K Number
K133894
Manufacturer
Date Cleared
2014-03-18

(88 days)

Product Code
Regulation Number
880.5440
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Surflo Winged Infusion Set with Filter and Needle Protection (Surshield) is intended to access the peripheral vascular system, for single-dose or short-term intravenous administration of fluids using a syringe or other compatible/appropriate devices. Additionally, after withdrawal of the needle from the patient's vein, the shield cover can be manually activated to cover the needle to minimize risk of accidental needle stick.

Device Description

The 23G and the 25G Terumo Surflo Winged Infusion Sets with Filter and Needle Protection (Surshield), are sterile, single use devices consisting of a needle attached to a winged hub, microbore tubing, adapter with integrated 20 um filter and adapter cap, and a hinged shield cover that attaches to the wing just below the needle-to-wing junction. The shield cover can be turned 180 degrees on the hinge. As the needle is removed from the patient's vessel, the user's finger actively pushes the shield cover until it latches onto the needle using a one- or two- handed technique. An audible click is noted upon activation. The shield cover is designed to allow the user's finger to remain behind the needle point so that the risk of needle stick injury is minimized. The shield cover is transparent for easy confirmation of the needle held in it. The device possesses a 350 mm length microbore tubing.

AI/ML Overview

The provided text describes the acceptance criteria and a summary of verification activities for the SURFLO® Winged Infusion Set with Filter and Needle Protection (Surshield®). This device is a medical product, and the study described is a series of engineering and performance tests, not a clinical study involving human patients or algorithmic performance. Therefore, many of the requested categories (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, training set information) are not applicable to this type of device submission.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

TESTACCEPTANCE CRITERIAReported Device Performance (Implied)
1. Visual inspectionWhen examined by normal vision or under 2.5 fold magnification, all components of the set are smooth, clean and free of flash, surface irregularities, detachable particles, and moulding or processing defectsMeets criteria
2. Effective tubing lengthThe effective tubing length is 350 ± 20mmMeets criteria
3. Dead space volumeDead space volume for sets with standard tubing: ≤ 0.40 ml; Dead space volume for sets with microbore tubing: ≤ 0.20 mlMeets criteria
4. Air leakage (= integrity of set)No leakageMeets criteria
5. Air leakage luer adaptorNo leakage at the luer connection (according to ISO 594-1 and ISO 594-2)Meets criteria
6. Liquid leakage luer adaptorNo leakage at the luer connection (according to ISO 594-1 and ISO 594-2)Meets criteria
7. Conical fittingThe conical fitting shall be within level 1-3 (according to ISO 594-1 and ISO 594-2)Meets criteria
8. Separation force of Luer slip tip connectionThe fitting remains attached (according to ISO 594-1)Meets criteria
9. Separation force of Luer lock connectionThe fitting remains attached (according to ISO 594-2)Meets criteria
10. Stress cracking on Luer conical fittingNo stress cracking on the conical fitting (according to ISO 594-1 and ISO-594-2)Meets criteria
11. Unscrewing torque of Luer lock fittingsThe fitting remains attached when the fitting between the hub and a reference fitting is tested according to ISO 594-2Meets criteria
12. Ease of assemblyA satisfactory fit is achieved when tested according to ISO 594-2Meets criteria
13. Overriding resistance of Luer lock systemsWhen the fitting between the hub and a reference fitting is tested, the reference fitting shall not override the threads or lugs of the fitting under test (according to ISO 594-2)Meets criteria
14. Torque resistance cap - adapterThe torque force required to unscrew the cap from the adapter does not exceed 9 N.cmMeets criteria
15. Bonding strength tube - adapterThe force required for separating the tube from the adapter is 15 N minimum.Meets criteria
16. Bonding strength tube - wing hubThe force required for separating the tube from the wing hub is 15 N minimum.Meets criteria
17. Air flow choke testAir bubbles escape out of the set when immersed under water and connected to an airline of 20 kPa (0.2 bar) effective pressure, to create an airflow through the set.Meets criteria
18. Flow rateFlow rate for sets with standard tubing: 23G: ≥ 2.9 ml/min & 25G: ≥ 1.6 ml/min; Flow rate for sets with microbore tubing: 23G: ≥ 1.7 ml/min & 25G: ≥ 1.5 ml/minMeets criteria
19. Bonding strength Surshield protectorThe bonding strength between the Surshield protector and the SV-set is minimum 4 NMeets criteria
20. Misalignment of Surshield ProtectorThe angle measured axially from the cannula between the horizontally positioned wings and the vertically positioned Surshield protector does not exceed 15°.Meets criteria
21. Peel behaviour of blisterThe blister is peeled by hand with minimal fibres and no paper splittingMeets criteria
22. Peel strengthThe peel strength of the pack seal is minimum 0.15.kN/mMeets criteria
23. Package integrityThe integrity of the unit pack is minimum 4 kPaMeets criteria
24. Kink stability of tubingFree of deformation and kinkingMeets criteria
25. Filter performance testingSame or better than predicateMeets criteria

