    K Number
    K243061
    Manufacturer
    Date Cleared
    2024-11-04

    (38 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SonoStik Guide Wire Introducer is used to facilitate the introduction of a guide wire and catheter through the skin into a vein or artery of the peripheral vasculature. The SonoStik Guide Wire Introducer is not intended for use in the coronary arteries or neurovasculature.

    Device Description

    The SonoStik Guide Wire Introducer is intended to facilitate the introduction of a guide wire and catheter through the skin into a vein or artery of the peripheral vasculature. The device consists of a plastic housing containing an Introducer Wheel, Advancing Wheels, and a guide wire. The mechanism of action of the SonoStik Guide Wire Introducer is accomplished by the Introducer Wheel and the two Advancing Wheels inside the plastic housing. The SonoStik Guide Wire Introducer housing has a "male end" connector capable of mating with the back end of a compatible needle introducer and accurately positions the guide wire through the plastic housing and into the back end of the needle. Once mated, the Introducer and cannula/needle are used to insert the needle into the tissue or vessel. The Introducer Wheel is turned, advancing the guide wire into the vessel lumen. The device has a transparent tube at the proximal end of the device which enables the user to visualize guide wire advancement as they engage the introducer wheel.

    The SonoStik Guide Wire Introducer is compatible with and intended to be used with the MedSource Labs VeroTrue Conventional IV Catheter (K193278 catheter and needle set in 18-, 20-, and 22-gauge sizes.

    AI/ML Overview

    The provided text is a 510(k) summary for the SonoStik Guide Wire Introducer. It describes a medical device, its intended use, and the non-clinical tests performed to demonstrate its substantial equivalence to a predicate device.

    However, the questions you asked are largely related to software as a medical device (SaMD) or AI/ML-enabled medical devices, specifically in the context of diagnostic performance evaluation using methodologies like MRMC studies, ground truth establishment by experts, and analysis of metrics like sensitivity, specificity, or AUC.

    The SonoStik Guide Wire Introducer is a physical medical device, not software or AI. Therefore, the information typically requested for AI/ML device acceptance criteria and performance studies (e.g., sample size for test/training sets, number of experts for ground truth, adjudication methods, MRMC studies, effect size of human reader improvement with AI assistance, standalone algorithm performance, type of ground truth like pathology or outcomes data) is not relevant and not present in this document.

    The document focuses on the physical and functional aspects of the device.

    Here's what can be extracted from the document regarding the device's evaluation, though it won't directly answer most of your detailed questions about AI/ML device performance:

    1. A table of acceptance criteria and the reported device performance:

    The document broadly states: "Testing found the SonoStik Guide Wire Introducer met all applicable acceptance criteria associated with the introduction of a guide wire and catheter through the skin into a vein or artery of the peripheral vasculature."

    The specific acceptance criteria are not detailed in a table, but the types of tests performed indicate the areas of evaluation:

    Acceptance Criterion (Inferred from tests)Reported Device Performance
    BiocompatibilityMet applicable acceptance criteria. (Implies no adverse biological reactions)
    Bacterial EndotoxinMet applicable acceptance criteria. (Implies acceptable levels of endotoxins, indicating sterility or appropriate manufacturing)
    Packaging PerformanceMet applicable acceptance criteria. (Ensures product integrity, sterility, and protection during transport and storage, specifically noting "pouch seal were lengthened")
    Product Performance (Functionality)Met applicable acceptance criteria associated with the introduction of a guide wire and catheter through the skin into a vein or artery of the peripheral vasculature. (Implies successful guide wire advancement, compatibility with specified needles/catheters, and proper mechanical function after worst-case sterilization simulation, environmental conditioning, and 1-year accelerated aging). "Design control testing demonstrated that these changes do not negatively impact the performance of the proposed device when compared to the predicate."
    Material SafetyImplicitly met through biocompatibility and endotoxin testing, and by the nature of non-clinical testing for equivalence. Nitinol guidewire (new material) tested.
    Sterilization EffectivenessImplicitly met by bacterial endotoxin testing and performance after "worst-case sterilization simulation"
    Shelf-life/AgingMet acceptance criteria after "1-year accelerated aging".

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified. The document only mentions "worst-case non-clinical testing." For physical device performance, this usually means testing a sufficient number of units to demonstrate statistical confidence or to meet engineering specification requirements, but exact numbers are not provided.
    • Data Provenance: Not applicable in the context of clinical data for AI/ML devices. The testing was laboratory-based non-clinical performance and safety testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This is for a physical device, not an AI/ML diagnostic algorithm that requires expert interpretation for ground truth. Performance was assessed mechanically and through established laboratory methods.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This concept is relevant for establishing a reference standard from multiple expert opinions in diagnostic studies, not for the physical performance testing of a medical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a physical device, not an AI or imaging device where MRMC studies are performed to evaluate diagnostic accuracy and reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. There is no algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not applicable in the AI/ML sense. "Ground truth" for this device would be its ability to physically perform its intended function (e.g., smoothly advance a guidewire, maintain integrity) as measured by engineering specifications and industry standards, and demonstrating biocompatibility. This is determined by validated test methods and comparison to a legally marketed predicate device.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. No training set for an algorithm.

    In summary: The provided document is a 510(k) summary for a physical catheter introducer device. It describes its components, intended use, and the non-clinical (laboratory) testing performed to demonstrate its substantial equivalence to a previously cleared device. The detailed questions you've posed are designed for evaluating AI/ML-enabled medical devices, which have very different performance validation requirements than the SonoStik Guide Wire Introducer.