Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K211564
    Date Cleared
    2021-10-19

    (152 days)

    Product Code
    Regulation Number
    870.1650
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Sterile High-pressure Angiographic Syringes for Single-use

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline: they shall be used with US legally marketed angiographic injectors.

    Device Description

    The subject devices are identical to all models of predicate devices of K192657. It includes disposable syringes, connection tube, J shape tube and spike. Labeling of subject device now includes pediatrics in the patient population.

    AI/ML Overview

    This document is a 510(k) summary for Sterile High-pressure Angiographic Syringes for Single-use. It primarily focuses on demonstrating substantial equivalence to a predicate device (K192657) and does not contain information about a study proving specific acceptance criteria in the context of device performance metrics like accuracy, sensitivity, or specificity for AI/ML technologies.

    This type of submission typically demonstrates the device meets a set of performance standards or specifications, rather than a "study that proves the device meets the acceptance criteria" in the way a clinical trial or AI performance study would. The acceptance criteria here relate to physical, chemical, and biological properties, as well as compatibility.

    Therefore, many of the requested details, such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance studies, and training set details, are not applicable to this type of device submission.

    Here's the available information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document lists various non-clinical tests and standards that the device complies with, along with some performance specifications.

    ItemAcceptance Criteria (Standard/Test/Specification)Reported Device Performance
    Product CodeDXTDXT
    Regulation NumberCFR 870.1650CFR 870.1650
    Indications for UseFor injection of contrast media or saline with US legally marketed angiographic injectors.Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline; they shall be used with US legally marketed angiographic injectors.
    Mode of operationPower-driven operation, single usePower-driven operation, single use
    SterilityEO SterilizedEO Sterilized
    Single UseYesYes
    Max. Withstanding Pressure (Syringe)300psi, 400psi, 1200psi300psi, 400psi, 1200psi (per specific models, e.g., 100101 at 400psi, 300101 at 1200psi)
    Max. Withstanding Pressure (Connection Tube)300psi, 400psi, 1200psi300psi, 400psi, 1200psi (per specific models)
    Max. Withstanding Pressure (Spike)400 psi400 psi (for specific models)
    Syringe PerformanceISO 7886ISO 7886
    Luer ConnectorISO 594-1; ISO 594-2ISO 594-1; ISO 594-2
    CompatibilityPass (with specific injectors outlined in Tables 1, 2, 3)Pass
    BiocompatibilityNo Cytotoxicity, No Irritation, No Sensitization, No Pyrogen, No Acute Toxicity, No HemolysisNo Cytotoxicity, No Irritation, No Sensitization, No Pyrogen, No Acute Toxicity, No Hemolysis
    Endotoxin Limit20 EU per device20 EU per device
    EO/ECH Residue LimitLimited Contact: ≤ 24h, EO: 0.6mg/day, ECH: 1.28mg/day for 10kg patient (Children) as per ISO 10993-7EO/ECH residual test has been conducted on the models in this submission and test results have demonstrated EO/ECH residue level meet the residue limit for 10kg children group as per ISO 10993-7. (This is a specific test mentioned to justify the change in labeling regarding pediatric use compared to the predicate).
    Sterilization ResidualsISO 10993-7:2008 (AMD1:2019)Complies with standard
    Seal StrengthASTM F88/F88M-15Complies with standard
    Seal Leak DetectionASTM F1929-15Complies with standard
    Bacterial EndotoxinsUSP 41-NF36 2018Complies with standard
    Sterile Hypodermic Syringes (Manual Use)ISO 7886-1:2017Complies with standard
    Sterile Hypodermic Syringes (Power-driven)ISO 7886-2:1996Complies with standard
    Conical fittings (Luer taper)ISO594-1:1986, ISO594-2:1998Complies with standards
    In Vitro CytotoxicityISO 10993-5:2009Complies with standard (No Cytotoxicity)
    Irritation and Skin SensitizationISO 10993-10:2010Complies with standard (No Irritation, No Sensitization)
    Interactions with BloodISO 10993-4:2017Complies with standard (No Hemolysis)
    Hemolytic PropertiesASTM F756:2017Complies with standard (No Hemolysis)
    Pyrogen TestUSP 41 NF 36Complies with standard (No Pyrogen)
    Sterilization ProcessISO 11135:2014Complies with standard
    Particulate MatterUSPComplies with standard
    Shelf life5 years5 years

