    Device Name :

    Radifocus Optitorque Angiographic Catheter

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Radifocus Optitorque Angiographic Catheter is indicated for use in cardiac and vascular procedures. It is designed to deliver radiopaque media, guide wires, catheters, and therapeutic agents to selected sites in the vascular system. The different shapes are designed to selectively engage arteries from access sites such as the femoral, radial, and brachial artery.

    Device Description

    The Radifocus Optitorque Angiographic Catheter is comprised of a two-layer construction featuring stainless steel mesh sandwiched between layers of polyurethane and polyamide elastomers. The shaft inner layer and outer layer contain barium sulfate for visibility and contrast under fluoroscopy. A soft tip is attached to the distal portion of some 4 Fr and all 5 Fr and 6 Fr catheters except those with a pigtail design; a soft tube is attached to the distal portion of 5 Fr and 6 Fr catheters with a pigtail design; the 4 Fr products are available with or without a soft tip. Constructed of flexible, supple polyurethane that is permanently welded to the catheter shaft, the soft tip and soft tube are designed to minimize trauma to the vessel wall. There is no change to the design of the device as a result of this submission.

    The device is offered in lengths of 65-120 cm. French sizes and shaft inner diameters are as follows:

    French SizeShaft Inner Diameter
    41.05mm
    51.22mm
    61.32mm

    It is a disposable device intended for single use only. This device is individually packaged and sterilized by ethylene oxide gas.

    AI/ML Overview

    The provided 510(k) summary describes a medical device, the Radifocus Optitorque Angiographic Catheter, and its performance testing. However, it does not include information about AI/ML device performance, human reader studies, or ground truth establishment relevant to AI/ML. The entire document focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance and biocompatibility testing for a material change.

    Therefore, many of the requested items related to AI/ML device performance, sample sizes for test/training sets, expert qualifications, and ground truth establishment cannot be extracted from this document.

    Here's the information that can be extracted, along with explanations for the missing items:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific Test/StandardReported Device Performance/Results
    ISO StandardsRadio-detectability (ISO 10555-1:2013 Section 4.2, ASTM F640-12)"All samples tested met the standard applicable to each test."
    Surface (ISO 10555-1:2013 Section 4.4)"All samples tested met the standard applicable to each test."
    Peak tensile force (ISO 10555-1:2013 Section 4.6)"All samples tested met the standard applicable to each test."
    Freedom from leakage (ISO 10555-1:2013 Section 4.7)"All samples tested met the standard applicable to each test."
    Power injection (ISO 10555-1:2013 Section 4.10, Annex F and G)"All samples tested met the standard applicable to each test."
    Distal tip (ISO 10555-1:2013 Section 4.12)"All samples tested met the standard applicable to each test."
    Flow rate (ISO 10555-1:2013 Section 6.3)"All samples tested met the standard applicable to each test."
    Burst pressure (ISO 10555-1:2013 Section 6.3)"All samples tested met the standard applicable to each test."
    Internal StandardsShaft stiffness (Stiffness of braided-area)"Meets acceptance criteria"
    Torque transmission"Meets acceptance criteria"
    Torque strength"Meets acceptance criteria"
    Tensile strength (Shaft)"Meets acceptance criteria"
    Bending strength"Meets acceptance criteria"
    Cleanliness"Meets acceptance criteria"
    Strength of hub junction"Meets acceptance criteria"
    Tensile Strength of soft tip"Meets acceptance criteria"
    Product dimensions (I.D.; O.D.; Tip I.D.; Tip O.D.; Length)"Meets acceptance criteria"
    Embolic cube testing"Meets acceptance criteria"
    BiocompatibilityCytotoxicity (non-aged and accelerated aged (3 years))Results of the testing demonstrate that the device remains biocompatible throughout its shelf life. A full list of specific acceptance criteria is not provided, but the overall conclusion indicates meeting standards.
    Sensitization (non-aged)Results of the testing demonstrate that the device remains biocompatible throughout its shelf life.
    Intracutaneous Reactivity (non-aged)Results of the testing demonstrate that the device remains biocompatible throughout its shelf life.
    Systemic Toxicity (acute) (non-aged)Results of the testing demonstrate that the device remains biocompatible throughout its shelf life.
    Pyrogenicity (non-aged)Certified to be non-pyrogenic in the unopened and undamaged package. Performed in accordance with ISO 10993-11 (2006), ISO 10993-12 (2012), and USP 35, NF 30, 2012 Pyrogen Test (USP Rabbit Test).
    Hemolysis (non-aged and accelerated aged (3 years))Results of the testing demonstrate that the device remains biocompatible throughout its shelf life.
    Thrombogenicity (non-aged)Results of the testing demonstrate that the device remains biocompatible throughout its shelf life.
    Complement Activation (Immunology) (non-aged)Results of the testing demonstrate that the device remains biocompatible throughout its shelf life.
    Chemical Extraction (non-aged and accelerated aged (3 years))Results of the testing demonstrate that the device remains biocompatible throughout its shelf life.
    SterilizationSterility Assurance Level (SAL)Validated to provide a SAL of 10^-6 per ISO 11135:2014.
    Ethylene Oxide (EO) ResidualsEO will not exceed 4 mg per device and Ethylene Chlorohydrin (ECH) will not exceed 9 mg per device, meeting requirements for limited exposure devices per ISO 10993-7.
    Risk AnalysisISO 14971 Compliance"A Product Risk Analysis was conducted... and it was determined that any new risks were adequately captured and mitigated."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This document describes non-clinical performance and biocompatibility testing for a physical medical device. It does not refer to "test sets" in the context of AI/ML models or clinical data. The testing was performed on "samples" of the device, both non-aged and accelerated aged (3 years). The exact number of samples for each test is not specified, only that "All samples tested met the standard applicable to each test."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable to the information provided. The document concerns a physical medical device and its non-clinical performance, not an AI/ML device that requires expert-established ground truth from clinical images or data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable to the information provided. The document concerns a physical medical device and its non-clinical performance, not an AI/ML device that requires adjudication of expert labels.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such study was conducted or reported in this document. This filing is for a conventional medical device, not an AI/ML diagnostic aid. The document explicitly states: "This 510(k) does not include data from clinical tests."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device described is not an algorithm or AI/ML product.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical performance tests, the "ground truth" or acceptance criteria were derived from recognized consensus standards (ISO, ASTM, USP) and the manufacturer's own internal standards for device functionality and safety. For biocompatibility, it was based on ISO 10993 standards and FDA G95-1 guidance.

