Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K221604
    Manufacturer
    Date Cleared
    2022-08-22

    (81 days)

    Product Code
    Regulation Number
    878.4850
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    SafetiHeel, MediHeel, Novaplus

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is used to obtain capillary blood samples from the heel of newborn and neonatal infants in the healthcare setting by qualified healthcare providers.

    Device Description

    The SafetiHeel Safety Lancet is single use only blood lancet used to puncture the skin to obtain a drop of blood for diagnostic purposes. It is a disposable blood lancet intended for a single use and the device is comprised of a blade encased in a housing made with plastic material commonly used for medical devices. The mechanism of the device includes an integral sharps injury prevention feature which allows the device to be used only once and then renders it inoperable and incapable of further use. The SafetiHeel Safety Lancet is a sterile device and the sterilization is done using the Gamma Irradiation means. The SafetiHeel device is marketed under model numbers SH-065, SH-100 and SH-150. The device is also marketed under trade names such as MediHeel and Novaplus in the USA market. The models for the MediHeel include model number 1003, 1004, 1005 and 1006. For the Novaplus device, the model numbers are 10041 and 10051.

    AI/ML Overview

    The document provided describes the SafetiHeel, MediHeel, and Novaplus blood lancets, which are intended for obtaining capillary blood samples from the heel of newborn and neonatal infants. The submission aims to demonstrate substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study information provided:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are essentially the "Predicate Device (specification)" column, and the reported device performance is indicated in the "SafetiHeel performance," "MediHeel performance," and "Novaplus performance" columns.

    SafetiHeel Device Performance

    TestModelAcceptance Criteria (Predicate Device Specification)Reported Device Performance (SafetiHeel)Meets Criteria?
    Cutting Depth Test(mm)(mm)
    SH-0650.65 to 0.850.63 to 0.80Yes (close)
    SH-0851.00 to 1.201.04 to 1.19Yes
    SH-1001.40 to 1.601.42 to 1.59Yes
    SH-1501.60 to 1.801.66 to 1.79Yes
    Trigger Force Test(grams)(grams)
    SH-065450 to 750475 to 575Yes
    SH-085450 to 750494 to 578Yes
    SH-100450 to 750496 to 565Yes
    SH-150450 to 750494 to 586Yes
    Blade Retraction TestAll ModelsBlade must retract into housing after incision.All tested samples retracted the blade.Yes

    MediHeel Device Performance

    TestPredicate Device ModelAcceptance Criteria (Predicate Device Specification)Reported Device Performance (MediHeel)MediHeel ModelMeets Criteria?
    Cutting Depth Test(mm)(mm)
    SH-0650.65 to 0.850.67 to 0.811003Yes
    SH-0851.00 to 1.201.05 to 1.161004Yes
    SH-1001.40 to 1.601.45 to 1.581005Yes
    SH-1501.60 to 1.801.65 to 1.751006Yes
    Trigger Force Test(grams)(grams)
    SH-065450 to 750475 to 5881003Yes
    SH-085450 to 750496 to 5911004Yes
    SH-100450 to 750488 to 5851005Yes
    SH-150450 to 750492 to 5961006Yes
    Blade Retraction TestAll ModelsBlade must retract into housing after incision.All tested samples retracted the blade.All ModelsYes

    Novaplus Device Performance

    TestPredicate Device ModelAcceptance Criteria (Predicate Device Specification)Reported Device Performance (Novaplus)Novaplus ModelMeets Criteria?
    Cutting Depth Test(mm)(mm)
    SH-0851.00 to 1.201.06 to 1.1510041Yes
    SH-1001.40 to 1.601.44 to 1.5710051Yes
    Trigger Force Test(grams)(grams)
    SH-085450 to 750492 to 59710041Yes
    SH-100450 to 750486 to 58910051Yes
    Blade Retraction TestAll ModelsBlade must retract into housing after incision.All tested samples retracted the blade.All ModelsYes

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample size for the test set beyond stating that "The samples of all the models tested show..." for the Blade Retraction Test, and presenting ranges for Cutting Depth and Trigger Force. This implies that multiple units were tested for each model.

    Data Provenance: The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. Given the nature of a 510(k) submission and the tests described (physical performance tests on the devices themselves), it is highly likely these were prospective laboratory or engineering tests conducted on manufactured devices, rather than clinical studies involving human patients or existing datasets.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to the type of performance testing described. The "ground truth" for these physical performance tests (cutting depth, trigger force, blade retraction) is established by direct measurement against known engineering specifications, not by expert interpretation.


    4. Adjudication method for the test set

    This information is not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies involving human readers/interpreters to resolve discrepancies when establishing ground truth for diagnostic decisions. The tests here are physical measurements of device specifications.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is a physical blood lancet, not an AI-powered diagnostic or assistance tool that would involve human readers or AI assistance.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a physical blood lancet, not an algorithm.


    7. The type of ground truth used

    The ground truth for these performance tests is based on engineering specifications and direct physical measurements. Specifically:

    • Cutting Depth: Measured depth of blade protrusion.
    • Trigger Force: Measured force required to activate the blade.
    • Blade Retraction: Visual and/or mechanical confirmation of the blade returning into the housing.
    • Sterility: Adherence to established sterility assurance levels (SAL) based on ISO standards.

    8. The sample size for the training set

    This information is not applicable. The device is a physical product, not an algorithm that requires a training set.


    9. How the ground truth for the training set was established

    This information is not applicable. As stated above, there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1