Search Results
Found 2 results
510(k) Data Aggregation
(88 days)
SURFLO WINGED INFUSION SET WITH FILTER & NEEDLE PROTECTION (SURSHIELD)
The Surflo Winged Infusion Set with Filter and Needle Protection (Surshield) is intended to access the peripheral vascular system, for single-dose or short-term intravenous administration of fluids using a syringe or other compatible/appropriate devices. Additionally, after withdrawal of the needle from the patient's vein, the shield cover can be manually activated to cover the needle to minimize risk of accidental needle stick.
The 23G and the 25G Terumo Surflo Winged Infusion Sets with Filter and Needle Protection (Surshield), are sterile, single use devices consisting of a needle attached to a winged hub, microbore tubing, adapter with integrated 20 um filter and adapter cap, and a hinged shield cover that attaches to the wing just below the needle-to-wing junction. The shield cover can be turned 180 degrees on the hinge. As the needle is removed from the patient's vessel, the user's finger actively pushes the shield cover until it latches onto the needle using a one- or two- handed technique. An audible click is noted upon activation. The shield cover is designed to allow the user's finger to remain behind the needle point so that the risk of needle stick injury is minimized. The shield cover is transparent for easy confirmation of the needle held in it. The device possesses a 350 mm length microbore tubing.
The provided text describes the acceptance criteria and a summary of verification activities for the SURFLO® Winged Infusion Set with Filter and Needle Protection (Surshield®). This device is a medical product, and the study described is a series of engineering and performance tests, not a clinical study involving human patients or algorithmic performance. Therefore, many of the requested categories (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, training set information) are not applicable to this type of device submission.
Here's a breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
TEST | ACCEPTANCE CRITERIA | Reported Device Performance (Implied) |
---|---|---|
1. Visual inspection | When examined by normal vision or under 2.5 fold magnification, all components of the set are smooth, clean and free of flash, surface irregularities, detachable particles, and moulding or processing defects | Meets criteria |
2. Effective tubing length | The effective tubing length is 350 ± 20mm | Meets criteria |
3. Dead space volume | Dead space volume for sets with standard tubing: ≤ 0.40 ml; Dead space volume for sets with microbore tubing: ≤ 0.20 ml | Meets criteria |
4. Air leakage (= integrity of set) | No leakage | Meets criteria |
5. Air leakage luer adaptor | No leakage at the luer connection (according to ISO 594-1 and ISO 594-2) | Meets criteria |
6. Liquid leakage luer adaptor | No leakage at the luer connection (according to ISO 594-1 and ISO 594-2) | Meets criteria |
7. Conical fitting | The conical fitting shall be within level 1-3 (according to ISO 594-1 and ISO 594-2) | Meets criteria |
8. Separation force of Luer slip tip connection | The fitting remains attached (according to ISO 594-1) | Meets criteria |
9. Separation force of Luer lock connection | The fitting remains attached (according to ISO 594-2) | Meets criteria |
10. Stress cracking on Luer conical fitting | No stress cracking on the conical fitting (according to ISO 594-1 and ISO-594-2) | Meets criteria |
11. Unscrewing torque of Luer lock fittings | The fitting remains attached when the fitting between the hub and a reference fitting is tested according to ISO 594-2 | Meets criteria |
12. Ease of assembly | A satisfactory fit is achieved when tested according to ISO 594-2 | Meets criteria |
13. Overriding resistance of Luer lock systems | When the fitting between the hub and a reference fitting is tested, the reference fitting shall not override the threads or lugs of the fitting under test (according to ISO 594-2) | Meets criteria |
14. Torque resistance cap - adapter | The torque force required to unscrew the cap from the adapter does not exceed 9 N.cm | Meets criteria |
15. Bonding strength tube - adapter | The force required for separating the tube from the adapter is 15 N minimum. | Meets criteria |
16. Bonding strength tube - wing hub | The force required for separating the tube from the wing hub is 15 N minimum. | Meets criteria |
17. Air flow choke test | Air bubbles escape out of the set when immersed under water and connected to an airline of 20 kPa (0.2 bar) effective pressure, to create an airflow through the set. | Meets criteria |
18. Flow rate | Flow rate for sets with standard tubing: 23G: ≥ 2.9 ml/min & 25G: ≥ 1.6 ml/min; Flow rate for sets with microbore tubing: 23G: ≥ 1.7 ml/min & 25G: ≥ 1.5 ml/min | Meets criteria |
19. Bonding strength Surshield protector | The bonding strength between the Surshield protector and the SV-set is minimum 4 N | Meets criteria |
20. Misalignment of Surshield Protector | The angle measured axially from the cannula between the horizontally positioned wings and the vertically positioned Surshield protector does not exceed 15°. | Meets criteria |
21. Peel behaviour of blister | The blister is peeled by hand with minimal fibres and no paper splitting | Meets criteria |
22. Peel strength | The peel strength of the pack seal is minimum 0.15.kN/m | Meets criteria |
23. Package integrity | The integrity of the unit pack is minimum 4 kPa | Meets criteria |
24. Kink stability of tubing | Free of deformation and kinking | Meets criteria |
25. Filter performance testing | Same or better than predicate | Meets criteria |
The document states: "The Surflo Winged Infusion Sets with Filter and Needle Protection (Surshield), meet all acceptance criteria as indicated in table above."