    8. The sample size for the training set

    This is not applicable. There is no AI/ML model or "training set" relevant to this device.

    9. How the ground truth for the training set was established

    This is not applicable. There is no AI/ML model or "training set" relevant to this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082736
    Date Cleared
    2008-09-24

    (6 days)

    Product Code
    Regulation Number
    870.1200
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    RADIFOCUS OPTITORQUE ANGIOGRAPHIC CATHETER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Radifocus Optitorque Angiographic Catheter is intended for cardiac and vascular procedures. It is designed to deliver radiopaque media, guide wires, catheters, and therapeutic agents to selected sites in the vascular system. The different shapes are designed to selectively engage arteries from access sites such as the femoral, radial, and brachial artery.

    Device Description

    The Radifocus Optitorque Angiographic Catheter is a two-layer construction comprised of a stainless steel mesh sandwiched between layers of polyurethane polyamide elastomer. The polyurethane polyamide elastomer contains barium sulfate for visibility and contrast under fluoroscopy. The Catheter has a "soft-tip" whose purpose is to minimize trauma to the vessel wall. The soft-tip is a flexible, supple polyurethane tip that is permanently welded to the catheter shaft.

    AI/ML Overview

    The provided text is a 510(k) summary for the Radifocus® Optitorque™ Angiographic Catheter. It states that the device is substantially equivalent to a previously cleared device (K992051) and thus does not contain a typical study with acceptance criteria and performance metrics for a new device. Instead, the justification for substantial equivalence relies on the fact that the device is identical in design, materials, and principle of operation to the predicate device, and a risk/hazard analysis was conducted.

    Therefore, many of the requested details about acceptance criteria, performance, sample sizes, expert involvement, and ground truth establishment are not applicable or cannot be extracted from this document, as a new clinical performance study was not required or submitted for this specific 510(k).

    Here's an attempt to address the points based on the provided text, indicating where information is not available:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Mechanical/Physical Specifications
    Available Sizes4 - 6 Fr.
    Catheter Length650 – 1200 mm
    Maximum Injection Pressure4 Fr.: 750 psi 5, 6 Fr.: 1000 psi
    Material BiocompatibilityBlood contacting materials are biocompatible (tested in accordance with ISO-10993).
    Sterilization EfficacySterilized to a SAL of 10^-6 (validated in accordance with EN ISO 11135-1).
    Overall Performance for New IndicationJustified that the performance is substantially equivalent to the predicate device (K992051).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided as a formal performance study on a test set (clinical or otherwise with specific sample sizes) for this new 510(k) was not conducted. The submission relies on substantial equivalence to a predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided as a specific clinical performance study requiring expert ground truth establishment for this 510(k) was not conducted.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided as a specific clinical performance study requiring adjudication for this 510(k) was not conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study was not conducted, as this is a medical device (angiographic catheter), not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is an angiographic catheter and does not involve an algorithm. The "Principle of Operation / Technology" states it "is operated manually or by a manual process."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not applicable to this type of device submission. The performance justification relies on compliance with standards for material safety, sterility, and mechanical/physical specifications rather than a comparative ground truth for diagnostic accuracy.

    8. The sample size for the training set

    This information is not applicable as this is a medical device (angiographic catheter) and does not involve a "training set" in the context of an algorithm or AI model development.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reasons as point 8.