Search Results
Found 3 results
510(k) Data Aggregation
(29 days)
React 71 Catheter
The React™ 71 Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.
The React™ 71 Catheter is a single lumen, flexible, variable stiffness composite catheter with a nitinol structure that is jacketed with various durable polymer outer layers. A lubricious, polytetrafluoroethylene liner and Engage™ inner layer is used to create a structure that has both proximal stiffness and distal flexibility. The React™ 71 Catheter is also designed with an encapsulated radiopaque distal platinumiridium marker band which is used for visualization under fluoroscopy. The React™ 71 Catheter is introduced into the vasculature through the Split-Y Introducer Sheath. The proximal end of the React™ 71 Catheter is designed with a green thermoplastic elastomer strain relief and a clear hub. The distal end of the React™ 71 Catheter is coated with a Surmodics Serene™ Coating.
The provided text describes a 510(k) submission for the React™ 71 Catheter, focusing on demonstrating substantial equivalence to a predicate device after a change in device length. The information primarily covers non-clinical bench testing.
Here's an analysis of the provided information against your requested categories:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria | Reported Device Performance |
---|---|
The length of the subject device shall be compatible with minimum length 136 cm guide catheter. | The subject device met the acceptance criteria. |
The subject device should be able to navigate to the distal M1 segment of the MCA over a microcatheter. | The subject device met the acceptance criteria. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not explicitly stated. The text mentions "non-clinical bench testing was conducted," but does not provide details on the number of devices tested for each criterion.
- Data Provenance: Not specified. "Non-clinical bench testing" typically implies laboratory-based tests. The country of origin of the data is not mentioned. The data is prospective in the sense that new tests were conducted for this submission, but it's not clinical human data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This section is not applicable as the study described is non-clinical bench testing, not a study involving human or animal data requiring expert ground truth for interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This section is not applicable as the study described is non-clinical bench testing against predefined performance criteria, not a study involving subjective interpretations requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. This submission is for a percutaneous catheter, not an AI-powered diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
No, a standalone algorithm performance study was not done. This submission is for a physical medical device (catheter), not an algorithm or software.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical bench tests, the "ground truth" is defined by the acceptance criteria established in the test methods, based on FDA Guidance 1757 and FDA Recognized Consensus Standards 5-125 & 5-129. These are objective engineering and performance standards, not clinical ground truths like pathology or expert consensus.
8. The sample size for the training set
This section is not applicable. This is a 510(k) submission for a physical medical device, not a machine learning or AI algorithm that requires a training set.
9. How the ground truth for the training set was established
This section is not applicable for the reasons stated in point 8.
Ask a specific question about this device
(103 days)
React 71 Catheter
The React™ 71 Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.
The React™ 71 Catheter is a single lumen, flexible, variable stiffness composite catheter with a nitinol structure that is jacketed with various durable polymer outer layers. A lubricious, polytetrafluoroethylene liner and Engage™ inner layer is used to create a structure that has both proximal stiffness and distal flexibility. The React™ 71 Catheter is also designed with an encapsulated radiopaque distal platinum-iridium marker band which is used for visualization under fluoroscopy. The React™ 71 Catheter is introduced into the vasculature through the Split-Y Introducer Sheath. The proximal end of the React™ 71 Catheter is designed with a green thermoplastic elastomer strain relief and a clear hub. The distal end of the React™ 71 Catheter is coated with a Surmodics Serene™ Coating.
The provided document describes the React™ 71 Catheter. Based on the content, here's a breakdown of the acceptance criteria and the study performed:
1. Table of Acceptance Criteria and Reported Device Performance
The document lists various tests, their methods, and states that the React™ 71 Catheter "met the acceptance criteria" or similar statements for each. Specific quantitative acceptance criteria are generally not provided in this summary, but the results confirm compliance.
