". However, the actual acceptance criteria (e.g., minimum percentage of sequestration, specific vacuum pressure ranges, specific diversion volume ranges) are not detailed in this document. The "PASS" result indicates the device met unstated criteria.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for the listed performance tests.
    • Data Provenance: Not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable/Not provided. The performance tests described (e.g., sequestration, vacuum performance) are technical/engineering tests, not clinical evaluations requiring expert interpretation of ground truth in the context of diagnostic accuracy.

    4. Adjudication method for the test set:

    • Not applicable/Not provided. These are technical tests, not diagnostic interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a blood collection system, not an AI diagnostic device that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a mechanical blood collection system, not an algorithm.

    7. The type of ground truth used:

    • For the listed performance tests (Sequestration, Vacuum performance, Diversion Volume, Positive Pressure Maximum), the "ground truth" would be established by the physical and functional specifications of the device, measured by engineering testing methods. It is not expert consensus, pathology, or outcomes data in a clinical diagnostic sense. The document states "all design verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met."

    8. The sample size for the training set:

    • Not applicable. This is a mechanical device, not an AI/machine learning model that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable.

    In summary, the provided document focuses on substantiating equivalency to a predicate device through general functional and safety testing, rather than presenting detailed clinical performance study results against specific, quantitatively defined acceptance criteria for diagnostic accuracy or contamination rates with detailed study methodologies. The "PASS" results for functional requirements confirm the device meets its design specifications but lacks the granular data requested in your prompt.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1