K Number
K111189
Date Cleared
2012-12-20

(602 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Syntec Dental Implant is intended to be surgically placed in the bone of the upper or lower jaw arches. To provide support for prosthetic devices, such as to restore the patient's chewing function, and artificial teeth. It is intended for delayed loading.

Device Description

The Syntec Dental Implant provides Titanium Alloy (commercially followed by ASTM F136) intended to be surgically placed in the bone of the upper or lower jaw arches. To provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. It consists of implant, cover screw, healing abutment screw and abutments. The Syntec Dental Implant is for one and two stage surgical procedures. Its materials, dimensions, and intended use are similar to devices currently marked worldwide. The range of implant in diameter is from 3.4mm to 4.8mm and in length from 8.0mm to 14.0mm. The range of abutment in diameter is from 4.5mm to 6.5mm, and in length from 5.5mm to 12.0mm.

AI/ML Overview

The Syntec Dental Implant K111189 510(k) summary does not contain information on an AI/ML device.

Here's a breakdown of the provided information, specifically addressing the criteria you've outlined, even though it's for a traditional medical device (dental implant):

1. A table of acceptance criteria and the reported device performance

Since this is a submission for a traditional dental implant and not an AI/ML device, the acceptance criteria are based on mechanical properties, biocompatibility, and sterilization, rather than performance metrics like sensitivity, specificity, or AUC. The "acceptance criteria" are implied by demonstrating substantial equivalence to predicate devices through various bench tests conforming to international standards.

Testing ItemStandard and Regulations AppliedReported Device Performance (Implied Acceptance)
Static StrengthISO 14801"Substantially equivalent to predicate devices" (result of biomechanical testing)
Fatigue TestingISO 14801"Substantially equivalent to predicate devices" (result of biomechanical testing)
Biocompatibility (Raw Material)ISO-10993"Certificate of Raw material" submitted, implying compliance.
Sterilization (Gamma Irradiation)ISO 11137"Sterilization Validation of Gamma Irradiation" submitted, implying compliance.
Sterilization (Moist Heat)ISO 17665-1, ISO 17665-2, ISO 11737-1, ISO 11737-2"Sterilization Validation of Moist Heat" and "Sterilization Process Validation" submitted, implying compliance.
Surface Analysis (Morphology)SEMData submitted, implying acceptable morphology.
Surface Analysis (Coating Thickness)EN ISO 4288Data submitted, implying acceptable coating thickness.
Surface Analysis (Ca/P ratio)EN ISO 11885-E22Data submitted, implying acceptable Ca/P ratio.
Surface Analysis (Residue)ASTM F1854/ DIN EN ISO 4288Data submitted, implying acceptable residue levels.
CytotoxicityDIN EN ISO 10993-5Data submitted, implying acceptable cytotoxicity.
SolubilityFDA guidance (2/21/97)Data submitted, implying acceptable solubility.
Analysis of Raw MaterialASTM F1185 and ASTM F1609Data submitted, implying compliance with material specifications.
Adhesive Tensile StrengthASTM F-1147Data submitted, implying acceptable adhesive tensile strength.
Shear StrengthASTM F-1044Data submitted, implying acceptable shear strength.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the 510(k) summary. For bench testing of physical devices, "sample size" typically refers to the number of units tested per experiment, which is not detailed here. The studies are bench tests, not clinical studies with human participants. Data provenance is implied to be from the internal testing of Syntec Scientific Corporation in Taiwan.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable to a traditional dental implant submission. "Ground truth" in this context would be defined by the physical characteristics and performance measured against the standards, not by expert interpretation of data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable for a traditional dental implant submission based on bench testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML device, and no MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is established by the international standards and regulations (e.g., ISO 14801, ISO-10993, ISO 11137, etc.) referenced in the bench testing section. The device's physical and mechanical properties are measured and compared directly to the requirements outlined in these standards.

8. The sample size for the training set

Not applicable. This is not an AI/ML device.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/ML device.

{0}------------------------------------------------

K111189

DEC 2 0 2012

510(k) SUMMARY

The summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

Submitter's name :Syntec Scientific Corporation
Address :2, Kung San Rd,Chuan Shing Industrial Zone,Shen Kang, Chang Hua,Taiwan
Tel :886-4-7987099
Fax:886-4-7987077
Date :December 17, 2012
Contact person :Chia-Ching Lee
Name of the device :Syntec Dental Implant
Trade or proprietary name :Syntec Dental Implant
Common name:Dental implant, Dental implant abutments
Classification name:Endosseous dental implant, Endosseous dental implant abutments
Produce Code :DZE, NHA
Regulation Number :872.3640, 872.3630
Class :II
Predicate device:Dentium Company Limited Implantium, K041368Implantium Prosthetics, K052957

Description of the Device:

The Syntec Dental Implant provides Titanium Alloy (commercially followed by ASTM F136) intended to be surgically placed in the bone of the upper or lower jaw arches. To provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. It consists of implant, cover screw, healing abutment screw and abutments. The Syntec Dental Implant is for one and two stage surgical procedures. Its materials, dimensions, and intended use are similar to devices currently marked worldwide. The range of implant in diameter is from 3.4mm to 4.8mm and in length from 8.0mm to 14.0mm. The range of abutment in diameter is from 4.5mm to 6.5mm, and in length from 5.5mm to 12.0mm.