Test Description | Test Method Summary | Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|---|---|
Biocompatibility | |||
Cytotoxicity (Elution Method) | The test article extract showed no evidence of causing cell lysis or toxicity and had a Grade 0 (No Reactivity). | Grade . | Met acceptance criteria for endotoxin levels |
Packaging Visual Inspection | Evaluated per ASTM F1886. | Met acceptance criteria for visual integrity | Met the acceptance criteria. |
Bubble Leak | Evaluated per ASTM F2096. | Met acceptance criteria for leak integrity | Met the acceptance criteria. |
Seal Strength | Evaluated per ASTM F88. | Met acceptance criteria for seal strength | Met the acceptance criteria. |
Bench Performance (Conducted Testing) | |||
Bioburden | Evaluated per ISO 11737-1. | Met acceptance criteria for bioburden | Met the acceptance criteria. |
Visual Inspection | Inspected under x2.5 magnification. | Met acceptance criteria for visual integrity | Met the acceptance criteria. |
Dimensional Measurements | Proximal ID, distal ID, proximal OD, distal OD, usable length, total length, coating length, and distal tip length measured. | Met acceptance criteria for specified dimensions | Met the acceptance criteria. |
Tip Buckling | Evaluated for maximum compressive force. | Met acceptance criteria for tip buckling force | Met the acceptance criteria. |
Kink Resistance | Evaluated for maximum kink diameter. | Met acceptance criteria for kink diameter | Met the acceptance criteria. |
Particulate | Evaluated per USP . | Met acceptance criteria for particulate levels | Met the acceptance criteria. |
Coating Lubricity | Evaluated for average frictional forces. | Met acceptance criteria for frictional forces | Met the acceptance criteria. |
Tensile Strength | Evaluated per ISO 10555-1. Annex B. | Met acceptance criteria for tensile strength | Met the acceptance criteria. |
Liquid Leak | Evaluated per ISO 10555-1. Annex C. | Met acceptance criteria for liquid leak | Met the acceptance criteria. |
Corrosion Resistance | Evaluated per ISO 10555-1. Annex A. | Met acceptance criteria for corrosion | Met the acceptance criteria. |
Hub Aspiration Resistance | Evaluated per ISO 10555-1. Annex D. | Met acceptance criteria for hub air aspiration | Met the acceptance criteria. |
Radiopacity | Markerband length and wall thickness measured; confirmed via fluoroscopy. | Met acceptance criteria for radiopacity | Met the acceptance criteria. |
Luer Standards | Evaluated per ISO 594-1 and ISO 80369-7. | Met acceptance criteria for luer standards | Met the acceptance criteria. |
Compatibility | Inspected for visual damage when delivering and retrieving interventional devices. | Met acceptance criteria for no visual damage/functionality | Met the acceptance criteria. |
Torque to Failure | Evaluated for transmission of proximal torque to the distal tip. | Withstood typical clinical torsional forces | Was able to withstand torsional forces typical of clinical use. |
Dynamic Pressure | Evaluated for the amount of pressure it can withstand. | Withstood typical clinical pressures | Was able to withstand pressures typical of clinical use. |
Coating Integrity | Evaluated for coating coverage and lubricity. | Remained coated and lubricious | Remained coated and lubricious. |
Usability | Evaluated for maneuverability and flexibility against a predicate device. | Met acceptance criteria for maneuverability and flexibility. | Met the acceptance criteria. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify exact sample sizes for each test. For the biocompatibility study on thrombogenicity, it mentions a "Canine Model" with "four (4) hours (±30 minutes) without systemic anticoagulation," and comparing a "control article" to the "test article." This implies a very small comparative animal study, rather than a large test set for AI evaluation.
The provenance of data (e.g., country of origin, retrospective/prospective) is not provided as the tests are primarily bench and animal studies (biocompatibility) conducted in controlled laboratory settings for device performance and safety.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable. This is a medical device submission based on product performance and biocompatibility testing, not an AI/software submission requiring expert ground truth for classification or diagnosis.
4. Adjudication Method for the Test Set
Not applicable. As this is not an AI/software submission involving human interpretation or adjudication for ground truth, this information is not relevant.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size
No. This is a medical device submission for a catheter, not an AI system that would typically undergo MRMC studies.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This is a physical medical device, not an algorithm. The "standalone" performance here refers to the device's physical and biological performance in bench and animal tests.
7. The Type of Ground Truth Used
The "ground truth" for the tests performed is based on established scientific and engineering standards and methods. For example:
- Biocompatibility: In vitro and in vivo testing (e.g., cell cultures, guinea pigs, mice, canine model) following ISO 10993 standards.
- Performance Data – Bench: Standardized laboratory tests (e.g., ISO, ASTM, USP) for physical, mechanical, and material properties.
8. The Sample Size for the Training Set
Not applicable. This is a physical medical device. There is no concept of a "training set" as understood in AI/machine learning. The device design and manufacturing processes are refined through engineering development, not machine learning training.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for an AI model. For the device itself, the "ground truth" for its design and performance requirements is established through regulatory standards, predicate device characteristics, and engineering specifications.
Ask a specific question about this device
(103 days)
Riptide Aspiration System (React 71 Catheter)
The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute iscoke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
The React™ 71 Catheter is a single lumen, flexible, variable stiffness composite catheter with a nitinol structure that is jacketed with a durable polymer outer layer. A lubricious, polytetrafluoroethylene and Engage liner is used to create a structure that has both proximal stiffness and distal flexibility. The React™ 71 Catheter is also designed with an encapsulated radiopaque distal platinum-iridium markerband which is used for visualization under fluoroscopy. The React™ 71 Catheter is introduced into the vasculature through the split y-introducer sheath.
The proximal end of the React™ 71 Catheter is designed with a thermoplastic elastomer strain relief and a clear hub. The distal end of the React™ 71 Catheter is designed with a hydrophilic coating. The React™ 71 Catheter is navigated to the intended treatment site and positioned proximal to the site of occlusion.
Based on the provided text, the Riptide™ Aspiration System (React™ 71 Catheter) is a medical device. The acceptance criteria and the studies performed are detailed in the "510(k) Summary: K182101."
