Search Filters

Search Results

Found 27 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    Strauss Surgical Cystoscopes & Hysteroscopes (STS-2040-302ASA HD-endoscope, STS-2040-302AA HD-endoscope
    , STS-2040-302BA HD-endoscope, STS-2040-302BSA HD-endoscope, STS-2040-302ESA HD-endoscope, STS-2040-302CA
    HD-endoscope, STS-2029-302AA HD-endoscope, STS-2029-302BA HD-endoscope, STS-2040-298DSA-RWF-endoscope
    , STS-2040-298ASA-RWF-endoscope, STS-2040-300GSA-RWF-endoscope, STS-2040-303CA-RWF-endoscope, STS-2040
    -280ASA-OLY-endoscope, STS-2040-280ESA-OLY-endoscope, STS-2040-283BSA-OLY-endo

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Strauss Surgical Cystoscopes are intended to illuminate and visualize the male urethra, prostate, and bladder for the purpose of performing diagnostic and surgical procedures.

    Strauss Surgical Hysteroscopes are intended to illuminate and visualize the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.

    Device Description

    The cystoscopes/hysteroscopes described herein are rigid endoscopes for visualizing the urethra, the urinary bladder or uterus, fallopian tube ostium and the cervical canal during the performance of endoscopic procedures in urology or gynecology.

    A rigid endoscope consists of a fiber optic cable and sensitive image transmission system with eyepiece. The fiber optic cable is used to illuminate the site inside the body. The connector for connecting the light guide to the light source is situated at the proximal end of the endoscope. The adapters required to connect the light guide are included in the scope of delivery.

    The distal end of the endoscope features an objective lens that captures the image from inside the body. The image is sent through the image transmission system to the eyepiece.

    AI/ML Overview

    The provided document is a 510(k) clearance letter from the FDA for a medical device called "Strauss Surgical Cystoscopes & Hysteroscopes." This type of document primarily focuses on establishing substantial equivalence to a predicate device rather than detailing specific acceptance criteria and a study proving those criteria are met for the newly cleared device itself.

    Crucially, the document explicitly states under the "Non-Clinical and/or Clinical Tests Summary & Conclusions" section that:

    "The subject and predicate devices have identical technological characteristics. Therefore, no performance testing was necessary to demonstrate that the subject device is equivalent to the predicate device in terms of safety and performance."

    This means that a study proving the device meets individual acceptance criteria, as typically understood for new device performance validation, was not performed or not deemed necessary by the FDA for this 510(k) clearance due to the identical technological characteristics with a previously cleared predicate device.

    Therefore, I cannot provide the requested information regarding specific acceptance criteria and the study that proves the device meets them because the document clearly states such testing was not performed for this clearance.

    However, I can still address some of your points based on the information provided, even if it's to state the absence of the requested detail:


    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not ApplicableNot Applicable
    (No specific performance testing was conducted for this 510(k) clearance based on the document.)(No specific performance testing was conducted or reported for this 510(k) clearance based on the document.)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not applicable, as no performance testing was conducted.
    • Data Provenance: Not applicable, as no performance testing was conducted.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable, as no performance testing was conducted.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable, as no performance testing was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a rigid endoscope for visualization, not an AI-powered diagnostic tool. Furthermore, no performance testing was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a rigid endoscope, not an algorithm. Furthermore, no performance testing was conducted.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable, as no performance testing was conducted.

    8. The sample size for the training set

    • Not applicable, as no performance testing was conducted (and this is not an AI device that would typically have a training set).

    9. How the ground truth for the training set was established

    • Not applicable, as no performance testing was conducted (and this is not an AI device that would typically have a training set).

    Summary of Document's Key Information regarding Performance:

    The FDA clearance for the Strauss Surgical Cystoscopes & Hysteroscopes (K251652) was based on substantial equivalence to existing predicate devices (K150158 Schoelly Cystoscopes/Hysteroscopes and accessories). The manufacturer asserted, and the FDA accepted, that the subject device and the predicate device have identical technological characteristics. Therefore, no new performance testing (non-clinical or clinical) was deemed necessary or performed to demonstrate safety and effectiveness for this specific 510(k) submission. The clearance relies on the established safety and performance of the predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243220
    Manufacturer
    Date Cleared
    2025-07-03

    (269 days)

    Product Code
    Regulation Number
    868.2375
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Onera STS 2 (ONERA STS 2)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Onera STS 2 measures and records multiple physiological parameters from a patient during a sleep study, which are used by clinicians to decide on the diagnosis of sleep disorders.

    Onera STS 2 is intended to be used on a patient who has been prescribed a polysomnography study by a healthcare professional. The device is designed to be used under the direction of a physician or trained technician but applied by a layperson.

    The recorded data will be made available to the healthcare professional to assist in the diagnosis of sleep disorders.

    The device is intended to be used for adults.

    The device is not a life supporting physiological monitor.

    Device Description

    Onera Sleep Test System 2 (Onera STS 2) is a hardware, wearable system for measuring physiological signals during a sleep study. The device can be used in the home (Home Healthcare Environment) as well as in Professional Healthcare Facilities.

    The device measures EEG, EOG, EMG, ECG, respiratory signals, cannula based respiratory flow, oxygen saturation, activity, position, ambient light level and sound pressure level.

    The ECG signal is not intended for diagnosis of cardiac disorders, except for the manual determination of arrhythmia during polysomnography studies.

    The Onera STS 2 does not provide any automated output like heart rate, assessments of arrhythmia, heart rate variability, or other related heart rate measurement functions.

    Onera STS 2 consists of five Sensors applied on the forehead, upper chest area, abdomen and lower leg area (both left and right leg).

    The Sensors should be placed on intact skin. During the night measurement it is not needed to inspect the application sites. The Sensors encrypt the recorded signals and upload the measurement data during the night to the Patient App installed on the patient's phone via Bluetooth. The Patient App securely transfers the data to the cloud interface for further processing once all data has been uploaded. The output of the system is represented by a file in EDF format which contains the recorded (physiological) signals. EDF files can be read by any application software that accepts such files as input.

    It is not possible to relocate the Sensors during a sleep study since the Sensor is single-use and for one sleep study only.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Onera STS 2, based on the provided FDA 510(k) clearance letter:


    Onera STS 2: Acceptance Criteria and Performance Study Summary

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document primarily focuses on the SpO2 measurement accuracy as a key performance metric with specific acceptance criteria.

