Search Filters

Search Results

Found 12 results

510(k) Data Aggregation

    K Number
    K031750
    Date Cleared
    2003-10-24

    (141 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    SAFESKIN* BLUE ZONE POWDER-FREE VINYL EXAM GLOVE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a medical device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    The powder-free vinyl exam gloves are disposable devices intended to be worn by healthcare and similar personnel to prevent contamination between such personnel and the patient. The powder-free vinyl exam gloves possess the following technological characteristics (as compared to ASTM or equivalent standards): Dimensions, Physical Properties, Freedom from pinholes, Powder Free.

    AI/ML Overview

    Here's an analysis of the provided information regarding the SAFESKIN* BLUE ZONE Powder-Free Vinyl Exam Glove, structured to address your request:

    Acceptance Criteria and Device Performance Study for SAFESKIN BLUE ZONE Powder-Free Vinyl Exam Glove*

    This submission details the safety and effectiveness of a powder-free vinyl exam glove, demonstrating its substantial equivalence to a predicate device. The primary method of proving equivalence and meeting acceptance criteria relies on compliance with established ASTM standards for patient examination gloves and ISO standards for biocompatibility.

    1. Table of Acceptance Criteria and Reported Device Performance

    Characteristic / Study TitleAcceptance Criteria (Standard)Reported Device Performance
    DimensionsMeets ASTM D 5250-00 ("Standard Specification for Poly(vinyl chloride) Gloves for Medical Application")Meets ASTM D 5250-00
    Physical PropertiesMeets ASTM D 5250-00Meets ASTM D 5250-00
    Freedom from PinholesMeets ASTM D 5250-00 and ASTM D 5151-99Meets ASTM D 5250-00
    Meets ASTM D 5151-99
    Powder FreeMeets ASTM D 6124-01 and ASTM D 5250-00Meets ASTM D 6124-01
    Meets ASTM D 5250-00
    Biocompatibility:(ISO 10993, "Biological Evaluation of Medical Devices")
    ISO Skin Irritation StudyPart 10, "Tests for Irritation and Sensitization" (Pass/Fail)Passed (Test Animal: Rabbit)
    Murine Local Lymph Node Assay (LLNA)Part 10, "Tests for Irritation and Sensitization" (Pass/Fail)Passed (Test Animal: Mouse)
    USP and ISO Systemic Toxicity Study ExtractPart 11, "Tests for Systemic Toxicity(ISO)" (Pass/Fail)Passed (Test Animal: Mouse)

    2. Sample Size Used for the Test Set and Data Provenance

    The provided summary does not explicitly state exact sample sizes for each specific test (e.g., number of gloves tested for dimensions, pinholes, or powder). However, compliance with ASTM standards generally implies specific sampling plans and test methodologies. The data provenance is not explicitly mentioned by country of origin, but the testing was conducted by the Kimberly-Clark Corporation for submission to the US FDA. The studies are prospective in nature, as they are conducted specifically to demonstrate compliance for a new device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This type of medical device (exam glove) does not typically involve human expert interpretation of results in the way, for example, an imaging AI would. The "ground truth" for the acceptance criteria (ASTM and ISO standards) is established by the test methods and objective measurements outlined within those standards. Therefore, no human experts as typically defined in an AI/clinical study context were used to establish the ground truth for the performance test set. Expert committees within ASTM and ISO develop and maintain these standards.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the acceptance criteria are based on objective measurements against established technical standards, not on subjective expert consensus requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret data, often with or without AI assistance, to assess diagnostic accuracy or efficiency. For a patient examination glove, the efficacy is determined by its physical properties and biocompatibility, not by human interpretation of its function in a diagnostic context.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, a standalone (algorithm-only) performance study was not done. This device is a physical product, not a software algorithm. The performance evaluation is based on direct physical and chemical testing of the glove.