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • The document does not specify sample sizes for the individual non-clinical tests (e.g., number of syringes tested for pressure, number of samples for biocompatibility).
    • Data provenance: Not explicitly stated, but the submission is from Shenzhen Boon Medical Supply Co., Ltd. in China, suggesting the tests were likely conducted in China. The submission itself is dated August 13, 2021, and received August 23, 2021, implying these are retrospective test results for device validation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • This is not applicable. The "ground truth" for this medical device (angiographic syringes) involves meeting engineering specifications and biological safety standards, not diagnostic interpretations that require expert consensus. The compliance is against defined standards (e.g., ISO, ASTM, USP) and internal specifications.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • This is not applicable for this type of device. Adjudication methods are typically used in clinical studies for diagnostic accuracy, where disagreements between readers or algorithms need resolution. Here, the tests yield objective pass/fail results against established criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This is not applicable. The device is a physical medical instrument (syringe), not an AI/ML diagnostic or assistive tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • This is not applicable. The device is a physical medical instrument (syringe), not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" in this context is defined by international and national standards (e.g., ISO 10993, ASTM F88, USP 41) and the device's own engineering specifications (e.g., maximum pressure resistance, volume, material properties). Compliance is measured against these objective criteria.

    8. The sample size for the training set

    • This is not applicable. There is no AI/ML component, therefore no "training set."

    9. How the ground truth for the training set was established

    • This is not applicable. There is no AI/ML component, therefore no "training set" ground truth.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    Sterile High-pressure Angiographic Syringes for single-use

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The proposed device is intended for the injection of contrast media or saline. This syringe use with US legally marketed angiographic injectors.

    Device Description

    The proposed devices are available in packs, which may include different configurations of syringes and accessories. The syringes are plastic, disposable syringes, which are available in various models and configurations. They are intended to be used with an U.S. legally marketed angiography injector, compatibilities are shown in Table 1. The connection tubes, which is used to connect the syringe and the catheter. The tubes are available in two configurations, which are Type Y and Type T tube. The different between these tubes is tube shape, which is available in Y shape for Y tube and T shape for T tube. J shape tube, which is used to draw contrast media/ saline into the syringe barrel. Spikes, which are used to draw contrast media/saline into the syringe barrel.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device: "Sterile High-pressure Angiographic Syringes for Single-use". This type of submission focuses on demonstrating "Substantial Equivalence" to a legally marketed predicate device, rather than proving novel effectiveness or safety through extensive clinical trials.

    As such, the document does not describe a study to prove the device meets acceptance criteria in the way you've outlined for an AI-powered diagnostic device. Instead, it details non-clinical tests to show the proposed device performs comparably to the predicate and meets relevant engineering standards.

    Therefore, many of your requested points regarding Acceptance Criteria and Study Proof (especially related to AI, MRMC, expert consensus, and ground truth establishment) are not applicable to this 510(k) submission for a non-AI medical device.

    However, I can extract the relevant information from the document regarding the "acceptance criteria" (in the sense of standards compliance and comparative performance) and the "study" (non-clinical tests) that proves these criteria are met.

    Here's a breakdown based on the provided document:


    Device: Sterile High-pressure Angiographic Syringes for Single-use

    Device Type: Non-AI Medical Device (Angiographic Syringes)

    Study Purpose: To demonstrate Substantial Equivalence (SE) to a predicate device (Shenzhen BaoAn Medical Supplies Co., Ltd. Sterile High-pressure Angiographic Syringes for Single-use, K151960) through non-clinical testing and comparison.


    1. A table of acceptance criteria and the reported device performance

    For a non-AI device like this syringe, "acceptance criteria" primarily refer to adherence to recognized international standards and demonstration of comparable performance to a predicate device. "Reported device performance" is the outcome of non-clinical tests demonstrating compliance.