    ParameterAcceptance CriteriaReported Device Performance
    SpO2 Accuracy (70-100% SpO2 range)±3% (ISO 80601-2-61:2019 Clause 201.12.1.101.1)±2.5%

    Note: While other parameters are listed as "Identical" to the predicate, specific numerical acceptance criteria for those parameters are not explicitly stated in this document beyond their qualitative equivalence.

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size: 12 healthy subjects
      • 9 male, 3 female
      • Aged between 23 and 46 years old
    • Data Provenance: The study was conducted in an "independent research laboratory." The country of origin is not explicitly stated in the provided text. The study involved "induced hypoxia," indicating a prospective, controlled experimental design.

    3. Number and Qualifications of Experts for Ground Truth

    • The document does not mention the use of experts to establish ground truth for the SpO2 accuracy test.
    • The ground truth for SpO2 was established by "laboratory co-oximeter" measurements of arterial hemoglobin oxygen (SaO2) values from blood samples.

    4. Adjudication Method for the Test Set

    • The document does not describe any adjudication method. The SpO2 accuracy was determined by direct comparison of the device's SpO2 measurements to SaO2 values from a laboratory co-oximeter.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted or described in this document. The device primarily measures physiological parameters, and the study focused on the accuracy of these measurements rather than human reader interpretation with or without AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    • Yes, a standalone performance study was conducted for SpO2 measurement accuracy. The device's SpO2 readings were directly compared to reference SaO2 values without human intervention in the SpO2 measurement process itself. The Onera STS 2 is described as measuring and recording parameters, with the output as an EDF file to be read by other software. The SpO2 accuracy assessment is specifically for the device's measurement capability.

    7. Type of Ground Truth Used

    • Objective Measurement (Laboratory Co-oximeter): For SpO2 accuracy, the ground truth was established by arterial hemoglobin oxygen (SaO2) values determined from blood samples using a laboratory co-oximeter, which is considered a gold standard for blood oxygen saturation.

    8. Sample Size for the Training Set

    • The document does not provide information regarding the sample size for a training set. This is likely because the performance study described (SpO2 accuracy) is a validation of the device's sensor capabilities, not an evaluation of a machine learning algorithm that would typically require a training set. The device outputs raw physiological signals in EDF format for clinicians to interpret, rather than providing automated diagnoses based on an internal algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • As no training set is mentioned or implied for the device's core functionality (measuring and recording parameters for clinician interpretation), this information is not applicable and not provided in the document.
    Ask a Question

    Ask a specific question about this device

    K Number
    K240293
    Device Name
    STS
    Manufacturer
    Date Cleared
    2024-09-17

    (229 days)

    Product Code
    Regulation Number
    868.1880
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    STS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The STS device is intended to measure lung function in adult patients while at rest (including spirometry and lung volumes). The STS device is to be used by either a physician, respiratory therapist, or technician. The STS device is intended to be used in a professional healthcare environment.

    Device Description

    The STS is a handheld spirometer intended to measure lung function in adult patients while at rest (including spirometry and lung volumes). The STS is to be used by either a physician, respiratory therapist, or technician. The STS device is pulmonary function testing device that measures both flow/volume (spirometry), lung volume and resistance/compliance parameters. It is a multi-use device that should be used with a compatible single-use, disposable mouthpiece which incorporates a viralbacterial filter protecting the patient from the internal components of the device is battery operated allowing for approximately 40 operating hours between charges. The measurement results, which are transmitted via Bluetooth, are displayed on the physician's computer via the STS software/App.

    AI/ML Overview

    The provided text is a 510(k) summary for the STS device, a pulmonary function data calculator. While it mentions a clinical study, the details provided are limited and do not fully address all requested points regarding acceptance criteria and study specifics for proving the device meets those criteria.

    Based on the provided document, here's what can be extracted and what remains unknown:

    1. Table of acceptance criteria and reported device performance:

    The document primarily focuses on demonstrating substantial equivalence to a predicate device rather than explicitly stating acceptance criteria values for performance. However, an implicit acceptance criterion is a "high correlation" and specific statistical thresholds mentioned in the clinical study summary.

    Acceptance Criteria (Implicit)Reported Device Performance
    High correlation between STS and Body Plethysmography (BP) for resistance and compliance parametersCorrelation coefficients between 0.75 and 0.97
    P-value not to exceed a certain threshold (implied to be low for significance)P-value did not exceed 3%
    Repeatability coefficients below a certain thresholdRepeatability coefficients less than 0.3L
    Within-subject standard deviation below a certain thresholdWithin-subject standard deviation values less than 0.15L

    2. Sample size used for the test set and the data provenance:

    • Sample Size: 161 subjects (61 females, 100 males, implicitly). This test set was used for the clinical study comparing STS to Body Plethysmography.
    • Data Provenance: Not explicitly stated (e.g., country of origin). It is a prospective study as subjects were "enrolled" and "tested by both STS and BP," suggesting data was collected specifically for this study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document states that Body Plethysmography (BP) is considered the "gold standard" for measuring lung volumes. Therefore, the BP measurements served as the ground truth.
    • It does not specify the number of experts or their qualifications involved in performing or interpreting the Body Plethysmography tests. It's implied that standard clinical practice for BP was followed, which would involve trained medical professionals, but this is not detailed.

    4. Adjudication method for the test set:

    • No information on adjudication methods is provided. The comparison is directly between the STS device measurements and the "gold standard" Body Plethysmography measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done or reported. This device is a pulmonary function data calculator and not an AI-assisted diagnostic imaging tool where human reader performance with and without AI assistance would typically be evaluated. The study focuses on the device's accuracy in measuring physiological parameters when compared to a gold standard.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The clinical study described evaluates the "performance of STS technique in calculating resistance and compliance parameters" by comparing it with Body Plethysmography.
    • The STS device itself is a "pulmonary function testing device that measures both flow/volume (spirometry), lung volume and resistance/compliance parameters." This implies a standalone performance evaluation of the device's measurements against the gold standard. However, the device "is to be used by either a physician, respiratory therapist, or technician" and its results are "displayed on the physician's computer via the STS software/App," indicating there's a human-in-the-loop for operation and interpretation. The study evaluates the measurement accuracy, which can be seen as a form of standalone performance for the measurement part of the device.

