Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K122225
    Manufacturer
    Date Cleared
    2013-04-26

    (275 days)

    Product Code
    Regulation Number
    882.4460
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterile Radiolucent Skull Pins are designed for the fixation of a compatible head-holder unit to the patient's skull. This allows the fixation of the head and neck during craniotomy and subsequent neurosurgical procedures. These pins reduce the incidence of artifacts in images acquired intra- operatively using CT, MR (Conditional) or fluoroscopy

    Device Description

    The Sterile Radiolucent Skull Pins are part of a fixation system which provides rigid immobilization of the patient's skull during surgery. The Pins are the components which get in direct contact with the patient and are made of Carbon-Fiber Reinforced PEEK. The Pins are designed to penetrate the skull tabula externa to provide rigid fixation to compatible head-holder units while significantly reducing imaging artifacts.
    The Sterile Radiolucent Skull Pins are intended for use by physicians or their assistants in a hospital environment only. The product may not be used repeatedly and is a single-use device accordingly. A compatible head- holder unit is required for the safe and effective use of the device.
    The use scenario is primarily rigid fixation of the patient's skull during surgery e.g. percutaneous craniotomy. The device requires the same user actions as the listed predicate device, namely insertion and removal from the head-holder unit. A valid fixation setup with a compatible head- holder unit consists of exactly three sterile pins.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Sterile Radiolucent Skull Pins, based on the provided text:

    Acceptance Criteria and Device Performance

    Test ProtocolAcceptance Criteria (Stated)Reported Device Performance
    Mechanical Performance TestingPins have the required performance and properties to meet the intended use.Verified in cadaver studies and non-clinical mechanical tests.
    Compatibility Testing with Head-Holder UnitsCompatible with specified head-holder units for safe and effective rigid head fixation.Verified for head holder units listed in the instruction leaflet.
    Image Artifact Testing (MR Image Artifacts)Minimize artifacts.Worst artifacts for spin echo sequence = 1.45mm / 1.47mm. Worst artifacts for gradient echo sequence = 2.14mm / 1.94mm.
    Image Artifact Testing (CT Artifacts)Minimize artifacts (implied comparison to predicate).Brainlab Radiolucent Skull Pins = 490 HU (Significantly lower than predicate and other devices).
    Magnetically Induced Displacement Force on Medical Device in MR Environment (ASTM F2052-06e1)No significant magnetically induced displacement force.No magnetically induced displacement force was detectable.
    Magnetically Induced Torque on Medical Devices in MR Environment (ASTM F2213-06)Magnetically induced torque below a certain threshold (implied safe for use).Magnetically induced torque of $10\mu Nm$ (worst case torque due to gravity of $\approx 1.077mNm$).
    Radio Frequency Induced Heating Near Passive Implants during MR Imaging - 1.5 Tesla (ASTM F2182-11a)Temperature rise within safe limits (implied: <5°C).Temperature rise of less than 2.6°C (with a background temperature increase of $\approx 1.5$ °C) at 2.3 W/kg SAR.
    Radio Frequency Induced Heating Near Passive Implants during MR Imaging - 3 Tesla (ASTM F2182-11a)Temperature rise within safe limits (implied: <5°C).Temperature rise of less than 5.9°C (with a background temperature increase of $\approx 2.8$ °C) at 2.5 W/kg SAR.
    Cleaning, Sterilization, and Packaging ValidationsSuccessfully completed to pre-approved protocol; mechanical properties remain same.Successfully completed. Mechanical properties remained the same.
    Clinical Performance (Post-Market Clinical Study)Safe and effective for rigid head fixation; substantially equivalent to predicate device.Users found the product safe and effective.

    Important Note: The document explicitly states, "All tests were passed according to the predetermined acceptance criteria stemming from the intended use." While some criteria are quantitatively defined (e.g., specific artifact measurements, temperature rise), others are more qualitative (e.g., "required performance," "safe and effective").

    Study Details

    1. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Clinical Study (Patients): 11 patients (male and female, aged over 5 years old). The document does not specify the country of origin but mentions Brainlab is a German manufacturer. It is described as a "clinical study," which implies prospective data collection, though it also refers to it as a "Post-Market Clinical Study" in non-US markets, suggesting a retrospective collection of real-world use for feedback.
      • Clinical Study (Cadavers): Not explicitly stated, but "a series of mechanical tests on human cadavers" implies a test set of cadavers.
      • Non-Clinical Data: Not explicitly specified (e.g., number of pins or test repetitions for mechanical, imaging, and MR compatibility tests).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Clinical Study (Patients): "Users" (physicians or their assistants in a hospital environment) provided feedback on safety and effectiveness. Their specific number or qualifications are not provided beyond being medical professionals.
      • Other tests: Ground truth for mechanical, imaging, and MR compatibility tests was based on standardized protocols (ASTM) and internal comparative analysis, rather than expert consensus on individual "cases" in the traditional sense of diagnostic imaging.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not applicable in the context described. The clinical study involved user feedback, not an adjudication of diagnostic labels. Mechanical and imaging tests rely on measurements against standards/benchmarks.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • No MRMC study was performed or described. This device is a physical medical device (skull pins) and not an AI/software device intended to assist human readers in image interpretation.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Not applicable, as this is a physical medical device, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      • Clinical Study (Patients): User feedback on "safe and effective to use for rigid head fixation" serves as the ground truth. This is a form of performance outcome/user experience data.
      • Cadaver Study: The ground truth was based on the "required performance and properties to meet the intended use" as determined by mechanical tests. This is engineering/performance specification truth.
      • Non-Clinical Data (Mechanical, Imaging, MR): Ground truth was established by adherence to recognized industry standards (ASTM) and comparative testing against a predicate device and other commercially available devices to demonstrate improvement in artifact performance.
    7. The sample size for the training set

      • Not applicable. This is a physical medical device, not an AI/machine learning algorithm requiring a "training set." The design and development of the device would have involved internal testing and iterations, but not in the sense of a machine learning training set.
    8. How the ground truth for the training set was established

      • Not applicable, as there is no training set in the context of this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1