{1}------------------------------------------------

Indications for Use:

The Syntec Dental Implant is intended to be surgically placed in the bone of the upper or lower jaw arches. To provide support for prosthetic devices, such as to restore the patient's chewing function, and artificial teeth. It is intended for delayed loading.

Predicate DeviceApplicant
SponsorDentium Co., Ltd.Syntec Scientific Corporation
Device nameDentium Company Limited ImplantiumImplantium ProstheticsSyntec Dental Implant
510(k) numberK041368K052957K111189
Similarities:
Regulation number/name/ class/ productcodeClass II §872.3640, 872.3630/Endosseous dental implant,Endosseous dental implant abutment/DZE, NHAClass II §872.3640, 872.3630/Endosseous dental implant,Endosseous dental implant abutment/DZE, NHA
Intended for UseThe Dentium Co., Ltd Implantium is intended to besurgically placed in the bone of the upper or lower jawarches to provide support for prosthetic devices, such asartificial teeth, and to restore the patient's chewingfunction. Implantium prosthetics is intended for use as anaid in prosthetic rehabilitation.The Syntec Dental Implant is intended to be surgicallyplaced in the bone of the upper or lower jaw arches. Toprovide support for prosthetic devices, such as to restorethe patient's chewing function, and artificial teeth. It isintended for delayed loading.
Diameters of ImplantFrom 3.4mm to 4.8mmFrom 3.4mm to 4.8mm
Lengths of ImplantFrom 8.0mm to 14.0mmFrom 8.0mm to 14.0mm
Diameters of AbutmentFrom 4.0mm to 6.5mmFrom 4.5mm to 6.5mm
Lengths of AbutmentFrom 5.5mm to 12.0mmFrom 5.5mm to 12.0mm
Single Use?YesYes
Mechanical PropertiesGoodGood
Differences:
MaterialsPure titanium and titanium alloyTitanium alloy
SterilitySterileImplant and cover screw are sterile. Other components arenon-sterile.

Device Comparison Table

{2}------------------------------------------------

Bench Testing:

Overview all bench Testing:

Testing ItemStandard and regulations appliedSubmitted on
Biomechanical testing
Static strengthISO 14801November 25, 2011
Fatigue testingISO 14801November 25, 2011
Biocompatibility
Certificate of Raw materialISO-10993April 26, 2011
Sterilization :
Sterilization Validation of Gamma IrradiationISO 11137November 25, 2011
Sterilization Validation of Moist HeatISO 17665-1, ISO 17665-2, ISO 11737-1November 09, 2012
Sterilization Process Validationand ISO 11737-2
surface analysis
MorphologySEMAugust 7, 2012
Coating thicknessEN ISO 4288August 7, 2012
Ca/P ratioEN ISO 11885-E22August 7, 2012
Residue analysisASTM F1854/ DIN EN ISO 4288August 7, 2012
CytotoxicityDIN EN ISO 10993-5August 7, 2012
SolubilityFDA guidance (2/21/97)August 7, 2012
Analysis of raw materialASTM F1185 and ASTM F1609August 7, 2012
Adhesive Tensile StrengthASTM F-1147November 09, 2012
Shear StrengthASTM F-1044November 09, 2012

Brief discussion of Biomechanical Studies: The experiments were according ISO 14801 . (Dentistry - Implant - Dynamic Fatigue Test for Endosseous Dental Implants). In these experiments, the static strength and fatigue life were evaluated. The result of Dental Implant is substantially equivalent to predicate devices.

Conclusion:

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification Syntec concludes that Dental is substantially equivalent to predicate devices as described herein.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of an eagle.

December 20, 2012

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Ms. Chia-Ching Lee Regulatory Affairs Specialist Syntec Scientific Corporation 2, Kung San Road Chuan Shing Industrial Zone Shen Kang, Chang Hua Taiwan

Re: K11189

Trade/Device Name: Syntec Dental Implant Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: December 17, 2012 Received: December 19, 2012

Dear Ms. Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Lee

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Anthony De ma

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATION FOR USE

Indications for Use

510(k) Number (if known): k | | | 89 Device Name: Syntec Dental Implant Indications for Use:

The Syntec Dental Implant is intended to be surgically placed in the bone of the upper or lower jaw arches. To provide support for prosthetic devices, such as to restore the patient's chewing function, and artificial teeth.

AND/OR Over-The-Counter Use x Prescription Use _ (21 CFR 801 Subpart C) ( Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Runner DDS, MA2012.12.2013:32:08 -05'00'
------------------------------------------------------

(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices

510(k) Number:K111189
-------------------------

. 02

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.