    7. The type of ground truth used:

    • The ground truth used was "gold standard" Body Plethysmography (BP) measurements of lung volumes. While BP is a clinical measurement, it serves as the established reference standard in this context.

    8. The sample size for the training set:

    • The document does not provide any information about the sample size used for the training set. This is a 510(k) submission, and often, proprietary training data details are not included in the public summary.

    9. How the ground truth for the training set was established:

    • The document does not provide any information on how the ground truth for the training set was established, as details about the training set itself are absent.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223573
    Manufacturer
    Date Cleared
    2023-08-18

    (261 days)

    Product Code
    Regulation Number
    868.2375
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Onera Sleep Test System (Onera STS)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Onera STS measures and records multiple physiological parameters from a patient during a sleep study which are used by clinicians to make a decision on the diagnosis of sleep disorders.

    Onera STS is intended to be used on a patient, who has been prescribed a sleep study by a healthcare professional. The device is designed to be used under the direction of a physician or trained technician but applied by a layperson. The recorded data will be made available to a healthcare professional to assist in the diagnosis of sleep disorders.

    The device is intended to be used for adults.

    Device Description

    Onera STS is a wearable system for measuring (physiological) signals during a sleep study. The device can be used in home (Home Healthcare Environment) as well as Professional Healthcare Facilities, to perform the sleep study.

    Onera STS consists of four sensors applied on the forehead, upper chest area, abdominal and lower leg area.

    The sensors measure EEG, EOG, EMG, ECG, bioimpedance based respiratory effort, bioimpedance based respiratory flow, cannula based respiratory flow, oxygen saturation, activity, position, and sound pressure level.

    The study preparation and data retrieval are done in a professional/clinical environment by the trained operator. The device is provided to the user by the trained operator.

    The device is not a life supporting physiological monitor.

    AI/ML Overview

    The provided text is a 510(k) summary for the Onera Sleep Test System (Onera STS). It details the device, its intended use, and a comparison to a previously cleared predicate device (K210593). The primary change in this submission (K223573) is an extension of the device's use time from 8 hours to 16 hours.

    It's crucial to understand that this document is a premarket notification asserting substantial equivalence, not a detailed clinical study report proving performance against specific acceptance criteria for diagnostic accuracy. The FDA's review for a 510(k) focuses on safety and effectiveness in comparison to a predicate device, rather than requiring new, full-scale clinical trials for every parameter.

    Therefore, the information regarding acceptance criteria and performance studies within this document is limited to demonstrating that the change (extended wear time) does not introduce new safety or effectiveness concerns, and that the device still meets regulatory standards. There is no information provided about a study that assesses the device's diagnostic accuracy or performance against specific, quantifiable "acceptance criteria" for detecting sleep disorders. The device is described as measuring and recording parameters for clinicians to use, not as an automated diagnostic tool itself.

    Given this context, I will address your points based on the available information in the provided text.


    Acceptance Criteria and Device Performance (Based on provided 510(k) Summary)

    The FDA 510(k) summary primarily focuses on demonstrating that the modified device (Onera STS with 16-hour wear time) is substantially equivalent to its predicate device (Onera STS with 8-hour wear time) and that the change does not introduce new safety or effectiveness concerns. There are no explicit "acceptance criteria" for diagnostic accuracy or performance of the device in the context of detecting sleep disorders presented in this document. The "performance testing" mentioned is related to meeting general safety and performance standards, and verifying that the extended wear time does not degrade existing performance or safety.

    Table of Comparison (Not "Acceptance Criteria" for Diagnostic Performance):

    The table below summarizes the comparison between the proposed device and the predicate, highlighting the only stated difference: extended use time. This is a demonstration of substantial equivalence rather than meeting specific diagnostic acceptance criteria.

    CharacteristicProposed device Onera STSCurrently cleared Onera STS (Predicate)ResultImplication for "Acceptance"
    Recording time16 hours8 hoursExtended use timeThe primary change, supported by "new skin irritation testing and 16-hour wear time testing as part of design verification and validation activities" to ensure continued safety and performance at extended duration. This is the core "acceptance" point for this 510(k).
    Regulation Number21 CFR 868.237521 CFR 868.2375IdenticalRegulatory classification is the same.
    Product CodeMNRMNRIdenticalProduct code is the same.
    Indications GeneralMeasures and records multiple physiological parameters from a patient during a sleep study which are used by clinicians to make a decision on the diagnosis of sleep disorders. Intended for patients prescribed a sleep study; applied by layperson under direction of physician/trained technician. Recorded data assists healthcare professional.Measures and records multiple physiological parameters from a patient during a sleep study which are used by clinicians to make a decision on the diagnosis of sleep disorders. Intended for patients prescribed a sleep study; applied by layperson under direction of physician/trained technician. Recorded data assists healthcare professional.IdenticalIntended use remains the same.
    Indications – Patient populationAdultsAdultsIdenticalTarget population remains the same.
    Indications – EnvironmentHome and professional environments.Home and professional environments.IdenticalUsage environment remains the same.
    Indications - LimitationsNot intended to monitor or diagnose the patient and does not issue alarms.Not intended to monitor or diagnose the patient and does not issue alarms.IdenticalDevice limitations remain the same.
    Operating principleMeasuring of electrophysiological and other (sound, flow, position) signals via a range of sensors. Recording of the data. Making the data available for display on a suitable platform.Measuring of electrophysiological and other (sound, flow, position) signals via a range of sensors. Recording of the data. Making the data available for display on a suitable platform.IdenticalCore technology and function are unchanged.
    EnergyBattery powered devices.Battery powered devices.IdenticalPower source unchanged.
    MaterialsPatches are included with the device and found biocompatible.Patches are included with the device and found biocompatible.IdenticalMaterials remain the same, biocompatibility verified.
    Measured parameters (EEG, EOG, EMG head, EMG leg, SpO2, ECG, Respiratory effort via bioimpedance, Respiratory flow via nasal cannula, Sound pressure)All listed as identical. ECG listed as "Similar" for proposed device vs predicate but this appears to be a formatting error in the FDA document as it is then listed as "Identical" across the line. Assuming it is identical.All listed as identical.Identical (or similar as noted for ECG, likely formatting error)The parameters measured are unchanged.
    Derived parameters (Position, Activity)IdenticalIdenticalIdenticalDerived parameters are unchanged.
    Operating temperature10°C - 40°C10°C - 40°CIdenticalOperating range unchanged.
    Operating relative humidity10% - 90%10% - 90%IdenticalOperating range unchanged.