2. Sample size used for the test set and the data provenance
The document does not specify the exact sample sizes (number of units tested) for each individual test. It only states that "All necessary verification and validation tests have been performed." Data provenance is not explicitly mentioned as "country of origin," but the submission is from "TERUMO EUROPE N.V." in Leuven, BELGIUM, so the testing likely occurred in Europe. The tests are prospective in nature, as they are performed on the newly designed devices to verify their performance against established criteria.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. These are engineering and performance tests of a physical medical device, not a diagnostic algorithm requiring expert interpretation of medical images or data. The "ground truth" is defined by the technical specifications and international standards (e.g., ISO 594-1, ISO 594-2, EN ISO 11135-1, EN 556-1, EN ISO 10993-7).
4. Adjudication method for the test set
Not applicable. There is no adjudication method described as these are quantitative and qualitative engineering tests against predefined specifications.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-assisted diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an algorithm.
7. The type of ground truth used
The ground truth for these tests is established by international standards and technical specifications for sterile medical devices for intravascular administration. Examples include:
- ISO 594-1: Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment — Part 1: General requirements
- ISO 594-2: Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment — Part 2: Lock fittings
- EN ISO 11135-1: Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- EN 556-1: Sterilization of Medical Devices - Requirements for medical devices to be designated "STERILE" - Part - I: Requirements for terminally sterilized medical devices
- EN ISO 10993-7: Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
- FDA General Program Memorandum #G95-1: Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and testing"
8. The sample size for the training set
Not applicable. There is no "training set" as this is not a machine learning or AI device.
9. How the ground truth for the training set was established
Not applicable.
Ask a specific question about this device
(88 days)
SURFLO WINGED INFUSION SET WITH FILTER AND NEEDLE PROTECTION (SURSHIELD)
The Surflo Winged Infusion Set with Filter and Needle Protection (Surshield) is intended to access the peripheral vascular system for intravenous administration of fluids using a syringe, or other compatible/appropriate devices and may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused and duration of therapy. Additionally, after withdrawal of the needle from the patient's vein, the shield cover can be manually activated to cover the needle to minimize risk of accidental needle stick.
The Terumo Surflo Winged Infusion Set with Filter and Needle Protection (Surshield) is a sterile, single use device consisting of a needle attached to a winged hub, microbore tubing, adapter with integrated 20 um filter and adapter cap, and a hinged shield cover that attaches to the wing just below the needle-to-wing junction. The shield cover can be turned 180 degrees on the hinge. As the needle is removed from the patient's vessel, the user's finger actively pushes the shield cover until it latches onto needle using a one- or two- handed technique. An audible click is noted upon activation. The shield cover is designed to allow the user's finger to remain behind the needle point so that the risk of needle stick injury is minimized. The shield cover is transparent for easy confirmation of the needle held in it. The Surflo Winged Infusion Sets with Filter and Needle Protection (Surshield) are available with a cannula with a needle size of 23 gauge (0.6 mm) or 25 gauge (0.5 mm) and a length of 19 mm (3/4"). The device possesses 350 mm length tubing.
Here's an analysis of the provided text regarding the "SURFLO® Winged Infusion Set with Filter and Needle Protection (Surshield™)":
It's important to note that the provided 510(k) summary does not contain details about specific acceptance criteria or a dedicated study proving performance against those criteria in the way one might expect for a software or AI/ML device. This document is for a physical medical device (an infusion set), and the "study" mentioned is primarily focused on demonstrating substantial equivalence to previously cleared devices through material compatibility, sterilization validation, and shelf-life testing, rather than an AI-driven performance study.