    Acceptance Criteria (Standard / Property)Basis / Test MethodReported Device Performance (Result & SE Justification)
    Performance
    Compliance with Syringe StandardsISO 7886-1 (Sterile hypodermic syringes for single use), ISO 594-1 (Conical fittings with a 6% (Luer) taper), ISO 594-2 (Lock Fitting)Complied with ISO 7886-1, ISO 594-1, and ISO 594-2. This demonstrates performance equivalent to the predicate.
    Seal StrengthASTM F88/F88M-15 (Standard test method for seal strength of flexible barrier materials)Test results met the standard.
    Seal Leaks (Porous Medical Package)ASTM F1929-15 (Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration)Test results met the standard.
    Connection Tube PerformanceISO 594-1 and ISO 594-2Performance test results of connection tube meet the criteria. Even with more product specifications (various lengths), the performance is maintained.
    J-shape Tube LeakageNon-standardized test, implied to ensure functionality.Performance test results showed no leaks.
    Biocompatibility
    Biocompatibility AssessmentISO 10993-1: 2018 (Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process)Overall biocompatibility demonstrated; differences in materials (e.g., Piston, Filter membrane) from predicate are deemed not to affect SE due to successful biocompatibility testing (Analysis 2).
    CytotoxicityISO 10993-5:2009 (Tests for In Vitro cytotoxicity)No Cytotoxicity. (Same as predicate)
    IrritationISO 10993-10:2010 (Tests for irritation and skin sensitization)No Irritation. (Same as predicate)
    SensitizationISO 10993-10:2010 (Tests for irritation and skin sensitization)No Sensitization. (Same as predicate)
    Systemic ToxicityISO 10993-11:2017 (Tests for systemic toxicity)No Acute Toxicity. (Same as predicate)
    HemolysisASTM F 756-17 (Standard practice for assessment of hemolytic properties of material)No Hemolysis. (Same as predicate)
    PyrogenicityUSP 41-NF36:2018 (Pyrogen Test (USP Rabbit Test))No Pyrogen. (Same as predicate)
    Sterility & Particulate Matter
    Bacterial Endotoxins LimitUSP 41-NF36 2018 (Bacterial Endotoxins Limit)Test results met the standard.
    Sterility Assurance (Post-sterilization)ISO 11737-2:2009 (Sterilization of medical devices- Microbiological methods- Part 2: Test of sterility performed in the definition, validation and maintenance of a sterilization process)Test results met the standard. (EO sterilized, same as predicate).
    Ethylene Oxide Sterilization ResidualsISO 10993-7:2008 (Biological evaluation of medical devices- Part 7: Ethylene oxide sterilization residuals)Test results met the standard. (EO sterilized, same as predicate).
    Particulate MatterUSP (Particulate matters)Complied with USP 788. (Same as predicate)
    Material & Physical Compatibility
    Syringe Volume & Configuration Range Compatibility with Predicate's PerformanceComparison of various single and dual shot volumes (50ml to 200ml) and accessory types (connection tubes, J-shape tube, spikes) with the predicate device.While the proposed device has a wider range of sizes/configurations than the predicate, non-clinical tests confirmed that all these configurations meet the performance criteria of relevant ISO standards. This difference is deemed not to affect SE. (Analysis 1)
    Patient-Contact Material CompatibilityComparison of materials used for various components (Barrel, Piston, Connection Tube, Luer Connectors, Spike, Filter Membrane, J-shape tube) with the predicate. Explicitly, changes in Piston (Medical rubber vs. Polyisoprene rubber) and Filter Membrane (Medical rubber vs. Polypropylene) from the predicate.Differences in Piston and Filter Membrane materials exist. However, the contact classification (blood path, indirect limited contact) remains the same as the predicate. Extensive biocompatibility testing (Cytotoxicity, Irritation, Sensitization, Acute Toxicity, Hemolysis, Pyrogen) on the
    proposed device's materials showed no adverse effects. Therefore, these material differences are not considered to affect substantial equivalence. (Analysis 2)
    Labeling21 CFR Part 801 (Medical Device Labeling)Complied with 21 CFR Part 801. (Same as predicate)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not explicitly state the number of units/samples used for each non-clinical test (e.g., how many syringes were tested for seal strength, or how many batches for endotoxin). It only states that "Non clinical tests were conducted to verify that the proposed device met all design specifications." The specific number of test replicates/samples would typically be detailed in the full test reports, which are not part of this summary document.
    • Data Provenance: The tests are non-clinical (laboratory/bench testing) performed to verify compliance with standards. The company "Youwo (Guangzhou) Medical Device Co., Ltd." is based in Guangzhou, P.R. China, implying the testing was conducted in China, likely at their own facilities or certified labs in China. The data is entirely lab-generated, not from human subjects. The nature of the testing is prospective, in the sense that the tests were performed specifically for this submission on new device samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This is a non-AI, non-diagnostic medical device. There is no concept of "ground truth" established by human experts in the context of diagnostic interpretation. The "ground truth" for the performance of these syringes is defined by the objective pass/fail criteria of the recognized engineering and biocompatibility standards (e.g., is the seal strong enough? Is there pyrogen presence?). Compliance is measured by laboratory equipment and established protocols, not expert human review of an image or signal.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. As per point 3, there's no human diagnostic interpretation to adjudicate. Test results are objective measurements against pre-defined thresholds from standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This device is not an AI diagnostic tool, nor is it intended to assist human readers in interpreting medical cases. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an algorithm or AI device. Standalone performance refers to the device's ability to meet its specifications independently, which is what the non-clinical tests implicitly demonstrate (e.g., a syringe's seal strength is measured independently of human operation).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Standards-Defined Ground Truth: The "ground truth" here is the adherence to internationally recognized performance and safety standards (e.g., ISO, ASTM, USP). The device is "true" if it passes the specified tests according to the criteria defined within those standards. For example, a "true" biocompatible device produces no cytotoxicity according to ISO 10993-5. This is determined by laboratory measurements and chemical/biological assays, not clinical outcomes or expert diagnostician consensus.