    Study Details (Based on provided 510(k) Summary)

    The provided document describes a 510(k) "premarket notification" which asserts substantial equivalence, not a dedicated clinical study for diagnostic accuracy. Therefore, information related to diagnostic performance studies, AI algorithms, human readers, or specific "acceptance criteria" for disease diagnosis is largely absent.

    1. A table of acceptance criteria and the reported device performance:

      • As explained above, the "acceptance criteria" discussed are for substantial equivalence and compliance with general safety and performance standards for a medical device (e.g., biocompatibility for extended wear, electrical safety, EMC).
      • The reported performance for the primary change (extended wear time) is implied by the statement: "The extension is supported by new skin irritation testing and 16-hour wear time testing as part of design verification and validation activities." This testing demonstrated that the device performs safely and effectively for the increased duration. No quantitative performance metrics related to diagnostic accuracy (e.g., sensitivity, specificity for detecting specific sleep disorders) are provided, as the device records data for clinicians to use for diagnosis, it does not autonomously diagnose.
    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • The document states: "No clinical data was required to support equivalence." This means a dedicated clinical test set for diagnostic accuracy was not part of this 510(k) submission.
      • The extended wear time and skin irritation testing would have involved engineering and biocompatibility testing, likely using a limited number of human subjects or in-vitro/in-vivo models. Specific sample sizes and provenance for these verification/validation activities are not detailed in this summary.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable/Not provided. Since "No clinical data was required," there was no clinical test set for which ground truth needed to be established by experts for diagnostic performance.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable/Not provided.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. The document does not describe the device as incorporating AI for automated diagnosis or interpretation to assist human readers. The device records physiological parameters for clinicians to interpret.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. The device's stated function is to "measure and record multiple physiological parameters... which are used by clinicians to make a decision on the diagnosis of sleep disorders." It is not presented as a standalone diagnostic algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not applicable for diagnostic ground truth. The "ground truth" for this 510(k) was the predicate device's cleared performance and safety standards, combined with verification and validation of the extended wear time not introducing new risks.
    8. The sample size for the training set:

      • Not applicable. This device is not described as involving machine learning or AI that would require a "training set" for diagnostic functions.
    9. How the ground truth for the training set was established:

      • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210593
    Manufacturer
    Date Cleared
    2022-04-06

    (401 days)

    Product Code
    Regulation Number
    868.2375
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Onera Sleep Test System (Onera STS)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Onera STS measures and records multiple physiological parameters from a sleep study which are used by clinicians to make a decision on the diagnosis of sleep disorders.

    Onera STS is intended to be used on a patient, who has been prescribed a sleep study by a healthcare professional. The device is designed to be used under the direction of a physician or trained technician but applied by a layperson.

    The recorded data will be made available to a healthcare professional to assist in the diagnosis of sleep disorders.

    The device is intended to be used for adults.

    Device Description

    Onera STS is a wearable system for measuring (physiological) signals during a sleep study. The device can be used in home (Home Healthcare Environment) as well as Professional Healthcare Facilities, to perform the sleep study.

    Onera STS consists of four sensors applied on the forehead, upper chest area, abdominal and lower leg area.

    The sensors measure EEG, EOG, EMG, ECG, bioimpedance based respiratory effort, bioimpedance based respiratory flow, cannula based respiratory flow, oxygen saturation, activity, position, and sound pressure level.

    The study preparation and data retrieval are done in a professional/clinical environment by a dedicated trained operator.

    The device is not a life supporting physiological monitor

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets the acceptance criteria, based on the provided FDA 510(k) summary for the Onera Sleep Test System (K210593):

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) summary primarily focuses on demonstrating substantial equivalence to a predicate device rather than explicitly stating pre-defined "acceptance criteria" in a quantitative table format for all parameters. However, we can infer some key performance metrics and their proven results:

    Measured Parameter / AspectAcceptance Criteria (Implied/Stated)Reported Device PerformanceStudy Type
    BiocompatibilityNo cytotoxic potentialExposure of L929 mammalian cell cultures to test item extracts shows no cytotoxic potential.Lab Testing (Cytotoxicity)
    Negligible irritantElectrode and enclosure did not produce any primary dermal reaction after exposure to the skin of New Zealand White Rabbits.Lab Testing (Irritation or Intracutaneous reactivity)
    No sensitization potentialElectrode and enclosure did not induce any skin reaction scores at the challenge exposure following an induction phase when applied topical to albino guinea pigs.Lab Testing (Sensitization)
    SpO2 Measurement Accuracy±3% in the range 70-100% (as per ISO 80601-2-61:2019)Onera STS SpO2 showed an accuracy of ±3% in the range 70-100%.Clinical Study (SpO2 accuracy validation)
    Sleep Staging Agreement (Kappa)Substantial agreement (implied for substantial equivalence)Cohen's kappa = 0.69 (overall agreement with predicate device). When Stage N1 was removed, kappa reached 0.81. This is generally considered "substantial" to "almost perfect" agreement depending on interpretation.Clinical Study (Comparative sleep study testing)
    Sleep Staging AccuracyHigh accuracy (implied for substantial equivalence)Wake: Accuracy 94.08 ± 4.34%, Specificity 97.87 ± 2.00%, Sensitivity 61.92 ± 21.70%
    Stage N1: Accuracy 89.62 ± 4.23%, Specificity 95.25 ± 3.02%, Sensitivity 27.19 ± 12.11%
    Stage N2: Accuracy 84.69 ± 4.29%, Specificity 81.55 ± 8.05%, Sensitivity 88.32 ± 4.80%
    Stage N3: Accuracy 95.60 ± 1.80%, Specificity 98.26 ± 2.05%, Sensitivity 76.60 ± 18.58%
    REM: Accuracy 94.70 ± 3.27%, Specificity 95.95 ± 2.66%, Sensitivity 88.12 ± 14.46%Clinical Study (Comparative sleep study testing)
    Sleep Parameter CorrelationHigh correlation (implied for substantial equivalence)Total sleep time: 0.77
    Sleep efficiency: 0.65
    Sleep latency: 0.95
    REM onset latency: 0.58
    Wake after sleep onset: 0.55
    Minutes in Stage N1: 0.67
    N2: 0.69
    N3: 0.65
    REM: 0.91
    Wake: 0.64Clinical Study (Comparative sleep study testing)