The document states: "The Surflo Winged Infusion Sets with Filter and Needle Protection (Surshield), meet all acceptance criteria as indicated in table above."

2. Sample size used for the test set and the data provenance
The document does not specify the exact sample sizes (number of units tested) for each individual test. It only states that "All necessary verification and validation tests have been performed." Data provenance is not explicitly mentioned as "country of origin," but the submission is from "TERUMO EUROPE N.V." in Leuven, BELGIUM, so the testing likely occurred in Europe. The tests are prospective in nature, as they are performed on the newly designed devices to verify their performance against established criteria.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. These are engineering and performance tests of a physical medical device, not a diagnostic algorithm requiring expert interpretation of medical images or data. The "ground truth" is defined by the technical specifications and international standards (e.g., ISO 594-1, ISO 594-2, EN ISO 11135-1, EN 556-1, EN ISO 10993-7).

4. Adjudication method for the test set
Not applicable. There is no adjudication method described as these are quantitative and qualitative engineering tests against predefined specifications.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an algorithm.

7. The type of ground truth used
The ground truth for these tests is established by international standards and technical specifications for sterile medical devices for intravascular administration. Examples include:

  • ISO 594-1: Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment — Part 1: General requirements
  • ISO 594-2: Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment — Part 2: Lock fittings
  • EN ISO 11135-1: Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
  • EN 556-1: Sterilization of Medical Devices - Requirements for medical devices to be designated "STERILE" - Part - I: Requirements for terminally sterilized medical devices
  • EN ISO 10993-7: Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
  • FDA General Program Memorandum #G95-1: Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and testing"

8. The sample size for the training set
Not applicable. There is no "training set" as this is not a machine learning or AI device.

9. How the ground truth for the training set was established
Not applicable.

{0}------------------------------------------------

MAR 1 8 2014

510(k) SUMMARY

Prepared for:

TERUMO EUROPE N.V. Interleuvenlaan 40, 3001 Leuven, BELGIUM

Prepared by/Contact Person:

Mrs. M.J. Aerts - Manager Regulatory Affairs Tel. (+32) 16 38 13 53 Fax (+32) 16 40 02 49

Date prepared:

December 2013

Note: Class III Summary & Certification is not applicable for the submission of a Class II device

{1}------------------------------------------------

l I. 510(k) SUMMARY

  • l . Device Name
  • Reason for Submission 2.
    1. Intended Use
    1. Description
    1. Substantial Equivalence
    1. Additional Safety Information
  • Summary of Verification Activities 7.
  • Additional Safety Information 8.
    1. Conclusion

{2}------------------------------------------------

510(k) SUMMARY as required by 807.92 Summary of Safety & Effectiveness Information

Submitter information

Prepared for:TERUMO EUROPE N.V.Interleuvenlaan 40,3001 Leuven,BELGIUM
Prepared by/Contact person:Mrs. M.J. Aerts - Manager Regulatory AffairsTel. (+32) 16 38 13 53Fax (+32) 16 40 02 49
Date prepared:December 2013

Device Name l .

:

Proprietary Name

SURFLO® Winged Infusion Set with Filter and Needle Protection (Surshield')

Classification Name

Intravascular Administration Set (-FPA)

21 CFR, Section 880.5440

Classification: Class II

2. Reason for Submission

This premarket notification [510(k)] is being submitted for the 25G Surflo Winged Infusion Set with Filter and Needle Protection (Surshield), to provide supporting information that despite of some changes in material, design, process and labeling to the devices cleared under K070547, the proposed devices are safe and effective and substantially equivalent to the devices cleared under K070547.

3. Intended Use

The Surflo Winged Infusion Set with Filter and Needle Protection (Surshield) is intended to access the peripheral vascular system, for single-dose or short-term intravenous administration of fluids using a syringe or other compatible/appropriate devices. Additionally, after withdrawal of the needle from the patient's vein, the shield cover can be manually activated to cover the needle to minimize risk of accidental needle stick.