    8. The sample size for the training set

    • Not Applicable. This is not an AI/machine learning device. There is no training set.

    9. How the ground truth for the training set was established

    • Not Applicable. This is not an AI/machine learning device. No ground truth for a training set was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K192657
    Date Cleared
    2020-08-07

    (317 days)

    Product Code
    Regulation Number
    870.1650
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Sterile High-pressure Angiographic Syringes for Single-use

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline; they shall be used with an US legally marketed angiographic injectors.

    Device Description

    The proposed device is intended for the injection of contrast media or saline. It includes disposable syringes, connection tube, J shape tube and spike.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device (Sterile High-pressure Angiographic Syringes for Single-use). It describes non-clinical testing performed to demonstrate substantial equivalence to a predicate device, rather than a study proving the device meets acceptance criteria in a clinical setting.

    Therefore, I cannot provide information regarding:

    • Acceptance Criteria and Reported Device Performance for clinical outcomes: The document focuses on performance specifications (e.g., pressure, sterility, biocompatibility) of the device components, not clinical performance metrics or specific acceptance criteria for a clinical study.
    • Sample size for the test set and data provenance: No clinical test set is described. The "test set" here refers to samples used for non-clinical engineering and biological tests.
    • Number of experts and qualifications, and adjudication method: These pertain to clinical studies with a ground truth established by experts, which is not applicable here.
    • Multi-reader multi-case (MRMC) comparative effectiveness study: This is a type of clinical study not performed or referenced in this document.
    • Effect size of human readers with/without AI assistance: This is irrelevant as the device is a syringe and connection tube, not an AI-assisted diagnostic tool.
    • Standalone (algorithm only without human-in-the-loop performance): This is also irrelevant for this type of medical device.
    • Type of ground truth used: Ground truth, typically defined by expert consensus or pathology in a clinical context, is not applicable here. The "ground truth" for these non-clinical tests is adherence to established engineering and biological standards.
    • Sample size for the training set and how the ground truth for the training set was established: These concepts are related to machine learning and AI, which are not relevant to the described device.

    What the document does provide in relation to "acceptance criteria" and "study that proves the device meets the acceptance criteria" (within the scope of a 510(k) for a physical medical device):

    The document details a series of non-clinical tests performed to demonstrate that the proposed device meets established performance standards and is substantially equivalent to a predicate device. These standards effectively act as the "acceptance criteria" for the device's design and manufacturing.