    2. Sample Size Used for the Test Set and Data Provenance

    • SpO2 Accuracy Validation:
      • Sample Size: 10 healthy volunteers.
      • Data Provenance: Not explicitly stated regarding country of origin, but described as a controlled study meeting ISO standards. It is a prospective study as it involved active data collection from subjects for the purpose of validating the device.
    • Comparative Sleep Study Testing (Primary Clinical Test Set):
      • Sample Size: 32 patients.
      • Data Provenance: Not explicitly stated regarding country of origin or if it was retrospective or prospective. However, the description states "concurrently applied to 32 patients undergoing a routine sleep study," which implies a prospective, clinical data collection for the validation purpose.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • SpO2 Accuracy Validation: The ground truth for SpO2 accuracy is typically established using a reference oximeter that has been calibrated and validated according to specific standards (e.g., arterial blood gas measurements). The document does not specify the number or qualifications of experts involved in the direct "ground truth" establishment, as it's a technical parameter validated against a standard reference.
    • Comparative Sleep Study Testing:
      • Number of Experts: Not explicitly stated. The document mentions "Studies were scored blinded by a qualified sleep professional." This phrasing suggests at least one, but possibly more, qualified sleep professionals were involved.
      • Qualifications: "Qualified sleep professional." Specific details like years of experience or board certification (e.g., diplomate of the American Board of Sleep Medicine) are not provided.

    4. Adjudication Method for the Test Set

    • SpO2 Accuracy Validation: No explicit adjudication method is mentioned, as SpO2 validation against a reference is a direct measurement comparison.
    • Comparative Sleep Study Testing: The scoring was done "blinded" by a qualified sleep professional. It's not explicitly stated if multiple professionals scored and then adjudicated differences (e.g., 2+1, 3+1). The wording "a qualified sleep professional" might suggest single-reader scoring relative to the predicate device, or it might be a simplification of a more complex process. More specific adjudication methods (e.g., consensus, expert panel review) are not described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study and Effect Size

    • No explicit MRMC comparative effectiveness study involving human readers improving with AI vs. without AI assistance was described. This study is focused on the device's ability to record and process physiological parameters similarly to a predicate device, and the sleep staging/scoring is done by a human professional based on the device's output. The device itself is not described as providing AI-assisted interpretations that then improve human reader performance. Its purpose is to provide the raw data for such interpretation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • The study is essentially a standalone performance evaluation of the device's data collection and preliminary processing capabilities, compared against a predicate device. The "scoring" is done by a human, but the device provides the signals for that scoring. Therefore, the device's performance in generating these signals is evaluated independently of a human's final diagnostic decision. The agreement metrics (Cohen's kappa, accuracy, sensitivity, specificity, correlation) reflect the device's ability to produce data that leads to similar scores as the predicate device.

    7. Type of Ground Truth Used

    • SpO2 Accuracy Validation: The ground truth for SpO2 was based on a reference standard (likely a co-oximeter or another highly accurate method) as per ISO 80601-2-61:2019, generally derived from arterial blood gas measurements.
    • Comparative Sleep Study Testing: The ground truth was established by comparing the Onera STS data to the data from the predicate device (Embletta MPR), with both sets of data scored by a "qualified sleep professional." This means the ground truth for sleep staging and physiological scoring is expert scoring of data from a legally marketed device. It is not stated to be pathology or direct outcomes data, but rather a comparative performance to a widely accepted clinical standard.

    8. Sample Size for the Training Set

    • The provided document does not mention a specific "training set" or its size for the Onera STS device. This implies that the device's underlying algorithms (if any, separate from standard signal processing) were not developed using a distinct, large, and formally defined training dataset in the way a deep learning AI model would be. The focus is on the device's performance in measuring and recording parameters, and the comparison to an existing predicate device.

    9. How the Ground Truth for the Training Set Was Established

    • Since a dedicated "training set" is not described, the method for establishing its ground truth is also not applicable or not provided in this document. The device appears to rely on established physiological measurement principles and signal processing, validated against a predicate device and relevant standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K170624
    Device Name
    TalarLift STS
    Date Cleared
    2017-08-11

    (162 days)

    Product Code
    Regulation Number
    888.3040
    Why did this record match?
    Device Name :

    TalarLift STS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TalarLift STS arthroereisis implant is designed for use in the treatment of the hyperpronated foot and stabilization of the subtalar joint. It is designed to block the posterior and inferior displacement of the talus, allowing normal subtalar joint motion while blocking excessive pronation and the resulting sequelae. The TalarLift STS is intended for the following pathological conditions resulting from disease, injury, or other trauma:
    • Hypermobile pes valgus;
    • Posterior tibial tendon dysfunction;
    • Severe pronation;
    • Subtalar instability;
    • Hypermobile flexible congenital flat foot.

    Device Description

    The TalarLift STS consists of a threaded implant, designed to be inserted between the posterior and middle facets of the subtalar joint, and corresponding instrumentation to facilitate insertion. The system consists of three configurations, conical, cylindrical, and anatomic. All three devices are manufactured from titanium alloy (Ti-6Al-4V) complying with ASTM F136.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the TalarLift STS device, which is a subtalar spacer. It describes the device, its intended use, and claims substantial equivalence to predicate devices. This document does not describe an AI/ML powered device, nor does it contain information about acceptance criteria or a study proving that an AI/ML device meets those criteria.

    Therefore, I cannot provide the requested information about acceptance criteria or a study that proves an AI/ML device meets those criteria based on the provided text. The document primarily focuses on non-clinical testing for mechanical properties and pyrogenicity of a physical implant. It explicitly states, "No clinical studies were performed."

    Ask a Question

    Ask a specific question about this device

    K Number
    K170851
    Manufacturer
    Date Cleared
    2017-08-07

    (138 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Anterior Spine Truss System (STS) Interbody Fusion Device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Anterior Spine Truss System (STS) Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the device must be used with supplemental fixation and must be used with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s).

    Device Description

    The device is an open architecture truss design mathematically formulated to provide structural support with open space throughout the implant for bone growth and fusion. The 4WEB additive manufacturing process provides a hierarchical surface roughness. The implant is made from Ti6Al4V alloy. The device is available in a variety of sizes and lordotic angles to accommodate the patient's anatomy.

