Search Results
Found 1 results
510(k) Data Aggregation
(15 days)
The RadialSeal™ Introducer Kit is used to facilitate placing a catheter through the skin into the radial artery.
The RadialSeal™ Introducer Kit consists of an introducer (a sheath and a dilator), and either a needle or a stainless steel guidewire or an IV catheter with a plastic guidewire which has a pre-shaped distal tip. The kit is used to facilitate placing a catheter through the radial artery. The thin-walled introducer sheath is coated for reduced friction and has a hemostasis valve to minimize blood loss and air intake. The side port on the introducer sheath hub has a 3-way stopcock for flushing and infusion. The dilator and sheath cannula contain radiopaque material for visualization under fluoroscopy.
An entry needle is provided to facilitate entry into the vessel. The entry needle is offered in either a stainless steel bare version, or as part of an IV catheter. A guidewire (0.021" diameter) is inserted through the bare entry needle or the cannula of the IV catheter, and maintains access to the puncture site upon removal of the needle and before insertion of the introducer (sheath and dilatorassembly).
A dilator facilitates the entry of the sheath introducer by forming an atraumatic transition from the skin through the subcutaneous tissue to the vessel. The introducer is inserted over the guidewire and through the skin into the radial artery. Once the introducer is situated in the vessel, the dilator and guidewire are removed to allow access by the treatment device(s).
The introducers are available in French sizes of 4, 5, and 6, and in lengths of 10, 16, and 25cm. The 0.054 cm (0.021" diameter) guidewire is available in either 45 cm (17.7 in) or 80 cm (31.5 in) length, and is comprised of either stainless steel spring coil or plastic (poly-jacketed Nitinol) material.
This document, a 510(k) Summary for the RadialSeal™ Introducer Kit (K181855), focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study to prove the device meets acceptance criteria specific to algorithm performance. Therefore, many of the requested categories related to AI model evaluation will not be applicable.
Here's an analysis based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state quantitative acceptance criteria in a tabular format for each test, nor does it provide specific numerical results for the RadialSeal™ Introducer Kit for most tests. Instead, it generally states that "results indicated that no new issues of safety and effectiveness were raised and the results were within the predetermined acceptance criteria" and that "The devices performed equivalently."
However, it does list the types of performance tests conducted and the standards they followed. A summary of the reported performance is general equivalency to the predicate device.
| Test Category | Method / Standard | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| Sheath Performance Testing | Implied to meet relevant ISO 11070 requirements and internal standards, showing no new safety/effectiveness issues. | Performed equivalently to predicate device. Results within predetermined acceptance criteria. | |
| Surface | ISO 11070 Sec 4.3 | N/A (implied to meet standard) | Equivalency to predicate. |
| Corrosion Resistance | ISO 11070 Sec 4.4 | N/A (implied to meet standard) | Equivalency to predicate. |
| Radio-detectability | ISO 11070 Sec 4.5 | N/A (implied to meet standard) | Equivalency to predicate. |
| Dimensions- Size Designation | ISO 11070 Sec 7.2 | N/A (implied to meet standard) | Equivalency to predicate. |
| Sheath Introducer leakage | ISO 11070 Sec 7.3 | N/A (implied to meet standard, freedom from leakage) | Equivalency to predicate. |
| Hemostasis | ISO 11070 Sec 7.4 | N/A (implied to meet standard, freedom from leakage through valve) | Equivalency to predicate. |
| Peak Tensile Force | ISO 11070 Sec 7.6 | N/A (implied to meet standard) | Equivalency to predicate. |
| Sheath to dilator fit, rollback test, Puncture model test, Kink Angle, Radius of Curvature | ISO 11070 Sec A.1 | N/A (implied to meet standard) | Side by side testing showed equivalent performance to predicate. |
| Valve Insertion Force | Internal Standard | N/A (implied to meet standard) | Side by side testing showed equivalent performance to predicate. |
| Introducer Insertion Force | Internal Standard | N/A (implied to meet standard) | Side by side testing showed equivalent performance to predicate. |
| External surface sliding force | Internal Standard | N/A (implied to meet standard) | Side by side testing showed equivalent performance to predicate. |
| Hydrophilic coating separation resistance (cycle testing) | Internal Standard | N/A (implied to meet standard) | Side by side testing showed equivalent performance to predicate. |