    Here's a summary of the non-clinical acceptance criteria (standards/tests) and the reported performance:

    1. Table of Acceptance Criteria and the Reported Device Performance (Non-Clinical)

    Acceptance Criteria (Standard/Test)Reported Device Performance
    Biocompatibility
    ISO 10993-7:2008 (Ethylene oxide sterilization residuals)Complies (implied by "Non clinical tests were conducted to verify... met all design specifications")
    ISO 10993-5:2009 (Tests for In Vitro cytotoxicity)No Cytotoxicity
    ISO 10993-10:2010 (Tests for irritation and skin sensitization)No Irritation, No Sensitization
    ISO 10993-4:2017 (Selection of tests for interactions with blood) & ASTM F756:2017 (Hemolytic Properties)No Hemolysis
    **USP 41 NF 36 ** (Pyrogen Test)No Pyrogen
    Endotoxin Limit (USP 41-NF36 2018 )20 EU per device
    Functionality & Performance
    ISO 7886-1:2017 (Sterile hypodermic syringes for manual use)Complies with test items (Overall length of scale, syringe dimension, nozzle lumen, dead space, liquid leakage, air leakage, flow characteristics)
    ISO 7886-2:1996 (Sterile hypodermic syringes for use with power-driven syringe pumps)Complies with test items (Lubricant weight, Graduated capacity tolerance, overall length of scale, syringe dimension, nozzle lumen, dead space, liquid leakage, air leakage and flow characteristics)
    ISO 594-1:1986 (Conical fittings with a 6% (Luer) taper - General Requirements)Complies with test items (Gauging test, liquid leakage, separation force and stress cracking)
    ISO 594-2:1998 (Conical fittings with a 6% (Luer) taper - Lock Fitting)Complies with test items (Liquid leakage, air leakage, separation force, unscrew torque, ease of assembly, resistance to overriding and stress cracking)
    Compatibility Test Report (Injectors, syringe, connection tube, spike/J shape tube)Each device meets performance under maximum sustained pressure specification. Syringe, connection tube, and spike can withstand stated maximum pressure.
    Sterilization & Packaging
    ASTM F88/F88M-15 (Seal strength of flexible barrier materials)Complies
    ASTM F1929-15 (Detecting Seal Leaks in Porous Medical Package by Dye Penetration)Complies
    ISO 11134:2014 (Ethylene Oxide Sterilization)Complies
    **USP ** (Particular Matter in Injections)Complies
    Shelf Life Testing5-year shelf life demonstrated after accelerated aging.

    2. Sample size used for the test set and the data provenance:

    • The document does not specify the exact sample sizes (number of units) used for each non-clinical test. It generally states that "Non clinical tests were conducted."
    • The data provenance is from non-clinical laboratory testing performed by the manufacturer (Shenzhen Boon Medical Supply Co., Ltd) in China, as per the sponsor identification. These are not human clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Ground truth in this context refers to established engineering and biological standards (e.g., ISO, ASTM, USP), not expert clinical adjudication. Compliance with these standards is determined by trained technicians/engineers through validated testing methods.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This concept relates to multi-expert clinical review, which is not relevant for non-clinical device testing verifying adherence to standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done:

    • No. This is not relevant to a sterile syringe product.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is not an algorithm.

    7. The type of ground truth used:

    • For this device, the "ground truth" for proving it meets "acceptance criteria" is adherence to recognized international and national standards for medical device performance, biocompatibility, and sterility (e.g., ISO 7886, ISO 594, ISO 10993, ASTM F88, USP monographs).

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device requiring a training set.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182289
    Date Cleared
    2018-10-10

    (48 days)

    Product Code
    Regulation Number
    870.1650
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Sterile High-pressure Angiographic Syringes for Single-use

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline; they shall be used with an US legally marketed angiographic injectors.

    Device Description

    The proposed devices include 2 models of plastic, disposable syringes and 5 connection tubes of different specifications. They are intended to be used with an U.S. legally marketed angiography injector, compatibilities are shown in Table 1. The connection tubes are used to connect the syringe and the catheter. The tubes are also available in three configurations, which are straight tube (used with single shot syringe), type Y and type T tube (used with dual shot syringe), they can withstand maximum pressure of 400 psi. The proposed devices are used with the J shape tube / spike cleared in K151960.

    AI/ML Overview

    The provided FDA 510(k) summary (K182289) describes the substantial equivalence of Sterile High-pressure Angiographic Syringes for Single-use to a predicate device, focusing on mechanical and material properties rather than AI/algorithm performance. Therefore, most of the requested information regarding AI acceptance criteria, study design (MRMC, standalone), ground truth establishment, and training/test set details for AI models is not applicable to this document.

    The document primarily addresses the physical and functional characteristics of a medical device (syringes and connection tubes) and its compliance with relevant international standards.