Note: This is the same intended use as the predicate device, Surflo Winged Infusion Set with Filter and Needle Protection (Surshield), cleared under K070547.

4. Description

The 23G and the 25G Terumo Surflo Winged Infusion Sets with Filter and Needle Protection (Surshield), are sterile, single use devices consisting of a needle attached to a winged hub, microbore tubing, adapter with integrated 20 um filter and adapter cap, and a hinged shield cover that attaches to the wing just below the needle-to-wing junction.

The shield cover can be turned 180 degrees on the hinge. As the needle is removed from the patient's vessel, the user's finger actively pushes the shield cover until it latches onto the needle using a one- or two- handed technique. An audible click is noted upon activation. The shield cover is

{3}------------------------------------------------

designed to allow the user's finger to remain behind the needle point so that the risk of needle stick injury is minimized. The shield cover is transparent for easy confirmation of the needle held in it.

The device possesses a 350 mm length microbore tubing.

  • ડ . Substantial Equivalence
    The 23G and the 25G Surflo Winged Infusion Sets with Filter and Needle Protection (Surshield), manufactured by Terumo Europe N.V., submitted in this 510(k) file are substantially equivalent in intended use, description/specifications, technology/principles of operation, materials and performance to the following cleared devices:

Surflo Winged Infusion Set with Filter and Needle Protection (Surshield) (K070547) (SV-S23FL35VS & SV-S25FL35VS), manufactured by Terumo Europe N.V.

Differences between the devices do not raise any significant issues of safety and effectiveness.

6. Summary of Technological Modifications compared to Predicate Devices:

The technological modifications made to the subjected device compared to the predicate devices are summarized in the following table:

Predicate deviceModification summary (Proposed device)
Surflo Winged Infusion Set with Filterand Needle Protection (Surshield)(K070547)(SV-S23FL35VS & SV-S25FL35VS)SV-S23FL35 & SV-S25FL35:
a) Change of code
b) Change of tube supplier (material remains identical)
c) Change of adapter design
d) Change of process for embedding of filter in adapter from insert mouldingto ultrasonic welding
e) Change of adapter cap design
f) Change of adapter cap material from PP to different grade of PP.
g) Connection wing hub - tube from tetrahydrofurane solvent bonding tocyclohexanone solvent bonding
h) Connection tube - adapter: from solvent to glue bonding
i) Connection Surshield protector from cyano-acrylate glue to UV curing glue
j) Change of specification for dead space volume
k) Update of labeling

7. Summary of Verification Activities:

All necessary verification and validation tests have been performed for the Surflo Winged Infusion Sets with Filter and Needle Protection (Surshield), to assure substantial equivalence to the predicate Summary of all verification activities including acceptance criteria is given in the devices. following table:

{4}------------------------------------------------

TESTACCEPTANCE CRITERIA
1. Visual inspectionWhen examined by normal vision or under 2.5 fold magnification, all components of the setare smooth, clean and free of flash, surface irregularities, detachable particles, and mouldingor processing defects
2. Effective tubing lengthThe effective tubing length is 350 ± 20mm
3. Dead space volumeDead space volume for sets with standard tubing: ≤ 0.40 mlDead space volume for sets with microbore tubing: ≤ 0.20 ml
4. Air leakage (= integrity of set)No leakage
5. Air leakage luer adaptorNo leakage at the luer connection (according to ISO 594-1 and ISO 594-2)
6. Liquid leakage luer adaptorNo leakage at the luer connection (according to ISO 594-1 and ISO 594-2)
7. Conical fittingThe conical fitting shall be within level 1-3 (according to ISO 594-1 and ISO 594-2)
8. Separation force of Luer slip tipconnectionThe fitting remains attached (according to ISO 594-1)
9. Separation force of Luer lockconnectionThe fitting remains attached (according to ISO 594-2)
10. Stress cracking on Luer conicalfittingNo stress cracking on the conical fitting (according to ISO 594-1 and ISO-594-2)
11. Unscrewing torque of Luerlock fittingsThe fitting remains attached when the fitting between the hub and a reference fitting is testedaccording to ISO 594-2
12. Ease of assemblyA satisfactory fit is achieved when tested according to ISO 594-2
13. Overriding resistance of Luerlock systemsWhen the fitting between the hub and a reference fitting is tested, the reference fitting shallnot override the threads or lugs of the fitting under test (according to ISO 594-2)
14. Torque resistance cap -adapterThe torque force required to unscrew the cap from the adapter does not exceed 9 N.cm
15. Bonding strength tube -adapterThe force required for separating the tube from the adapter is 15 N minimum.
16. Bonding strength tube - winghubThe force required for separating the tube from the wing hub is 15 N minimum.
17. Air flow choke testAir bubbles escape out of the set when immersed under water and connected to an airline of20 kPa (0.2 bar) effective pressure, to create an airflow through the set.
18. Flow rateFlow rate for sets with standard tubing: 23G: ≥ 2.9 ml/min & 25G: ≥ 1.6 ml/minFlow rate for sets with microbore tubing: 23G: ≥ 1.7 ml/min & 25G: ≥ 1.5 ml/min
19. Bonding strength SurshieldprotectorThe bonding strength between the Surshield protector and the SV-set is minimum 4 N
20. Misalignment of SurshieldProtectorThe angle measured axially from the cannula between the horizontally positioned wings andthe vertically positioned Surshield protector does not exceed 15°.
21. Peel behaviour of blisterThe blister is peeled by hand with minimal fibres and no paper splitting
22. Peel strengthThe peel strength of the pack seal is minimum 0.15.kN/m
23. Package integrityThe integrity of the unit pack is minimum 4 kPa
24. Kink stability of tubingFree of deformation and kinking
25. Filter performance testingSame or better than predicate