    AI/ML Overview

    This document describes a 510(k) premarket notification for a medical device called the "Anterior Spine Truss System (STS) Interbody Fusion Device." This device is an intervertebral body fusion device made from Ti6Al4V alloy, designed with an open architecture truss to support bone growth and fusion. It's intended for skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1, and must be used with supplemental fixation and bone graft.

    Based on the provided text, the document focuses on demonstrating substantial equivalence to predicate devices, primarily through engineering performance testing (mechanical, MR compatibility). It does NOT describe a clinical study involving human patients, nor a study on an Artificial Intelligence (AI) device.

    Therefore, I cannot provide information on acceptance criteria and study details related to an AI/Machine Learning device's performance against human readers or standalone performance, expert ground truth establishment, or sample sizes for training/test sets in the context of an AI study.

    The acceptance criteria and "study" described in this document are related to the physical properties and mechanical performance of the implied medical device itself, not an AI component.

    Here's what can be inferred about the "acceptance criteria" and "study" from the provided text, related to the physical device:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't provide a direct table of specific acceptance limits alongside the exact numerical results for each test. Instead, it lists the performance standards that the device was tested against. The "reported device performance" is implicitly that the device met these standards, allowing for a finding of substantial equivalence.

    Acceptance Criteria CategorySpecific Standard/TestImplied Performance Statement
    Mechanical PerformanceASTM F2077Met (Static & dynamic axial compression, static & dynamic compression shear, static torsion)
    ASTM F2267-04Met (Subsidence Testing)
    MR CompatibilityASTM F2119Met (MR Image Artifact)
    ASTM F2052Met (MR Induced Displacement Force)
    ASTM F2213Met (MR Induced Torque)
    ASTM F2182Met (MR Induced Heating)
    OtherExpulsion testingMet (per accepted industry standard)

    2. Sample size used for the test set and the data provenance:

    • Sample size: Not applicable in the context of a clinical study or AI model test set. For mechanical and MR testing, "samples" would refer to the number of physical devices tested to statistically demonstrate compliance with the standards. This specific number is not provided in the document, but it's typically a small number of devices (e.g., 5-10 per test) as per the individual ASTM standards.
    • Data provenance: Not applicable. These are engineering tests performed on the device itself, not on patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable as there is no "ground truth" in the context of a clinical or AI study. The "ground truth" for the device's performance is established by the engineering standards themselves and the results of the physical tests.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This is not applicable as there are no expert readers or interpretations to adjudicate in these engineering and MR compatibility tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. This document describes an interbody fusion device, not an AI or imaging diagnostic device that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable. This document describes a physical medical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this device's performance is compliance with established engineering and material standards (ASTM standards), demonstrated through physical testing of the device.

    8. The sample size for the training set:

    This is not applicable as there is no training set mentioned for an AI model.

    9. How the ground truth for the training set was established:

    This is not applicable as there is no training set mentioned for an AI model.

    In summary, this document is a 510(k) submission for a physical medical device (an interbody fusion device) and does not involve AI or machine learning technology, nor does it describe a clinical study with human patients in the typical sense of evaluating diagnostic or treatment effectiveness through patient outcomes. The "study" refers to the engineering performance testing to demonstrate the device meets predefined physical and material requirements.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150213
    Date Cleared
    2015-05-08

    (98 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Subtalar Spacer System (STS)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The STS screw is indicated for use in treatment of the hyperpronated foot and stabilization of the subtalar joint. It is designed to block the posterior and inferior displacement of the allowing normal subtalar joint motion while blocking excessive pronation and the resulting sequela.

    • Severely pronated foot;
    • Walking intemperance;
    • Calcaneal stance position greater than 50;
    • Manually correctable deformities;
    • Mid-tarsal breech (arch pain);
    • Forefoot varus greater than 10°.
    Device Description

    The Subtalar Spacer System (STS) a threaded implant designed to be inserted between the posterior and middle facets of the subtalar joint. It is manufactured from Ti-6A1-4V per ASTM F136 and is available in 6 sizes. STS has a center cannula design for use with a guide wire to facilitate proper placement.

    AI/ML Overview

    This FDA 510(k) summary (K150213) is for a medical device called the Subtalar Spacer System (STS). Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    This document does not present explicit acceptance criteria in the typical format of a table with quantitative metrics (e.g., sensitivity, specificity, accuracy, or specific mechanical performance thresholds).

    Instead, the submission relies on demonstrating substantial equivalence to predicate devices. This means the primary "acceptance criterion" is that the new device is as safe and effective as a legally marketed predicate device. The performance is assessed by comparing the new device's technological characteristics and showing no new questions of safety or effectiveness are raised.

    Acceptance Criterion (Implicit)Reported Device Performance
    Safety and Effectiveness (comparable to predicate device)"The safety and effectiveness of the STS implant is adequately supported within this premarket notification."
    "The design characteristics of the subject devices do not raise any new types of questions of safety or effectiveness."
    "From the evidence submitted in this 510(k), the subject devices can be expected to perform at least as well as the predicate systems and are substantially equivalent."
    Identical Indications for UseThe STS implant and the predicate Ortho-Pro STS Screw have "identical indications."
    Same Overall FeaturesThe STS implant and the predicate Ortho-Pro STS Screw have the "same overall features."
    Identical MaterialThe STS implant and the predicate Ortho-Pro STS Screw are "identical in material" (Ti-6Al-4V per ASTM F136).
    Technological Characteristics Comparison (Minor Differences)The differences identified (Total length and Pitch) are presented as not impacting substantial equivalence. The document states, "Testing rationales were provided to support the substantial equivalence of the subject device and show that no new worst-case devices are introduced in this system." This implies that the differences in length and pitch were evaluated and deemed not to introduce new risks or affect performance adversely compared to the predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable in the context of this 510(k) submission. This submission does not detail a "test set" in the sense of a clinical trial or a performance study with a specific number of cases/patients. Instead, it relies on non-clinical (bench) testing and comparison to predicate devices.
    • Data Provenance: The data provenance is primarily from non-clinical evidence (bench testing and technical characteristic analysis) supporting the comparison to predicate devices. No human data (clinical or retrospective/prospective study) is described.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. As no clinical "test set" with ground truth determination is described, there's no mention of experts for this purpose.

