Search Results
Found 1 results
510(k) Data Aggregation
(396 days)
Sterile Biliary Stone Retrieval Balloon Catheter, Retrieval Balloon / short-wire compatible
The Retrieval Balloon Catheter is indicated for use endoscopically to remove stones from biliary system, or to facilitate injection of contrast medium while occluding the duct with the balloon.
The proposed device Retrieval Balloon Catheter includes Sterile Biliary Stone Retrieval Balloon Catheter (hereafter referred as Category 1) and Retrieval Balloon / short-wire compatible (hereafter referred as Category 2). Category 1 is commonly used in traditional ERCP surgery with a long guidewire (4.5m) while Category 2 adopts short-wire design which is compatible with a short guidewire (2.6m). For specifications of Category 2, the main feature of the short wire design is the C-shaped groove on the sidewall of catheter which is used to separate guidewire from the proposed device. The guidewire can be locked in place using Guidewire Locking Device to maintain guidewire access. Then the exchange of various devices can be performed without concern over wire displacement.
The proposed device Retrieval Balloon Catheter is sterile, single-use endoscopic device, and is indicated for use endoscopically to remove stones from biliary system, or to facilitate injection of contrast medium while occluding the duct with the balloon.
The proposed Retrieval Balloon Catheter is comprised of a natural latex balloon mounted at the distal end of a Pebax catheter with three internal lumens for ballooning, guidewire and contrast medium. The internal lumen for ballooning is used to inflate/deflate the balloon. Multiple syringes are included with the packaging, allowing balloon inflation to a specific diameter.
Following insufflation, the balloon surface lies flat against the bile duct wall, enabling efficient and complete cleaning of the bile duct. A separate lumen is designed for a 0.035 inch guidewire and the guidewire is able to be loaded either from the front or the back. Another separate lumen is designed for contrast medium. There are for injection-distal or injection-proximal options offered to support physician preference and facilitate procedural needs.
The balloon can be inflated to 9 mm, 12 mm, 15 mm, 18 mm diameters using the pre-measured syringes and a single balloon can be inflated to two or three distinct sizes if a different diameter is needed without having to exchange devices. During stone removal process, the balloon can be filled to removal multiple stones in the biliary tract.
There are two radiopaque bands placed at the distal and proximal ends of the balloon providing high fluoroscopic visualization of the balloon location.
This document describes the performance testing and acceptance criteria for the "Sterile Biliary Stone Retrieval Balloon Catheter, Retrieval Balloon / short-wire compatible" (Device Name: Retrieval Balloon Catheter), submitted under 510(k) K211021.
It is important to note that this submission does not involve Artificial Intelligence (AI) or machine learning. Therefore, information concerning human reader improvement with AI, standalone algorithm performance, or training set details are not applicable to this medical device submission. The device is a physical medical instrument, not a diagnostic AI tool.
1. Acceptance Criteria and Reported Device Performance
The submission states that "Performance testing was conducted to demonstrate the performance of the proposed device and confirmed that the proposed device works as intended with the compatible devices. Additionally, the results of the tests below were evaluated as substantially equivalent to the predicate device."
While specific numerical acceptance criteria and precise performance values are not detailed in this public summary, the document indicates that the following tests were conducted and the proposed device met the acceptance criteria and supported substantial equivalence to the predicate device.
Test Category | Acceptance Criteria (Implicit: Met requirements for substantial equivalence to predicate) | Reported Device Performance |
---|---|---|
Bench Tests | To confirm intended performance and substantial equivalence to predicate device (K102082) | All tests were conducted, and results supported substantial equivalence to predicate. Specific numerical values are not provided in this summary. |
Dimension | Met acceptance criteria. | |
Endoscope Compatibility | Met acceptance criteria. | |
Infusion Patency | Met acceptance criteria. | |
Guidewire Matching | Met acceptance criteria. | |
Balloon Fatigue | Met acceptance criteria. | |
Luer Connector | Met acceptance criteria. | |
Connection Strength | Met acceptance criteria. | |
Shelf-Life & Packaging | Compliance with ASTM F1980-16, ISO 11607-1:2019, ISO 11607-2:2019. Initial shelf life of 12 months. | Conducted based on accelerated aging. Initial results support 12 months. Two-year aging test to be performed for longer stability. |
Sterilization | Compliance with ISO 11135:2014+A1:2018 (EO Sterilization, SAL: 10^-6) | Validation carried out, results met requirements (SAL: 10^-6). |
Biocompatibility | Compliance with ISO 10993-1:2009 and FDA guidance. | All listed tests conducted, results met requirements. |
Testing Method | Test Results | |
Cytotoxicity | Met acceptance criteria. | |
Sensitization | Met acceptance criteria. | |
Irritation | Met acceptance criteria. | |
Acute Systemic Toxicity | Met acceptance criteria. | |
Pyrogen | Met acceptance criteria. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify the exact sample sizes for each of the bench tests, shelf-life tests, or biocompatibility tests. It states that "Performance testing was conducted" and "results of the tests below were evaluated as substantially equivalent."
- Data Provenance: The tests are described as bench tests and laboratory evaluations conducted by the manufacturer, Micro-Tech (Nanjing) Co., Ltd., which is located in Nanjing, Jiangsu Province, China. The data is retrospective in the sense that it was collected prior to the submission for regulatory review. It is not patient or clinical data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and their Qualifications
This is not applicable as the described tests are laboratory-based performance and safety evaluations of a physical medical device, not a diagnostic AI tool requiring expert image interpretation or clinical consensus for ground truth. The "ground truth" for these tests relates to engineering specifications, material properties, and biological safety standards. These are established by validated test methods and established standards (e.g., ISO, ASTM).
4. Adjudication Method for the Test Set
This is not applicable for the same reason as above. Adjudication methods like "2+1" or "3+1" are relevant for establishing ground truth in clinical data (e.g., for diagnostic accuracy studies involving human readers or AI), not for bench testing of a physical device.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and what was the effect size of how much human readers improve with AI vs. without AI assistance.
This is not applicable. This submission is for a physical medical device (Retrieval Balloon Catheter) and does not involve AI or machine learning. Therefore, no MRMC study or evaluation of human reader improvement with AI assistance was performed or is relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This is not applicable. As stated, this device is a physical medical instrument and does not involve an algorithm or AI.
7. The Type of Ground Truth Used
The "ground truth" for the performance evaluations of this device is based on:
- Engineering Specifications: Meeting predefined dimensional tolerances, force requirements, fluid flow rates, and fatigue limits.
- International Standards: Compliance with recognized standards such as ISO 11607 (packaging), ISO 11135 (sterilization), and ISO 10993 (biocompatibility).
- Predicate Device Equivalence: The primary "ground truth" for regulatory clearance is demonstrating substantial equivalence to a legally marketed predicate device (Extractor™ Pro Retrieval Balloon Catheter cleared under K102082) in terms of technology, safety, and performance.
8. The Sample Size for the Training Set
This is not applicable. There is no "training set" as this is not an AI/machine learning device.
9. How the Ground Truth for the Training Set was Established
This is not applicable. There is no "training set" to establish ground truth for.
Ask a specific question about this device
Page 1 of 1