Therefore, many of the requested fields regarding AI/ML-specific performance metrics (like sample size for test/training sets, number of experts for ground truth, MRMC studies, standalone algorithm performance) are not applicable to this document.
However, I can extract the information that is present and explain why other requested details are absent.
Analysis of SURFLO® Winged Infusion Set with Filter and Needle Protection (Surshield™) 510(k) Summary
1. Table of Acceptance Criteria and Reported Device Performance
As mentioned, this document does not present explicit performance acceptance criteria in the format typically seen for novel diagnostic or AI devices (e.g., sensitivity, specificity thresholds). Instead, the "performance" demonstrated is through substantial equivalence and compliance with relevant standards.
Acceptance Criterion (Implied by Regulatory Compliance) | Reported Device Performance |
---|---|
Sterility Assurance Level (SAL) | Achieved 10⁻⁶ SAL per EN 550 and ISO 11135 |
Ethylene Oxide Residual Levels | In compliance with ISO 10993-7 |
Biocompatibility of Blood Contacting Materials | Materials demonstrated to be biocompatible per ISO 10993-1 |
Shelf-Life/Expiration Dating | Established at 5 years |
Needle Protection Feature Functionality | Substantially equivalent to predicate device (K052887) |
Overall Functionality (Infusion Set) | Substantially equivalent to predicate device (K033350) |
2. Sample Size for the Test Set and Data Provenance
- Not Applicable in the context of an AI/ML device's "test set."
- For the physical device, "testing" would involve laboratory evaluations for sterility, material properties, and shelf-life. The sample sizes for these types of tests are not specified in this summary but would typically follow established validation protocols for manufacturing and quality control.
- Data Provenance: The studies (sterilization, biocompatibility, shelf-life) are conducted by the manufacturer, Terumo Europe N.V. The country of origin of the data is implicitly Belgium (where Terumo Europe N.V. is located). These are more akin to prospective laboratory or manufacturing validation studies rather than retrospective clinical data analysis.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Not Applicable. This is not a device that relies on expert interpretation for its "ground truth" performance in a clinical context (like image analysis). Ground truth relates to objective measurements for sterility, chemical residuals, and material compatibility standards.
4. Adjudication Method for the Test Set
- Not Applicable. There is no "adjudication" in the sense of resolving discrepancies between expert opinions on a test set, as this is not an interpretive device.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No. An MRMC study is relevant for diagnostic devices where human readers interpret outputs (e.g., radiologists reading images with and without AI assistance). This device is an intravenous administration set, not a diagnostic tool requiring human interpretation for its primary function.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not Applicable. This is a physical medical device, not an algorithm.
7. The Type of Ground Truth Used
The ground truth used for relevant testing includes:
- Standardized Sterility Assurance Level (SAL): Defined by international standards (EN 550, ISO 11135) to be 10⁻⁶.
- Reference Limits for Chemical Residues: Defined by international standards (ISO 10993-7) for ethylene oxide residuals.
- Biocompatibility Standards: Defined by international standards (ISO 10993-1).
- Performance Specifications: The core "ground truth" for the device's functionality and safety features is its ability to meet the design specifications and be substantially equivalent to predicate devices that have already demonstrated safety and effectiveness for their intended use.
8. The Sample Size for the Training Set
- Not Applicable. This refers to AI/ML model training, which is not relevant for this device.
9. How the Ground Truth for the Training Set was Established
- Not Applicable. See point 8.
Summary of Device and Regulatory Context:
The 510(k) for the SURFLO® Winged Infusion Set with Filter and Needle Protection (Surshield™) is a submission for a Class II physical medical device. The primary goal of this type of submission is to demonstrate substantial equivalence to predicate devices already on the market. This is achieved by showing that the new device has the same intended use, similar technological characteristics, and that any differences do not raise new questions of safety or effectiveness.
The "studies" referenced are primarily:
- Comparison to Predicate Devices: Demonstrating similarity in design and function to previously cleared infusion sets.
- Compliance with International Standards: Validation of manufacturing processes (sterilization), material safety (biocompatibility, chemical residuals), and performance (shelf-life).
This is distinctly different from the type of information requested for a software as a medical device (SaMD) or AI/ML-powered diagnostic device, where performance metrics like sensitivity, specificity, and results from clinical studies (often involving human readers) are central to the submission.
Ask a specific question about this device
Page 1 of 1