Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K111189
    Date Cleared
    2012-12-20

    (602 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    SYNTEC DENTAL IMPLANT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Syntec Dental Implant is intended to be surgically placed in the bone of the upper or lower jaw arches. To provide support for prosthetic devices, such as to restore the patient's chewing function, and artificial teeth. It is intended for delayed loading.

    Device Description

    The Syntec Dental Implant provides Titanium Alloy (commercially followed by ASTM F136) intended to be surgically placed in the bone of the upper or lower jaw arches. To provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. It consists of implant, cover screw, healing abutment screw and abutments. The Syntec Dental Implant is for one and two stage surgical procedures. Its materials, dimensions, and intended use are similar to devices currently marked worldwide. The range of implant in diameter is from 3.4mm to 4.8mm and in length from 8.0mm to 14.0mm. The range of abutment in diameter is from 4.5mm to 6.5mm, and in length from 5.5mm to 12.0mm.

    AI/ML Overview

    The Syntec Dental Implant K111189 510(k) summary does not contain information on an AI/ML device.

    Here's a breakdown of the provided information, specifically addressing the criteria you've outlined, even though it's for a traditional medical device (dental implant):

    1. A table of acceptance criteria and the reported device performance

    Since this is a submission for a traditional dental implant and not an AI/ML device, the acceptance criteria are based on mechanical properties, biocompatibility, and sterilization, rather than performance metrics like sensitivity, specificity, or AUC. The "acceptance criteria" are implied by demonstrating substantial equivalence to predicate devices through various bench tests conforming to international standards.

    Testing ItemStandard and Regulations AppliedReported Device Performance (Implied Acceptance)
    Static StrengthISO 14801"Substantially equivalent to predicate devices" (result of biomechanical testing)
    Fatigue TestingISO 14801"Substantially equivalent to predicate devices" (result of biomechanical testing)
    Biocompatibility (Raw Material)ISO-10993"Certificate of Raw material" submitted, implying compliance.
    Sterilization (Gamma Irradiation)ISO 11137"Sterilization Validation of Gamma Irradiation" submitted, implying compliance.
    Sterilization (Moist Heat)ISO 17665-1, ISO 17665-2, ISO 11737-1, ISO 11737-2"Sterilization Validation of Moist Heat" and "Sterilization Process Validation" submitted, implying compliance.
    Surface Analysis (Morphology)SEMData submitted, implying acceptable morphology.
    Surface Analysis (Coating Thickness)EN ISO 4288Data submitted, implying acceptable coating thickness.
    Surface Analysis (Ca/P ratio)EN ISO 11885-E22Data submitted, implying acceptable Ca/P ratio.
    Surface Analysis (Residue)ASTM F1854/ DIN EN ISO 4288Data submitted, implying acceptable residue levels.
    CytotoxicityDIN EN ISO 10993-5Data submitted, implying acceptable cytotoxicity.
    SolubilityFDA guidance (2/21/97)Data submitted, implying acceptable solubility.
    Analysis of Raw MaterialASTM F1185 and ASTM F1609Data submitted, implying compliance with material specifications.
    Adhesive Tensile StrengthASTM F-1147Data submitted, implying acceptable adhesive tensile strength.
    Shear StrengthASTM F-1044Data submitted, implying acceptable shear strength.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the 510(k) summary. For bench testing of physical devices, "sample size" typically refers to the number of units tested per experiment, which is not detailed here. The studies are bench tests, not clinical studies with human participants. Data provenance is implied to be from the internal testing of Syntec Scientific Corporation in Taiwan.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable to a traditional dental implant submission. "Ground truth" in this context would be defined by the physical characteristics and performance measured against the standards, not by expert interpretation of data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable for a traditional dental implant submission based on bench testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device, and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is established by the international standards and regulations (e.g., ISO 14801, ISO-10993, ISO 11137, etc.) referenced in the bench testing section. The device's physical and mechanical properties are measured and compared directly to the requirements outlined in these standards.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1