    7. The Type of Ground Truth Used

    The ground truth used for evaluating the device's performance is established industry standards and objective test methods. Specifically:

    • ASTM D 5250-00, ASTM D 5151-99, ASTM D 6124-01: These standards define the acceptable physical and chemical properties of patient examination gloves and provide the test methods to determine compliance.
    • ISO 10993 Parts 10 and 11: These international standards define the test methods and acceptable outcomes for biological evaluation (biocompatibility) of medical devices.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical product and does not involve a "training set" in the context of machine learning or AI algorithms. The manufacturing process is subject to quality control, but there isn't a training dataset for an algorithm.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for a software algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K012815
    Date Cleared
    2001-11-13

    (83 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    SAFESKIN HEALTHTOUCH POWDER-FREE EXAM GLOVE WITH VITAMIN E AND ALOE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A medical glove intended to be worn on the hands of healthcare and similar personnel to prevent contamination between such personnel and the patient.

    Device Description

    The powder-free latex exam glove with vitamin E and aloe meets the current specifications for ASTM D 3578-00a. The powder-free latex exam gloves with vitamin E and aloe are disposable devices intended to be worn by healthcare and similar personnel to prevent contamination between such personnel and the patient.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device, specifically the "SAFESKIN HEALTHTOUCH" Powder-Free Latex Exam Glove with Vitamin E and Aloe. It outlines the device's characteristics and its equivalence to a predicate device, as well as the FDA's approval.

    Here's an analysis of the acceptance criteria and the "study" that proves the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    CharacteristicsAcceptance Criteria (Standards)Reported Device PerformanceComments
    DimensionsMeets ASTM D 3578-00aMeets ASTM D 3578-00aConfirmed
    Physical PropertiesMeets ASTM D 3578-00aMeets ASTM D 3578-00aConfirmed
    Freedom from pinholesMeets ASTM D 3578-00aMeets ASTM D 3578-00aConfirmed
    Meets ASTM D 5151-99Meets ASTM D 5151-99Confirmed
    Powder FreeMeets ASTM D 6124-00aMeets ASTM D 6124-00aConfirmed
    Meets ASTM D 3578-00aMeets ASTM D 3578-00aConfirmed
    Protein ContentMeets ASTM D 5712-99 (
    Ask a Question

    Ask a specific question about this device

    K Number
    K011713
    Date Cleared
    2001-07-05

    (31 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    SAFESKIN NEON NITRILE - POWDER FREE NEON EXAM GLOVE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A medical glove intended to be worn on the hands of healthcare and similar personnel to prevent contamination between such personnel and the patient.

    Device Description

    The powder free neon nitrile exam glove is a medical glove intended to be worn on the hands of healthcare and similar personnel to prevent contamination between such personnel and the patient. The powder free neon nitrile exam glove possesses the following technological characteristics (as compared to ASTM or equivalent standards): Dimensions Meets ASTM D 3578-00, Physical Properties (except % elongation) Meets ASTM D 3578-00, Freedom from pinholes Meets ASTM D 3578-00 Meets ASTM D 5151-99, Powder Free Meets ASTM D 6124-00 Meets ASTM D 3578-00. Biocompatability: Primary Skin Irritation in Rabbits Passes, Guinea Pig Sensitization Passes.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the Kimberly-Clark "Safeskin Neon Nitrile" Powder Free Neon Nitrile Exam Glove. It outlines the acceptance criteria and performance of the device based on established standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance
    DimensionsASTM D 3578-00Meets ASTM D 3578-00
    Physical Properties (except % elongation)ASTM D 3578-00Meets ASTM D 3578-00
    Freedom from pinholesASTM D 3578-00 and ASTM D 5151-99Meets ASTM D 3578-00, Meets ASTM D 5151-99
    Powder FreeASTM D 6124-00 and ASTM D 3578-00Meets ASTM D 6124-00, Meets ASTM D 3578-00
    Biocompatibility: Primary Skin Irritation in Rabbits(Implied "Passes" for safety)Passes
    Biocompatibility: Guinea Pig Sensitization(Implied "Passes" for safety)Passes

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for testing each characteristic. However, the tests are conducted in accordance with ASTM standards (e.g., ASTM D 3578-00, ASTM D 5151-99, ASTM D 6124-00), which would typically specify sample sizes for robust testing.