1. Table of Acceptance Criteria and Reported Device Performance:
The document lists numerous tests with their acceptance criteria (implicitly, as the results consistently state "met the acceptance criteria") and the reported device performance. Due to the extensive list, a summarized table is provided below, grouping similar types of tests.
Test Category | Specific Test | Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|---|---|
Biocompatibility | Cytotoxicity | No evidence of cell lysis/toxicity, Grade criteria | Met the acceptance criteria. |
Bioburden | Met ISO 11737-1 criteria | Met the acceptance criteria. | |
Packaging | Visual Inspection | Met ASTM F1886 criteria | Met the acceptance criteria. |
Bubble Leak | Met ASTM F2096 criteria | Met the acceptance criteria. | |
Seal Strength | Met ASTM F88 criteria | Met the acceptance criteria. | |
Performance (Bench) | Visual Inspection | Met criteria for x2.5 magnification | Met the acceptance criteria. |
Dimensional Measurements | Met specified dimensions | Met the acceptance criteria. | |
Tip Buckling | Withstand maximum compressive force | Met the acceptance criteria. | |
Kink Resistance | Met maximum kink diameter | Met the acceptance criteria. | |
Particulate | Met USP criteria | Met the acceptance criteria. | |
Coating Lubricity | Met average frictional forces criteria | Met the acceptance criteria. | |
Tensile Strength | Met ISO 10555-1, Annex B criteria | Met the acceptance criteria at hub and shaft. | |
Liquid Leak | Met ISO 10555-1, Annex C criteria | Met the acceptance criteria. | |
Corrosion Resistance | Met ISO 10555-1, Annex A criteria | Met the acceptance criteria. | |
Hub Aspiration Resistance | Met ISO 10555-1, Annex D criteria | Met the acceptance criteria. | |
Radiopacity | Met criteria for markerband length, wall thickness, and fluoroscopy confirmation | Met the acceptance criteria. | |
Luer Standards | Met ISO 594-1 and ISO 80369-7 criteria | Met the acceptance criteria. | |
Torque to Failure | Withstand typical clinical torsional forces | Was able to withstand torsional forces typical of clinical use. | |
Dynamic Pressure | Withstand typical clinical pressures | Was able to withstand pressures typical of clinical use. | |
Coating Integrity | Remained coated and lubricious | Remained coated and lubricious. | |
System Performance | Recanalization | Met specified criteria for recanalization | Met the acceptance criteria. |
Vacuum Pressure | Met specified criteria for vacuum pressure | Met the acceptance criteria. | |
Suction Flow Rate | Met specified criteria for suction flow rate | Met the acceptance criteria. | |
Lumen Collapse | Remained patent | Remained patent. | |
Usability | Met criteria for maneuverability, flexibility, aspiration, and clot retrieval | Met the acceptance criteria. |
2. Sample Size for the Test Set and Data Provenance:
The document does not explicitly state sample sizes for each bench or animal study. It mentions a "canine model" for the thrombogenicity test and a "porcine model" for the animal performance data. Specific numbers of animals or test articles used in these studies are not provided.
The data provenance is non-clinical bench and animal testing. There is no mention of human subject data. Therefore, it is entirely retrospective or prospective non-clinical data performed in a laboratory/animal setting. The country of origin of the data is not specified, but the submission is to the U.S. FDA, suggesting the studies were conducted to meet U.S. regulatory standards.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:
This information is not applicable as the studies are entirely non-clinical (bench and animal testing). The "ground truth" is based on the objective measurements and observations from the physical and biological tests conducted according to established standards (e.g., ISO, ASTM, USP).
4. Adjudication Method for the Test Set:
This information is not applicable as the studies are entirely non-clinical. Adjudication typically refers to expert consensus or review in clinical studies.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable. The device is a physical medical device (catheter), not an AI software. The studies described are non-clinical bench and animal tests, not an MRMC comparative effectiveness study involving human readers or AI assistance.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:
This information is not applicable. The device is a physical medical device, not an algorithm or AI system.
7. The type of ground truth used:
The ground truth used for the non-clinical studies is based on:
- Objective Measurements: Such as dimensional measurements, pressure readings, flow rates, tensile strength, Luer standards, etc.
- Standardized Test Methods: Adherence to established international (ISO) and national (ASTM, USP, ANSI/AAMI) standards for evaluating materials, sterility, packaging, and performance.
- Biological Endpoints: In animal models, these include observations for systemic toxicity, sensitization, irritation, pyrogenicity, and direct measurement of thrombogenic potential.
- Visual Inspection and Functional Assessment: E.g., for coating integrity, recanalization, lumen patency, and usability by trained personnel.
8. The Sample Size for the Training Set:
This information is not applicable. This is a submission for a physical medical device. The concepts of "training set" and "testing set" are primarily relevant to machine learning and AI, which are not involved here. The non-clinical studies evaluate the device's inherent physical and biological properties.
9. How the ground truth for the training set was established:
This information is not applicable for the same reasons as #8.
Ask a specific question about this device
Page 1 of 1