    However, based on the non-clinical tests described, we can infer a form of acceptance criteria and the "study" (non-clinical testing) that proves the device meets these criteria.

    Here's an interpretation based on the provided document:

    Acceptance Criteria and Device Performance (Non-AI Medical Device)

    Acceptance Criteria (Inferred from Standards & Claims)Reported Device Performance (Non-Clinical Test Conclusion)
    Resistant Liquid Leak Pressure: Must withstand a certain pressure without leakage. (Implicitly, the proposed device seeks to demonstrate a higher resistance than the predicate)Syringe: 400 psi (Tested)
    Connection Tube: 400 psi (Tested)
    Material & Design Compatibility: Conformance to specified luer taper standards for secure connections.ISO 594-1:1986 (Conical fittings, Part 1: General requirements) - Complies
    ISO 594-2:1998 (Conical fittings, Part 2: Lock fittings) - Complies
    Syringe Functionality (Manual Use): Conformance to general requirements for single-use hypodermic syringes for manual use.ISO 7886-1:2017 (Sterile hypodermic syringes for single use, Part 1: Syringes for manual use) - Complies
    Syringe Functionality (Power-Driven Use): Conformance to requirements for single-use hypodermic syringes for use with power-driven pumps.ISO 7886-2:1996 (Sterile hypodermic syringes for single use, Part 2: Syringes for use with power-driven syringe pumps) - Complies
    Sterility: EO sterilized.EO Sterilized
    Single Use: Designed for single use.Yes
    Biocompatibility: Conformance to standards for biological evaluation of medical devices.ISO 10993 series Standards - Conforms to requirements

    Explanation of "Study" Proving Acceptance Criteria (Non-Clinical Testing):

    The "study" here refers to non-clinical bench testing conducted to verify the device's technical specifications and compliance with relevant standards.

    1. Sample Size Used for the Test Set and Data Provenance:

      • Sample Size: Not explicitly stated as a number of units tested. The document mentions "The test samples of syringe 100104 and 100114 with accessories are tested for liquid leakage and flow characteristics." This implies a series of physical tests on multiple units of the specified models. Regulatory submissions for devices of this nature typically involve testing a statistically representative sample size, but the exact number isn't provided in this summary.
      • Data Provenance: The tests were conducted internally by the manufacturer, Shenzhen Baoan Medical Supplies Co., Ltd. (China). The data is retrospective in the sense that it was collected as part of product development and validation prior to submission.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • Not Applicable. For a medical device like a syringe, "ground truth" is established through engineering specifications, material science, and performance standards, not through expert consensus on medical images or diagnoses. The "truth" is whether the device meets the physical and functional requirements (e.g., does it leak at 400 psi? Does it conform to luer taper dimensions?). These are objective measurements not requiring expert clinical adjudication.
    3. Adjudication Method for the Test Set:

      • Not Applicable. As mentioned above, the tests are objective, measurable physical and mechanical tests. There is no subjective interpretation requiring adjudication by multiple readers or experts.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

      • No. An MRMC study is a clinical study design typically used for diagnostic devices (like imaging AI) to assess the impact of a new technology on human reader performance. This document concerns a physical device (syringe), not a diagnostic tool or AI.
    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No. This is not an AI/algorithm. This question is irrelevant to the device described.
    6. The Type of Ground Truth Used:

      • Engineering Specifications and International Standards: The ground truth for this device's performance is defined by adherence to established international standards (ISO 594-1, ISO 594-2, ISO 7886-1, ISO 7886-2, ISO 10993 series) and the manufacturer's own design specifications (e.g., target resistant liquid leak pressure). This is a physical/mechanical "ground truth."
    7. The Sample Size for the Training Set:

      • Not Applicable. There is no "training set" as this is not an AI/machine learning device. The design and manufacturing processes are based on engineering principles and established device standards, not trained data.
    8. How the Ground Truth for the Training Set was Established:

      • Not Applicable. Again, this is not an AI/machine learning device. "Ground truth" for manufacturing and design would be derived from engineering drawings, material specifications, and regulatory requirements rather than labeled training data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K151960
    Date Cleared
    2016-01-07

    (175 days)

    Product Code
    Regulation Number
    870.1650
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Sterile High-pressure Angiographic Syringes for Single-use

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline; they shall be used with an US legally marketed angiographic injectors.