    The data provenance is not specified in terms of country of origin, but the testing would have been conducted by Kimberly-Clark or an accredited lab on their behalf. The data is retrospective in the sense that the tests were performed on finished product samples prior to the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this type of device and study. The "ground truth" for glove performance is established by standardized testing protocols outlined in ASTM standards, not by expert consensus on individual cases.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving interpretation of medical images or diagnostic results where expert agreement is needed to establish ground truth. For medical gloves, performance is objectively measured against specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret data, and the effect of AI on their performance is being evaluated. This device is an examination glove, and its effectiveness is determined by its physical and chemical properties, not by human interpretation of its output.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in a sense, the device's performance is standalone. The tests performed (e.g., dimensions, physical properties, freedom from pinholes, powder-free status, biocompatibility) are objective measurements of the glove itself, without human intervention in its function. The "algorithm" here is the glove's design and manufacturing process, and its performance is assessed directly.

    7. The Type of Ground Truth Used

    The ground truth used is established by objective, standardized measurements and criteria defined within recognized industry standards, specifically ASTM standards (e.g., ASTM D 3578-00, ASTM D 5151-99, ASTM D 6124-00). For biocompatibility, the ground truth is established through standard biological evaluation tests (e.g., primary skin irritation, guinea pig sensitization) designed to assess material safety.

    8. The Sample Size for the Training Set

    This information is not applicable. Medical gloves do not typically involve "training sets" in the context of machine learning or AI algorithms. The design and manufacturing processes are refined through engineering and material science, not through iterative training on data.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no "training set" in the context of this device. The physical and chemical properties of the glove materials and manufacturing processes are developed based on established scientific principles and prior knowledge of material performance and regulatory requirements.

    Ask a Question

    Ask a specific question about this device

    K Number
    K993117
    Manufacturer
    Date Cleared
    2000-03-02

    (164 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    SAFESKIN PFE-XTRA POWDER-FREE LATEX EXAMINATION GLOVES, PROTEIN CONTENT 50 MCG OR LESS, FOR USE WITH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A medical examination glove intended to be worn on the hands of healthcare personnel and similar personnel to prevent contamination between such personnel and the patient and for use with chemotherapeutic agents.

    Device Description

    The Safeskin PFE-XTRA Examination Glove is a disposable device intended to be worn by healthcare and similar personnel to prevent contamination between such personnel and the patient and for use with chemotherapeutic agents.

    AI/ML Overview

    The Safeskin PFE-XTRA Latex Examination Glove underwent testing to demonstrate its substantial equivalence to predicate devices and adherence to relevant standards.

    1. Acceptance Criteria and Reported Device Performance

    CharacteristicsStandards (Acceptance Criteria)Reported Device Performance
    DimensionsMeets ASTM D 3578-99Meets ASTM D 3578-99
    Physical PropertiesMeets ASTM D 3578-99Meets ASTM D 3578-99
    Freedom from pinholesMeets ASTM D 3578-99Meets ASTM D 3578-99
    Freedom from pinholesMeets ASTM D 5151Meets ASTM D 5151
    Powder-FreeMeets ASTM D 6124 (4 mg/glove maximum)Meets ASTM D 6124 (4 mg/glove maximum)
    Primary Skin Irritation in RabbitsPassesPasses
    Guinea Pig SensitizationPassesPasses

    2. Sample Size and Data Provenance for Test Set

    The document does not specify the exact sample sizes used for each individual test (Dimensions, Physical Properties, Freedom from pinholes, Powder-Free, Biocompatibility). However, the tests are laboratory-based, adhering to predefined ASTM standards.

    • Data Provenance: The data is generated from laboratory testing against established ASTM (American Society for Testing and Materials) standards. This indicates a prospective testing methodology. The country of origin of the data is not explicitly stated, but the submission is to the FDA in the USA.