{5}------------------------------------------------

The Surflo Winged Infusion Sets with Filter and Needle Protection (Surshield), meet all acceptance criteria as indicated in table above. None of the obtained data raises any new issue of safety and effectiveness.

8. Additional Safety Information

The sterility of the Surflo Winged Infusion Sets with Filter and Needle Protection (Surshield) is assured by using a validated sterilization method qualified in accordance with the requirements of EN ISO 11135-1 "Sterilization of health care products - Ethylene oxide - Part 1:Requirements for development, validation and routine control of a sterilization process for medical devices" to a sterility assurance level (SAL) of 106 as required by EN 556-1: "Sterilization of Medical Devices -Requirements for medical devices to be designated "STERILE" - Part - I: Requirements for terminally sterilized medical devices".

Ethylene oxide residual levels resulting from EtO sterilization are in compliance with EN ISO 10993-7: "Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals".

The 23G and the 25G Surflo Winged Infusion Sets with Filter and Needle Protection (Surshield) are Externally Communicating devices, Circulating Blood, Limited Exposure (24 hrs). The devices' blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and testing".

The expiration dating for the 23G and the 25G Surflo Winged Infusion Sets with Filter and Needle Protection (Surshield) have been established at 5 years which is the same as the cleared Surflo Winged Infusion Set with Filter and Needle Protection (Surshield) (K070547).

9. Conclusion

The Surflo Winged Infusion Sets with Filter and Needle Protection (Surshield) manufactured by Terumo Europe N.V., and submitted in this 510(k) file are substantially equivalent in intended use, description, specifications, and technology/principles of operation, materials and performance to the following cleared devices:

Surflo Winged Infusion Set with Filter and Needle Protection (Surshield) (K070547)

Differences between the devices do not raise any new issues of safety or effectiveness.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, with a staff and intertwined snakes. The image is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 18, 2014

TERUMO EUROPE N.V. Mrs. M.J. Aerts Manager Regulatory Affairs Interleuvenlaan 40, 3001 Leuven, BELGIUM

Re: K133894

Trade/Device Name: SURFLO® Winged Infusion Set With Filter And Needle Protection (Surshield®) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: December 18, 2013 Received: December 20, 2013

Dear Mrs. Aerts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2 - Mrs. Aerts

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Mary S. Runner -S

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indication for Use

510(k) Number (if known): K133894

Device Name: Surflo® Winged Infusion Set with Filter and Needle Protection (Surshield®) Indication For Use:

The Surflo Winged Infusion Set with Filter and Needle Protection (Surshield) is intended to access the peripheral vascular system, for single-dose or short-term intravenous administration of fluids using a syringe or other compatible/appropriate devices. Additionally, after withdrawal of the needle from the patient's vein, the shield cover can be manually activated to cover the needle to minimize risk of accidental needle stick.

Prescription Use _ X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use _ (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Olgally signed by Saljad H. Syed -S
Ole calls, Government, Sailad H 2-5-5-1-12000 18 1431401 18 143304 4150

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.