Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K210635
    Date Cleared
    2021-07-27

    (146 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Route 92 Medical Full Length 088 Access System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Route 92 Medical Full Length 088 Access System is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system.

    Device Description

    The Route 92 Medical Full Length 088 Access System is comprised of a Support Catheter and a Delivery Catheter. The Support Catheter is a single-lumen, variable stiffness catheter. The Delivery Catheter is a hubbed, single-lumen variable stiffness catheters are hydrophilically coated. The Route 92 Medical Full Length 088 Access System is intended for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system. The devices are provided sterile and non-pyrogenic and are intended for single use only.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the Route 92 Medical Full Length 088 Access System. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel effectiveness through clinical studies. Therefore, the information requested about acceptance criteria and studies proving the device meets those criteria in the context of efficacy for a new therapeutic claim, a multi-reader multi-case study, or standalone algorithm performance, is not directly applicable in the same way it would be for a de novo submission or a PMA.

    Instead, the document details non-clinical (biocompatibility and performance) testing to show the new device performs similarly safely and effectively to its predicate.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" in a quantitative, numerical format for each performance test, but rather implies them via "All samples met the pre-determined acceptance criteria" or similar statements. The "Reported device performance" is summarized as "PASS" for all tests, indicating that they met these unquantified criteria.

    TestTest MethodAcceptance Criteria (Implied)Reported Device Performance
    Dimensional VerificationDevice dimensions were measured to confirm conformance to the specificationsConformance to specificationsPASS
    Luer IntegrityTested per ISO 80369-7:2016Met ISO 80369-7:2016 requirementsPASS
    Tensile StrengthThe tensile strength of the catheter sections and bonds was testedMet pre-determined acceptance criteria for tensile strengthPASS
    Kink ResistanceTest specimen segments were formed into a defined bend diameter to evaluate kink resistanceMet pre-determined acceptance criteria for kink resistancePASS
    Torsion ResistanceThe test specimens were rotated to evaluate integrity after rotationMaintained integrity after rotation, met pre-determined acceptance criteriaPASS
    Tip FlexibilityTest specimens were tested for tip flexibilityMet pre-determined acceptance criteria for tip flexibilityPASS
    Air LeakageTested per ISO 10555-1:2013 Annex D.Met ISO 10555-1:2013 Annex D requirementsPASS
    Liquid LeakageTested per ISO 10555-1:2013 Annex C.Met ISO 10555-1:2013 Annex C requirementsPASS
    Static BurstTested per ISO 10555-1:2013 Annex F.Met ISO 10555-1:2013 Annex F requirementsPASS
    Dynamic BurstMechanical integrity was maintained up to the specified pressuresMaintained mechanical integrity up to specified pressures, met acceptance criteriaPASS
    Hydrophilic Coating IntegrityThe integrity of the hydrophilic coating was evaluated after multiple insertion and withdrawal cycles.Maintained integrity after multiple cycles, met acceptance criteriaPASS
    Particulate RecoveryAfter multiple insertion and withdrawal cycles, the effluent water rinsed and flushed from the devices and model was tested per USP . Particulates were characterized for size ranges ≥10 µm, ≥25 µm, ≥50 µm, ≥100 µm, ≥200 µm, ≥500 µm and ≥1000 µmMet USP requirements for particulate matter, met acceptance criteriaPASS
    Simulated Use TestingDeliverability and compatibility with accessory devices were evaluated in a neurovascular modelDemonstrated deliverability and compatibility in a neurovascular model, met acceptance criteriaPASS
    Packaging IntegrityISO 11607-1 Part 1
    ISO 11607-2 Part 2Met ISO 11607-1 and ISO 11607-2 requirementsPASS

    Biocompatibility Testing:

    TestConclusions (Reported Performance)
    Cytotoxicity - ISO MEM ElutionThe test article is non-cytotoxic. (Met requirements, grade
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1