| Hydrophilic coating particulate evaluation (In vitro particulate test) | FDA Guidance, USP 788 | N/A (implied to meet standard) | Side by side testing showed equivalent performance to predicate. |
| Dimensional Verification | ISO 11070 Sec 7.2 | N/A (implied to meet standard) | Equivalency to predicate. |
| Hub | ISO 11070 Sec 7.5, ISO 594-1, ISO 594-2 | N/A (implied to meet standard) | Equivalency to predicate. |
| Stainless Steel Needle Performance Testing | Implied to meet relevant ISO 11070 requirements. | N/A (implied to meet relevant ISO standards for all needle tests) | |
| IV Catheter Performance Testing | Implied to meet relevant ISO 11070 and ISO 594-1 standards. | N/A (implied to meet relevant ISO standards for all catheter tests) | |
| IV Needle Performance Testing | Implied to meet relevant ISO 11070 and ISO 594-2 standards. | N/A (implied to meet relevant ISO standards for all IV needle tests) | |
| Plastic Guidewire Performance Testing | Implied to meet relevant ISO 11070 standards. | N/A (implied to meet relevant ISO standards for all plastic guidewire tests) | |
| Stainless Steel Guidewire Performance Testing | Implied to meet relevant ISO 11070 standards. | N/A (implied to meet relevant ISO standards for all SS guidewire tests) | |
| Dilator Performance Testing | Implied to meet relevant ISO 11070 and ISO 594-2 standards. | N/A (implied to meet relevant ISO standards for all dilator tests) | |
| Biocompatibility | ISO 10993-1 | No new issues of safety or effectiveness. | Testing completed per ISO 10993-1. Device classified as externally communicating, circulating blood, limited contact (<24 hours). |
| Sterilization | ISO 11135-1:2007, TIR28:2009 | SAL level of 10^-6 | Sterilized by ethylene oxide to SAL level of 10^-6. |
| Packaging | ASTM F2096, ASTM F88, ASTM F1886, ISTA 2A, ASTM D4169 | Maintain integrity for supported shelf life. | Package integrity testing included these standards. |
| Shelf Life | N/A | Support minimum 6 months shelf life. | Supported 6 months shelf life. |
| Animal Testing | GLP Acute Animal Porcine study | Devices perform equivalently. | Both new and predicate devices performed equivalently. |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: The document does not specify exact sample sizes for each bench or animal test. It mentions that "side by side testing between the predicate device and RadialSeal was performed" and implicitly suggests multiple units were tested to demonstrate equivalency.
- Data Provenance: The bench testing refers to compliance with ISO standards and internal standards. The animal study was a "GLP Acute Animal Porcine study" conducted at "American Preclinical Services." This implies the animal data is from a controlled, prospective animal study conducted in the USA (presumably, given "American Preclinical Services"). All data appears to be prospective as it was generated to support the 510(k) submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This is not applicable as this is a medical device (introducer kit) and not an AI/software device that requires ground truth derived from expert consensus for performance evaluation. The "ground truth" for this type of device is defined by passing established physical, chemical, and biological performance standards and demonstrating equivalence to a predicate device.
4. Adjudication Method for the Test Set
Not applicable for a physical medical device. The "adjudication" is compliance with international and internal standards, and comparison to a predicate device's performance.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
This is not applicable as this is a physical medical device, not an AI or software device.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable as this is a physical medical device.
7. The Type of Ground Truth Used
For this device, the "ground truth" is defined by:
- Compliance with established international and national standards (e.g., ISO 11070, ISO 10993-1, ISO 594-1, ISO 594-2, ASTM, USP).
- Demonstrating functional equivalence to a legally marketed predicate device (Terumo Glidesheath Slender® K142183) through direct comparative testing.
- The safety and effectiveness of the device as assessed through non-clinical (bench) and pre-clinical (animal) testing to ensure it raises no new issues compared to the predicate.
8. The Sample Size for the Training Set
Not applicable as this is a physical medical device and not an AI/software device that requires a training set. The "training" in manufacturing would relate to process validation and quality control, not data training.
9. How the Ground Truth for the Training Set was Established
Not applicable as this is a physical medical device and not an AI/software device.
Ask a specific question about this device
Page 1 of 1