Search Results
Found 1 results
510(k) Data Aggregation
(114 days)
Scopis Extended Instrument Set EM
The Scopis Extended Instrument Set EM is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to the following ENT procedures:
- Transsphenoidal access procedures;
- Intranasal procedures:
- . Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies:
- . ENT related anterior skull base procedures.
The Scopis Extended Instrument Set EM is a set of accessories for the Scopis Hybrid Navigation Unit EM and is intended to localize the patient, track the position of the patient, display the position of navigated instruments on a model of the patient's anatomy based on preoperative images (CT or MRI) using electromagnetic tracking technology. The position of the instruments and the patient are localized within an electromagnetic field produced by a field generator. The navigation of instruments relative to the patient's anatomy is established via registration of the patient's anatomy to the image set via fiducial markers, anatomical landmarks, or surface matching. The position of navigated instruments is then displayed on the model from the image set.
The Scopis Extended Instrument Set EM consists of:
-
- Navigation Software (NOVA AR)
-
- Navigated instruments with integrated localizers
-
- Patient tracker with integrated localizer
-
- Non-tracked accessories for mechanical fixation
The Scopis Extended Instrument Set EM is a set of accessories for the Scopis Hybrid Navigation Unit EM, intended to localize the patient, track the position of the patient, and display the position of navigated instruments on a model of the patient's anatomy based on preoperative CT or MRI images using electromagnetic tracking technology.
Here's an analysis of the acceptance criteria and the study proving the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criteria (from predicate devices) | Reported Device Performance (Scopis Extended Instrument Set EM) |
---|---|
Fiagon predicate K133573: Navigation Accuracy of 0.9 mm ± 0.34 mm (for Precision Pointer equivalent) | Scopis Hybrid Navigation System EM (K161491) Navigation Accuracy (Precision Pointer EM): 0.49 mm ± 0.27 mm |
Fiagon predicate K141456: Bench Accuracy of 0.7 mm to 1.2 mm ± 0.29 mm to 0.42 mm (for Registration pointer equivalent) | Scopis Extended Instrument Set EM Bench Accuracy (Registration pointer EM-D): 0.64 mm ± 0.27 mm |
2. Sample Size Used for the Test Set and Data Provenance:
The document only states "bench testing of the navigation accuracy was performed". It does not specify the sample size (number of measurements or trials) for the test set or the data provenance (e.g., country of origin, retrospective or prospective).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:
Not applicable. The ground truth for the test set appears to be based on physical measurements of navigation accuracy on a test bench, not on expert consensus of medical images.
4. Adjudication Method for the Test Set:
Not applicable. The "study" appears to be a bench test involving physical accuracy measurements, not a human-adjudicated test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document describes bench testing for navigation accuracy, not a study involving human readers or AI assistance.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:
Yes, the described performance testing is a standalone assessment of the device's accuracy. The "bench testing of the navigation accuracy" assesses the algorithm and hardware's ability to precisely track and display instrument positions, independent of a human operator's performance in a surgical setting.
7. Type of Ground Truth Used:
The ground truth used for the performance testing was physical measurements of navigation accuracy. This likely involves comparing the device's reported position of an instrument to its actual physical position, as measured by a highly accurate reference system on a test bench.
8. Sample Size for the Training Set:
The document does not provide any information regarding the sample size for a training set. This is a medical device clearance, and the focus is on the safety and effectiveness of the device as a whole, not specifically on the details of an AI model's training if one were present as a distinct component. The "Navigation Software (NOVA AR)" is listed, but its development methodology (e.g., if it uses machine learning and thus a training set) is not detailed.
9. How the Ground Truth for the Training Set Was Established:
The document does not provide any information on how ground truth for a training set was established. As mentioned in point 8, details about the software's development, especially if it involved machine learning requiring a training set, are not present in this document.
Ask a specific question about this device
Page 1 of 1