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    This device is not an AI/ML powered device, therefore, the concept of establishing ground truth with experts in the typical sense for image interpretation or diagnosis does not apply. The "ground truth" here refers to the objective measurements and pass/fail criteria defined by the ASTM standards themselves. These standards are developed through a consensus process by experts in the field of materials science and medical devices.

    4. Adjudication Method for Test Set

    Not applicable. The tests are based on objective measurements and adherence to specific ASTM standard specifications, rather than subjective interpretations requiring adjudication.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is not an AI/ML powered device, and no MRMC study was performed or required. The device is a physical medical examination glove. Clinical data were explicitly stated as "not needed for examination gloves" [9].

    6. Standalone Performance (Algorithm Only)

    Not applicable. This is a physical device, not an algorithm.

    7. Type of Ground Truth Used (Test Set)

    The ground truth for the performance claims is derived from established ASTM standards for medical examination gloves. These standards specify objective physical, chemical, and biological properties that the gloves must meet.

    8. Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI/ML model that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no training set for a physical product, the concept of establishing ground truth for a training set does not apply.

    Ask a Question

    Ask a specific question about this device

    K Number
    K994081
    Manufacturer
    Date Cleared
    2000-01-18

    (46 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    SAFESKIN TACTYLON PF POWDER-FREE SURGEON'S GLOVES

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A surgeon's glove is intended to be worn on the hands of operating room personnel to prevent contamination between such personnel and the patient during a surgical procedure.

    Device Description

    Safeskin Tactylon® PF powder-free surgical gloves are sterile disposable devices intended to be worn by operating room personnel to prevent contamination between such personnel and the patient during a surgical procedure. Safeskin® Tactylon PF powder-free surgical gloves possess the following technological characteristics (as compared to ASTM or equivalent standards): Dimensions Meets ASTM D 3577, Physical Properties Meets ASTM D 3577, Freedom from pinholes Meets ASTM D 3577 Meets ASTM D 5151, Powder-Free Meets ASTM D 6124 2 mg/glove maximum.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a medical device, specifically a surgical glove. It does not describe an AI/ML powered device, nor does it detail a clinical study with an AI component. Therefore, much of the requested information regarding AI acceptance criteria, sample sizes for AI training/test sets, expert ground truth, MRMC studies, or standalone algorithm performance is not applicable.

    However, I can extract information related to the acceptance criteria and the study (or lack thereof) to prove the device meets these criteria as presented in the document for the physical device.


    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsAcceptance Criteria (Standards)Reported Device Performance
    DimensionsMeets ASTM D 3577Meets ASTM D 3577
    Physical PropertiesMeets ASTM D 3577Meets ASTM D 3577
    Freedom from pinholesMeets ASTM D 3577 & Meets ASTM D 5151Meets ASTM D 3577 & Meets ASTM D 5151
    Powder-FreeMeets ASTM D 6124 (2 mg/glove maximum)Meets ASTM D 6124 (2 mg/glove maximum)
    Primary Skin IrritationPasses in Rabbits (Biocompatibility)Passes
    Guinea Pig SensitizationPasses (Biocompatibility)Passes

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for the individual tests (e.g., how many gloves were tested for pinholes, or how many rabbits/guinea pigs were used). It implicitly refers to compliance with ASTM standards, which would dictate appropriate sampling methods.

    The data provenance is not explicitly stated as "country of origin" for testing, but the company is based in San Diego, CA, USA, and the submission is to the US FDA. The testing described is intrinsic to the product's physical and biocompatibility characteristics, not clinical data provenance in the sense of patient records.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable as this is a physical device (surgical glove) meeting manufacturing standards, not a diagnostic or AI device requiring expert-established ground truth from medical images or data. The "ground truth" here is the pass/fail criteria defined by ASTM and biocompatibility standards.

    4. Adjudication method for the test set

    Not applicable. The tests are based on objective measurements against engineering and biological standards, not subjective expert assessment requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical device submission and does not involve AI or human readers for diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a physical device submission and does not involve an algorithm.