    4. Adjudication Method for the Test Set

    Not applicable. No "test set" requiring adjudication is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. This document explicitly states "N/A" under "Substantial Equivalence - Clinical Evidence," indicating that a clinical study, including an MRMC study, was not performed or deemed necessary for this 510(k) submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical implant (Subtalar Spacer System), not an algorithm or AI-driven system.

    7. The Type of Ground Truth Used

    Not applicable in the conventional sense of clinical ground truth (e.g., pathology, outcomes data). For a 510(k) based on substantial equivalence, the "ground truth" is established through:

    • Predicate Device Performance: The safety and effectiveness profile of the legally marketed predicate devices.
    • Engineering/Bench Test Data: Non-clinical testing data to ensure mechanical integrity and comparable performance to the predicate, often against established standards (e.g., ASTM F136 for material).

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical implant, not an AI/ML model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. No training set is involved.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143258
    Manufacturer
    Date Cleared
    2015-04-09

    (147 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    PLIF STS, TLIF STS, OLIF STS (Spinal Truss System)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PLIF STS, TLIF STS and OLIF STS (Spine Truss System) are indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation and must be used with autograft bone. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s).

    Device Description

    The 4WEB PLIF STS, TLIF STS and OLIF STS (Spine Truss System) consists of a series of titanium implants that are designed to provide mechanical support to the spine while biologic fusion occurs. The device is an "open architecture" design consisting of trusses mathematically designed to provide maximum support with the greatest amount of open space throughout the implant for bone growth and fusion. The implant is made from Ti6Al4V alloy. The device is available in multiple sizes to accommodate varied patient's anatomy.

    AI/ML Overview

    This document is focused on the regulatory clearance of a medical device, specifically a Spine Truss System, based on substantial equivalence to predicate devices, rather than an AI/ML powered medical device. Therefore, the requested information on acceptance criteria and studies proving an AI device meets those criteria, as well as details like sample sizes for test/training sets, expert qualifications, adjudication methods, and MRMC studies, are not applicable and cannot be extracted from the provided text.

    The document discusses preclinical testing for mechanical performance of the spine truss system, which is different from validating an AI algorithm's performance.

    Here's what can be extracted related to the device and its assessment, framed as closely as possible to your request, but highlighting the difference in context:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Performance Standards)Reported Device Performance (Summary of Preclinical Testing)
    Meet requirements of static axial compression per ASTM F2077Performed as intended, demonstrating substantial equivalence to predicate devices.
    Meet requirements of static compression shear per ASTM F2077Performed as intended, demonstrating substantial equivalence to predicate devices.
    Meet requirements of dynamic axial compression per ASTM F2077Performed as intended, demonstrating substantial equivalence to predicate devices.
    Meet requirements of dynamic compressive shear per ASTM F2077Performed as intended, demonstrating substantial equivalence to predicate devices.
    Meet subsidence requirements per ASTM F2267-04Performed as intended, demonstrating substantial equivalence to predicate devices.
    Meet expulsion testing requirementsPerformed as intended, demonstrating substantial equivalence to predicate devices.

    Important Note: These are mechanical performance criteria for an implantable device, not performance metrics for an AI algorithm. The reporting states the devices "performed as intended, demonstrating substantial equivalence," which is the regulatory standard for this type of device, implying the test results fell within acceptable ranges defined by the ASTM standards for equivalence. Specific numerical performance values are not provided in this summary.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set: Not applicable in the context of an AI device. For the mechanical testing of the Spine Truss System, the sample size (number of devices tested) for each specific test (static axial compression, dynamic axial compression, etc.) is not explicitly stated in this summary.
    • Data Provenance: Not applicable in the context of an AI device. The tests are preclinical mechanical tests performed on physical devices, not clinical data sets from patients of a specific country.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • This is not applicable as the device is a spinal implant, not an AI diagnostic tool. Ground truth in this context refers to the defined parameters of the ASTM mechanical tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable as this is not an AI diagnostic study requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a physical medical implant, not an AI-assisted diagnostic tool. No MRMC study was conducted or is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the preclinical mechanical testing, the "ground truth" is defined by the objective quantitative measurements and pass/fail criteria established by the ASTM standards (e.g., ASTM F2077, ASTM F2267-04) for intervertebral body fusion devices.

    8. The sample size for the training set:

    • Not applicable. This is not an AI device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K113539
    Date Cleared
    2013-05-03

    (520 days)

    Product Code
    Regulation Number
    872.3640
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    REVOIS PRO IMPLANT STSYEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    REVOIS PRO Implant System is an implant system recommended for surgical placement in the edentulous or partially edentulous jaw bone (upper or lower jaw bone) to create support for prosthetic devices such as single artificial teeth, fixed or removable bridges, or dentures.

    The titanium implant can be applied either in a one-stage surgical procedure with immediate loading (with good primary stability and appropriate occlusal loading) or in a two-stage surgical procedure (after osseointegration of the implant).

    Device Description

    The REVOIS® PRO Implant System is a self-contained, modular dental implant system for placement into the jaw bone (upper or lower jaw bone) to support prosthetic devices for dental restoration. The system is designed for one-stage or two-stage surgical procedures.

    The REVOIS® PRO Implant System is composed of a titanium screw type cylindrical implant. It is a screw form implant with cylindrical thread and a progressive thread near the apex. The double thread of the implant body is contiguous with the triple thread (fine thread) at the implant neck. Together with the conical implant body of the REVOIS®: PRO implant, the progressive thread makes sure that the implant is well anchored in the host bone of the alveolar crest. The incremental pitch of the thread upwards evenly distributes the forces so that implants heal well and can be loaded immediately.

    The implant system is delivered in a sterile insert blister and comprises an implant, an impression abutment, an impression screw and a cover screw (sterile packed as a set in a blister packaging). The impression screw serves as a connector between impression abutment and implant so that the pre-assembled parts fit tightly into the inlay for sterility reasons. The system offers implants in various diameters and lengths (3.8; 4.3; 5.0 mm diameter; 9; 11; 13; 15 mm lengths). For ease of identification the implants are color coded according to diameter.

