Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K150758
    Date Cleared
    2015-06-18

    (87 days)

    Product Code
    Regulation Number
    880.5860
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Safety Disposable Insulin Syringe

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Safelock Disposable Insulin Syringe is intended to inject U-100 insulin into the human body and aid in the prevention of accidental needle stick injuries.

    Device Description

    Safelock Disposable Insulin Syringes are provided sterile, single use, which consist of five pieces components: (1) a syringe barrel calibrated in units of insulin (U-100) with a permanently attached single lumen needle on it, (2) a needle cap, (3) plunger and (4) protective end cap and (5) an additional safety mechanism installed at the needle end of the syringe. The devices are available in different combination of syringe volumes and needle sizes.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and the study that proves the device meets them, structured as requested:

    Device: Safelock Disposable Insulin Syringe

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided summary details various non-clinical tests performed according to recognized ISO and ASTM standards. The reported performance is generally stated as "met all design specifications" or "met the pre-established criteria." Specific quantitative acceptance criteria or performance metrics are largely absent from this summary but are implied by adherence to the referenced standards.

    Acceptance Criteria CategoryStandard/GuidanceReported Device Performance
    Physical, Mechanical, and Chemical (Syringe)ISO 8537:2007 (Clauses 5, 6.1, 6.2, 7, 8, 9.1, 9.2, 10.1, 10.2, 11.1, 11.2, 13.2, 14.1, 14.2, 14.3)Met all design specifications (implied by adherence to ISO 8537:2007 requirements)
    Freedom from extraneous matterClause 5 of ISO 8537:2007Met
    Limits for acidity and alkalinityClause 6.1 of ISO 8537:2007Met
    Limits for extractable metalsClause 6.2 of ISO 8537:2007Met
    Lubrication of syringes and needlesClause 7 of ISO 8537:2007Met
    Range of sizesClause 8 of ISO 8537:2007Met
    ScaleClause 9.1 of ISO 8537:2007Met
    Numbering of scaleClause 9.2 of ISO 8537:2007Met
    DimensionsClause 10.1 of ISO 8537:2007Met
    Finger gripsClause 10.2 of ISO 8537:2007Met
    Piston/plunger assemblyClause 11.1 of ISO 8537:2007Met
    Fit of piston in barrelClause 11.2 of ISO 8537:2007Met
    Needle tubing for syringesClause 13.2 of ISO 8537:2007Met
    Dead spaceClause 14.1 of ISO 8537:2007Met
    Freedom from leakage at needleClause 14.2 of ISO 8537:2007Met
    Liquid and air leakage past pistonClause 14.3 of ISO 8537:2007Met
    Physical, Mechanical, and Chemical (Needle)ISO 9626:1991/AMD-1:2001 (Clauses 3-11)Met all design specifications (implied by adherence to ISO 9626:1991 requirements)
    MaterialsClause 3 of ISO 9626:1991/AMD-1:2001Met
    Surface finishClause 4 of ISO 9626:1991/AMD-1:2001Met
    CleanlinessClause 5 of ISO 9626:1991/AMD-1:2001Met
    Limits for acidity and alkalinityClause 6 of ISO 9626:1991/AMD-1:2001Met
    Size designationClause 7 of ISO 9626:1991/AMD-1:2001Met
    DimensionsClause 8 of ISO 9626:1991/AMD-1:2001Met
    StiffnessClause 9 of ISO 9626:1991/AMD-1:2001Met
    Resistance to breakageClause 10 of ISO 9626:1991/AMD-1:2001Met
    Resistance to corrosionClause 11 of ISO 9626:1991/AMD-1:2001Met
    Sterile Barrier Packaging TestingASTM F88/F88M-09, ASTM F1140/F1140M-13Met all design specifications
    Seal strengthASTM F88/F88M-09Met
    Internal pressureASTM F1140/F1140M-13Met
    Sterilization and Shelf Life TestingISO 10993-7:2008, USP37-NF 32,Verified claimed shelf life based on tests on accelerated aged samples
    EO residueISO 10993-7:2008Met
    ECH residueISO 10993-7:2008Met
    Bacteria Endotoxin LimitUSP37-NF 32,Met
    Shelf Life EvaluationPhysical, Mechanical Chemical, Package and Sterility Test on accelerated aged samplesVerified claimed shelf life of the device
    BiocompatibilityISO 10993 series standardsNot required; materials identical to previously cleared K113091
    Safety Feature TestingFDA Guidance for Industry and FDA Staff: Medical Device with Sharps Injury Prevention Feature, issued on August 9, 2005Met pre-established criteria; force to activate and detach safety mechanism passed

    2. Sample Size Used for the Test Set and the Data Provenance

    The provided text does not specify the sample size for any of the non-clinical tests (e.g., number of syringes tested for each ISO standard criterion, number of packages for ASTM, number of devices for safety feature testing).

    The provenance of the data is that it was generated through non-clinical bench testing performed by the manufacturer, Jiangyin Caina Technology Co., Ltd. (located in China). The summary reports on tests performed on the "proposed device." It does not indicate retrospective or prospective in the typical sense of clinical studies, as these are lab-based evaluations.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of those Experts

    This section is not applicable as the evaluation relies on direct measurement and adherence to objective standards (ISO, ASTM, USP) rather than expert interpretation of a ground truth in a clinical or image-based context. The "ground truth" is defined by the parameters and limits set forth in the referenced standards.

    4. Adjudication Method for the Test Set

    This section is not applicable. The tests are objective measurements against established standard criteria, not subjective assessments requiring adjudication among multiple reviewers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable as the device is a medical syringe with a safety feature, not an AI-powered diagnostic or assistive technology for human readers. There is no mention of AI in this submission.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable for the same reason as point 5. The device does not involve an algorithm.

    7. The Type of Ground Truth Used

    The ground truth for most tests is based on objective, internationally recognized performance standards (ISO, ASTM, USP) which define acceptable limits for physical, mechanical, chemical, and sterility properties. For the simulated clinical study and safety feature testing, the "ground truth" was derived from "pre-established criteria" and FDA guidance for sharps injury prevention devices, which likely includes defined force thresholds or functional performance checks.

    8. The Sample Size for the Training Set

    This is not applicable. The device is a physical medical device and does not involve AI or machine learning models that require a training set.

    9. How the Ground Truth for the Training Set was Established

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1