    7. The type of ground truth used

    The "ground truth" for this device's performance is established by:

    • ASTM (American Society for Testing and Materials) Standards: These are consensus standards for material properties and testing methods (e.g., D 3577 for rubber surgical gloves, D 5151 for detecting pinholes, D 6124 for powder residue).
    • Biocompatibility Standards: These standards (implied by "Passes" for skin irritation and sensitization) ensure the material is safe for contact with human tissue.

    8. The sample size for the training set

    Not applicable. This is a physical device and the concept of a "training set" for an AI model does not apply. The manufacturing process is analogous to "training" in the sense of quality control and process optimization, but not in the AI context.

    9. How the ground truth for the training set was established

    Not applicable. As noted above, there is no AI training set. The "ground truth" for the device's acceptable quality is defined by adherence to established, external manufacturing and material standards (ASTM, biocompatibility).

    Ask a Question

    Ask a specific question about this device

    K Number
    K992830
    Manufacturer
    Date Cleared
    1999-10-27

    (65 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    SAFESKIN POLYMER-COATED NITRILE EXAMINATION GLOVES, POWDER-FREE SHIELDMASTER NITRILE GLOVE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A medical glove intended to be worn on the hands of healthcare and similar personnel to prevent contamination between such personnel and the patient.

    Device Description

    The Safeskin Polymer-Coated Nitrile Examination Glove are sterile disposable devices intended to be worn by healthcare and similar personnel to prevent contamination between personnel and the patient. The Safeskin Polymer-Coated Nitrile Examination Glove possesses the following technological characteristics (as compared to ASTM or equivalent standards): Dimensions Meets ASTM D 3578-99, Physical Properties Meets ASTM D 3578-99, Freedom from pinholes Meets ASTM D 5151, Powder-Free Meets ASTM D 6124 2 mg/glove maximum.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the "Safeskin Polymer-Coated Nitrile Examination Glove." This document focuses on demonstrating substantial equivalence to a predicate device by meeting established performance standards for examination gloves.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard)Reported Device Performance
    ASTM D 3578-99 DimensionsMeets ASTM D 3578-99 Dimensions
    ASTM D 3578-99 Physical PropertiesMeets ASTM D 3578-99 Physical Properties (with the exception of elongation)
    ASTM D 3578-99 Freedom from pinholesMeets ASTM D 3578-99 Freedom from pinholes
    ASTM D 5151 (Freedom from pinholes)Meets ASTM D 5151
    ASTM D 6124 Powder-FreeMeets ASTM D 6124 (2 mg/glove maximum)
    Biocompatibility (Primary Skin Irritation in Rabbits)Passes
    Biocompatibility (Guinea Pig Sensitization)Passes

    Note: The text explicitly states that the device meets ASTM D 3578-99 with the exception of elongation. It then reiterates that it meets "ASTM D 3578-99 Physical Properties" but implies this exception still applies, as it was mentioned earlier in the same paragraph.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size used for the tests to meet the ASTM standards or biocompatibility tests. It also does not mention the country of origin of the data or whether it was retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth and Qualifications of Those Experts

    This information is not applicable to this type of device submission. The "ground truth" for examination gloves is defined by adherence to established industry performance standards (ASTM standards) and biocompatibility tests, not by expert consensus on interpretations.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods are typically used in studies involving human interpretation (e.g., medical imaging) to resolve discrepancies in expert opinions. Here, the assessment is against objective physical and chemical standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not performed. This type of study is relevant for AI-powered diagnostic devices where the AI assists human readers. This device is a medical glove.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, a standalone performance assessment is not applicable. This device is a simple physical barrier, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth used is based on established industry performance standards (ASTM standards) and biocompatibility testing results. These standards define the acceptable physical properties, freedom from defects, and biological safety of the examination glove.

    8. The Sample Size for the Training Set

    This information is not applicable as this device does not involve a training set. Machine learning algorithms, which require training sets, are not used in the development or evaluation of this simple medical device.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this device.