    The dental abutment portions of the REVOIS® : PRO implant system include several abutments described subsequently:

    • Healing Abutments of diameter 4.4 mm and 5.2 mm in height 2 mm, 3mm and 5 mm each, made of Titanium Grade 5
    • Standard abutment (incl. retaining screw) of diameter 4,4 mm in height 5 mm and 7 mm each, made of Titanium Grade 5
    • Individual aesthetic abutment of diameter 4,4 mm, made of CERAMICOR®, a gold- platinium alloy
    • Impression abutment, multi (pre-mounted with impression screw) made of Titanium Grade 5
    • Impression abutment open tray of diameter 4.4 mm, made of Titanium Grade 5
    • Locator* abutment, made of Titanium Grade 5
    • Locatore Replacement set, consists of several components made of plastic and Titanium Grade 5
    • Locator® impression coping, made of the Aluminium Alloy 6061
    • Dalbot ball attachment of diameter 4.4 mm and in height 3 mm and 5mm, made of Titanium Grade 5

    The REVOIS® PRO Implant System is provided with a variety of prosthetic components and tool which are definitely 510(K)-exempt or still cleared by the submission K063106.

    The main components of the implant system are made of Grade V Titanium (implant, abutment). The materials comply with the ASTM standards ASTM F067 (implant) and ASTM F0136-2a (abutment). The implant surface is blasted with luminium oxide particles (Corundum) and then acid-etched for micro-roughness.

    The REVOIS® PRO titanium implant is sterile packed for single use only. The sterility is indicated by the color on the sterility indicator attached to the inner Tyvek® foil.

    AI/ML Overview

    The acceptance criteria and study proving the device meets them are not explicitly detailed in the provided text in the format typically expected for AI/medical device performance evaluation reports (e.g., sensitivity, specificity, accuracy against a gold standard).

    This document is a 510(k) summary for a dental implant system (REVOIS® PRO Implant System), which primarily focuses on demonstrating substantial equivalence to a previously cleared predicate device, not on presenting novel performance study results for a new AI algorithm.

    Here's a breakdown based on the provided text, addressing your questions where possible:


    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't provide a typical "acceptance criteria" table with performance metrics like accuracy, sensitivity, or specificity. Instead, the "acceptance criteria" for this 510(k) submission are implicitly met by demonstrating substantial equivalence to predicate devices in:

    CategoryAcceptance Criteria (Implicit for Substantial Equivalence)Reported Device Performance (as stated in the summary)
    Intended UseMust be for surgical placement in edentulous or partly edentulous jaw bone to support prosthetic devices (single artificial teeth, fixed/removable bridges, dentures).The REVOIS® PRO Implant System is recommended for surgical placement in the edentulous or partly edentulous jaw bone (upper or lower jaw bone) to create support for prosthetic devices such as single artificial teeth, fixed or removable bridges, or dentures. Can be applied in one-stage (immediate loading with good primary stability) or two-stage (after osseointegration) surgical procedures. This matches the predicate device.
    Technological Characteristics (Design, Materials, Performance)Must be essentially similar in design, materials, and performance aspects, or any differences must not affect safety or effectiveness.Device Design: Titanium screw-type cylindrical implant with cylindrical and progressive threads. Double thread of implant body contiguous with triple (fine) thread at implant neck. Conical implant body. Available in various diameters (3.8, 4.3, 5.0 mm) and lengths (9, 11, 13, 15 mm). Color-coded. Materials: Grade V Titanium (implant, abutment), complying with ASTM F067 and ASTM F0136-2a. Individual aesthetic abutment from CERAMICOR® (gold-platinum alloy). Locator® impression coping from Aluminum Alloy 6061. Surface: Blasted with aluminum oxide particles and then acid-etched for micro-roughness. Performance: Stability of implant body and thread tested via fatigue testing (ISO 14801) and bench testing with an animal model. Tests performed with precursor REVOIS® with comparable results.
    Safety and EffectivenessMust be as safe and effective as the predicate device."The differences outlined in the SE comparison evaluation/discussion between the REVOIS® system and the predicate device, do not affect the safety or effectiveness of the REVOIS® PRO Implant System." (Based on comparative analysis and non-clinical testing data listed).

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not applicable in the context of this 510(k) summary for a dental implant. The "test set" here refers to non-clinical performance testing of mechanical properties:
      • Fatigue Testing was done according to ISO 14801. The standard specifies the number of samples required for fatigue testing, which typically involves a set number of test pieces (e.g., 5-10 samples per condition) to establish fatigue limits. The exact number of implants tested is not specified in the summary but would follow the ISO standard.
      • Bench Testing with Animal Model was also performed. The summary does not specify the number of animals or implants used in this animal model bench test.
    • Data Provenance: The document does not explicitly state the country of origin for the non-clinical tests' data, nor whether it was retrospective or prospective. Given that RIEMSER Arzneimittel AG is a German company, it's highly probable the testing was conducted in Germany or Europe. Non-clinical tests like fatigue testing are inherently prospective at the time of execution.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This question is not applicable to this 510(k) summary. The "ground truth" for a dental implant system like this is established through objective engineering and biological testing (mechanical fatigue, biocompatibility, and animal/human clinical studies for osseointegration and loading), not expert consensus on image interpretation or diagnostic accuracy. The listed studies (e.g., FRIATEC, 1998; Cia et al., 2003) are scientific publications and clinical trials, where the "ground truth" for success or failure would be determined by clinical outcomes, radiographic assessment, or histological analysis, as defined by the methodologies of those studies.


    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set:

    Not applicable. This is not a study assessing diagnostic accuracy requiring expert adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a dental implant, not an AI-powered diagnostic tool. Therefore, no MRMC study or AI assistance evaluation was conducted.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Not applicable. This device is a dental implant.


    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

    For the REVOIS® PRO Implant System, the "ground truth" for its performance and safety is derived from:

    • Mechanical Integrity/Fatigue Testing: Compliance with ISO 14801 standards, which inherently defines "ground truth" for mechanical performance.
    • Biocompatibility: Conformance to ASTM material standards (ASTM F067, ASTM F0136-2a).
    • Bench Testing with Animal Models: Histological and biomechanical outcomes observed in animal models (e.g., osseointegration, bone-implant contact) serve as a form of "ground truth" for biological response before human trials.
    • Clinical Outcomes Data (from referenced studies): The numerous referenced studies (e.g., Cia et al., 2003; Striezel et al., 1998) would have established their "ground truth" through clinical success rates, survival rates, bone loss measurements, clinical evaluations by dentists/surgeons, and patient outcomes over time.

    8. The Sample Size for the Training Set:

    Not applicable. This device is a physical dental implant, not an AI algorithm requiring a training set.


    9. How the Ground Truth for the Training Set Was Established:

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 3