Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K140137
    Manufacturer
    Date Cleared
    2014-03-25

    (62 days)

    Product Code
    Regulation Number
    876.1075
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    SAFETY BIOPSY NEEDLE SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Safety Biopsy Needle System (Safety Biopsy Needle and Safety Coaxial Needle with Finger Guard) is intended to be used for soft tissue and tumor biopsy of such organs as the liver, spleen, kidney, prostate, lung, breast, and lymph nodes.

    When used for breast biopsy, the product is for diagnosis only. The extent of histological abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histological abnormality; e.g., malignancy. When sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal of using standard surgical procedures.

    Device Description

    The Safety Biopsy Needle System consists of two main components: The Safety Biopsy Needle and the Safety Coaxial Needle with Finger Guard.

    The Safety Biopsy Needle is composed of two primary assemblies: the biopsy needle and the safety mechanism. The material used to manufacture both the biopsy needle and the safety mechanism is polycarbonate. The needle sheath is made of polyethylene. The available gauges for the safety biopsy needle ranges in size from 14 to 20, with lengths ranging up to 20 centimeters.

    The Safety Coaxial Needle with Finger Guard is composed of two primary assemblies, the coaxial needle cannula and the stylet. The material used to manufacture the assembly and introducer cannula, and for trocar stylet, is stainless steel. The needle sheath is made of polyethylene. The available gauges for the safety coaxial needle with finger guard ranges in size from 15 to 19, with lengths ranging up to 17 centimeters.

    The Safety Biopsy Needle System is sterilized via Ethylene Oxide (EtO) and is designed for single use only.

    AI/ML Overview

    This document describes the 510(k) summary for the Safety Biopsy Needle System. Here's a breakdown of the requested information based on the provided text:

    Acceptance Criteria and Device Performance Study

    The primary "study" proving the device meets acceptance criteria is a series of non-clinical bench tests and biocompatibility tests. The acceptance criteria are implicitly met by the successful completion of these tests, demonstrating the device's functional integrity, safety features, and material compatibility.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present acceptance criteria in a quantitative table format with corresponding performance values. Instead, it lists the types of performance tests conducted and then states a general conclusion about their successful outcome.

    Implicit Acceptance Criteria (based on tests conducted) and Reported Device Performance:

    Acceptance Criteria Category/TestReported Device Performance
    Functional Performance:
    Ability to extract a biopsy tissue samplePerforms according to intended use
    Average Tissue LengthPerforms according to intended use
    Edge of Cut Sample EvaluationPerforms according to intended use
    Force to Arm the Safety Biopsy NeedlePerforms according to intended use
    Force to Activate the Safety Biopsy NeedlePerforms according to intended use
    Multiple Arming-Activation of the Safety Biopsy NeedlePerforms according to intended use
    Penetration Force of the Safety Biopsy NeedlePerforms according to intended use
    Force to advance the Depth Stop on the Safety Coaxial NeedlePerforms according to intended use
    Force to penetrate tissue on the Safety Coaxial NeedlePerforms according to intended use
    Safety Features Performance (Anti-Needle Stick):
    Force to overcome the Anti-Needle Stick Feature Safety Biopsy NeedlePerforms according to intended use
    Force to Break Safety Tape Safety Biopsy NeedlePerforms according to intended use
    Force to detach Safety Tape from AdapterPerforms according to intended use
    Force to detach Safety Tape from SpoolPerforms according to intended use
    Force to peel Safety Tape after being affixed to each otherPerforms according to intended use
    Force to overcome the Anti-Needle Stick Feature Safety Coaxial NeedlePerforms according to intended use
    Force to Break Safety Tape Safety Coaxial NeedlePerforms according to intended use
    Force to detach Safety Tape from Trocar HubPerforms according to intended use
    Simulated Clinical Use:
    Simulated Clinical Use EvaluationPerforms according to intended use
    Sharps Injury Prevention EvaluationPerforms according to intended use
    Biocompatibility:
    ISO 10993-1:2009 (Biological Evaluation of Medical Devices Part 1)Device meets the requirements of ISO 10993-1:2009
    ISO 10993-7:2008 (Ethylene Oxide Sterilization Results)Performs according to intended use (implied by overall summary)
    Sterilization:
    ISO 11135:2007 (Sterilization of Healthcare Products Ethylene oxide)Performs according to intended use (implied by overall summary)
    Risk Management:
    ISO 14971:2007 (Medical Devices - Risk Management for Medical Devices)Performs according to intended use (implied by overall summary)

    The summary explicitly states: "All of these performance tests demonstrate the device performs according to its intended use and meets the performance specifications." and "The biocompatibility tests conducted on the Safety Biopsy Needle System demonstrate that the device meets the requirements of ISO 10993-1:2009."

    2. Sample size used for the test set and the data provenance

    The document lists "non-clinical bench tests" and "biocompatibility tests" as the performance evaluation methods. It does not specify the sample sizes used for these tests. Data provenance is not applicable in the sense of patient data, as these are bench tests. The tests are conducted on the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. For bench tests, the "ground truth" is typically the established engineering and safety standards, and the adherence of the device to its design specifications. Expert consensus in a clinical diagnostic sense (e.g., radiologists interpreting images) is not relevant for this type of device and testing.

    4. Adjudication method for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 typically refer to the process of resolving discrepancies among multiple human reviewers for clinical data. This is a technical bench test evaluation, not a clinical study involving human judgment on diagnostic results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a biopsy needle system, not an AI or diagnostic imaging device. Therefore, MRMC studies involving human readers and AI assistance are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical instrument, not a software algorithm.

    7. The type of ground truth used

    The ground truth used for these performance tests is engineering specifications, established international standards (e.g., ISO, AAMI/ANSI), and the device's intended use parameters. For example, "Force to activate the Safety Biopsy Needle" would be measured against a predetermined, acceptable force range specified in the design. For biocompatibility, the ground truth is the pass/fail criteria defined by ISO 10993-1:2009.

    8. The sample size for the training set

    Not applicable. This device is a physical medical instrument evaluated by bench testing, not a machine learning model requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable (as above).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1