    Summary of the Study:

    The study proving the device meets acceptance criteria is a series of standardized laboratory tests. These tests evaluate the physical properties, integrity (freedom from pinholes), powder content, and biocompatibility of the "Safeskin Polymer-Coated Nitrile Examination Glove" against the specified ASTM (American Society for Testing and Materials) standards. The document states that the glove "will meet the current specifications" for ASTM D 3578-99 (with the exception of elongation), ASTM D 5151, and ASTM D 6124, and "Passes" biocompatibility tests. The specific details of these tests (e.g., number of gloves tested for pinholes, the exact methodology for physical property measurement) are not provided in this summary but would be part of the full submission. The document concludes that these performance test data support a determination of substantial equivalence to a predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K992162
    Manufacturer
    Date Cleared
    1999-08-30

    (66 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    SAFESKIN PURPLE POWDER-FREE NITRILE EXAMINATION GLOVES FOR USE WITH CHEMOTHERAPEUTIC DRUGS LABELING CLAIM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A medical glove intended to be worn on the hands of healthcare and similar personnel to prevent contamination between such personnel and the patient and for use with chemotherapeutic agents.

    Device Description

    Safeskin Purple Nitrile Examination Gloves are disposable devices intended to be worn by healthcare and similar personnel to prevent contamination between such personnel and the patient and for use with chemotherapeutic agents.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Safeskin Purple Nitrile Examination Gloves:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsStandards / Acceptance CriteriaReported Device Performance
    DimensionsMeets ASTM D 6319-99Meets ASTM D 6319-99
    Physical PropertiesMeets ASTM D 6319-99Meets ASTM D 6319-99
    Freedom from pinholesMeets ASTM D 6319-99
    Meets ASTM D 5151Meets ASTM D 6319-99
    Meets ASTM D 5151
    Powder-FreeMeets ASTM D 6124
    2 mg/glove maximumMeets ASTM D 6124
    2 mg/glove maximum
    BiocompatibilityPrimary Skin Irritation in RabbitsPasses
    Guinea Pig SensitizationPasses
    Chemotherapeutic Agent PermeationASTM F 739-96Passes ASTM F 739-96

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for the testing, nor does it detail the provenance (e.g., country of origin, retrospective/prospective) of the data for each test. The testing is implied to be for product validation against ASTM standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of experts to establish a "ground truth" for the test set in the way one might for diagnostic algorithm performance. The device's performance is assessed against established ASTM standards, which themselves are developed by expert consensus in their respective fields but not as part of this device-specific submission.

    4. Adjudication Method for the Test Set

    Not applicable. The "adjudication method" typically refers to how discrepancies in expert opinions are resolved to establish ground truth for a diagnostic study. Here, the device is evaluated against predefined, objective ASTM standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI assistance on human reader performance, which is not applicable to a device like examination gloves. The device is a standalone product.

    6. Standalone Performance Study

    Yes, a standalone performance study was done. The entire submission details the performance of the Safeskin Purple Nitrile Examination Glove itself against established industry (ASTM) and internal standards without human intervention influencing the device's function. The results indicate that the device "will perform according to the glove performance standards referenced in Section 7 above, thereby meeting ASTM standards, FDA requirements, and the labeling claims for the product."

    7. Type of Ground Truth Used

    The "ground truth" used for this device's evaluation is primarily based on established industry standards (ASTM) for physical properties, dimensions, freedom from pinholes, powder content, biocompatibility, and chemotherapeutic agent permeation. There is no mention of pathology, expert consensus (in the context of clinical reads), or outcomes data as the primary ground truth for these product characteristics.

    8. Sample Size for the Training Set

    Not applicable. The concept of a "training set" is relevant for machine learning algorithms. This submission is for a physical medical device (examination gloves) and does not involve AI or machine learning models that require training data. The "training data" here would metaphorically be the experience and research that went into the glove's design, but not in the computational sense.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K992062
    Manufacturer
    Date Cleared
    1999-08-25

    (68 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    SAFESKIN STERILE PURPLE NITRILE EXAMINATION GLOVES

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A medical glove intended to be worn on the hands of healthcare and similar personnel to prevent contamination between such personnel and the patient.

