(275 days)
Sterile Radiolucent Skull Pins are designed for the fixation of a compatible head-holder unit to the patient's skull. This allows the fixation of the head and neck during craniotomy and subsequent neurosurgical procedures. These pins reduce the incidence of artifacts in images acquired intra- operatively using CT, MR (Conditional) or fluoroscopy
The Sterile Radiolucent Skull Pins are part of a fixation system which provides rigid immobilization of the patient's skull during surgery. The Pins are the components which get in direct contact with the patient and are made of Carbon-Fiber Reinforced PEEK. The Pins are designed to penetrate the skull tabula externa to provide rigid fixation to compatible head-holder units while significantly reducing imaging artifacts.
The Sterile Radiolucent Skull Pins are intended for use by physicians or their assistants in a hospital environment only. The product may not be used repeatedly and is a single-use device accordingly. A compatible head- holder unit is required for the safe and effective use of the device.
The use scenario is primarily rigid fixation of the patient's skull during surgery e.g. percutaneous craniotomy. The device requires the same user actions as the listed predicate device, namely insertion and removal from the head-holder unit. A valid fixation setup with a compatible head- holder unit consists of exactly three sterile pins.
Here's a breakdown of the acceptance criteria and the study information for the Sterile Radiolucent Skull Pins, based on the provided text:
Acceptance Criteria and Device Performance
Test Protocol | Acceptance Criteria (Stated) | Reported Device Performance |
---|---|---|
Mechanical Performance Testing | Pins have the required performance and properties to meet the intended use. | Verified in cadaver studies and non-clinical mechanical tests. |
Compatibility Testing with Head-Holder Units | Compatible with specified head-holder units for safe and effective rigid head fixation. | Verified for head holder units listed in the instruction leaflet. |
Image Artifact Testing (MR Image Artifacts) | Minimize artifacts. | Worst artifacts for spin echo sequence = 1.45mm / 1.47mm. Worst artifacts for gradient echo sequence = 2.14mm / 1.94mm. |
Image Artifact Testing (CT Artifacts) | Minimize artifacts (implied comparison to predicate). | Brainlab Radiolucent Skull Pins = 490 HU (Significantly lower than predicate and other devices). |
Magnetically Induced Displacement Force on Medical Device in MR Environment (ASTM F2052-06e1) | No significant magnetically induced displacement force. | No magnetically induced displacement force was detectable. |
Magnetically Induced Torque on Medical Devices in MR Environment (ASTM F2213-06) | Magnetically induced torque below a certain threshold (implied safe for use). | Magnetically induced torque of $10\mu Nm$ (worst case torque due to gravity of $\approx 1.077mNm$). |
Radio Frequency Induced Heating Near Passive Implants during MR Imaging - 1.5 Tesla (ASTM F2182-11a) | Temperature rise within safe limits (implied: |
§ 882.4460 Neurosurgical head holder (skull clamp).
(a)
Identification. A neurosurgical head holder (skull clamp) is a device used to clamp the patient's skull to hold head and neck in a particular position during surgical procedures.(b)
Classification. Class II (performance standards).