    Device Description

    The proposed devices are available in packs, which may include different configurations of syringes and accessories. The syringes are plastic, disposable syringes, which are available in various models and configurations. They are intended to be used with an U.S. legally marketed angiography injector, compatibilities are shown in Table 1.

    AI/ML Overview

    The provided document is a 510(k) summary for the Shenzhen Baoan Medical Supplies Co., Ltd. Sterile High-pressure Angiographic Syringes for Single-use. This type of submission is for medical devices, and the "acceptance criteria" and "device performance" primarily refer to compliance with recognized standards and biocompatibility, rather than clinical efficacy or diagnostic accuracy.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document lists various international and national standards that the device complies with, which serve as the acceptance criteria for its non-clinical performance and safety. The reported device performance is that it "complies with" these standards.

    Acceptance Criteria (Standards Complied With)Reported Device Performance
    ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residualsComplies
    ASTM F 88/F88M-09 Standard test method for seal strength of flexible barrier materialsComplies
    ASTM F1140/F1140M-13 Standard test methods for internal pressurization failure resistance of unrestrained packagesComplies
    ASTM F1929-12 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye PenetrationComplies
    USP37-NF32 Bacterial Endotoxins LimitComplies
    ISO 11737-2:2009 Sterilization of medical devices - Microbiological methods - Part 2: Test of sterilityComplies
    ISO 7886-1:1993 Sterile hypodermic syringes for single use - Part 1: Syringes for manual useComplies
    ISO 7886-2:1996 Sterile hypodermic syringes for single use - Part 2: Syringes for use with power-driven Syringe PumpsComplies
    ISO594:1986 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1Complies
    ISO594:1998 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2Complies
    ISO 10993-4:2002 AMD 2006 Biological evaluation of medical devices -- Part 4: Selection of tests for interactions with bloodComplies
    ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicityComplies
    ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationComplies
    ISO 10993-11:2006 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicityComplies
    ASTM F 756-08, Standard practice for assessment of hemolytic properties of materialComplies
    USP37-NF32 Pyrogen testComplies
    Biocompatibility requirements of ISO 10993 series StandardsConforms to the requirements (for both proposed and predicate devices)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states "Non clinical tests were conducted to verify that the proposed device met all design specifications..." and lists various standards for these tests. However, it does not specify the sample sizes used for these non-clinical tests (e.g., number of syringes tested for seal strength or internal pressurization). The data provenance is generally from laboratory testing performed by the manufacturer to demonstrate compliance with the listed standards. No country of origin is explicitly stated for the testing itself, but the manufacturer is based in China. The data is prospective in the sense that the manufacturer conducted these tests to support the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This type of information is not applicable to this 510(k) submission. The document focuses on the engineering and material safety aspects of a medical device (angiographic syringes), not on diagnostic accuracy where expert ground truth would be established. The "ground truth" here is compliance with established engineering and biological safety standards, which are verified through objective testing methods rather than expert consensus on images or case outcomes.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set:

    Not applicable. Adjudication methods like "2+1" typically refer to the process of resolving discrepancies in expert interpretations (e.g., in reading medical images). Since no expert interpretations are involved in establishing the "ground truth" for these engineering and biological tests, no such adjudication method was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical study is included in this submission." This type of study is relevant for assessing the clinical utility or diagnostic effectiveness of devices, often involving human readers interpreting medical data. This syringe device is a non-active, sterile, single-use product, and its evaluation focuses on its physical and biological integrity, not its interaction with clinical interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    Not applicable. This device is a physical medical product (angiographic syringe), not an algorithm or software. Therefore, the concept of "standalone performance" for an algorithm is not relevant here.

    7. The Type of Ground Truth Used:

    The "ground truth" for this device's evaluation is primarily based on compliance with recognized national and international consensus standards for medical device physical performance, sterility, and biocompatibility. This involves objective measurements against predefined thresholds and specifications outlined in standards like ISO 7886, ISO 10993, ASTM F88, etc.

    8. The Sample Size for the Training Set:

    Not applicable. The concept of a "training set" is relevant for machine learning algorithms. This submission is for a physical medical device, not an AI/ML product. The manufacturer did not develop this device using machine learning.

    9. How the Ground Truth for the Training Set was Established:

    The concept of a training set and its ground truth is not applicable for this device submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1