    Device Description

    Safeskin Sterile Purple Nitrile Examination Gloves are sterile disposable devices intended to be worn by healthcare and similar personnel to prevent contamination between such personnel and the patient.

    AI/ML Overview

    Here's an analysis of the provided information, focusing on the acceptance criteria and study details for the Safeskin Sterile Purple Nitrile Examination Gloves:

    Acceptance Criteria and Device Performance

    CriteriaAcceptance StandardDevice Performance
    DimensionsASTM D 6319-99Meets ASTM D 6319-99
    Physical PropertiesASTM D 6319-99Meets ASTM D 6319-99
    Freedom from pinholesASTM D 6319-99Meets ASTM D 6319-99
    ASTM D 5151Meets ASTM D 5151
    Powder-FreeASTM D 6124Meets ASTM D 6124
    Powder-Free (quantitative)2 mg/glove maximumMeets 2 mg/glove maximum
    Primary Skin Irritation in RabbitsPassesPasses
    Guinea Pig SensitizationPassesPasses

    Study Details

    It's important to note that the provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than a comprehensive clinical study report. Therefore, some of the requested information might not be explicitly detailed, especially concerning aspects typically found in clinical trials for more complex devices.

    1. Sample sizes used for the test set and the data provenance:

      • The document does not explicitly state specific sample sizes used for testing each characteristic (e.g., number of gloves tested for dimensions, pinholes, etc.). It generally states that the device "meets" the respective ASTM standards.
      • The provenance of the data (country of origin, retrospective/prospective) is not specified. However, the tests are laboratory-based conformance tests to established standards, not human clinical trials.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable in this context. The "ground truth" for examination gloves is defined by the objective, quantifiable parameters set forth in the ASTM standards (e.g., measurements, tensile strength, pinhole detection methods). Expert opinion is typically not used for these types of performance evaluations.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • This question is not applicable. Adjudication methods like 2+1 or 3+1 are used in studies where there's subjective interpretation (e.g., image reading). For objective performance testing against standards, the results are typically binary (pass/fail) based on direct measurement or observation.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This is not applicable. An MRMC study with AI assistance is relevant for diagnostic imaging devices where human interpretation is involved. This device is an examination glove, and its performance is assessed via physical and biological safety testing, not human reading of diagnostic output.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • This is not applicable. The device is a physical product (examination glove), not a software algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for the device's performance is established by the specified objective criteria and test methods within the referenced ASTM standards (e.g., ASTM D 6319-99, ASTM D 5151, ASTM D 6124). For biocompatibility, it's based on standard animal model tests (rabbit skin irritation, guinea pig sensitization) designed to detect a biological response.
    7. The sample size for the training set:

      • This information is not applicable. Training sets are used for machine learning algorithms. This device is a manufactured product requiring conformance testing to established hardware standards, not algorithmic training.
    8. How the ground truth for the training set was established:

      • This information is not applicable, as there is no training set mentioned or implied for this type of device.

    Overall Conclusion from the Provided Document:

    The device, Safeskin Sterile Purple Nitrile Examination Gloves, meets the acceptance criteria by demonstrating conformance to established industry standards (ASTM D 6319-99, D 5151, D 6124) for dimensions, physical properties, freedom from pinholes, and powder content. Additionally, it passed biocompatibility tests (primary skin irritation in rabbits and guinea pig sensitization). The study demonstrating this is primarily a series of laboratory-based conformance tests against these published standards. Clinical data was deemed not necessary for this type of disposable medical glove. The FDA's 510(k) clearance signifies that the device is substantially equivalent to a legally marketed predicate device based on these performance characteristics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K991951
    Manufacturer
    Date Cleared
    1999-07-26

    (46 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    SAFESKIN SYNTHETIC BINYO POWDER-FREE POLYMER COATED EXAMINATION GLOVES (GREEN, WHITE)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A medical glove intended to be worn on the hands of healthcare and similar personnel to prevent contamination between such personnel and the patient.

    Device Description

    Safeskin Synthetic Powder-Free Examination Gloves are disposable medical gloves intended to be worn by healthcare and similar personnel to prevent contamination between such personnel and the patient.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets them:

    The provided document is a 510(k) summary for a medical device, specifically Safeskin Synthetic Powder-Free Examination Gloves. It focuses on demonstrating substantial equivalence to a predicate device rather than a comprehensive study proving novel performance. Therefore, many of the requested fields related to clinical studies and AI performance will not be applicable.


    1. A table of acceptance criteria and the reported device performance

    CharacteristicsAcceptance Criteria (Standards)Reported Device Performance
    DimensionsMeets ASTM D 5250Meets ASTM D 5250
    Physical PropertiesMeets ASTM D 5250Meets ASTM D 5250
    Freedom from pinholesMeets ASTM D 5250; Meets ASTM D 5151Meets ASTM D 5250; Meets ASTM D 5151
    Powder-FreeMeets ASTM D 6124 (2 mg/glove maximum)Meets ASTM D 6124 (2 mg/glove maximum)
    Primary Skin Irritation in RabbitsPassPasses
    Guinea Pig SensitizationPassPasses

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for testing to demonstrate compliance with ASTM standards. The data provenance is also not specified, only that "The performance test data that support a determination of substantial equivalence are described above." This typically implies testing conducted by the manufacturer. It is a retrospective summary of tests performed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. The ground truth for this type of device (examination gloves) is established through adherence to standardized physical and chemical testing methods (ASTM standards), not through expert consensus on images or clinical outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are typically used in clinical studies or image-based diagnostic evaluations. Compliance with ASTM standards involves objective measurements and defined pass/fail criteria, not subjective adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical glove, not an AI-powered diagnostic tool. Therefore, MRMC studies and AI assistance are irrelevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a medical glove, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth is established by objective measurements against established industry standards (ASTM D 5250-99, ASTM D 5151, ASTM D 6124) as well as biocompatibility testing standards (Primary Skin Irritation in Rabbits, Guinea Pig Sensitization).

    8. The sample size for the training set

    Not applicable. This device is a physical product (medical glove), not a machine learning model, so there is no concept of a "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K991317
    Manufacturer
    Date Cleared
    1999-05-21

    (32 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    POWDER-FREE LATEX SURGEON'S GLOVE SAFESKIN PFS PLUS CONTAINS 50MLGM OR LESS OF TOTAL WATER EXTRACTABLE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A surgeon's glove is intended to be worn on the hands of operating room personnel to prevent contamination between such personnel and the patient during a surgical procedure.

    Device Description

    Safeskin® PFS PLUS powder-free latex surgical gloves are sterile disposable devices intended to be worn by operating room personnel to prevent contamination between such personnel and the patient during a surgical procedure.

    AI/ML Overview

    I am sorry, but I cannot provide information on the acceptance criteria or study details regarding the performance of the device based on the provided document. The document is a 510(k) summary and FDA clearance letter for a medical device (Safeskin® PFS PLUS powder-free latex surgical gloves), which focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed performance study data in the format of acceptance criteria and results.

    Specifically, the document states:

    • "[8] The performance test data that support a determination of substantial equivalence are described above." This refers to a general statement that the device "Meets ASTM D 3577 Dimensions," "Meets ASTM D 3577 Physical Properties," "Meets ASTM D 3577 Freedom from pinholes," "Meets ASTM D 5151 Powder-Free Less than 2 mg/glove," and "Passes Biocompatability Primary Skin Irritation in Rabbits" and "Passes Guinea Pit Sensitization." However, specific numerical acceptance criteria or the reported device performance values (e.g., exact dimensions, specific tensile strength, etc.) are not provided.
    • "[9] Clinical data are not needed for surgical gloves." This indicates that no clinical study was conducted for this submission.

    Therefore, I cannot extract the requested information like sample size, ground truth details, expert qualifications, or comparative effectiveness study results from this document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2